K133788 · Cook Incorporated · GEX · Mar 18, 2015 · General, Plastic Surgery
Device Facts
Record ID
K133788
Device Name
COOK HOLMIUM LASER FIBER
Applicant
Cook Incorporated
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 18, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Used for incision/excision, ablation, and coagulation (hemostasis) when attached to a cleared Ho:Y AG laser system comprised of any standard SMA-type comector and an output power of up to 100 watts, for the indications for which the system has been cleared.
Device Story
Cook Holmium Laser Fiber is a surgical accessory for Ho:YAG laser systems; transmits laser energy for incision, excision, ablation, and coagulation. Device consists of optical fibers with SMA-type connectors; available in multi-use (color-coded) and single-use (red connector) configurations. Used in clinical settings by physicians; fiber attaches to cleared laser systems to deliver energy to target tissue. Benefits include precise tissue interaction for surgical procedures. Performance verified through tensile, bend radius, and energy transmission testing; biocompatibility confirmed per ISO 10993-1. Sterilization validated for steam and Sterrad cycles.
Clinical Evidence
Bench testing only. Testing included tensile strength, bend radius, energy transmission (max 20% loss), accelerated aging (3-year equivalent), biocompatibility (ISO 10993-1), laser system compatibility, and sterilization/cleaning validation.
Technological Characteristics
Optical fiber for Ho:YAG laser energy delivery; SMA-type connector; multi-use (color-coded) and single-use (red connector) variants; compatible with laser systems up to 100W; biocompatible materials per ISO 10993-1; steam and Sterrad sterilization compatible.
Indications for Use
Indicated for incision, excision, ablation, and coagulation (hemostasis) when used with cleared Ho:YAG laser systems (up to 100W) for the indications for which the host laser system is cleared.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 18, 2015
Cook Incorporated Mr. David E. Chadwick 750 Daniels Way P.O. Box 489 Bloomington, Indiana 47402
Re: K133788
Trade/Device Name: Cook Holmium Laser Fiber Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Instrument, Surgical, Powered Regulatory Class: Class II Product Code: GEX Dated: March 2, 2015 Received: March 3, 2015
Dear Mr. Chadwick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K133788
Device Name Cook Holmium Laser Fiber
#### Indications for Use (Describe)
Used for incision/excision, ablation, and coagulation (hemostasis) when attached to a cleared Ho:Y AG laser system comprised of any standard SMA-type comector and an output power of up to 100 watts, for the indications for which the system has been cleared.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, with a registered trademark symbol next to it. Below the word "COOK" is the word "MEDICAL" in white. The background of the logo is a red color.
#### 5. 510(k) Summary
COOK INCORPORA 50 DANIELS WAY, P.O. MINGTON IN 47 W.COOKM
# Cook Incorporated Cook Holmium Laser Fiber 510(k) Summary 21 CFR 807.92
17 March 2015
Date Prepared:
#### Submitted By:
Applicant: Address:
Phone Number: Fax Number:
Contact: Contact Address:
Contact Phone Number: Contact Fax Number:
Cook Incorporated 750 Daniels Way P.O. Box 489 Bloomington, IN 47402
1 (800) 468-1379 (812) 332-0281
David E. Chadwick Cook Incorporated 750 Daniels Way P.O. Box 489 Bloomington, IN 47402
800-346-2686 or 812-339-2235 812-332-0281
#### Device Information:
Trade name: Common name: Classification: Regulation: Product Code:
Cook Holmium Laser Fiber Laser Instrument, Surgical, Powered Class II 21 CFR §878.4810 GEX
## Predicate Device:
Cook Holmium Laser Fiber
K124030, July 1, 2013
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image features the logo for Cook Medical. The logo consists of the word "COOK" in large, white, sans-serif font, with a registered trademark symbol to the right of the word. Below "COOK" is the word "MEDICAL" in a smaller, white, sans-serif font. The background is a solid red color.
## Indications for Use:
The Cook Holmium Laser Fiber is intended for incision/excision, ablation, and coagulation (hemostasis) when attached to a cleared Ho: Y AG laser system comprised of any standard SMA-type connector and an output power of up to 100 Watts, for the indications for which the system has been cleared.
## Device Description:
The Cook Holmium Laser Fibers are supplied sterile in peel-open packages. The multiuse fibers have a color coded connector to identify the four different fiber sizes, and will be sold individually. The single-use fibers have a red connector and come in sizes, and will be sold in boxes of three.
## Comparison to Predicate Device:
The proposed devices are substantially equivalent to the predicate in terms of intended use, duration of use, principles of operation, and technological characteristics.
#### Discussion of Tests and Test Results:
The device was subjected to the following tests to assure reliable design and performance under the specified testing parameters. These tests were comprised of:
- 1. Tensile Testing Testing shows the tensile force during proper clinical use should not fracture the materials and/or bonds.
- 2. Bend Radius Testing Testing shows that the fibers met the minimum bend radius requirements.
- 3. Energy Transmission Testing Testing shows that fibers transmit laser energy with not more than 20% loss from laser output to fiber output.
- 4. Accelerated Aged Testing Testing shows that devices accelerated-aged to the equivalent of three years meet performance requirements for tensile, bend radius, and energy transmission testing.
- 5. Biocompatibility Testing Testing, in conformance with ISO 10993-1, shows the device is biocompatible.
- 6. Laser System Compatibility Testing Testing shows that the Cook Holmium Laser Fiber is compatible with other cleared Ho: Y AG laser systems.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Cook Medical logo. The logo is white text on a red background. The word "COOK" is in a large, bold font, and the word "MEDICAL" is in a smaller font below it.
- 7. Steam Sterilization Validation Testing shows that the Cook Holmium Laser Fiber is compatible with recommended autoclave cycles for sterility and functionality.
- 8. Sterrad Sterilization Validation Testing shows that the fiber is compatible with recommended Sterrad cycles for sterility and functionality.
- 9. Cleaning Validation Testing shows that the fiber is compatible with recommended cleaning procedures.
# Conclusions Drawn from the Tests:
The results of these tests provide reasonable assurance that the device has been designed and tested to assure conformance to the requirements for its intended use.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.