K133283 · Candela Corp. · GEX · Dec 26, 2013 · General, Plastic Surgery
Device Facts
Record ID
K133283
Device Name
CANDELA GENTLEMAX FAMILY OF LASER SYSTEMS
Applicant
Candela Corp.
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 26, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The GentleMAX Family of Laser Systems is indicated for the following: 755nm The GentleMAX Family of Laser Systems is indicated for temporary hair reduction. Stable long-term or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. Treatment of benign pigmented lesions. Treatment of wrinkles. The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias. 1064nm The GentleMAX Family of Laser Systems is indicated for the removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to. lentigos (age spots), solar lentigos (sun spots), café au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques. The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of wrinkles.
Device Story
Dermatology laser system; dual-head (Alexandrite 755nm and Nd:YAG 1064nm); optical fiber delivery; fingerswitch/footswitch controlled. Inputs: user-selected laser parameters. Operation: high-intensity light penetrates skin; therapeutic heat delivery; Dynamic Cooling Device (DCD) provides cryogen spray burst for epidermal protection. Output: laser pulses for hair reduction, vascular/pigmented lesion treatment, and soft tissue coagulation. Used in clinical settings by trained professionals. Output allows clinicians to perform dermatological procedures; benefits include hair reduction and lesion clearance.
Clinical Evidence
No clinical data required for these modifications. Substantial equivalence supported by non-clinical performance testing including electromagnetic compatibility, electrical safety, and software verification and validation.
Technological Characteristics
Flashlamp-excited, solid-state Alexandrite (755nm) and Nd:YAG (1064nm) laser. Optical fiber delivery system with replaceable handpieces. Spot sizes: 1.5-18mm. Pulse duration: 0.25-100ms. Max fluence: 53 J/cm² (Alex), 80 J/cm² (YAG). Dynamic Cooling Device (DCD) uses cryogen spray. Standalone system.
Indications for Use
Indicated for hair reduction (temporary/permanent), treatment of benign pigmented lesions, wrinkles, and dermatological vascular lesions (e.g., port-wine stains, hemangiomas, telangiectasias, PFB, rosacea, leg/spider veins). Applicable to all skin types (Fitzpatrick I-VI) including tanned skin. Also indicated for soft tissue coagulation/hemostasis and reduction of red pigmentation in hypertrophic/keloid scars.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Section 7 - 510(k) Summary
## 510(k) Summary GentleMAX Family of Laser Systems
### General Information:
This Special 510(k) is to provide notification of substantial equivalence for the modified GentleMAX Family of Laser Systems (GentleMax Pro Laser System) to the previously cleared GentleMAX Family of Laser Systems manufactured by Candela Corporation.
| Submitted by: | Candela Corporation<br>530 Boston Post Road<br>Wayland, MA 01778-1886 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Sam Wade,<br>Global Vice President, Regulatory Affairs<br>Tel: 508-358-7400 x330<br>Fax: 508-358-5602 |
| Date prepared: | October 23, 2013 |
| Trade Name: | GentleMAX Family of Laser Systems |
| Common Name: | Dermatology Laser System |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery an<br>in dermatology (21 CFR 878.4810, Product Code GEX) |
| Predicate Devices: | GentleMAX Family of Laser Systems (K112715) |
## Intended Use / Indications for Use:
The GentleMAX Family of Laser Systems is indicated for the following:
#### 755nm
The GentleMAX Family of Laser Systems is indicated for temporary hair reduction. Stable long-term or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
Treatment of benign pigmented lesions.
Treatment of wrinkles.
The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias.
## GentleMAX Family of Laser Systems
7-1
{1}------------------------------------------------
### Section 7 - 510(k) Summary
#### 1064nm
The GentleMAX Family of Laser Systems is indicated for the removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to. lentigos (age spots), solar lentigos (sun spots), café au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.
The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
Treatment of wrinkles.
#### Description:
The Candela GentleMAX Family of Laser Systems contains two separate laser heads (Alexandrite and Nd: YAG), which produce laser light outputs of 755 nm and 1064 nm, respectively. The output of each laser head is optically combined on the laser rail, so that their beam paths are identical as they exit the laser system. This allows the use of a single delivery system which can output either 755 nm or 1064 nm wavelengths. The laser system creates a beam of high intensity light that penetrates deep into the skin tissue where it delivers a controlled amount of therapeutic heat. The Dynamic Cooling Device (DCD) protects the upper layers of the skin with a cooling burst of cryogen.
#### Technological Characteristics:
The Candela GentleMAX Family of Laser Systems delivers laser energy through an optical fiber handpiece delivery system, which can output either 755 nm or 1064 nm wavelengths. The output of this laser is delivered to the area of treatment by means of a lens coupled user replaceable optical fiber with a treatment handpiece attached to its distal end. A trigger switch (fingerswitch or footswitch) is used to control the delivery of laser pulses. The Dynamic Cooling Device provides a short burst of cryogen spray prior to firing the laser pulse. The laser output energy is delivered via an optical fiber to a handpiece, which produces circular beams with diameters of 1.5, 3, 6, 8, 10, 12, 15 and 18 millimeters on the skin. The cryogen, which is housed within the laser enclosure, is delivered via a hose to a nozzle located in the handpiece.
{2}------------------------------------------------
## Section 7 – 510(k) Summary
| Product | Modified GentleMAX Family of<br>Laser Systems (GentleMax Pro<br>Laser System) | GentleMAX Family of Laser<br>Systems | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| 510(k) | KXXXXXXX | K112715 | |
| Manufacturer | Candela Corporation | Candela Corporation | |
| Product Code | GEX | GEX | |
| Regulation<br>Number. | 21 CFR 878.4810 | 21 CFR 878.4810 | |
| Device Class | II | II | |
| Indications for<br>Use | 755nm | 755nm | |
| | The GentleMAX Family of Laser<br>Systems is indicated for temporary<br>hair reduction. Stable long-term or<br>permanent reduction through<br>selective targeting of melanin in<br>hair follicles. Permanent hair<br>reduction is defined as long-term<br>stable reduction in the number of<br>hairs regrowing after a treatment<br>regime. On all skin types<br>(Fitzpatrick I- VI) including tanned<br>skin. Permanent hair reduction is<br>defined as the long-term, stable<br>reduction in the number of hairs<br>regrowing when measured at 6, 9,<br>and 12 months after the completion<br>of a treatment regime.<br>Treatment of benign pigmented<br>lesions.<br>Treatment of wrinkles.<br>The photocoagulation of<br>dermatological vascular lesions<br>(such as port-wine stains,<br>hemangiomas, telangiectasias.<br><br>1064nm<br>The GentleMAX Family of Laser<br>Systems is indicated for the<br>removal of unwanted hair, for<br>stable long term or permanent hair<br>reduction and for treatment of PFB.<br>Permanent hair reduction is defined<br>as the long-term, stable reduction in<br>the number of hairs regrowing<br>when measured at 6 9 and 12 | The GentleMAX Family of Laser<br>Systems is indicated for temporary<br>hair reduction. Stable long-term or<br>permanent reduction through<br>selective targeting of melanin in<br>hair follicles. Permanent hair<br>reduction is defined as long-term<br>stable reduction in the number of<br>hairs regrowing after a treatment<br>regime. On all skin types<br>(Fitzpatrick I- VI) including tanned<br>skin. Permanent hair reduction is<br>defined as the long-term, stable<br>reduction in the number of hairs<br>regrowing when measured at 6, 9,<br>and 12 months after the completion<br>of a treatment regime.<br>Treatment of benign pigmented<br>lesions.<br>Treatment of wrinkles.<br>The photocoagulation of<br>dermatological vascular lesions<br>(such as port-wine stains,<br>hemangiomas, telangiectasias.<br><br>1064nm<br>The GentleMAX Family of Laser<br>Systems is indicated for the<br>removal of unwanted hair, for<br>stable long term or permanent hair<br>reduction and for treatment of PFB.<br>Permanent hair reduction is defined<br>as the long-term, stable reduction in<br>the number of hairs regrowing<br>whan measured at 6 9 and 12 | |
| months after the completion of a<br>treatment regime. The lasers are<br>indicated on all skin types<br>Fitzpatrick I-VI including tanned<br>skin. Photocoagulation and<br>hemostasis of pigmented and<br>vascular lesions such as but not<br>limited to port wine stains,<br>hemangioma, warts, telangiectasia,<br>rosacea, venus lake, leg veins and<br>spider veins. Coagulation and<br>hemostasis of soft tissue. Benign<br>pigmented lesions such as, but not<br>limited to, lentigos (age spots),<br>solar lentigos (sun spots), café au<br>lait macules, seborrheic keratosis,<br>nevi, chloasma, verrucae, skin tags,<br>keratosis, tattoos (significant<br>reduction in the intensity of black<br>and/or blue-black tattoos) and<br>plaques.<br>The laser is indicated for pigmented<br>lesions to reduce lesion size, for<br>patients with lesions that would<br>potentially benefit from aggressive<br>treatment, and for patients with<br>lesions that have not responded to<br>other laser treatments.<br>The laser is also indicated for the<br>reduction of red pigmentation in<br>hypertrophic and keloid scars<br>where vascularity is an integral part<br>of the scar.<br>Treatment of wrinkles. | | months after the completion of a<br>treatment regime. The lasers are<br>indicated on all skin types<br>Fitzpatrick I-VI including tanned<br>skin. Photocoagulation and<br>hemostasis of pigmented and<br>vascular lesions such as but not<br>limited to port wine stains,<br>hemangioma, warts, telangiectasia,<br>rosacea, venus lake, leg veins and<br>spider veins. Coagulation and<br>hemostasis of soft tissue. Benign<br>pigmented lesions such as, but not<br>limited to, lentigos (age spots),<br>solar lentigos (sun spots), café au<br>lait macules, seborrheic keratosis,<br>nevi, chloasma, verrucae, skin tags,<br>keratosis, tattoos (significant<br>reduction in the intensity of black<br>and/or blue-black tattoos) and<br>plaques.<br>The laser is indicated for pigmented<br>lesions to reduce lesion size, for<br>patients with lesions that would<br>potentially benefit from aggressive<br>treatment, and for patients with<br>lesions that have not responded to<br>other laser treatments.<br>The laser is also indicated for the<br>reduction of red pigmentation in<br>hypertrophic and keloid scars<br>where vascularity is an integral part<br>of the scar.<br>Treatment of wrinkles. | |
| Laser type | Flashlamp-excited, Solid state<br>Alexandrite and Nd:YAG laser | Flashlamp-excited, Solid state<br>Alexandrite and Nd:YAG laser | |
| Wavelength | 755nm/1064nm | 755nm/1064nm | |
| Pulse duration | 0.25 - 100 ms | 0.25 - 300 ms | |
| Maximum<br>fluence | 53 J/cm² (ALEX)<br>80 J/cm² (YAG) | 53 J/cm² (ALEX)<br>80 J/cm² (YAG) | |
| Spot size | 1.5, 3, 6, 8, 10, 12, 15, 18mm | 1.5, 3, 6, 8, 10, 12, 15, 18mm | |
| Pulse repetition<br>rate | 10 Hz, maximum | 10 Hz, maximum | |
| Pulsing control | Fingerswitch or footswitch | Fingerswitch or footswitch | |
| Product<br>dimensions | 42" x 18" x 27" | 45" x 26.5" x 38" | |
| (HxWxL) | | | |
| Product<br>Weight | 260 lbs | 341 lbs | |
GentleMAX Family of Laser Systems
7-3
{3}------------------------------------------------
{4}------------------------------------------------
Section 7 - 510(k) Summary
#### Performance Data:
Minor software and hardware modifications were made to support the changes for the modified GentleMAX Family of Laser Systems (GentleMax Pro Laser System). The testing performed to evaluate these modifications included electromagnetic compatibility (EMC), electrical safety, and software verification and validation. All performance testing demonstrated that the GentleMax Pro Laser System performs according to specifications and functions as intended.
## Clinical Data:
No clinical data was required for these modifications.
### Summary of Substantial Equivalence:
The modifications made to the GentleMAX Family of Laser Systems do not affect the indications for use or alter the fundamental scientific technology of the device, nor does it affect the mode of use. There are no labeling changes that affect the indications for use of the device. The modifications made to the GentleMAX Family of Laser Systems raises no new issues of safety or effectiveness. The modified GentleMAX Family of Laser Systems have the samc intended uses, utilizes similar operating principles, and matches key design aspects, including similar spot size, the same wavelength and the same maximum delivered fluence as the predicate devices. The non-clinical tests that were performed that our device is as safe, as effective, and performs as well as the predicate device. On the basis of similarities in methods of assembly, method of operation, and intended uses, as confirmed by the performance testing, Candela Corporation believes that the modified GentleMAX Family of Laser Systems is substantially equivalent to the predicate device.
#### GentleMAX Family of Laser Systems
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Cemer - WO66-G609 Silver Spring, MD 20993-0002
December 26, 2013
Candela Corporation Mr. Sam Wade Global Vice President, Regulatory Affairs 530 Boston Post Road Wayland, Massachusetts 01778
Re: K133283
Trade/Device Name: Gentlemax Family of Laser Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: December 6, 2013 Received: December 9, 2013
Dear Mr. Wade:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{6}------------------------------------------------
#### Page 2 - Mr. Sam Wade
ﻪ ﻣﺮﺩﺍﺩ ﺍ
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.