SIDELAZE LASER BEAM DELIVERY ACCESSORY FOR CYNOSURE 1440NM WAVELENGTH LASERS
K121127 · Cynosure, Inc. · GEX · May 13, 2012 · General, Plastic Surgery
Device Facts
Record ID
K121127
Device Name
SIDELAZE LASER BEAM DELIVERY ACCESSORY FOR CYNOSURE 1440NM WAVELENGTH LASERS
Applicant
Cynosure, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
May 13, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
SideLaze800™ is an accessory to deliver optical energy for Cynosure 1440mm wavelength lasers and intended to be used with Cynosure 1440nm wavelength lasers' cleared indications for use, such as the surgical incision, excision, vaporization, ablation, and coagulation of soft tissue (including skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle, cartilage meniscus, mucous membrane, lymph vessels and nodes, organs and glands) and laser assisted lipolysis. SideLaze800™ may be used in combination with a Cynosure SMA-compatible laser system for the same indications.
Device Story
SideLaze800™ is a side-firing fiber optic delivery accessory for Cynosure 1440nm wavelength surgical lasers. It functions by directing laser energy laterally to the fiber axis, facilitating soft tissue procedures including incision, excision, vaporization, ablation, coagulation, and laser-assisted lipolysis. Used in clinical settings by physicians, the device attaches to existing laser systems to deliver optical energy to target tissues. The output allows for precise tissue interaction as determined by the host laser's parameters. It provides a specialized delivery method for existing cleared laser indications, potentially improving access or control during minimally invasive surgical or dermatological procedures.
Clinical Evidence
Bench testing only.
Technological Characteristics
Side-firing fiber optic delivery accessory for 1440nm wavelength lasers. Materials are equivalent to previously cleared Cynosure delivery systems (K091537, K102541). Conforms to Laser Performance Standard 21 CFR 1040. Compatible with Cynosure SMA-compatible laser systems.
Indications for Use
Indicated for surgical incision, excision, vaporization, ablation, and coagulation of soft tissue (skin, cutaneous/subcutaneous tissue, muscle, cartilage, mucous membrane, lymph vessels/nodes, organs, glands) and laser-assisted lipolysis using Cynosure 1440nm wavelength lasers.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Cynosure Multiwavelength Laser with 1440nm Wavelength (K091537)
Cynosure Cellulaze Laser and Cellulaze Delivery Kit (K102541)
Twister™ Side-Firing Optic Delivery System (K12987)
Submission Summary (Full Text)
{0}------------------------------------------------
K 121127
MAY 1 3 2012
# 510(k) Summary
.
| 510(K) Owner: | CYNOSURE, INC. |
|---------------|----------------------------------------------|
| | 5 Carlisle Road |
| | Westford, MA 01886 |
| | Telephone: 978-256-4200 |
| | Fax: 978 256 6556 |
| | Owner/Operator Number: 9006419 |
| Contact: | Irina Kulinets |
| | VP of Regulatory Affairs and Quality Systems |
| | CYNOSURE, INC. |
| | 5 Carlisle Road |
| | Westford, MA 01886 |
| | Telephone: 978-367-2350 |
| | Fax: 978 256 6556 |
| | Email: ikulinets@cynosure.com |
| Submitter: | Lori Kahler |
| | The RC Insight Group |
| | 461 Main Street, Suite 217 |
| | Pawtucket, RI 02860 |
| | Telephone: 202-715-6242 |
| | Fax: 202-686-2877 |
| | Email: lkahler@rcinsightgroup.com |
| Manufacturer: | CYNOSURE, INC. |
| | 5 Carlisle Road |
| | Westford, MA 01886 |
| | Telephone: 978-256-4200 |
| | Fax: 978-256-4200 |
| | Establishment Registration Number: 1222993 |
12 April 2012 Date Prepared:
Addition of SideLaze800™ Option -Special 510(k)
8
Confidential
{1}------------------------------------------------
Trade name: SideLaze800™, laser beam delivery accessory for Cynosure 1440nm wavelength lasers
Powered Laser Surgical Instrument (Laser Common name: for Surgery and Dermatology)
Classification name: 21 CFR 878.4810
GEX (Laser Surgical Instrument for Use in General and Plastic Product Code(s): Surgery and in Dermatology)
Classification: Class II
Predicate Devices (Claiming Substantial Equivalence):
Cynosure Multiwavelength Laser with 1440nm Wavelength K091537
Cynosure Cellulaze Laser and Cellulaze Delivery Kit K102541
Twister™ Side-Firing Optic Delivery System K12987
Summary Description of the Device:
SideLaze800™ is an optional side-firing fiber optic accessory to the currently cleared Cynosure Lasers with 1440nm wavelength. The addition of the Cynosure SideLaze800™ option to the cleared laser devices is intended to offer physicians a convenient accessory within cleared indications for use.
{2}------------------------------------------------
### Intended Use / Indications for Use:
SideLaze800™ is an accessory to deliver optical energy for Cynosure 1440mm wavelength lasers and intended to be used with Cynosure 1440nm wavelength lasers' cleared indications for use, such as the surgical incision, excision, vaporization, ablation, and coagulation of soft tissue (including skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle, cartilage meniscus, mucous membrane, lymph vessels and nodes, organs and glands) and laser assisted lipolysis. SideLaze800™ may be used in combination with a Cynosure SMA-compatible laser system for the same indications.
### Technological Characteristics:
The SideLaze800™ option contains equivalent components and patient-contact materials as the cleared delivery accessories of Cynosure Multiwavelength Laser with 1440nm wavelength (K091537) and Cynosure Cellulaze (K102541). The fundamental scientific technology of the modified device (addition of SideLaze800™ accessory) has not changed.
### Performance Standards:
This device conforms to the Laser Performance Standard (21 CFR 1040). No additional performance standards have been promulgated under Section 514 of the Food, Drug and Cosmetic Act for these devices.
## Performance Data:
Performance (bench) testing summaries relating to the addition of the SideLaze800™ option is included in this submission.
#### Substantial Equivalence:
The SideLaze800™ option of the Cynosure 1440nm wavelength lasers is as safe and effective as the accessories cleared with the Cynosure Multiwavelength Lasers with 1440nm wavelength. The SideLaze800™ accessory option does not represent any fundamental changes in technology or any changes in indicated use, and this accessory option does not raise any additional questions of safety and effectiveness. In summary, the addition of the SideLaze800™ option used to deliver optical energy for Cynosure 1440nm wavelength lasers described in this submission is substantially equivalent to the predicate devices.
Confidential
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Cynosure, Inc. % Ms. Irina Kulinets Vice President, Regulatory Affairs and Quality Systems 5 Carlisle Road Westford, Massachusetts 01886
MAY 1 3 2012
Re: K121127
Trade/Device Name: SideLaze800", Laser Beam Delivery Accessory for Cynosure 1440nm Wavelength Lasers
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
plastic surgery and in dermatolo
. Class II
Regulatory Class: Class II Product Code: GEX Dated: May 08, 2012 Received: May 10, 2012
Dear Ms. Kulinets:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability. warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 - Ms. Irina Kulinets
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
For Robert chriſtopher Dri
Mark N. Melkerson · Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
K,121127
510(k) Number (if known):
Device Name:
SideLaze800™, laser beam delivery accessory for Cynosure 1440nm wavelength lasers
Indications for Use:
SideLaze800™ is an accessory to deliver optical energy for Cynosure 1440nm wavelength lasers and intended to be used with Cynosure 1440nm wavelength lasers' cleared indications for use, such as the surgical incision, excision, vaporization, ablation, and coagulation of soft tissue (including skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle, cartilage meniscus, mucous membrane, lymph vessels and nodes, organs and glands) and laser assisted lipolysis. SideLaze800™ may be used in combination with a Cynosure SMA-compatible laser system for the same indications.
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | Neil RP Ogden for mxn |
|--------------------------------------------------------|-----------------------------------------------------------|
| | (Division Sign-Off) |
| | Division of Surgical, Orthopedic, and Restorative Devices |
| Addition of SideLaze800™ Option - Special 510(k) | Confidential | Page 6 of 18 |
|--------------------------------------------------|--------------|--------------|
|--------------------------------------------------|--------------|--------------|
| 510(k) Number | K121127 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.