MODIFIED STRAIGHT FIRE LASER FIBER

K100078 · Boston Scientific Corp · GEX · Feb 12, 2010 · General, Plastic Surgery

Device Facts

Record IDK100078
Device NameMODIFIED STRAIGHT FIRE LASER FIBER
ApplicantBoston Scientific Corp
Product CodeGEX · General, Plastic Surgery
Decision DateFeb 12, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Indications for Use

The Straight Fire Laser Fiber is designed for use with Ho:YAG and Nd:YAG lasers for indications that are cleared for these laser systems, including, but not limited to endoscopic, laparoscopic, and open surgical procedures involving vaporization, ablation, coagulation, hemostasis, excision, resection, incision of soft and cartilaginous tissue, and fragmentation of urinary and biliary calculi (Ho:YAG wavelength only). The fiber is designed for use with a standard SMA-905 connector that have been cleared for surgical use.

Device Story

Fiber optic laser energy delivery device; consists of silica core fiber jacketed with ETFE; features SMA 905 connector and polished, flat output tip. Used as integral component of cleared Ho:YAG and Nd:YAG laser systems (e.g., Lumenis VeraPulse). Operated by surgeons in endoscopic, laparoscopic, or open surgical settings. Device transmits laser energy to target tissue for surgical applications including vaporization, ablation, and lithotripsy. Output is controlled by the host laser system; fiber serves as passive delivery medium. Benefits include precise tissue interaction and calculus fragmentation during minimally invasive or open procedures.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Silica core fiber; ETFE jacket; SMA 905 connector; polished flat output tip. Passive energy delivery device for Ho:YAG and Nd:YAG lasers. No internal electronics or software.

Indications for Use

Indicated for patients requiring endoscopic, laparoscopic, or open surgical procedures involving vaporization, ablation, coagulation, hemostasis, excision, resection, or incision of soft and cartilaginous tissue, and fragmentation of urinary and biliary calculi (Ho:YAG only).

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 100078 # FEB 1 2 2010 Boston Scientific Corporation ្រុក - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - CONFIDENTIAL | 510(k) Summary for Modified Straight Fire Laser Fiber | | |--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | A. Sponsor | Boston Scientific Corporation<br>Urology and Women's Health Division<br>100 Boston Scientific Way<br>Marlborough, MA 01756 | | B. Contact | Janet A. McGrath<br>Specialist, Regulatory Affairs<br>508-683-4726<br>or<br>Donna Gardner<br>Director, Regulatory Affairs<br>508-683-4398 | | C. Device Name | | | Trade name: | To Be Determined<br>*Although the Trade Name is to be determined, for<br>purposes of this submission the proposed device<br>will be referred to as the Modified Straight Fire<br>Laser Fiber | | Common/usual name:<br>Classification Name: | Laser Instrument, Surgical, Powered<br>GEX - Laser surgical instrument for use in general<br>and plastic surgery and in dermatology<br>21 CFR 878.4810, Class II | | D. Predicate Device(s) | | | Trade name: | Straight Fire Holmium Laser Fiber | | Common/usual name: | Laser Instrument, Surgical, Powered | | Classification Name: | GEX - Laser surgical instrument for use in general<br>and plastic surgery and in dermatology<br>21 CFR 878.4810, Class II | | Premarket Notification: | Boston Scientific; K082928 | | Trade name: | SlimLineEZTM Laser Lithotripsy Fibers | | Common/usual name: | Laser Instrument, Surgical, Powered | | Classification Name: | GEX - Laser surgical instrument for use in general<br>and plastic surgery and in dermatology<br>21 CFR 878.4810, Class II | | Premarket Notification: | Lumenis; K011703 | | Traditional 510(k) | | | Modified Straight Fire Laser Fiber<br>February 1, 2010 | | P. 1 of 2 9 {1}------------------------------------------------ #### Boston Scientific Corporation ## E. Device Description The proposed modified Straight Fire Laser Fiber is a fiber optic laser energy delivery device consisting of a silica core fiber jacketed with ethelylene tetraflouroethylene (ETFE) and a SMA 905 connector. It is equipped with a polished, flat output tip. This fiber may be used in a variety of laser based surgical cases as an integral part of laser systems. For use with holmium (Ho:YAG) and neodymium (Nd:YAG) lasers with a standard SMA 905 connector that have been cleared for surgical use. Recommended lasers are Lumenis VeraPulse® PowerSuite™ Laser Systems. #### F. Intended Use The Straight Fire Laser Fiber is designed for use with Ho:YAG and Nd:YAG lasers for indications that are cleared for these laser systems, including, but not limited to endoscopic, laparoscopic, and open surgical procedures involving vaporization, ablation, coagulation, hemostasis, excision, resection, incision of soft and cartilaginous tissue, and fragmentation of urinary and biliary calculi (Ho:YAG wavelength only). The fiber is designed for use with a standard SMA-905 connector that have been cleared for surgical use. ## G. Technological Characteristics The proposed modified Straight Fire Laser Fiber has the same technological characteristics (i.e. SMA 905 connector, length of fiber connector type, fiber core, and strain relief) as the predicate device. #### H. Substantial Equivalence Utilizing FDA's Guidance for Industry and FDA Staff "Format for Traditional and Abbreviated 510(k)s" a direct comparison of key characteristics demonstrates that the proposed laser fiber is substantially equivalent to the predicate device in terms of intended use, technological characteristics, and performance characteristics. The proposed modified Straight Fire Laser Fiber is as safe, as effective, and performs as well as the predicate device. Traditional 510(k) Modified Straight Fire Laser Fiber February 1, 2010 10 p. 2/2 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 FEB 1 2 2010 Boston Scientific Corp. % Intertek Testing Services Mr. Daniel W. Lehtonen 2307 East Aurora Road Unit B7 Twinsburg, Ohio 44087 Re: K100078 Krood76 Trade/Device Name: Modified Straight Fire Laser Fiber Regulation Number: 21 CFR 878.4810 Regulation Number. 21 CFR 876.4616 Regulation Name: Laser surgerical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: February 2, 2010 Received: February 3, 2010 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Dear Mr. Lehtonen: We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section 310(t) premained is substantially equivalent (for the indications referenced above and have determined the device is substancements in interstat referenced above and have determined the devices marketed in interstate for use stated in the enclosure) to legally manced promises and Device Amendments, or to commerce prior to May 28, 1970, the chaculence with the provisions of the Federal Food. Drug, devices that have been reciassified in accordance who was and application (PMA). and Cosmetic Act (Act) that do not require approval sentrals provisions of the Act . The and Cosmetic Act (Act) that do not require approval controls provisions of the Act. The Act. The You may, therefore, market the device, stolect to the gentlations of the mained with and general controls provisions of the Act mender requirement as a sunst misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it in the states the started and main major resultstips affecting your d If your device is classified (see adver) micr ciass in (epicisms affecting your device can be may be subject to additional controls. Existing major regulation FDA may may be subject to additional controls. Existing major regarations on and one of the Sederal Register found in the Code of Federal Regulations, This 2017-11-2017 Publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issuance of a substance organies with other requirements of the Act that FDA has made a determination that your device and agencies. You must that FDA has made a determination that your device compiles Federal agencies. You must and the first of the states . . . : " :' : .' {3}------------------------------------------------ Page 2 - Mr. Daniel W. Lehtonen comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quant by over provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n 70d abblic upvw.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part note the for questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the r out may obtain only general reservers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, For Rh m Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Boston Scientific Corporation CONFIDENTIAL Indications for Use Statement K/000 510(k) Number (if Known): Device Name: Undetermined Indications For Use: The Straight Fire Laser Fiber is designed for use with Ho: Y AG and Nd: Y AG lasers for indications that are cleared for these laser systems, including, but not limited to endoscopic, laparoscopic, and open surgical procedures involving vaporization, ablation, coagulation, hemostasis, excision, resection, incision of soft and cartilaginous tissue, and fragmentation of urinary and biliary calculi (Ho:YAG wavelength only). The fiber is designed for use with a standard SMA-905 connector that have been cleared for surgical use. | Prescription Use | X | |------------------------|---| | (21 CFR 801 Subpart D) | | AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | | | | |------------------------------------------------------------------------------|----------------------------------------------------------------|---------------|---------| | | (Division Sign-Off) | | | | | Division of Surgical, Orthopedic,<br>and Restorative Devices. | | | | Traditional 510(k)<br>Modified Straight Fire Laser Fiber<br>February 1, 2010 | <table><tr><td>510(k) Number</td><td>1400078</td></tr></table> | 510(k) Number | 1400078 | | 510(k) Number | 1400078 | | | 7
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...