K131757 · Asclepion Laser Technologies GmbH · GEX · Jul 17, 2014 · General, Plastic Surgery
Device Facts
Record ID
K131757
Device Name
TATTOOSTAR EFFECT COMBO
Applicant
Asclepion Laser Technologies GmbH
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 17, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TattooStar Effect Combo is indicated for incision, excision, ablation and vaporization of soft tissue in general dermatology and the removal of tattoos, pigmented lesions, vascular lesions and unwanted hair and skin resurfacing procedures.
Device Story
TattooStar Effect Combo is a multi-wavelength laser system combining Nd:YAG (1064nm, 532nm, optional 585nm dye) and Ruby (694nm) lasers. Both operate as pulsed q-switched lasers; Nd:YAG also supports free-running mode. Device delivers laser energy via articulated mirror arm to target tissue. Used in clinical dermatology settings by physicians for soft tissue procedures, tattoo removal, pigmented/vascular lesion treatment, hair removal, and skin resurfacing. Output is controlled laser radiation; provider selects wavelength and parameters based on clinical indication. Device enables precise tissue ablation or pigment destruction, facilitating dermatological treatment and aesthetic improvement.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and performance parameters identical to the predicate devices.
Technological Characteristics
Combination Nd:YAG and Ruby laser system. Nd:YAG wavelengths: 1064nm, 532nm, 585nm (via dye handpiece). Ruby wavelength: 694nm. Delivery via articulated mirror arm. Q-switched pulse durations: 8ns (Nd:YAG), 40ns (Ruby). Nd:YAG free-running mode: 300µs. Repetition rates up to 10Hz (Nd:YAG) and 2Hz (Ruby). Fluence up to 37 J/cm² (free-running) and 25 J/cm² (q-switched).
Indications for Use
Indicated for incision, ablation, and vaporization of soft tissue in general dermatology, removal of tattoos, pigmented lesions, and skin resurfacing. Specific applications depend on wavelength: 1064nm (black/blue tattoos, skin resurfacing), 532nm (pigmented lesions, red tattoos), 585nm (sky blue tattoos), and 694nm (pigmented lesions, black/blue/green tattoos).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
## 510(k) SUMMARY ASCLEPION LASÈR TECHNOLOGIES GmbH TattooStar Effect Combo
JUL 17 2014 KI31757
This 510(k) summary of safety and effectiveness for the Asclepion Laser Technologies GmbH TattooStar Effect Combo is submitted in accordance with the requirements of 21 CFR 807.92 and follows Office of Device Evaluation Guidance concerning the organization and content of a 510(k) summary.
| Applicant: | ASCLEPION LASER TECHNOLOGIES GmbH<br>Bruesseler Str. 10<br>07747 Jena, Germany | |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|
| Contact Person: | Mrs. Antje Katzer<br>Product Manager and<br>International Regulatory Affairs Manager | |
| Phone:<br>Fax:<br>e-mail: | +49 3641 77 00 309<br>+49 3641 77 00 302<br>antje.katzer@asclepion.com | |
| Preparation Date: | July 15, 2014 | |
| Device Name: | TattooStar Effect Combo | |
| Common Name: | q-switch laser | |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery<br>and in dermatology<br>79-GEX<br>21 CFR 878.4810 | |
| Equivalent Devices: | TattooStar Effect Y<br>TattooStar R | K112669<br>K060787 |
| Device Description: | The TattooStar Effect Combo is a combination of a Nd:YAG laser<br>and a Ruby laser. Both lasers are pulsed q-switched lasers, the<br>Nd:YAG can optionally emit millisecond pulses in the Free-<br>running mode. The Nd:YAG laser emits wavelengths of 1064nm<br>and 532nm. The beam can be converted to 585nm by means of<br>an optional dye handpiece. The Ruby laser emits 694 nm<br>radiation. | |
{1}------------------------------------------------
The TattooStar Effect Combo is indicated for incision, Intended Use: excision, ablation and vaporization of soft tissue in general dermatology and the removal of tattoos, pigmented lesions, vascular lesions and unwanted hair and skin resurfacing procedures.
| | Proposed Modified Device | Un-Modified Predicate Device |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | TattooStar Effect Combo<br>Modul: R (Ruby laser) | TattooStar R<br>K060787 |
| Indications | Incision, excision, vaporization<br>and ablation of soft tissue, the<br>removal of tattoos and benign<br>pigmented lesions | Cutting, vaporization and<br>ablation of soft tissue, the<br>removal of tattoos and of benign<br>pigmented lesions |
| | There are the same treatment recommendations for both systems: | |
| | 694nm: Removal of pigmented lesions: 3 J/cm², 4-5.5 mm, 0.5-2 Hz, 40 ns<br>Removal of black, blue and green tattoo: 2-3 J/cm², 4-5.5 mm, 0.5-2 Hz, 40 ns | |
| Device Type | Ruby | Ruby |
| Delivery | Articulated mirror arm | Articulated mirror arm |
| Wavelength | 694 nm | 694 nm |
| Max. Energy per<br>Pulse | 1,15 J | 1,2 J |
| Max. Fluence on<br>skin | 25 J/cm² | 20 J/cm² |
| Pulse Duration<br>q-switch | 40 ns | 40 ns |
| Repetition Rate | Up to 2 Hz | Up to 2 Hz |
| Spot Sizes | 2 – 7 mm | 2,5 - 6 mm |
| | Proposed Modified Device | Un-Modified Predicate Device |
| Name | TattooStar Effect Combo<br>Modul: Y (Nd:YAG laser) | TattooStar Effect Y<br>Modul: Y (Nd:YAG laser)<br>K112669 |
| Indications | Incision, excision, vaporization and<br>ablation of soft tissue, the removal of<br>tattoos, pigmented lesions, vascular<br>lesions and hair and skin resurfacing<br>procedures | Incision, excision, vaporization<br>and ablation of soft tissue, the<br>removal of tattoos, pigmented<br>lesions, vascular lesions and hair<br>and skin resurfacing procedures |
| | There are the same treatment recommendations for both systems: | |
| | 1064 nm: Removal of black and blue tattoo color:<br>3-3.5 J/cm², 2-4 mm, 1-10 Hz, 8 ns | |
| | Skin resurfacing: | 1.5-2.5 J/cm², 5.5 mm,<br>10 Hz, 0.3 ms<br>1.5 J/cm²,<br>7 mm,<br>2-5 Hz, 8 ns |
| | 532 nm: Removal of pigmented lesions<br>(including but not limited to lentigo benigna,<br>hyperpigmented burn and boil scar, naevus Ota / Ito,<br>freckles, Becker naevi, Café-au-lait spots):<br>2-2.5 J/cm²,<br>4 mm,<br>1-5 Hz,<br>8 ns | |
| | Removal of red tattoo color: | 2 J/cm²,<br>4 mm,<br>1-5 Hz,<br>8 ns |
| | 585 nm: Removal of sky blue tattoo color: | 2-3 J/cm²,<br>2.5 mm,<br>1-5 Hz,<br>8 ns |
| Device Type | Nd :YAG | Nd :YAG |
| Delivery | Articulated mirror arm | Articulated mirror arm |
| Wavelength | 1064 / 532 / 585 nm | 1064 / 532 / 585 nm |
| Max.<br>Energy per<br>Pulse | 1064 nm: 0,8 J (q-switch mode)<br>1064 nm: 1,5 J (free running mode)<br>532 nm: 0,4 J (q-switch mode)<br>585 nm: 0,25 J (q-switch mode) | 1064 nm: 0,8 J (q-switch mode)<br>1064 nm: 1,5 J (free running mode)<br>532 nm: 0,4 J (q-switch mode)<br>585 nm: 0,25 J (q-switch mode) |
| Max.<br>Fluence on<br>skin | 20 J/cm² (q-switch mode)<br>37 J/cm² (free running mode) | 20 J/cm² (q-switch mode)<br>37 J/cm² (free running mode) |
| Pulse<br>Duration<br>q-switch | 8 ns (q-switch mode)<br>300 µs (free running mode) | 8 ns (q-switch mode)<br>300 µs (free running mode) |
| Repetition<br>Rate | Up to 10 Hz | Up to 10 Hz |
| Spot Sizes | 2 – 7 mm | 2 – 7 mm |
{2}------------------------------------------------
.
·
・
.
{3}------------------------------------------------
The TattooStar Effect Combo is substantially equivalent to Comparison to: the TattooStar Effect Y Laser system and to the TattooStar R with the same principles of operation, with the same parameters and with the same indications for use.
Nonclinical Performance Data: None
Clinical Performance Data: None
Conclusion: The TattooStar Effect Combo is another safe and effective device for the incision, excision, ablation and vaporization of soft tissue in general dermatology and the removal of tattoos, pigmented lesions, vascular lesions and unwanted hair and skin resurfacing procedures.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Control - WO66-G609 Silver Spring, MD 20993-0002
July 17, 2014
Asclepion Laser Technologies GmbH Mr. Antje Katze Regulatory Affairs Manager Bruesseler Street 10 Jena, Germanv 07747
Re: K131757
Trade/Device Name: TattooStar Effect Combo Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: June 20, 2014 Received: June 24, 2014
Dear Mr. Katzer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you: however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{5}------------------------------------------------
Page 2 - Mr. Antje Katze
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21
CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
#### 510(k) Number (if known) K131757
Device Name TattooStar Effect Combo
#### Indications for Use (Describe)
The TattooStar Effect Combo is indicated for incision, ablation and vaporization of soft tissue in general dermatology and the removal of tattoos, pigmented lesions and skin resurfacing procedures. Depending on the wavelength selected, the indications are as follows:
| Wavelength | Indication | Recommended Parameters for the Beginning |
|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| 1064 nm: | Removal of black and blue tattoo color | 3-3.5 J/cm^2, 2-4 mm, 1-10Hz, 8 ns |
| | Skin resurfacing | 1.5-2.5 J/cm^2, 5.5 mm, 10 Hz, 0.3 ms<br>1.5 J/cm^2, 7 mm, 2-5 Hz, 8 ns |
| 532 nm: | Removal of pigmented lesions<br>(including but not limited to lentigo benigna,<br>hyperpigmented bum and boil scar,<br>naevus Ota / Ito, freckles, Becker naevi,<br>Cafe-au-lait spots) | 2-2.5 J/cm^2 4mm, 1-5Hz, 8 ns |
| | Removal of red tattoo color | 2 J/cm^2, 4mm, 1-5Hz, 8ns |
| 585 nm: | Removal of sky blue tattoo color | 2-3 J/cm^2, 2.5mm, 1-5Hz, 8 ns |
| 694nm: | Removal of pigmented lesions | 3 J/cm^2, 4-5.5 mm, 0.5-2 Hz, 40 ns |
| | Removal of black, blue and green<br>tattoo color | 2-3 J/cm^2, 4-5.5 mm, 0.5-2 Hz, 40 ns |
| | Type of Use (Select one or both, as applicable) | |
| | Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Neil R Ogden -S 2014.07.17 11:18:23 -04'00' > < ---
{7}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995. 'DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.'
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.