K140249 · Advanced Technology Laser Co., Ltd. · GEX · Oct 8, 2014 · General, Plastic Surgery
Device Facts
Record ID
K140249
Device Name
LONG PULSE ND: YAG LASER SYSTEM
Applicant
Advanced Technology Laser Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 8, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Long pulse Nd:YAG Laser System is intended for use for : Benign pigmented lesions such as ,but not limited to , lemtigos( age spots), solar lentigos( sun spots), café au lait macules, seborrheic karatoses, nevi, chloasma, verrucae, skin tags, karatoses, tattoss( significant reduction in the intensity of black and/or blue black tattoos) and plaques. . Pigmented lesions to reduce lesion size, for patients with lesions that would potientlly benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part . of the scar.
Device Story
Flashlamp-excited Nd:YAG laser system; emits 1064 nm pulsed laser energy. Principle: selective photothermolysis; pulse duration controlled to be equal to or shorter than target thermal relaxation time to minimize collateral tissue heating. Used by clinicians in dermatology/plastic surgery settings. Input: user-selected parameters via interface; Output: laser energy delivered via handpiece. 1064 nm wavelength targets deeper hair follicles, vessels, and pigmented lesions. Clinical benefit: reduction of lesion size, pigmentation, and scar vascularity. System includes control unit, power source, emission/delivery system, cooling, and safety features.
Clinical Evidence
Bench testing only. Device performance verified against IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1 standards to demonstrate substantial equivalence.
Technological Characteristics
Flashlamp-excited Nd:YAG laser; 1064 nm wavelength; fluence up to 600J/cm2; spot sizes 1.5mm-12mm and 3mm*8mm; 510nm diode aiming laser (<5mW, Class III). Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-22 (laser safety), IEC 60825-1 (laser classification).
Indications for Use
Indicated for patients with benign pigmented lesions (e.g., lentigos, café au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, tattoos), pigmented lesions requiring aggressive treatment or unresponsive to other lasers, and hypertrophic or keloid scars with vascular components.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 8, 2014
Advanced Technology Laser Company, Ltd % Ms. Diana Hong Mid-Link Consulting Company, Ltd P.O. Box 120-119 Shanghai, 200120 CHINA
Re: K140249
Trade/Device Name: Long Pulse Nd:YAG Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: September 5, 2014 Received: September 8, 2014
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Binita S. Ashar -S 2014.10.08 12:44:48 -04'00' Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K140249
Device Name
Long Pulse Nd: Y AG Laser System
#### Indications for Use (Describe)
Long pulse Nd: Y AG Laser System is intended for use for :
· Benign pigmented lesions such as but not limited to , lemtigos( age spots), solar lentigos( sun spots), café au lait macules, sebortheic karatoses, nevi, chloasma, verucae, skin tags, karatoses, tattoss( significant reduction in the intensity of black and/or blue black tattoos) and plaques.
• Pigmented lesions to reduce lesion size, for patients with lesions that would potiently benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
· Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
#### Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Section 3 510(k) Summary
This 510(k) Summary of 510(k) information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K140249
- 1. Date of Submission: 09/10/2014
- Sponsor Identification 2.
Advanced Technology Laser Co., Ltd. 920 Jian-chuan Road, Bldg. A2, Level 5,Shanghai, 200240, China
Establishment Registration Number: 3007604279;
Contact Person: Mingxia Xi Position: Regulator Affair Director Tel: +86-21-5471 2151 Fax: +86-21-5471 2152 Email: xmx@at-laser.com
- 3. Submission Correspondent
Ms. Diana Hong& Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net
{4}------------------------------------------------
- Proposed Device Identification 4.
Trade Name: Long Pulse Nd: YAG Laser System; Common Name: Nd:YAG Dermatology Laser System; Model: SmoothTouch;
Regulatory Information: Classification Name: Powered Laser Surgical Instrument; Classification: II; Product Code: GEX; Regulation Number: 21CFR 878.4810; Review Panel: Laser Instrument, Surgical, Powered;
Intended Use Statement:
Long pulse Nd:YAG Laser System is intended for use for :
- Benign pigmented lesions such as ,but not limited to , lemtigos( age spots), solar lentigos( sun spots), café au lait macules, seborrheic karatoses, nevi, chloasma, verrucae, skin tags, karatoses, tattoss( significant reduction in the intensity of black and/or blue black tattoos) and plaques.
- . Pigmented lesions to reduce lesion size, for patients with lesions that would potientlly benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
- Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part . of the scar.
- Predicate Device Identification ર.
510(k) Number: K022923 Product Name: Gentle YAG Laser System Manufacturer: Candela Corporation
- 6. Device Description
The Long Pulse Nd:YAG Laser System is a flashlamp-excited, Nd:YAG (Neodymium-doped Yttrium Aluminum Garnet) laser system. Pulsed laser energy at a nominal wavelength of 10641m. This wavelength causes maximum energy absorption by targeting the treatment area and minimum absorption by surrounding skin structures. In addition, the laser pulse duration is controlled to be equal to or shorter than the thermal relaxation time of the target, to minimize heat transfer to surrounding tissues.
{5}------------------------------------------------
Deeper penetration and a more moderate hemoglobin absorption makes 1064 nm wavelength more useful for the deeper hair follicles and vessels. Further, with the right combination of parameters, the 1064 nm wavelength is also suitable for more superficial hair, telangiectasia and spider veins.
Base on this, Long Pulse Nd: Y AG Laser System is intended for 1) Benign pigmented lesions such as ,but not limited to , lemtigos (age spots), solar lentigos (sun spots), café au lait macules, seborrheic karatoses, nevi, chloasma, verrucae, skin tags, karatoses, tattoss (significant reduction in the intensity of black and/or blue black tattoos) and plaques. 2) Pigmented lesion size, for patients with lesions that would potiently benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. 3) Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
The Nd: Y AG Laser System consists of control system, user interface, power source , laser emission and delivery system, cooling system and safety features
- 7. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
IEC 60601-1:2005, Medical Electrical Equipment - Part 1: General requirements for safety.
IEC60601-1-2:2007, Medical Electrical Equipment - Part 1: General requirements for safety - 2, Collateral Standard: Electromagnetic compatibility - Requirements and tests.
IEC 60601-2-22:2007, Medical Electrical Equipment - Part 2: Particular requirements for the safety of diagnostic and therapeutic laser equipment.
IEC 60825-1: 2007, Safety of laser products - Part 1: Equipment classification, requirements
{6}------------------------------------------------
Section 3 510 Summary
#### 8. Substantially Equivalent (SE) Conclusion
The following table compares the DEVICE to the predicate device with respect to intended use, technological characteristics and principles of operation, etc.
| Item | Proposed Device(s) | Predicate Device |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Product Code | GEX | GEX |
| Regulation Number | 21CFR 878.4810 | 21CFR 878.4810 |
| Intended Use | Long pulse Nd:YAG Laser System is intended for use for :<br>Benign pigmented lesions such as ,but not limited to , lemtigos( age spots), solar lentigos( sun spots), café au lait macules, seborrheic karatoses, nevi, chloasma, verrucae, skin tags, karatoses, tattoss( significant reduction in the intensity of black and/or blue black tattoos) and plaques.<br>Pigmented lesions to reduce lesion size, for patients with lesions that would potientlly benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.<br>Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. | SAME |
| Energy Source | ND:YAG Laser | SAME |
| Waveform | 1064 nm | SAME |
| Fluence energy range | up to 600J/cm2 | SAME |
| Spot size | 1.5mm, 3mm, 8mm, 10mm, 12mm and 3mm*8mm | SIMILAR |
| Aiming laser | 510nm diode laser (<5mW), Class III | SIMILAR |
| Beam Delivery | Permanently attached umbilical cable and handpieces | SAME |
Table 3-1 Comparison of Technology Characteristics
The proposed device, Long Pulse Nd:YAG Laser System, is determined to be Substantially Equivalent (SE) to the predicate device, Gentle YAG Laser System, in respect of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.