K133158 · Beijing Toplaser Technology Co., Ltd. · GEX · Feb 28, 2014 · General, Plastic Surgery
Device Facts
Record ID
K133158
Device Name
ND: YAG LASER SYSTEM
Applicant
Beijing Toplaser Technology Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Feb 28, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Tattoo Removal Treatment of Vascular Lesions Treatment of Pigmented Lesions, Incision, Excision, Ablation, Vaporization of Soft Tissue for General Dermatology
Device Story
Qmetrx 1000 (Victory 11) is a Q-switched Nd:YAG laser system. Inputs: electrical power; user-selected parameters (wavelength, energy, spot size, fluence). Operation: flashlamp-pumped Nd:YAG crystal generates 1064 nm beam; frequency-doubling non-linear crystal generates 532 nm beam. Red diode aiming beam reflected via beamsplitter. Combined beams delivered via articulated arm to focusing handpiece. Used in clinical dermatology settings by physicians. Output: therapeutic laser energy for tissue interaction. Clinical benefit: removal of tattoos/lesions and soft tissue surgery. Internal microprocessor controls wavelength selection, power, and cooling.
Clinical Evidence
Bench testing only. No clinical data required. Device validated against design specifications and predicate device performance standards.
Technological Characteristics
Q-switched Nd:YAG laser; 1064 nm and 532 nm (KTP) wavelengths. Flashlamp-pumped. Articulated arm delivery with focusing handpiece. Internal microprocessor control. Standards: IEC 60825-1, IEC 60601-2-22, IEC 60601-1, IEC 60601-1-2, UL 60601-1.
Indications for Use
Indicated for tattoo removal (dark ink at 1064nm; red, sky blue, green at 532nm), treatment of vascular lesions (port wine birthmarks, telangiectasias, spider/cherry angioma, spider nevi at 532nm), treatment of pigmented lesions (Nevus of Ota at 1064nm; café-au-lait, solar/senile lentigines, Becker's nevi, freckles, Nevus spilus at 532nm), and incision, excision, ablation, and vaporization of soft tissue in general dermatology.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
## FEB 2 8 2014
### 510(k) Summary
# Nd: YAG Laser System Beijing Toplaser Technology Co., Ltd. (As required by 21 CFR 807.92) K Number: K133158
1. Date Prepared: Sep. 22, 2013
2. Sponsor Information Beijing Toplaser Technology Co., Ltd. East 30 Floor, Building M7, No.1 Jiuxianqiao East Road, Chaoyang District, Beijing 100015, China Contract Person: Zhang Xiaosong, General Manager Phone: +86-10-64354759 Fax: +86-10-64356591
3. Proposed Device Information Device Common or Usual Name: Nd: YAG Laser Device Trade or Proprietary Name: Qmetrx 1000 (Victory 11) Nd: YAG Laser System Classification Name: Laser instrument, Surgical, Powered Regulation Number: 21 CFR 878.481 0 Product Code: GEX Panel: General and Plastic Surgery Model: QMetrx1000 (Victory-11)
4. Predicate Device Medlite C6 Q-Switched Nd: YAG Laser (K014234) Medlite™ C3 Q-Switched Nd: YAG Laser (K011677) Manufactured by Continuum Electro-Optics, Inc.
#### 5. Device Description
The Nd: YAG Laser System is based on the Q-Switched Nd:YAG (1064 nm) and frequency doubled KTP Nd:YAG (532 nm) laser technology. There is one optical cavity containing the Nd:YAG crystal. The frequency doubled KTP
{1}------------------------------------------------
Nd:YAG wavelength is achieved by directing the Nd:YAG laser beam through a frequency doubling non-linear crystal. The Nd:YAG laser is activated by means of the use of flashlamp. A red diode aiming beam is reflected onto a coaxial beam path using a beamsplitter assembly. The combined therapeutic and aiming beams are guided by articulated arm to a focusing handpiece.
Both laser wavelengths share a common power supply, control system, and cooling system. The internal microprocessor can be directed to select either the Nd:YAG or the KTP Nd:YAG laser wavelength. The physician is able to select the desired wavelength and the related output energy ,spot size ,fluence via control panel.
6. Intended use and Indications of Use
- · Tattoo Removal
- · Treatment of Vascular Lesions
- · Treatment of Pigmented Lesions,
• Incision, Excision, Ablation, Vaporization of Soft Tissue for General Dermatology
| 1064nm | 532nm |
|--------------------------------|------------------------------------------------------------------------------------------------|
| Tattoo Removal | Tattoo Removal |
| Dark ink: blue and black | Light ink: red Light ink: sky blue and green |
| Treatment of Pigmented Lesions | Treatment of Vascular Lesions |
| Nevus of ota | Port wine birthmarks Telangiectasias Spider angioma Cherry angioma Spider nevi |
| | Treatment of Pigmented Lesions |
| | Café-au-lait birthmarks Solar lentiginos Senile lentiginos Becker's nevi Freckles Nevus spilus |
#### 7. Substantial Equivalence
The product Nd:YAG Laser System, model QMetrx1000(Victory-11) shares the similar indications for use, design features, functional features, same safety compliance. Therefore the product Nd:YAG Laser System, model
{2}------------------------------------------------
QMetrx1000(Victory-11) is substantially equivalent to its predicate device cited above without raising new safety and/or effectiveness issues.
#### 8. Testing
The Nd: YAG Laser System is designed, tested and will be manufactured in accordance with both mandatory and voluntary standards, including:
IEC 60825-1: Safety of laser products - Part 1: Equipment classification, . requirements and user's quide.
IEC 60601-2-22: Medical Electrical Equipment - Part 2: Particular . requirements for the safety of diagnostic and therapeutic laser equipment.
IEC 60601-1: Medical Electrical Equipment - Part 1: General requirements . for safety.
IEC 60601-1-2: Medical Electrical Equipment -Part 1: General requirements . for safety-2, Collateral Standard: Electromagnetic compatibility - Requirements and tests.
UL 60601-1:2003 R6.03 .
Non-Clinical Conclusion:
Laboratory testing was conducted to validate and verify' that the proposed device, QMetrx1000(Victory-11) Nd: YAG Laser System met all design specifications and was substantially equivalent to the predicate device. No Clinical Information is required.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure with three flowing lines representing the body and arms.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 28, 2014
Beijing Toplaser Techonology Company, Ltd. Mr. Zhang Xiaosong General Manager East 3rd Floor, Building M7, No. 1 Jiuxianqiao East Road Chaoyang District, Beijing 100015 CHINA
Re: K133158
Trade/Device Name: Ometrx 1000 (Victory 11) Nd: YAG Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: November 28, 2013 Received: December 2, 2013
Dear Mr. Xiaosong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Mr. Zhang Xiaosong
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Felipe Aguel
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# 4. Indications for Use Statement
### Indications for Use
510(k) Number (if known): K133158
Device Name: Qmetrx 1000 (Victory 11) Nd: YAG Laser System Indications For Use:
- · Tattoo Removal
- · Treatment of Vascular Lesions
- · Treatment of Pigmented Lesions,
- · Incision, Excision, Ablation, Vaporization of Soft Tissue for General Dermatology
| 1064nm | 532nm |
|--------------------------------|---------------------------------|
| Tattoo Removal | Tattoo Removal |
| ● Dark ink: blue and black | ● Light ink: red |
| | ● Light ink: sky blue and green |
| Treatment of Pigmented Lesions | Treatment of Vascular Lesions |
| ● Nevus of ota | ● Port wine birthmarks |
| | ● Telangiectasias |
| | ● Spider angioma |
| | ● Cherry angioma |
| | ● Spider nevi |
| | Treatment of Pigmented Lesions |
| | ● Café-au-lait birthmarks |
| | ● Solar lentiginos |
| | ● Senile lentiginos |
| | ● Becker's nevi |
| | ● Freckles |
| | ● Nevus spilus |
AND/OR Over-The-Counter Use __________________________________________________________________________________________________________________________________________________ Prescription Use YES (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Felipe Aguel Date: 2014.02.28
14:20:51 -05'00'
(Division Sign-Off) for BSA Division of Surgical Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________ Page 1 of ____________________________________________________________________________________________________________________________________________________________________