K113588 · Lutronic Corporation · GEX · Feb 22, 2012 · General, Plastic Surgery
Device Facts
Record ID
K113588
Device Name
SPECTRA LASER SYSTEM
Applicant
Lutronic Corporation
Product Code
GEX · General, Plastic Surgery
Decision Date
Feb 22, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SPECTRA Laser System in indicated for: the incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. 532nm Wavelength (nominal delivered energy of 585 nm and 650 nm with optional dye handpieces): -Tattoo removal: light ink (red, tan, purple, orange, sky blue, green) -Removal of Epidermal Pigmented Lesions -Removal of Minor Vascular Lesions including but not limited to telangiectasias -Treatment of Lentigines -Treatment of Café-Au-Lait -Treatment of Seborrheic Keratoses -Treatment of Post Inflammatory Hyper-Pigmentation -Treatment of Becker's Nevi, Freckles and Nevi Spilus 1064nm Wavelength: -Tattoo removal: dark ink (black, blue and brown) -Removal of Nevus of Ota -Removal or lightening of unwanted hair with or without adjuvant preparation. -Treatment of Common Nevi -Skin resurfacing procedures for the treatment of acne scars and wrinkle -Treatment of melasma.
Device Story
SPECTRA Laser System is a Q-switched Nd:YAG laser producing 1064 nm and 532 nm pulsed coherent light; optional dye handpieces convert 532 nm to 585 nm or 650 nm. Device operates by delivering laser energy to target chromophores (melanin, tattoo particles) in human tissue; rapid localized heating causes fragmentation of chromophores. Physician controls treatment fluence, pulse energy, and spot size via handpiece to optimize clinical effect. Used in clinical settings by physicians for dermatologic and surgical procedures. Output allows for precise tissue ablation, coagulation, and pigment/tattoo removal. Benefits include effective treatment of various pigmented lesions, vascular lesions, and skin resurfacing.
Clinical Evidence
No clinical data presented; bench testing only.
Technological Characteristics
Q-switched Nd:YAG laser; wavelengths 1064 nm and 532 nm; optional dye handpieces for 585 nm and 650 nm. Articulated arm delivery system. Energy source: electrical. Device is a surgical laser instrument.
Indications for Use
Indicated for patients requiring soft tissue incision, excision, ablation, vaporization, coagulation, or hemostasis in dermatology/general surgery. Specific applications include tattoo removal (various ink colors), epidermal pigmented/vascular lesions, lentigines, café-au-lait spots, seborrheic keratoses, post-inflammatory hyper-pigmentation, nevi (Becker's, Spilus, Common, Ota), hair removal/lightening, acne scar/wrinkle resurfacing, and melasma treatment.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Lutronic Corporation SPECTRA Laser System (K103455)
Palomar Medical Technologies, Inc. Q-YAG 5 (K061436)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summary for the Lutronic Corporation Spectra Laser System
K113588
FEB 2 2 2012
This 510(k) Summary is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.
## 1. General Information
| Submitter: | Lutronic Corporation<br>Room 403-1,2,3,4,5, 404 Ilsan Technotown<br>1141-1, Baekseok-dong, Ilsandong-gu,<br>Goyang-si, Gyeonggi-do, 410-722<br>Republic of Korea |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jhung Won Vojir, Ph.D<br>Chief Executive Officer<br>Lutronic, Inc.<br>Six Neshaminy Interplex, Suite<br>Trevose, PA 19053<br>Telephone: 215-205-2219<br>Fax: 609-488-6958 |
| Summary Preparation Date: | December 2, 2011 |
| Names | |
| Device Name: | SPECTRA Q-Switched Nd:YAG<br>Laser System with Dye Handpieces |
| Classification Name: | Laser Instrument, Surgical, Powered<br>Product Code: GEX |
#### 3. Predicate Devices
2
The SPECTRA Laser System is substantially equivalent to the Lutronic Corporation SPECTRA Laser System (K103455) and Palomar Medical Technologies, Inc. Q-YAG 5 (K061436).
## 4. Device Description
The SPECTRA Laser System produces a pulsed beam of coherent near infrared 11064 nm) and visible (532nm) light. This beam is directed to the treatment zone (100 Film) and risulated arm coupled to a handpiece. In addition, two dye
Panel: General & Plastic Surgery
{1}------------------------------------------------
handpieces are available that convert the 532 nm wavelength to 585 nm and 650 nm.
When the beam contacts human tissue, the energy in the beam is absorbed, resulting in a very rapid, highly localized temperature increase to the target chromospheres such as melanin and tattoo particles. This increases localized temperature of the chromospheres. The instantaneous temperature increase causes fragmentation of the chromospheres to smaller particles.
By directing the beam onto specific tissue locations, using different handpieces, and controlling the treatment fluence, the intensity of the temperature of the target can be varied. The physician can optimize the effect for different applications by controlling the energy of the laser pulse and the spot size of the treatment beam.
#### Indications for Use 5.
The SPECTRA Laser System is indicated for the incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis.
532nm Wavelength (nominal delivered energy of 585 nm and 650 nm with optional dye handpieces):
-Tattoo removal: light ink (red, tan, purple, orange, sky blue, green) -Removal of Epidermal Pigmented Lesions
-Removal of Minor Vascular Lesions including but not limited to telangiectasias
-Treatment of Lentigines
-Treatment of Café-Au-Lait
-Treatment of Seborrheic Keratoses
-Treatment of Post Inflammatory Hyper-Pigmentation i
-Treatment of Becker's Nevi, Freckles and Nevi Spilus
1064nm Wavelength:
-Tattoo removal: dark ink (black, blue and brown)
-Removal of Nevus of Ota
-Removal or lightening of unwanted hair with or without adjuvant preparation.
-Treatment of Common Nevi
-Skin resurfacing procedures for the treatment of acne scars and wrinkle
-Treatment of melasma.
## 6. Substantial Equivalence
The SPECTRA Laser System is substantially equivalent to the SPECTRA Laser System cleared in K103455 and the Palomar Q-YAG 5 Laser System cleared in Bystem freared the SPECTRA Laser System has the same indications statement as the
{2}------------------------------------------------
SPECTRA Laser System cleared in K103455 with the exception of the melasma indication for use which is being added for the 1064 nm wavelength in this 510(k). The SPECTRA Laser System has the same intended use as the Palomar Q-YAG 5 (K061436) with minor differences in the indications for use statement. Both the SPECTRA and the Palomar Q-YAG 5 are intended for the incision, excision, ablation, and vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. The SPECTRA Laser System has the same technological characteristics as the predicate devices. All three devices are Q-switched Nd: YAG lasers operating at wavelength of 1064 nm and 532 nm. Additionally, all three devices have dye handpieces that convert the 532 nm wavelength beam into 585 nm or 650 nm wavelengths. The SPECTRA Laser System has the exact technological characteristics as the previously cleared SPECTRA Laser System (K103455). The SPECTRA Laser System has similar technological characteristics to the Q-YAG 5 predicate. The SPECTRA Laser System is substantially equivalent to the predicate devices.
7. Performance Data None presented.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the perimeter. Inside the circle is an eagle emblem, which is a symbol of the United States. The eagle is depicted in a stylized manner, with its wings spread and its head turned to the right.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
FEB 2 2 2012
Lutronic Corporation % Lutronic, Inc. Jhung Won Vojir 6 Neshaminy Interplex, Suite 207 Trevose, Pennsylvania 19053
Re: K113588
Trade/Device Name: SPECTRA Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: II Product Code: GEX Dated: December 2, 2012 Received: December 5, 2012
Dear Jhung Won Vojir:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I va may , aronerols provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r hat FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
## Page 2 - Jhung Won Vojir
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark McMullen
Mark N. Melkerson Director Division of Surgical. Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
ા ૩૬૫૫૮ 510(k) Number (if known):
Device Name:__________________________________________________________________________________________________________________________________________________________________ SPECTRA Laser System
#### Indications for Use:
The SPECTRA Laser System in indicated for: the incision, excision, ablation, The of DOTTE Laber of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis.
532nm Wavelength (nominal delivered energy of 585 nm and 650 nm with optional dye handpieces):
-Tattoo removal: light ink (red, tan, purple, orange, sky blue, green)
-Removal of Epidermal Pigmented Lesions
-Removal of Minor Vascular Lesions including but not limited to telangiectasias
-Treatment of Lentigines
-Treatment of Café-Au-Lait
-Treatment of Seborrheic Keratoses
-Treatment of Post Inflammatory Hyper-Pigmentation
-Treatment of Becker's Nevi, Freckles and Nevi Spilus
1064nm Wavelength:
-Tattoo removal: dark ink (black, blue and brown)
-Removal of Nevus of Ota
-Removal or lightening of unwanted hair with or without adjuvant preparation.
-Treatment of Common Nevi
- Treatment or Sommedures for the treatment of acne scars and wrinkle
-Treatment of melasma
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
ાર્ ભ
Over The Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nilson Diniz firmino
Page 1 of 1
(Division Sign-Off Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K113588
Confidential and Proprietary Information for Lutronic Corporation
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.