K121580 · Clinicon Corp. · GEX · Sep 25, 2012 · General, Plastic Surgery
Device Facts
Record ID
K121580
Device Name
SURELASE CO2 LASER SYSTEM
Applicant
Clinicon Corp.
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 25, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Clinicon SureLase CO2 Laser System is intended for use in laser surgery procedures for incision, excision, vaporization, ablation, coagulation, or cauterization of soft tissue in specialties such as: general surgery, cosmetic surgery, dermatology, gynecology, head & neck surgery, neurosurgery, oral surgery, otorhinolaryngology, pediatric surgery, plastic & reconstructive surgery, podiatry, and urology
Device Story
SureLase is a mobile, RF-excited CO2 laser system (10.6 µm wavelength) for soft tissue surgery. System uses touch-screen interface to control 2-20W continuous wave, pulsed, and superpulse emission modes. Laser energy delivered via flexible fiber waveguide and handpiece/tip assembly. Includes calibration port for power verification and fiber transmission efficiency. Safety features: keyswitch, emergency stop, remote/fiber interlocks, beam shutter, and internal power/RF monitors. Operated by clinicians in surgical settings. Provides precise tissue interaction (incision/ablation) to facilitate surgical procedures; benefits include controlled tissue removal with coagulation.
Clinical Evidence
No clinical data. Bench testing only. Performance verified through verification/validation tests per product specifications and compliance with 21 CFR 1040, IEC 60601-1, IEC 60601-2-22, and IEC 60825-1.
Indicated for soft tissue incision, excision, vaporization, ablation, coagulation, or cauterization in general, cosmetic, dermatological, gynecological, head/neck, neuro, oral, ENT, pediatric, plastic/reconstructive, podiatric, and urological surgery. No specific age or gender restrictions provided.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "CLINICON" in a bold, sans-serif font. To the left of the word is a logo consisting of several geometric shapes, including triangles and parallelograms, arranged to form a stylized design. The overall impression is that of a company logo or brand name, possibly related to the medical or healthcare industry given the name "CLINICON".
Clinicon Corporation 3025 Industry Street Suite A Oceanside, CA 92054 760-439-1700 760-439-1798 (FAX)
SION FOCUSED"
# SECTION 5 - 510(k) Summary
510(k) Summary Preparation Date:
- 1. 510(K) Owner:
Fritz A. Brauer, President Clinicon Corporation 3025 Industry Street Suite A Oceanside, CA 92054 800-254-6426 / 760-439-1700 / 760-439-1798 (FAX)
- 2. 510(k) Contact:
Gerald S. Palecki Quality / Regulatory Consultant Palecki Enterprises LLP 1107 259th ST NW Stanwood, WA 98292 760-715-5954 gpalecki@palecki-enterprises.com
- 3. Device Trade Name: SureLase
Common Name: CO2 Laser System
Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology (21 CFR 878.4810).
> "A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide."
Classification: Class II
Product Code: GEX
- 4. Predicate Device(s)
The SureLase CO2 Laser System is equivalent in operating principles and intended uses to the Luxar LX-20SP and DEKA SmartXide CO2 Laser Systems.
- 5. Device Description and Function
The Clinicon SureLase laser system is a mobile platform that utilizes a radio frequency (RF) excited carbon dioxide (CO2) laser tube to produce an infrared beam at a nominal 10.6 µm wavelength at powers adjustable from 2 to 20 Watts Continuous Wave (CVV). Laser energy is conducted to the point of application by a flexible fiber waveguide and handpiece / tip assembly. Laser system operation is controlled by operator input on a touch-screen display panel. The RF laser drive is modulated to provide additional pulsed and superpulse emission modes selected from the laser system control panel.
K121580
SEP 25 2012
{1}------------------------------------------------
# SECTION 5 - 510(k) Summary
A "calibration port" on the side of the laser system allows checking and setting the power emitted from the distal laser aperture and serves as a check on fiber waveguide transmission efficiency. Laser system power, rates, and durations are adjustable as tabulated below:
| Parameter | SureLase |
|-------------------------------|-----------|
| CW Power (watts) | 2-20 |
| Pulsed Power (watts) | 2-20 |
| Gated Pulse Width (msec) | 5-500 |
| Gated Rep Rate (pps) | 1 or 2-20 |
| Superpulse Power (watts) | 2-10 |
| Gated Pulse Widths (msec) | 10-500 |
| Superpulse Pulse Rates (pps) | 160-375 |
| Gated Rep Rate (pps) | 1 or 2-30 |
| Peak Power (watts) | > 50 |
| Superpulse Pulse Width (usec) | 100 -800 |
The laser system has safety features complying with requirements in 21 CFR 1040, Performance Standards for Light Emitting Products; IEC 60601-2-22, Medical electrical equipment –Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment ; and IEC 60825-1, Safety of laser products -Part 1: Equipment classification and requirements. Major safety features are as follows:
System On-Off Keyswitch, Emergency Stop Switch, Remote Interlock, Fiber Interlock, Beam Blocking Shutter, Internal Laser Power Detector, RF Power Monitor, and required Laser Safety Labels and Labeling.
The laser system, components, and features are illustrated in Figure 1. Laser system physical characteristics are:
| Parameter | SureLase |
|----------------------|-------------------------|
| Dimensions | 40" H x 15" W x 15" D |
| Weight (Pounds / Kg) | ~ 47 / ~ 21.3 |
| Electrical Power | 115 VAC, 15 A, 50/60 Hz |
3
{2}------------------------------------------------
Delivery System: Flexible Fiber Waveguide; ~ 0.5mm ID.; Handpiece w/collet grip; wavequide tips.
- Purge Gas: Provision for external air or inert gas purge through the Fiber and Handpiece / Tip.
System Cooling: Air; two thermostatically controlled fans with over-temperature protection.
Mobility: 4 wheels and handgrip on console for convenient system positioning.
6. Intended use(s) of the Device
The Clinicon SureLase CO2 Laser System is intended for use in laser surgery procedures for incision, excision, vaporization, ablation, coagulation, or cauterization of soft tissue in specialties such as: general surgery, cosmetic surgery, dermatology, gynecology, head & neck surgery, neurosurgery, oral surgery, otorhinolaryngology, pediatric surgery, plastic & reconstructive surgery, podiatry, and urology
- 7. Technological Characteristics Comparison to Predicate Devices
The technological characteristics comparison to predicate devices is summarized in the following table.
| Characteristic | Clinicon Surelase | Luxar LX-20 SP | DEKA SmartXide |
|-----------------------|-----------------------------|---------------------------------------------------|-----------------|
| Predicate 510(k) | N/A | K953074 | K072159 |
| Laser Medium | CO2 | CO2 | CO2 |
| Wavelength (µm) | 10.6 | 10.6 | 10.6 |
| Laser Drive Source | RF | RF | DC |
| Output Power (W) | 2 - 20 | 2 - 20 | 2 - 30 |
| Pulsed Power (W) | 2 - 20 | 2 - 20 | 0.5 - 15 |
| Gated PW (ms) | 5 - 500 | 5 - 500 | 0.2 - 80 |
| Gated Rep. Rate (pps) | 1 or 2 - 20 | 2 - 20 | 5 - 100 |
| Super Pulse (W) | 2 - 10 | 2 - 10 | 0.5 - 10 |
| Super Pulse Peak (W) | > 50 | 50 | 320 |
| Beam Delivery System | Flexible Fiber<br>Waveguide | Articulated Arm or<br>Flexible Fiber<br>Waveguide | Articulated Arm |
| System HxWxD (in) | 40 x 15 x 15 | 36 x 14 x 14 | 47 x 19 x 22 |
{3}------------------------------------------------
## SECTION 5 - 510(k) Summary
| System Weight (lb) | 47 | 42 | 66 |
|--------------------|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| Mobility | 4 Wheels & Handle | 4 Wheels & Handle | 4 Wheels & Handle |
| Electrical Power | 115 VAC | 115 VAC | 115 VAC |
| Intended Use | Incision, Excision,<br>Vaporization,<br>Ablation,<br>Coagulation, or<br>Cauterization of<br>Soft Tissue | Incision, Excision,<br>Vaporization,<br>Ablation,<br>Coagulation, or<br>Cauterization of<br>Soft Tissue | Incision, Excision,<br>Vaporization,<br>Ablation,<br>Coagulation, or<br>Cauterization of<br>Soft Tissue |
#### 8. Performance Testing
The SureLase CO2 Laser System has been evaluated for performance equivalent to predicate devices through a combination of (1) verification and validation tests per product requirements and specifications and (2) inspections and tests derived from applicable regulations and safety standards as noted below.
- 21 CFR 1040, Performance Standards for Light-emitting Products
- IEC 60601-1, Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance
- IEC 60601-2-22, Medical electrical equipment Part 2-22: Particular requirements for basic safety and essential performance of surgical. cosmetic, therapeutic and diagnostic laser equipment
- IEC 60825-1, Safety of laser products Part 1: Equipment classification, requirements and user's guide
### 9. Animal and Clinical Testing
Device performance evaluation did not involve Animal of Clinical Testing.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Clinicon Corporation % Palecki Enterprise, LLP Mr. Gerald S. Palecki Quality/Regulatory Consultant 1107 259TH Street NW Stanwood, Washington 98292
SEP 25 2012
Re: K121580
Trade/Device Name: SureLase CO2 Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: August 8, 2012 Received: August 14, 2012
Dear Mr. Palecki:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
### Page 2 - Mr. Gerald S. Palecki
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark McMullen
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## SECTION 4 - Indications for Use
510(k) Number (if known): J/ (21580 Device Name: SureLase CO2 Laser System Indications for Use:
The Clinicon SureLase CO2 Laser System is intended for use in laser surgery procedures for incision, excision, vaporization, ablation, coagulation, or cauterization in specialties such as: general surgery, cosmetic surgery, dermatology, gynecology, head & neck surgery, neurosurgery, oral surgery, otorhinolaryngology, pediatric surgery, plastic & reconstructive surgery, podiatry, and urology.
#### Prescription Use (Part 21 CFR 801 Subpart D)
## AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sian-Off
Office of Surgical, Orthopedic, and Restorative Device
510(k) K121580
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.