K131473 · Quanta System Spa · GEX · Oct 24, 2013 · General, Plastic Surgery
Device Facts
Record ID
K131473
Device Name
(QUANTA SYSTEM) SURGICAL LASER FIBERS
Applicant
Quanta System Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 24, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Quanta System Surgical Laser Fibers are intended to be used in conjunction with any cleared surgical laser manufactured by Quanta System equipped with SMA 906 or compatible connector. Quanta System Surgical Laser Fibers are indicated for use in general surgical applications for: incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in a contact or non contact mode (with a compatible laser marketed for use in the desired application). Quanta System Surgical Laser Fibers (surgical fiber optic laser delivery devices) are indicated for use in general surgery, urology, gastroenterology, gynecology, dermatology, vascular surgery, neurosurgery, plastic surgery, ENT/otolaryngology, endovenous occlusion of the greater saphenous vein in the patient with superficial vein reflux and laser assisted lipolysis with an approved compatible laser marketed for use in the desired application. Quanta System Surgical Laser Fibers are also indicated for use in open or closed endoscopic applications where incision, excision, tissue dissection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors or lesions, tissue vaporization, hemostasis and or coagulation may be indicated with an approved compatible laser marketed for use in the desired application. Quanta System Surgical Laser Fibers are also intended as an aid for otologic procedures, for use in incision, excision, coagulation and vaporization of soft and fibrous tissue including osseous tissue with an approved compatible laser marketed for use in the desired application. Quanta System Surgical Laser Fibers are also indicated for use in lithotripsy with an approved compatible laser marketed for use in the desired application. Quanta System Surqical Laser Fibers are indicated for use with laser devices emitting radiation from 532 nm to 2100 nm, with pulsed and continuous wave (CW) emission mode. Quanta System Surgical Laser Fibers are indicated, but not limited, for use with Diode laser, Argon, KTP/532, Ho;YAG, Tm:YAG pulsed and continuous wave CW laser devices. Quanta System Surgical Laser Fibers may be used in surgical specialty or procedures for which compatible lasers have received regulatory clearance: for a complete information about applications, contraindications, precautions and warnings when using fiber optic it is necessary to refer to the applicable laser device User Manual.
Device Story
Quanta System Surgical Laser Fibers are fiber optic delivery systems; silica core/clad fibers; ETFE jacketed; 3.0m length; SMA 905 connector. Used in clinical settings (OR/clinic) by physicians to transmit laser energy from compatible Quanta System lasers to target tissue. Device acts as a passive conduit for laser radiation (532-2100 nm); enables contact or non-contact delivery for surgical applications including incision, ablation, hemostasis, and lithotripsy. Output is laser energy delivered to tissue; clinical decision-making relies on the physician's control of the laser source and fiber tip placement. Benefits include precise tissue interaction for various surgical specialties.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified via visual, mechanical, and functional optical testing; biocompatibility (ISO 10993-1); EtO sterilization validation (ISO 11135-1); packaging validation (ISO 11607-1); and re-processing validation (ISO 17664).
Technological Characteristics
Silica core/silica clad fiber; ETFE jacket; 3.0m length; SMA 905 connector; core diameters 150-1000 microns. Passive fiber optic delivery system. Sterilized by Ethylene Oxide (EtO). No electronic components or software.
Indications for Use
Indicated for patients requiring soft tissue incision, excision, vaporization, ablation, hemostasis, or coagulation; endovenous occlusion for superficial vein reflux; laser-assisted lipolysis; endoscopic resection of tumors/lesions; otologic procedures involving soft, fibrous, or osseous tissue; and lithotripsy. Compatible with lasers emitting 532-2100 nm (pulsed/CW). Contraindications defined by the specific compatible laser device used.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
MED-Fibers Fiber Optics Laser Deliv. System (K124003)
Submission Summary (Full Text)
{0}------------------------------------------------
KI31473
## 7. Traditional 510(k) SUMMARY
#### Introduction:
This document contains the 510(k) Summary for the Quanta System Surgical Laser Fibers . The content of this summary is based on the requirements of 21 CFR 807.92(c).
| Submitter: | Maurizio Bianchi<br>Regulatory Affairs Manager<br>Quanta System SPA |
|------------------------------------------|----------------------------------------------------------------------------------|
| | Email: Maurizio.bianchi@quantasystem.com<br>regulatory@quantasystem.com.local |
| | Phone: +39-0331-376797<br>Fax: +39-0331-367815 |
| Manufacturer name and<br>street address: | Quanta System SPA<br>Via IV Novembre, 116<br>Solbiate Olona (VA)<br>Italy, 21058 |
| | OCT 24 2013 |
| Establishment Registration<br>Number: | 3004378299 |
| Type of Submission: | Traditional 510(k) |
| Trade Name: | Quanta System Surgical Laser Fibers |
| Common Name: | Fiber Optic Laser Delivery System |
| Classification Name: | Surgical Laser Accessory<br>(for Powered Laser Surgical Instrument) |
| Device Classification: | Class II, 21 CFR 878.4810 |
| Device Classification Panel: | General & Plastic Surgery |
| Device Product Code: | GEX |
#### Legally Marketed Device
The Quanta System Surgical Laser Fibers is claimed to be substantially equivalent to these legally marketed devices:
- Laser Peripherals Bare Fibers (K972272)
- Laser Peripherals Reusable Bare Fibers (K011207)
- FiberTech Leoni Bare Fibers (single & reusable) (K050738)
- Lumenis SlimLineEZ Fiber Delivery Device (K011703)
- InnovaQuartz (AMS) Sure Flex Laser Fiber (K050108)
- BioLitec Radial Emitting Fiber Optic Deliv.System (K110080)
{1}------------------------------------------------
- BioLitec MeqaBeam Endo-ENT-Probe (K113858)
- BARD Medical Holmium Laser Fibers (K120926)
- Fiberoptic Fabrications Laser Delivery System (K120810)
- Cynosure SideLaze800Side-Fire Laser deliv. Dev. (K121127)
- MED-Fibers Fiber Optics Laser Deliv. System (K124003)
#### Performance Standards:
There are no mandatory performance standards for this device.
#### General Device Description:
The Quanta System Surgical Laser Fibers are surgical fiber optic laser delivery devices based on a silica core/silica clad fiber jacketed with ethylene tetraffluoroethylene (ETFE). The devices are 3.0 meters (9.8 ft) in length and are terminated with a laser specific SMA 905 connector plus a strain relief on the proximal end. Different distal tip configurations and various core diameter sizes (150, 200, 272, 365, 400, 800 and 1000 microns) are offered.
The Quanta System Surgical Laser Fibers are single-use sterile medical devices, supplied sterilized by Ethylene Oxide (EtO).
#### Summary of Performance and basic Safety testing:
A full plan of non-clinical performance tests was conducted in accordance with ISO 14971:2009. ISO 10993-1:2008 and FDA Blue Book Memo G95-1. This plan includes:
- visual, mechanical and functional(optical) tests per internal methods.
- Biocompatibility testing following ISO 10993-1:2008 and FDA Blue Book Memo G95-1.
- First EtO sterilization validation following ISO 11135-1:2007 and FDA Sterility Review Guidance K90
- Packaging validation following ISO 11607-1.2009 and FDA Sterility Review Guidance K90
- Re-processing validation following ISO 17664:2004 and FDA MDUFMA Validation Data in 510(k) for Reprocessed single-use medical device
#### Intended Use:
Quanta System Surgical Laser Fibers are intended to be used in conjunction with any cleared surgical laser manufactured by Quanta System equipped with SMA 906 or compatible connector.
Quanta System Surgical Laser Fibers are indicated for use in general surgical applications for: incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in a contact or non contact mode (with a compatible laser marketed for use in the desired application).
Quanta System Surgical Laser Fibers (surgical fiber optic laser delivery devices) are indicated for use in general surgery, urology, gastroenterology, gynecology, dermatology, vascular surgery, neurosurgery, plastic surgery, ENT/otolaryngology, endovenous occlusion of the greater saphenous vein in the patient with superficial vein reflux and laser assisted
{2}------------------------------------------------
lipolysis with an approved compatible laser marketed for use in the desired application. Quanta System Surgical Laser Fibers are also indicated for use in open or closed endoscopic applications where incision, excision, tissue dissection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors or lesions, tissue vaporization, hemostasis and or coagulation may be indicated with an approved compatible laser marketed for use in the desired application.
Quanta System Surgical Laser Fibers are also intended as an aid for otologic procedures, for use in incision, excision, coagulation and vaporization of soft and fibrous tissue including osseous tissue with an approved compatible laser marketed for use in the desired application.
Quanta System Surgical Laser Fibers are also indicated for use in lithotripsy with an approved compatible laser marketed for use in the desired application.
Quanta System Surqical Laser Fibers are indicated for use with laser devices emitting radiation from 532 nm to 2100 nm, with pulsed and continuous wave (CW) emission mode.
Quanta System Surgical Laser Fibers are indicated, but not limited, for use with Diode laser, Argon, KTP/532, Ho;YAG, Tm:YAG pulsed and continuous wave CW laser devices.
Quanta System Surgical Laser Fibers may be used in surgical specialty or procedures for which compatible lasers have received regulatory clearance: for a complete information about applications, contraindications, precautions and warnings when using fiber optic it is necessary to refer to the applicable laser device User Manual.
#### Substantial Equivalence:
Quanta System Surgical Laser Fibers have the same components and the same technological characteristics as the predicate devices. The fiber core and cladding are made from silica which is the same material used in all the predicate devices. The fiber is iacketed with ethylene tetraffluoroethylene (ETFE) which is the same patient-contacting material used in all the predicate devices.
Quanta System Surgical Laser Fibers do not have any technological difference compared to predicate devices.
Quanta System Surgical Laser Fibers have the same intended use as the predicate devices.
Nonclinical performance tests demonstrate that the device is safe and effective as the predicate devices.
Thus Quanta System Surgical Laser Fibers are substantially equivalent to the legally marketed predicate devices.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Quanta System SPA Mr. Maurizio Bianchi Regulatory Affairs Manager Via IV Novembre, 116 21058 Solbiate Olana (VA) Italy
October 24, 2013
Re: K131473
Trade/Device Name: Quanta System Surgical Laser Fibers Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: September 9, 2013 Received: September 11, 2013
Dear Mr. Bianchi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRFI does not evaluate information related to contract liability warranties. We remind you: however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Mr. Maurizio Bianchi
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
## 510(k) Number: K131473
#### Device Name: Quanta System S.P.A Surgical Laser Fibers
#### Indications For Use:
Quanta Systems Surgical Laser Fibers are intended to be used in conjunction with any cleared surgical laser manufactured by Quanta system equipped with SMA 905 connector.
Quanta Systems Surgical Laser Fibers are indicated for use in general surgery applications for: incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in a contact mode (with a compatible laser marketed for use in the desired application).
Quanta System Surgical Laser Fibers (surgical fiber optic laser delivery devices) are indicated for use in general surgery, urology, gastroenterology, dermatology, vascular surgery, neurosurgery, plastic surgery. ENT/ololaryngology, endovenous occlusion of the greater saphenous vein in the patient with superficial vein reflux and laser assisted lipolysis with an approved compatible laser marketed for use in the desired application.
Quanta System Surgical Laser Fibers are also indicated for use in open or closed endoscopic applications where incision, excision, tissue dissection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors or lesions, tissue vaporization, hemostasis and or coagulation may be indicated with an approved compatible laser marketed for use in the desired application.
Quanta System Surgical Laser Fibers are also intended as an aid for otologic procedures, for use in incision, excision, coagulation and vaporization of soft and fibrous tissue including osseous tissue with an approved compatible laser marketed for use in the desired application.
opproved on Surgical Laser Fibers are also indicated for use in lithotripsy with an approved compatible laser marketed for use in the desired application.
Quanta System Surgical Laser Fibers are indicated for use with laser devices emitting radiation from 532 nm to 2100 nm, with pulsed and continuous wave (CW) emission mode.
Quanta System Surgical Laser Fibers are indicated, but not limited, for use with Diode laser, Argon, KTP/532, Ho:YAG, Nd;YAG, Tm:YAG pulsed and continuous wave CW laser devices.
Quanta System Surgical Laser Fibers may be used in surgical specially or procedures for which compatible lasers have received regulatory clearance: for a complete information about applications, contraindications, precautions and warnings when using fiber optic it is necessary to refer to the applicable laser device User Manual.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Neil R Ogden 2013.10.24 15:45:00 -04'00'
(Division Sign-Off) for MXM Division of Surgical Devices 510(k) Number K131473
Page 1 of __ 1___
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.