K140458 · Diotech CO · GEX · Aug 26, 2014 · General, Plastic Surgery
Device Facts
Record ID
K140458
Device Name
DIOTECH LASER FIBER
Applicant
Diotech CO
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 26, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Diotech Laser Fiber is indicated for endovascular coagulation of the great saphenous vein in patients with superficial vein reflux, for the treatment of varicose veins and varicosities associated with superficial reflux of the great saphenous vein, and for the treatment of incompetence and reflux of superficial veins of the lower extremity. The Diotech Laser Fiber used with CW laser with wavelength range 810~1470nm, a power range 5~15W, using an SMA905 connector.
Device Story
Diotech Laser Fiber is a free-beam delivery device for endovascular laser coagulation. It consists of an optical fiber (silica core/cladding) and a catheter tube (PE, PP, polycarbonate). The device connects to a CW laser (810-1470nm, 5-15W) via an SMA905 connector. Used by physicians in clinical settings to treat lower extremity venous reflux; the fiber is inserted through the catheter into the vein. The fiber tip delivers laser energy diagonally (70-80° angle) to the vein wall to induce coagulation. Catheter markings (every 10mm) assist in depth control. The device is single-use and EO-sterilized. It benefits patients by providing a minimally invasive treatment for varicose veins.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993) and functional performance comparisons against the predicate device. No clinical data provided.
Indicated for patients with superficial vein reflux, varicose veins, varicosities associated with superficial reflux of the great saphenous vein, and incompetence/reflux of superficial veins of the lower extremity.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
AngioDynamics, Inc. NeverTouch 600 um Fiber (K071959)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of a series of overlapping human profiles, creating an abstract representation of people. The profiles are arranged in a way that suggests movement or progress.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 26, 2014
Plus Global Mr. Peter Chung 300 Atwood Street Pittsburgh, Pennsylvania 15213
Re: K140458 Trade/Device Name: Diotech Laser Fiber Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: July 15, 2014 Received: July 28, 2014
Dear Mr. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may; therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{2}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K140458
Device Name Diotech Laser Fibers
#### Indications for Use (Describe)
Diotech Laser Fibers (400um, 600um) is indicated for endovascular coagulation of the great saphenous vein in patients with superficial vein reflux, for the treatment of varicose veins and varicosties associated with superficial reflux of the great saphenous vein, and for the treatment of incompetence and reflux of superficial veins of the lower extremity. The Diotech Laser Fibers (400um, 600um) may be used with CW laser with wavelength range 810-1470mm, a power range 5-15W, using an SMA905 connector.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 1. Applicant Information
- 1) Company : Diotech Co.
- 2) Address : 301, Nakdong-daero, Saha-gu, Busan, Korea
- 3) Device : Diotech Laser Fibers
- 4) Phone Number : +82-51-292-6237
- 5) Fax Number : +82-51-292-6258
- 6) Homepage: http://www.diotech21.com
## 2. Device Information
- 1) Trade Name : Diotech Laser Fibers
- 2) Common Name : Diode Laser
- 3) Regulation Name : Laser surgical instrument for use in general and plastic surgery and in dermatology
- 4) Product code : GEX
- 5) Regulation number : 878.4810
- 6) Class of device : Class II
### 3. The legally marketed device to which we are claiming equivalence
- 1) 510(K) Number : K071959
- 2) Trade Name : AngioDynamics, Inc. NeverTouch 600 mm Fiber
- 3) Product Code : GEX
### 4. Description of device
Diotech Laser Fiber is free- beam delivery device that transits laser energy into intended direction. Fiber length is 2.9mters. And catheter tube length is 0.85meters. Catheter tube marked on per 10mm. This device can be connected with SMA905 connector and deliver to 810~1470m. This device can be connected with cleared surgical use laser diode. This device is single use only and sterilized by EO gas.
#### 5. Intended Use
Diotech Laser Fiber is indicated for endovascular cagulation of the great saphenous vein in patients with superficial vein reflux, for the treatment of varicose veins and varicosities associated with superficial reflux of the great saphenous vein, and for the treatment of incompetence and reflux of superficial veins of the lower extremity.
The Diotech Laser Fiber used with CW laser with wavelength range 810~1470nm, a power range 5~15W, using an SMA905 connector.
#### 6. Technological Characteristics
Diotech laser fiber consists of two parts that are optic fiber and catheter. The optic fiber also consists of connector, optical fiber, and the tip part. Fiber cable is made of the optical fiber that delivers the laser beam, and tip part is made of glass that can deliver the laser beam diagonally. And the fiber may be used any laser wavelength between 810~1470 m that have been cleared for surgical use diode laser. Outer diameter of fiber cable is 0.6mm and length is 2900 mm.
Catheter has a function that let the fiber cable be inserted into blood vessel as vein through its inner path. The catheter tube also consist of three parts as hub, tube and outer Inlet. Hub is used to connect with syringe. Tube is used for way that fiber cable may go through inner hole of tube. Outer Inlet is entrance hole for fiber to go through into tube. Catheter tube length is 0.85meters(850mm). It was marked on per 10mm to check depth to insert. The fiber core and cladding for the subject device are made from silica which is the same material used in the
{5}------------------------------------------------
predicate device. This device can be connected with SMA905 connector. This device is single use only and sterilized by EO gas.
## 7. Performance
Bench tests were performed. Bench testing included biocompatibility. The tests demonstrated that the device is as safe, as effective and performs in a substantially equivalent manner to the predicate device.
#### 8. Predicate device comparison table
| Trade Name | | | Diotech laser fiber | AngioDynamics, Inc. NeverTouch 600 um Fiber |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|-------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | | | This device is indicated for endovascular<br>coagulation of the great saphenous vein in<br>patients with superficial vein reflux, for the<br>treatment of varicose veins varicosities and<br>associated with superficial reflux of the great<br>saphenous vein, and for the treatment of<br>incompetence and reflux of superficial veins of<br>the lower extremity. | The AngioDynamics, Inc. NeverTouch 600gm<br>Fiber is indicated for endovascular coagulation of<br>the great saphenous vein in patients with superficial<br>vein reflux, for the treatment of varicose veins<br>varicosities and associated with superficial reflux<br>of the great saphenous vein, and for the treatment<br>of incompetence and reflux of superficial veins of<br>the lower extremity. |
| 510(k) Number | | | K140458 | K071959 |
| rediy | Design | | .1mm<br>0.6mm<br>10mm | |
| | Diameter of Optical<br>Fiber | | 0.6 mm(DRF-A600-1, DBF-A600-1)<br>0.4 mm(DBF-A400-1) | 0.6 mm |
| | Diameter of Glass Tube | | 1.1 mm | 0.905 mm |
| | Length of Glass Tube | | 10 mm | 5.8 mm |
| ទេវាទប្រពុជ | Design | | | |
| | Total Length | | 85 cm | 80 cm & 90 cm |
| | Operation Length | | 82 cm | 83 cm |
| | Hub | | 5 cm | Similar |
| | Outer Inlet | | 1.5 cm | Similar |
| | Fiber | Optic Fiber | Silica | Silica |
| sinə aid bitki növü. İstinadlar Rüspitaliya Şamaları Şahil Şamaları Şamaları Şamaları Şamaları Şamaları Şamaları Şamaları Şamaları Şamaları Şamaları Şamaları | | Connector | Stainless steel | Stainless steel & Plastic |
| | Catheter | Hub | PP | Unknown |
| | | Outer inlet | Polycarbonate | Unknown |
| | | Tube | PE | PE |
| Fiber transmission angle | | | 70~80° | 30~50° |
| Wavelength range | | | 810~1470nm | 810~1470nm |
{6}------------------------------------------------
| Maximum output power | 15W | Unknown |
|----------------------|-------------------|-------------------|
| Single use | Single use | Single use |
| Biocompatibility | ISO10993 | ISO10993 |
| Sterility | EO gas sterilized | EO gas sterilized |
## 9. Conclusion
The Device is investigated for function and effectiveness to compare the operation of function between Diotech laser fiber and predicate devices .
Comparison results demonstrate that the specifications and performance of the device are similar as functional and effective as the legally marketed predicate device.
Therefore, it is concluded that Diotech laser fiber is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.