K140005 · Alma Lasers , Ltd. · GEX · Jul 18, 2014 · General, Plastic Surgery
Device Facts
Record ID
K140005
Device Name
ALMA DIODE TABLETOP LASER
Applicant
Alma Lasers , Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 18, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Alma Diode Tabletop Laser is intended for use in dermatologic and general surgical procedures.
Device Story
The Alma Diode Tabletop Laser is a surgical laser system consisting of a main console, footswitch, and accessories. It utilizes diode laser modules at 810nm, 980nm, or 1470nm wavelengths to deliver laser energy to soft tissue via optical fibers. The device operates in continuous wave, single pulse, or pulsed modes. It is intended for use by clinicians in dermatologic and general surgical settings. The operator controls the device via an LCD touch screen interface. The laser energy is used for endovenous occlusion of the greater saphenous vein and laser-assisted lipolysis. By delivering controlled thermal energy to target tissues, the device facilitates surgical procedures, potentially benefiting patients through minimally invasive vein treatment and lipolysis.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and laser safety (IEC 60601-2-22, IEC 60825-1).
Technological Characteristics
Diode laser source; wavelengths 810nm, 980nm, 1470nm; max power 15W-30W; optical fiber delivery; continuous/pulsed operation; LCD touch screen interface; 635nm aiming beam; 100-240V AC power; compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1.
Indications for Use
Indicated for endovenous laser surgery for saphenous incompetent veins (810nm module); endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux and laser assisted lipolysis (980nm and 1470nm modules).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
# JUL 1 8 2014
## K140005
### Section 8 - 510(k) Summary or 510(k) Statement
#### l, General Information
Submitter:
Alma Lasers, Ltd, Halamish St. POB 3021 Caesarea Industrial Park, Caesarea, Israel 38900
Contact Person:
Kathy Maynor Consultant 352-586-3113 (cell)
Summary Preparation Date:
Dec 30, 2013
- II. Names
| <b>Device Names:</b> | The Alma Diode Tabletop Laser |
|----------------------|-------------------------------|
|----------------------|-------------------------------|
Surgical Powered Light Instrument, Primary Classification Names:
### III. Predicate Devices
. Predicate Device ___________________________________________________________________________________________________________________________________________________________ K # ** K100058 Quanta Diode Laser Family
### IV. Product Description
The Alma Diode Tabletop Laser is comprised of the following major components:
- 1. The main console unit
- 2. Footswitch.
- 3. Accessories
### V. Indications for Use
The Alma diode tabletop laser is intended for use in dermatologic and general surgical procedures.
{1}------------------------------------------------
### VI. Summary of Technical Characteristics
.
| | K13 | K100558 |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Alma diode tabletop laser | Quanta Diode Laser Family |
| Parameter | | |
| Product Code &<br>Regulation No. | GEX<br>21 CFR 878.4810 | GEX<br>21 CFR 878.4810 |
| Laser Wavelength [nm] | 1470 | 1470 |
| Max power | 15W | 15W |
| Light/Laser Source | Diode | Diode |
| Laser Delivery | Optical Fiber | Optical Fiber |
| Operation Mode | Continuous wave, single pulse, pulsed | Continuous wave, single pulse, pulsed |
| Pulse Duration | 10-990ms | 3ms - 2.5s |
| Bare fiber size | 200, 300, 320, 400, 600, 800, 1000 | 200, 300, 320, 400, 600, 800, 1000 |
| User Interface | LCD touch screen | LCD touch screen |
| Aiming beam | 635nm | 650nm |
| Electrical Requirements | 100-240. V AC 50-60 Hz, 6.3 A | 100-240, V AC 50-60 Hz, 6.3 A, single phase, |
| Indications for Use | The Alma 1470nm diode tabletop laser is indicated for use in endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The Alma 1470nm diode tabletop laser is further indicated for laser assisted lipolysis, | The QUANTA Diode Laser System is intended for delivery of laser light to soft tissue in the contact and non contact mode during surgical procedures including via endoscopes. The QUANTA Diode Laser System is generally indicated for use in endovenous occlusion of the greater saphenous vein in Patients with Superficial Vein Reflux. The QUANTA1470 Diode Laser is further indicated for laser assisted lipolysis |
.
.
.
Table 1: Salient Characteristics of the 1470nm module and the Predicate Devices
·
:
:
2/5
י
:
{2}------------------------------------------------
# Table 2: Salient Characteristics of the 980nm module and the Predicate Devices
・
| | K13 | K100558 |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Alma diode tabletop laser | Quanta Diode Laser Family |
| Parameter | | |
| Product<br>Code &<br>Regulation<br>No. | GEX<br>21 CFR 878.4810 | GEX<br>21 CFR 878.4810 |
| Laser<br>Wavelength<br>[nm] | 980nm | 980nm |
| Max power | 30W | 30W |
| Light/Laser<br>Source | Diode | Diode |
| Laser<br>Delivery | Optical Fiber | Optical Fiber |
| Operation<br>Mode | Continuous wave, single<br>pulse, pulsed | Continuous wave, single pulse,<br>pulsed |
| Pulse<br>Duration | 10-990ms | 3ms - 2.5s |
| Bare fiber<br>size | 200, 300, 320, 400, 600, 800,<br>1000 | 200, 300, 320, 400, 600, 800,<br>1000 |
| User<br>Interface | LCD touch screen | LCD touch screen |
| Aiming beam | 635nm | 650nm |
| Electrical<br>Requirements | 100-240, V AC 50-60 Hz, 6.3<br>A, | 100-240, V AC 50-60 Hz, 6.3 A.<br>single phase. |
| Indications<br>for Use | The Alma 980nm diode<br>tabletop laser is indicated<br>for use in endovenous<br>occlusion of the greater<br>saphenous vein in patients<br>with superficial vein<br>reflux. The Alma 980nm<br>diode tabletop laser is<br>further indicated for laser<br>assisted lipolysis, | The QUANTA Diode Laser<br>System is intended for delivery of<br>laser light to soft tissue in the<br>contact and non contact mode<br>during surgical procedures<br>including<br>via endoscopes. The QUANTA<br>Diode Laser System is generally<br>indicated for use in endovenous<br>occlusion of the greater<br>saphenous vein in Patients with<br>Superficial Vein Reflux. The<br>QUANTA980 Diode<br>Laser is further indicated for laser<br>assisted lipolysis. (This is a<br>subset of the cleared indications<br>for this product) |
·
・
.
: 3/5
·
{3}------------------------------------------------
# Table 3: Salient Characteristics of 810nm diode module and the predicate devices
·
વે
| | K13 | K100558 |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Alma diode tabletop laser | Quanta Diode Laser Family |
| Parameter | | |
| Product<br>Code &<br>Regulation<br>No. | GEX<br>21 CFR 878.4810 | GEX<br>21 CFR 878.4810 |
| Laser<br>Wavelength<br>[nm] | 810nm | 808nm |
| Max power | 30W | 30W |
| Light/Laser<br>Source | Diode | Diode |
| Laser<br>Delivery | Optical Fiber | Optical Fiber |
| Operation<br>Mode | Continuous wave, single<br>pulse, pulsed | Continuous wave, single pulse,<br>pulsed |
| Pulse<br>Duration | 10-990ms | 3ms - 2.5s |
| Bare fiber<br>size | 200, 300, 320, 400, 600, 800,<br>1000 | 200, 300, 320, 400, 600, 800,<br>1000 |
| User<br>Interface | LCD touch screen | LCD touch screen |
| Aiming beam | 635nm | 650nm |
| Electrical<br>Requirements | 100-240, V AC 50-60 Hz, 6.3<br>A. | 100-240, V AC 50-60 Hz, 6.3 A,<br>single phase, |
| Indications<br>for Use | The Alma 810nm diode<br>tabletop laser is indicated<br>for endoluminal or<br>endovenous laser surgery<br>for saphenous incompetent<br>veins. | The QUANTA Diode Laser<br>System is intended for delivery of<br>laser light to soft tissue in the<br>contact and non contact mode<br>during surgical procedures<br>including<br>via endoscopes. The QUANTA<br>808 Diode Laser System is<br>generally indicated for<br>endoluminal or endovenous laser<br>surgery for saphenous<br>incompetent veins. (This is a<br>subset of the cleared indications<br>for this product). |
·
.
4 / 5
{4}------------------------------------------------
#### VII. Safety and Effectiveness Information
The review of the indications for use and technical characteristics provided demonstrates that the Alma diode tabletop laser is substantially equivalent to the predicate devices.
The Diode Tabletop Laser was tested by a certified laboratory according to:
IEC 60601-1: 1988+A1:1991+A2:1995: Medical Electrical Equipment Part 1: General Requirements for basic safety and essential performance
IEC 60601-1-2: Medical electrical equipment: Part 1-2: General requirements for safety-Collateral Standard: Electromagnetic compatibility - Requirements and tests (2001 + A1(4))
IEC 60601-2-22:1995 Medical electrical equipment- Part 2-22: Particular requirements for the safety of diagnostic and therapeutic laser equipment 1995.
IEC 60825-1:2007 (2nd edition) Safety of Laser Products-Part 1: Equipment Classification and Requirements
### Conclusion VIII.
The Alma diode tabletop laser was found to be substantially equivalent to the predicate devices.
The Alma diode tabletop laser shares the same or similar indications for use, similar design features, and functional features with, and thus is substantially equivalent to the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings or feathers. The eagle is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 18, 2014
Alma Lasers Ltd. % Ms. Kathy Maynor Regulatory Consultant for Alma LTD 26 Rebecca Court Homosassa. Florida 34446
Re: K140005
Trade/Device Name: The Alma Diode Tabletop Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: April 29, 2014 Received: June 20, 2014
Dear Ms. Maynor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I DA has intatutes and regulations administered by other Federal agencies. You must or any reath all the Act s requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{6}------------------------------------------------
Page 2 - Ms. Kathy Maynor
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
| 510(k) Number (if known) | K140005 |
|--------------------------|---------|
|--------------------------|---------|
Device Name The Alma Diode Tabletop Laser
Indications for Use (Describe) Intended Use The Alma Diode Tabletop Laser is intended for use in dermatologic and general surgical procedures.
Indications for Use The Alma Diode Tabletop Laser includes three possible diode laser modules depending on the customer order.
Diode Laser Modules:
The indications for use for the 810nm Alma Diode Tabletop Laser include:
- The Alma 810nm diode tabletop laser is indicated for endovenous laser surgery for saphenous incompetent veins.
The indications for use for the 980 nm Alma Diode Tabletop Laser include:
-The Alma 980nm diode tabletop laser is indicated for use in endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The Alma 980nm diode tabletop laser is further indicated for laser assisted lipolysis.
The indications for use for the 1470nm Alma Diode Tabletop Laser include:
-The Alma 1470nm diode tabletop laser is indicated for use in endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux. The Alma 1470nm diode tabletop laser is further indicated for laser assisted lipolysis.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Neil R Ogden -S 2014.07.17 16:13:51 -04′00″
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.