OBERON: LASER SURGERY FIBER, ATRAUMATIC SURGERY FIBER, SIDE FIRE FIBER, RADIAL EMISSION FIBER, DENTAL SURGERY FIBER
K140470 · Oberon GmbH Fiber Technologies · GEX · Oct 15, 2014 · General, Plastic Surgery
Device Facts
Record ID
K140470
Device Name
OBERON: LASER SURGERY FIBER, ATRAUMATIC SURGERY FIBER, SIDE FIRE FIBER, RADIAL EMISSION FIBER, DENTAL SURGERY FIBER
Applicant
Oberon GmbH Fiber Technologies
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 15, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OBERON's, laser surgery fiber, is intended for delivery of laser light to soft tissue in the contact and non-contact mode during surgical procedures including via endoscopes and cystoscopes. The OBERON laser surgery fiber is intended for use with any cleared surgical laser with a SMA 905 or compatible connectors. The OBERON, laser surgery fibers, Side Fire Laser surgery fibers and Dental Probes are intended for delivery of laser light to soft tissue in the contact and non-contact mode during surgical procedures including via endoscopes and cystoscopes. The OBERON laser surgery fibers are intended for use with any cleared surgical laser with a SMA 905 connector. The OBERON laser surgery fibers laser are indicated for use in general surgery applications for: incision, vaporization, ablation, hemostasis or coagulation of soft tissue in a contact mode. It is also indicated for use in open or closed endoscopic applications where incision, tissue dissection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors or lesions, tissue vaporization, hemostasis and or coagulation may be indicated. (Diode Laser 532nm - 2100nm) QBERON laser surgery fiber is also indicated for use in lithotripsy with a compatible laser cleared for the desired application. It is indicated for use with Argon, KTP/532, Ho;YAG, Nd;YAG, 1.44YAG and Diode Lasers (532nm - 2100nm) with peak and continuous power from 1 - 100 Watt. OBERON laser surgery fibers are indicated for use in general surgery, urology, gastroenterology, gynecology, vascular surgery, neurosurgery, plastic surgery, ENT and endovenous occlusion of the greater saphenous vein in the patient with superficial vein reflux and laser assisted lipolysis with an approved compatible laser marketed for use in the desired application.
Device Story
OBERON Laser Surgery Fiber is a fiber optic delivery system for laser light; transmits energy from cleared surgical lasers to soft tissue. Input: laser energy from external source (Argon, KTP, Ho:YAG, Nd:YAG, Diode). Output: laser light delivered to target tissue via contact or non-contact modes. Used in clinical settings (OR, endoscopy, cystoscopy) by physicians. Device features various distal tip configurations (flat, shaped, ball, round, side fire, radial) to facilitate specific surgical tasks like incision, ablation, or lithotripsy. Benefits include precise tissue interaction and versatility across multiple surgical specialties. No active electronic components; passive delivery system.
Clinical Evidence
No clinical data. Bench testing only; performance is established by comparison to legally marketed predicate devices.
Technological Characteristics
Passive fiber optic delivery system; silica glass core and cladding; protective sheath. Distal tips: flat, shaped, ball, round, side fire, radial. Connectors: SMA 905. Core diameters: 200-600 microns. Compatible with lasers 532nm-2100nm, 1-100W power. Sterile.
Indications for Use
Indicated for patients requiring soft tissue surgery (incision, vaporization, ablation, hemostasis, coagulation, dissection, resection) or lithotripsy. Applicable across general surgery, urology, gastroenterology, gynecology, vascular surgery, neurosurgery, plastic surgery, and ENT. Includes endovenous occlusion for superficial vein reflux and laser-assisted lipolysis. Used with compatible lasers (532nm-2100nm) in contact or non-contact modes.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three faces in profile, stacked on top of each other. The faces are stylized and appear to be flowing or merging together.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 15, 2014
OBERON GmbH Fiber Technologies % Dr. Armin Kaus MED-Fibers Incorporated 7404 West Detroit Street, Suite 140 Chandler, Arizona 85226
Re: K140470
Trade Name: OBERON Laser Surgery Fiber Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: September 12, 2014 Received: September 17, 2014
Dear Dr. Kaus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
> Sincerely yours, Binita S. Ashar -S 2014.10.15 15:29:03 -04'00'
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K140470
Device Name OBERON Laser Surgery Fiber
#### Indications for Use (Describe)
The OBERON, laser surgery fiber, Side Fire laser surgery fibers are intended for delivery of laser light to soft tissue in the contact and non-contact mode during surgical procedures including via endoscopes and cystoscopes. The OBERON laser surgery fiber is intended for use with any cleared surgical laser with a SMA 905 connector. The OBERON laser surgery fibers are indicated for use in general surgery applications for: incision, vaporization, ablation, hemostasis or coagulation of soft tissue in a contact mode. It is also indicated for use in open or closed endoscopic applications where incision, excision, tissue dissection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors or lesions, tissue vaporization, hemostasis and or coagulation may be indicated. (Diode Laser 532nm- 2100mm) OBERON laser surgery fiber is also indicated for use in lithotripsy with a compatible laser cleared for the desired application. It is indicated for use with Argon, KTP/532, Ho;YAG, 1.44YAG and Diode Lasers (532mm - 2100mm) with peak and continuous power from 1-100 Watt. OBERON laser surgery fibers are indicated for use in general surgery, urology, gynecology, dermatology, vascular surgery, neurosurgery, plastic surgery, ENT and endovenous occlusion of the greater saphenous vein in the patient with superficial vein reflux and laser assisted lipolysis with an approved compatible laser marketed for use in the desired application.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510 (k) SUMMARY
This 510(k) summary of safety and effectiveness is submitted in accordance with all of the requirements and follows Office of Device Evaluation guidance concerning the organization of a 510 (k) summary.
### OBERON 's Laser Surgery Fiber
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
OBERON GmbH Fiber Technologies Freiheitstrasse 120 15745 Wildau GERMANY Phone: +49 3375 21500-0 Fax: +49 3375 21500-10
### Official Correspondent:
Dr. Armin Kaus, Ph.D. 7404 W. Detroit Street, STE 140 Chandler, AZ 85226
Date Prepared: October 7, 2014
# Name of Device and Name of Sponsor
OBERON Laser Surgery Fiber
OBERON GmbH Fiber Technologies Freiheitsstrasse 120 15745 Wildau, GERMANY
#### Common or Usual Name
Nd:YAG-, Ho:YAG-, KTP- and Diode Laser -- Laser Fiber Delivery Systems
# Classification Name
Surgical Laser Accessory Regulation: 21 CFR §878.4810 Classification: II Product Code: GEX
{4}------------------------------------------------
# Predicate Devices
# 1 - Per 21 CFR 807.92 item (6) Summary with comparison values
| | OBERON GmbH<br>The fiber optic laser delivery<br>devices are intended for<br>delivery of laser light to soft<br>tissue in the contact and<br>noncontact mode during<br>surgical procedures including<br>via endoscopes and<br>cystoscopes. The laser fiber<br>is intended for use with any<br>cleared surgical laser with an<br>SMA 905 and<br>compatible connectors. | MED-Fibers, Inc.<br>K124003<br>The surgical fiber optic<br>Laser delivery devices<br>are intended for delivery<br>of laser light to soft tissue<br>in the contact and non con-<br>tact mode during surgical<br>procedures. The delivery device<br>is intended for use with any<br>surgical lasers with cleared<br>laser with a compatible SMA<br>905 and compatible connectors. | Fiberoptic Fab<br>K120810<br>The fiber optic laser delivery<br>system is intended for delivery<br>of laser light to soft tissue in<br>the contact and non<br>contact mode during surgical<br>procedures. The fiber optic<br>delivery System is intended<br>for use with any cleared<br>surgical laser with an SMA<br>905. |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | | | |
| Delivery Systems | | | |
| Fiber | Flexible Fiber | Flexible Fiber | Flexible Fiber |
| Material | Quartz Glass | Quartz Glass | Quartz Glass |
| Core Diameter<br>Proximal end | 200 - 600 micron<br>SMA 905 and other<br>compatible connectors | 200 - 1000 micron<br>SMA 905 and other<br>compatible connectors | 200 - 1000 micron<br>SMA 905, SMA<br>906 and other compatible<br>connectors |
| Distal end | flat, shaped, ball,<br>round, side fire, radial | flat, side fire and different<br>shapes | flat, shaped, side fire |
| Sterile (Yes/No) | Yes | Yes | Yes |
| Max Recommended Power | Core Size: 200, 272 μ $20 W$<br>365 μ $40 W$<br>500 μ $100 W$<br>400, 600 μ HPCS $30 W$<br>600 µIRHCN $100 W$<br>800 and 1000μ N/A | up to $45 W$<br>up to $60 W$<br>up to $100 W$<br>up to $40 W$<br>up to $100 W$<br>up to $200W$ | up to $45 W$<br>up to $113 W$<br>N/A<br>N/A<br>up to $100 W$<br>up to $200 W$ |
| Indication for use | The laser delivery devices<br>are intended for use in<br>laser based surgical<br>application to deliver<br>the laser light to soft<br>tissue during surgical<br>applications. The delivery<br>device is cleared for<br>surgical lasers with<br>compatible SMA 905<br>connectors. It is indicated<br>for use in general surgery<br>applications for incision,<br>excision, vaporization,<br>ablation and or coagulation.<br>It is indicated for use in<br>soft tissue procedures with<br>cleared surgical laser systems<br>with different wavelength.<br>It is also indicated for use<br>with cleared compatible<br>laser systems in lithotripsy<br>for the desired application. | The surgical fiber optic laser<br>delivery devices, Endo-ENT<br>fibers, side fire Fibers and Endo<br>probes in the contact and non-<br>contact mode during surgical<br>procedures including via<br>endoscopes and cysto-scopes. The<br>surgical laser<br>fiber is intended for use with any<br>cleared surgical laser with an<br>SMA 905 and compatible<br>connectors. The surgical fibers<br>are indicated for use in general<br>surgery applications for incision,<br>excision, vaporization, ablation,<br>hemostasis, or coagulation. It is<br>also indicated for use In open or<br>closed endo- scopic applications<br>where incision, excision, tissue<br>dissection, excision of ex-<br>ternal tumors and lesions<br>complete or partial resection of<br>intern. organs, It is also intended<br>as an aid for otology procedures.<br>In addition all surgical devices<br>are intended for use in lithotripsy<br>applications with cleared laser<br>Systems. | The fiber optic laser delivery<br>system is indicated for use in<br>general surgery applications for<br>incision, excision vaporization,<br>ablation, hemostasis or<br>coagulation, of soft tissue in a<br>contact or non-contact mode.<br>It is also indicated for use in<br>open or closed endoscopic<br>applications, where incision,<br>excision, tissue dissection,<br>excision of external tumors or<br>lesions, complete or partial<br>resection of internal organs,<br>tumors or lesions, tissue vapo-<br>rization, hemostasis and or<br>coagulation may be indicated.<br>It is also indicated for use in<br>lithotripsy with a compatible<br>laser cleared for the desired<br>application. Its intended to use<br>for different cleared lasers with<br>peak and continuous<br>power from 1W - 200W. |
{5}------------------------------------------------
# Intended Use / Indications of Use
The OBERON's, laser surgery fiber, is intended for delivery of laser light to soft tissue in the contact and non-contact mode during surgical procedures including via endoscopes and cystoscopes. The OBERON laser surgery fiber is intended for use with any cleared surgical laser with a SMA 905 or compatible connectors.
The OBERON, laser surgery fibers, Side Fire Laser surgery fibers and Dental Probes are intended for delivery of laser light to soft tissue in the contact and non-contact mode during surgical procedures including via endoscopes and cystoscopes. The OBERON laser surgery fibers are intended for use with any cleared surgical laser with a SMA 905 connector. The OBERON laser surgery fibers laser are indicated for use in general surgery applications for: incision, vaporization, ablation, hemostasis or coagulation of soft tissue in a contact mode. It is also indicated for use in open or closed endoscopic applications where incision, tissue dissection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors or lesions, tissue vaporization, hemostasis and or coagulation may be indicated. (Diode Laser 532nm - 2100nm) QBERON laser surgery fiber is also indicated for use in lithotripsy with a compatible laser cleared for the desired application. It is indicated for use with Argon, KTP/532, Ho;YAG, Nd;YAG, 1.44YAG and Diode Lasers (532nm - 2100nm) with peak and continuous power from 1 - 100 Watt. OBERON laser surgery fibers are indicated for use in general surgery, urology, gastroenterology, gynecology,
{6}------------------------------------------------
vascular surgery, neurosurgery, plastic surgery, ENT and endovenous occlusion of the greater saphenous vein in the patient with superficial vein reflux and laser assisted lipolysis with an approved compatible laser marketed for use in the desired application.
# Technical Characteristics
OBERON's laser surgery fiber contains the same components and the same technological characteristics as the predicate devices. The fiber core and cladding or fibers with no cladding are made from silica which is the same material used in all the predicate devices. As mentioned, the optical fiber is made out of silica with a coaxially mounted protective sheath. The fiber distal tip can be several configurations and the fiber can be also used with hand pieces.
OBERON's laser surgery fiber has no differences in technology and as such does not raise any new questions on safety or efficacy. Various core diameter sizes (200, 272, 365, 400, 550, 600microns) are offered.
# Performance Data
The performance of the OBERON's laser surgery fiber is well established and documented so no performance testing is included. The OBERON's laser surgery fiber delivery system operates in the same manner as the predicate devices and performs with no difference as compared with the predicate devices.
# Substantial Equivalence
The OBERON's laser surgery fibers are as safe and effective for the indications for use as the Laser Peripherals, Leoni, Lumenis, MED-Fibers, Cynosure, Fiberoptic Fabrications and CermaOptec fiber optic laser delivery systems previously cleared thus the OBERON's laser surgery fibers are substantially equivalent.
| Signed: | |
|------------------------|--|
| Dr. Armin Kaus, Ph.D. | |
| Official Correspondent | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.