The QuickLase™DUALand QuickLase™810/980 Dental Lasers are intended for incision, excision, ablation, vaporization, and/or coagulation of oral soft tissues (including marginal and interdental gingival and epithelial lining of free gingiva). It is also intended for light activation for bleaching materials for teeth whitening, and laser assisted bleaching/whitening for teeth whitening.
Device Story
Device is a dental diode laser system using Gallium Aluminum Arsenide (GaAlAs) semiconductor diodes to emit infrared laser energy at 810nm or 980nm. Input is provided via membrane switch pad; operator selects parameters and activates laser using a footswitch. Device delivers energy through a 400-micron fiber optic cable. Includes a visible 640-650nm red aiming beam. Used by dental professionals in clinical settings for soft tissue procedures and teeth whitening. Output is laser energy for tissue interaction; provides hemostasis and ablation. Benefits include precise tissue management and reduced inflammation. Device is identical to previously cleared predicate systems, with only branding and housing cosmetic changes.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on identical technological characteristics and performance specifications to the predicate device (K072295). Bench testing confirmed that power densities and safety controls are identical to the predicate.
Technological Characteristics
Solid-state GaAlAs semiconductor diode laser; 810nm or 980nm wavelength; 0.1-10W power output. 400-micron fiber optic delivery system. Membrane switch interface. Visible 640-650nm red aiming beam (<5mW). Complies with 21 CFR 1040.10/1040.11, IEC 60601, and ISO 13485. Sterilizable delivery components.
Indications for Use
Indicated for incision, excision, ablation, vaporization, hemostasis, and treatment of oral soft tissue, including biopsies, frenectomy, gingivectomy, gingivoplasty, implant recovery, abscess drainage, leukoplakia, operculectomy, pulpotomy, gingival hypertrophy reduction, soft tissue crown lengthening, vestibuloplasty, curettage, tissue retraction, and removal of diseased/inflamed/necrotic soft tissue in periodontal pockets. Also indicated for light-activated teeth whitening.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
TwinWave and TidalWave by Lumen Development and Manufacturing, Inc. (K072295)
DioDent Micro 980/810 by Hoya Conio, Inc (K063384)
Odyssey Navigator Diode Laser by Ivoclar Vivadent Inc (K062258)
SIROLaser by Sriona Dental Systems, GmbH (K053161)
{0}------------------------------------------------
QuickLase Limited Traditional 510(K)-QuickLase™DUAL+ QuickLase™ 810/980
K100474/
## 510(K) Summary
し QuickLase Limited
MAR 1 6 2010
| Submitter: | QuickLase Limited<br>Rosalind Nahab<br>Director<br>18 Dover Street<br>Canterbury, Kent CT1 3HD<br>United Kingdom | MAR 1 6 2010 |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Preparer & Contact: | Calvin D. Ostler<br>Agent and E-contact<br>PO Box 1264<br>1094 West Greasewood Drive<br>Riverton, Utah 84065<br>Phone: 801-831-5331<br>Email: CalOstler@msn.com | |
| Preparation Date: | February 9, 2010 | |
| Device Trade Name: | QuickLase™DUAL+ Dental Laser<br>QuickLase™ (810nm) Dental Lasers<br>QuickLase™ (980nm) Dental Lasers | |
| Common Name: | Dental Diode Laser | |
| Classification Name: | Instrument, surgical, powered, laser | |
| Regulation Description: | 878.4810 Laser surgical instrument for use in general and<br>plastic surgery and dermatology<br>A. Identification.<br>(1) A carbon dioxide laser for use in general surgery and in<br>dermatology is a laser device intended to cut, destroy, or<br>remove tissue by light energy emitted by carbon dioxide<br>(2) An argon laser for use in dermatology is a laser device<br>intended to destroy or coagulate tissue by light energy<br>emitted by argon.<br>B. Classification<br>(1) Class II | |
| Product Code: | GEX | |
This document including any attachments may confidential information, privileged material or otherwise constitute non-public information by contract of all alle contain proprietary, contestial merinent pronogother than the intended pay is prohibited. I four niomation by contact of law. Any been mornition the Reserved the Research on one of the more of the more of the semination, verieted ins occanent in effor, please inniedately lectific and action of your of real of your your your versions is not authorized and may be inlay be in
{1}------------------------------------------------
| Device Class | 2 |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Review Panel | General & Plastic Surgery |
| Equivalent Device: | TwinWave and TidalWave by Lumen Development and<br>Manufacturing, Inc., a California Corporation, K072295<br>AND<br>DioDent Micro 980/810 by Hoya Conio, Inc, K063384<br>Odyssey Navigator Diode Laser by Ivoclar Vivadent Inc,<br>K062258<br>SIROLaser by Sriona Dental Systems, GmbH, K053161<br>LaserSmile by BioLase Technology, Inc., K030539 |
| Device Description: | The laser source of the QuickLase™DUALDental Lasers or<br>QuickLase™810/980 Dental Laser is an array of Gallium<br>Aluminum Arsenide (GaAlAs), solid state, semiconductor<br>diode lasers. The individual lasers comprising the array are<br>capable of producing invisible infrared laser energy at either<br>810 or 980 nanometers depending upon the concentration of<br>and construction of the epitaxial layers of the diode. The<br>delivery system consists of a fiber optic cable with a nominal<br>diameter of 400 microns. In all available configurations, that<br>portion of the delivery system that is handled by the operator<br>and/or in contact with the patient can be sterilized by heat,<br>steam, gas, chemicals or combinations thereof. Activation of<br>the laser occurs after the operator either selects or enters<br>the parameters for the procedure, verifies and authorizes the<br>parameters and depresses a footswitch which fires the laser.<br>The footswitch is an on/off switch. The interface for input is<br>electro-mechanical by way of a membrane switch pad. The<br>laser operates in continuous wave or pulse mode which is<br>selected and/or verified by the operator. A visible aiming<br>beam is provided in the form of a visible red laser which<br>emits not more than 5mW (user adjustable and verified) in<br>the 640-650 nanometer wavelength range. |
| Intended Use: | The QuickLase™DUALand QuickLase™810/980 Dental<br>Lasers are intended for incision, excision, ablation,<br>vaporization, and/or coagulation of oral soft tissues<br>(including marginal and interdental gingival and epithelial<br>lining of free gingiva). It is also intended for light activation |
nomarios by onlast of als costined the mornition that in the more one that in home of the provincial of por system.
There received this document in entremation contained nee
{2}------------------------------------------------
for bleaching materials for teeth whitening, and laser assisted bleaching/whitening for teeth whitening.
Comparison:
The QuickLaseDUAL, QuickLase810, and QuickLase980, Diode Dental Laser Systems in their various configurations are, in fact, the TwinWave, TidalWave810, and TidalWave980 Diode Dental Lasers Systems that were cleared by the FDA on December 18, 2007 under 510(K) Number: K072995. The applications having been submitted by Lumen Development and Manufacturing, Inc. a California Corporation. We, QuickLase Limited, entered into a joint development agreement for these devices and, by the terms of the contract, were given rights in Europe. We obtained CE testing and marking and launched these lasers under the tradename QuickLaseDUAL+, QuickLase810, and QuickLase980. These lasers have performed flawlessly in Europe in thousands of procedures over the past 2 years. Lumen Development and Manufacturing, Inc., a California Corporation, fell under hard times and is out of business. According to the terms in our contract we are now able to market the devices in the USA. We are, therefore, applying for 510(K) clearance from the FDA for marketing the devices in the USA. With a notable caveat; these lasers are not substantially equivalent to, they are identical in terms of safety and efficacy. The only difference between the approved devices under 510(K) number K072995 and the devices covered in this application are the brandname change from TwinWave, TidalWave 810, and TidalWave980 to, QuickLaseDUAL+, QuickLase810, and QuickLase980 respectively, and some cosmetic differences in the housings. The units operate exactly the same and produce exactly the same results.
None the less, we offer the exact same narrative in this Summary as was submitted in and cleared in 510(k) Number K072995:
This document including any attachments may confidential information, privileged material or other wise and in and in and in and in and in and in and in and in and in and nt in the booking any elterments hay online propriodary, somiterials by anyone other than the interest prohibited. If your in the interest probles of your homilation by contact of raw. Ally disernent and in the remation the risens of the recomment. Use, dissemination, and may and may and may and may any ve received in enot, press infinedicty fective information contained herein by united by sons is not authorized and may be unlawful.
{3}------------------------------------------------
## QuickLase Limited
Traditional 510(K)-QuickLase™DUAL+ QuickLase™ 810/980
The QuickLase™DUALDental Laser produces a power range of . 1-5 Watts of 980nanometer laser energy and .1-5watts of 810nanometer laser energy; a combination of .1-10watt of invisible infrared energy. The QuickLase™810/980 980 Dental Laser produces a power range of .1-10 Watts of 980nanometer invisible infrared energy. The QuickLase™810/980 810 Dental Laser produces a power range of .1-10 Watts of 810nanometer invisible infrared energy. An operator adjustable visible red aiming beam is incorporated by way of solid state 640-650nanometer range solid state lasers with a maximum output of 5 milliwatts. The energy is delivered through 400 micron diameter optical fiber. All four of the comparable devices deliver comparable energy density through identical or similar/comparable fiber delivery systems; therapeutic results are identical. The SiroLaser and DioDent Micro 980 are rated at 7 and 3watts (respectively) of 980nanometer laser energy. The LaserSmile and Navigator are rated at 10 and 3 watts (respectively) of 810nanometer laser energy. All comparable devices have aiming beams and similar physical specifications. All devices share similar methods of control, safety features and performance monitoring. The device complies with the performance requirements of
Performance Standards: 21 CFR 1040.10 and 1040.11, with permissible deviations relative to Laser Notice 50, Dated July 26, 2001. The device has been tested by Notified Body and are CE mark and thereby complies with other international standards including IEC 60601 and 13485 Standards.
Clinical Performance: None
Non-Clinical Performance: Tests confirm that the power densities for treatment are identical in all devices insuring that all devices provide the exact same efficacy. Similar power density and safety controls insure equivalent or superior safety performance to the comparison device. Please see attached data summary.
None requested at this time. Additional Information:
The QuickLase™DUAL Dental Laser or The Conclusions: QuickLase™810/980 Dental Laser are, in terms of Safety and Efficacy, exactly the same as the TwinWave and TidalWave devices cleared in 510(K) number K072295 and
This document including ary attachments may confidential information, privileged material or otherwise constitute non-public information by contract or law. Any use of this document and it contains by anyone other than the intended pary is prohibited. If you have received this document in eror, please immediately return I to QuickLase Limited and delete any copies from your system. Use, dissemination, distribution or reproduction of the information contained herein by unintended persons is not authorized and may be unlawful.
{4}------------------------------------------------
## QuickLase Limited Traditional 510(K)-QuickLase™DUAL+ QuickLase™ 810/980
are only different in Brandname and certain cosmetic differences in the housings. Additionally, the QuickLase™DUAL Dental Laser or The QuickLase™810/980 Dental Laser are substantially equivalent to the listed devices without raising any issues of safety or efficacy. The device shares similar intended uses, and similar functional and performance characteristics. The device is designed to and its components have been evaluated and/or tested to comply with or exceed relevant federal safety and performance standards.
This document including any attachments may confidential information, privileged material or otherwise constitute non-public information by contract or law. Any use of this document and it contains by anyone other than the intended party is prohibited. If you have received this document in error, please imited and delete any copies from your system. Use, disemination, distribution or repoduction of the information contained herein by unintended persons is not authorized and may be vilaw[i].
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
MAR 1 6 2010
QuickLase Limited % QuickLase - Engineering and Regulatory Consulting Mr. Calvin Ostler P.O. Box 1264, 1094 West Greasewood Drive Riverton, Utah 84065
Re: K100474 . Trade/Device Name: QuickLase™DUAL+ Dental Laser QuickLase™ (810nm) Dental Lasers
QuickLase™ (980nm) Dental Lasers Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and Plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: February 27, 2010 Received: March 09, 2010
Dear Mr. Ostler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Calvin Ostler
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
folh
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
510(k) Number: K100474
Device Name: QuickLase™DUAL+ Dental Laser, QuickLase™ 810 Dental Lasers, QuickLase™ 980 Dental Lasers
Indications for Use: For the incision, excision, ablation, vaporization, hemostasis, and treatment of oral soft tissue.
| Examples: | |
|------------------------------------|-------------------------------------------------------------------------|
| Excisional and incisional biopsies | Frenectomy and Frenotomy |
| Exposure of unerupted teeth | Gingival Troughing for crown impressions |
| Fibroma removal | Gingivectomy |
| Gingivoplasty | Ginigival incision and excision |
| Hemostasis and coagulation | Implant Recovery |
| Incision and drainage of abscess | Leukoplakia |
| Operculectomy | Oral papillectomies |
| Pulpotomy | Pulpotomy as an adjunct to root canal therapy |
| Reduction of gingival hypertrophy | Reduction of bacterial level (decontamination) and inflammation |
| Soft Tissue crown lengthening | Treatment of aphthous ulcers |
| Vestibuloplasty | Lesion (tumor) removal |
| Laser Soft Tissue Curettage | Treatment of canker sores, herpetic and other ulcers of the oral mucosa |
| Tissue Retraction | |
Removal of Diseased, Inflamed and necrotic soft tissue within the periodontal pocket Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithlium
Sulcular debridement (removal of necrotic, diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices ingluding gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility Light activation of bleaching materials for teeth whitening
Laser-assisted whitening/bleaching of teeth
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Explus Division of Surgical, Orthopedic, and Restorative Devic Page 1 of 1 This document including any attachments may confidential information, privileged material or otherwise constitute non-publi
information by contract or law. Any use of this document and it contains by anyone other than the intended party is prohibiled. If you have received this document in error, please imited and delete any copies from your system. Use, dissenination distribution, or reproduction of the information contained herein by unintended persons is not authorized and may be unlawful
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.