K091922 · Biolase Technology, Inc. · GEX · Jul 14, 2009 · General, Plastic Surgery
Device Facts
Record ID
K091922
Device Name
WATERLASE C100, MODEL 7100104
Applicant
Biolase Technology, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 14, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
This device may be used for the following indications (previously cleared under submissions K030523, K071363, and K081589): General Indications * - Class I, II, III, IV and V cavity preparation - Caries removal - Hard tissue surface roughening or etching - Enameloplasty, excavation of pits and fissures for placement of sealants * For use on adult and pediatric patients Root Canal Hard Tissue Indications - Tooth preparation to obtain access to root canal - Root canal preparation including enlargement - Root canal debridement and cleaning Root Canal Disinfection - Laser root canal disinfection after endodontic instrumentation Endodontic Surgery (Root Amputation) Indications - Flap preparation incision of soft tissue to prepare a flap and expose the bone. - Cutting bone to prepare a window access to the apex (apices) of the root(s). - Apicoectomy amputation of the root end. - Root end preparation for retrofill amalgam or composite. - Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues (i.e., granulation tissue) from around the apex NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation. Bone Surgical Indications - Cutting, shaving, contouring and resection of oral osseous tissues (bone) - Osteotomy Soft Tissue Indications including Pulpal Tissues* Incision, excision, vaporization, ablation and coagulation of oral soft tissues, including: - Excisional and incisional biopsies - Exposure of unerupted teeth - Fibroma removal - Flap preparation incision of soft tissue to prepare a flap and expose the bone. - Flap preparation incision of soft tissue to prepare a flap and expose unerupted teeth (hard and soft tissue impactions) - Frenectomy and frenotomy - Gingival troughing for crown impressions - Gingivectomy - Gingivoplasty - Gingival incision and excision - Hemostasis - Implant recovery - Incision and drainage of abscesses - Laser soft tissue curettage of the post-extraction tooth sockets and the periapical area during apical surgery - Leukoplakia - Operculectomy - Oral papillectomies - Pulpotomy - Pulp extirpation - Pulpotomy as an adjunct to root canal therapy - Root canal debridement and cleaning - Reduction of gingival hypertrophy - Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues (i.e., granulation tissue) from around the apex NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation. - Soft tissue crown lengthening - Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa - Vestibuloplasty * For use on adult and pediatric patient Laser Periodontal Procedures - Full thickness flap - Partial thickness flap - Split thickness flap - Laser soft tissue curettage - Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket - Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining junctional epithelium - Removal of granulation tissue from bony defects - Sulcular debridement (removal of diseased, inflamed or necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility) - Osteoplasty and osseous recontouring (removal of bone to correct osseous defects and create physiologic osseous contours) - Ostectomy (resection of bone to restore bony architecture, resection of bone for grafting, etc.) - Osseous crown lengthening
Device Story
Waterlase C100 is an Er,Cr:YSGG dental laser system. Device uses laser energy combined with water atomization spray to cut, remove, roughen, or etch hard tissues; soft tissue mode applies laser energy for incision, excision, or ablation with water spray for cooling/hydration. Flexible fiberoptic handpiece delivers laser energy; visible light pinpoints treatment area. Clinician adjusts power output, pulse energy, repetition rate, and air/water flow. Used in dental clinics by dentists for various oral procedures. Output allows precise tissue management, potentially reducing need for traditional mechanical tools and improving patient outcomes in periodontal and endodontic treatments.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and predicate comparison.
Indicated for adult and pediatric patients requiring dental hard and soft tissue procedures, including cavity preparation, caries removal, endodontic surgery, bone surgery, periodontal procedures, and soft tissue incision/excision/ablation.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Koq1922
# 510(k) SPECIAL Summary (As required by 21CFR807.92, 21CFR807.81(a)(3), FDA Memorandum #K97-1)
| Date Prepared: | June 25, 2009<br>JUL 1 4 2009 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Company: | Biolase Technology, Inc.<br>4 Cromwell<br>Irvine, CA 92618<br>Tel: (949) 361-1200<br>Fax: (949) 273-6680 |
| Contact: | Ms. Ioana M. Rizoiu<br>VP, Clinical R&D<br>Tel: (949) 226-8144<br>email: irizoiu@biolase.com |
| Trade Name: | Waterlase® C100 |
| Common Name: | Er,Cr:YSGG laser |
| Classification Name: | Surgical laser instrument |
| Classification Code: | 79 GEX, MXF, DZI a Class II device |
| Predicate Devices: | Waterlase®<br>Biolase Technology, Inc<br>K030523 (January 30, 2004), K071363 (February 12, 2008)<br>Waterlase® 3.0<br>Biolase Technology, Inc |
# DEVICE DESCRIPTION:
The Waterlase C100 dental laser system is a diverse device utilized to perform a variety of dental applications. For hard tissue procedures the Waterlase® C100 uses the Erbium, Chromium, Yttrium, Scandium, Gallium Gamet (Er,Cr.YSGG) laser in combination with advanced water atomization spray technology to cut, remove, roughen and etch tissues. Soft tissue procedures are performed using a different mode of operation where direct Er, Cr.YSGG laser energy is applied to incise, excise or ablate these tissues. In soft tissue procedures the water spray is applied for hydration, cooling or to keep tissues clean. For hard tissue applications the spray is part of the tissue removing process as well as hydration, cooling and keeping tissues clean.
A flexible fiberoptic handpiece delivers the Waterlase® C100 laser energy. A visible light emitted from the handpiece distal end pinpoints the area of treatment. In both hard and soft tissue applications the power output, pulse energy, repetition rate (frequency) and air and water flow rates are adjustable to specific user requirements.
{1}------------------------------------------------
## INDICATIONS FOR USE:
This device may be used for the following indications (previously cleared under submissions K030523, K071363, and K081589):
# General Indications *
- r Class I, II, III, IV and V cavity preparation
- Caries removal
- Hard tissue surface roughening or etching
- Enameloplasty, excavation of pits and fissures for placement of sealants
- * For use on adult and pediatric patients
# Root Canal Hard Tissue Indications
- 이 Tooth preparation to obtain access to root canal
- Root canal preparation including enlargement 트
- l Root canal debridement and cleaning Root Canal Disinfection
- 트 Laser root canal disinfection after endodontic instrumentation Endodontic Surgery (Root Amputation) Indications
- Flap preparation incision of soft tissue to prepare a flap and expose the bone. 트
- Cutting bone to prepare a window access to the apex (apices) of the root(s). l
- Apicoectomy amputation of the root end. 트
- Root end preparation for retrofill amalgam or composite. 트
- Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic 트 tissues (i.e., granulation tissue) from around the apex
NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.
# Bone Surgical Indications
- Cutting, shaving, contouring and resection of oral osseous tissues (bone) ■
- . Osteotomy
### Soft Tissue Indications including Pulpal Tissues*
Incision, excision, vaporization, ablation and coagulation of oral soft tissues, including:
- 트 Excisional and incisional biopsies
- I Exposure of unerupted teeth
- I Fibroma removal
- Flap preparation incision of soft tissue to prepare a flap and expose the bone. ■
- Flap preparation incision of soft tissue to prepare a flap and expose unerupted teeth . (hard and soft tissue impactions)
- I Frenectomy and frenotomy
- Gingival troughing for crown impressions I
- 미 Gingivectomy
- I Gingivoplasty
- I Gingival incision and excision
- 트 Hemostasis
- I Implant recovery
{2}------------------------------------------------
- .. Incision and drainage of abscesses
- Laser soft tissue curettage of the post-extraction tooth sockets and the periapical 프 area during apical surgery
- 트 Leukoplakia
- Operculectomy 바
- 트 Oral papillectomies
- l Operculectomy
- Oral papillectomies 트
- 트 Pulpotomy
- Pulp extirpation ■
- 트 Pulpotomy as an adjunct to root canal therapy
- Root canal debridement and cleaning 트
- 트 Reduction of gingival hypertrophy
- Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic 트 tissues (i.e., granulation tissue) from around the apex
NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.
- Soft tissue crown lengthening ■
- Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa B
- 트 Vestibuloplasty
- * For use on adult and pediatric patient Laser Periodontal Procedures
- s Full thickness flap
- l Partial thickness flap
- 트 Split thickness flap
- 트 Laser soft tissue curettage
- 트 Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket
- Removal of highly inflamed edematous tissue affected by bacteria penetration of ■ the pocket lining junctional epithelium
- Removal of granulation tissue from bony defects ■
- Sulcular debridement (removal of diseased, inflamed or necrosed soft I tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
- Osteoplasty and osseous recontouring (removal of bone to correct osseous defects . and create physiologic osseous contours)
- Ostectomy (resection of bone to restore bony architecture, resection of bone for 프 grafting, etc.)
- . Osseous crown lengthening
#### CONTRAINDICATIONS:
All clinical procedures performed with the Waterlase® C100 must be subject to the same clinical judgment and care as with traditional techniques. Patient risk must always be considered and fully understood before clinical treatment. The clinician must completely understand the patient's medical history prior to treatment. Exercise caution for general
{3}------------------------------------------------
medical conditions, which might contraindicate a local procedure. Such conditions may include, but are not limited to, allergy to local or topical anesthetics, heart disease, bleeding disorders, sleep apnea or an immune system deficiency. Medical clearance from the patient's physician is advisable when doubt exists regarding treatment.
#### CONCLUSION:
The indications included herein are the same as indications that have been previously cleared by the FDA for the predicate device. Substantial equivalency for the Waterlase C100 has been determined through comparison to previously cleared devices.
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight. The symbol is composed of several thick, curved lines that create the shape of the bird's body and wings.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Biolase Technology, Inc. % Ms. Ioana M. Rizoiu VP, Clinical R & D 4 Cromwell Irvine, California 92618
JUL 1 4 2009
Re: K091922
Trade/Device Name: Waterclase® C100 Regulation Number: 21 CFR 878.4810
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: II
Product Code: GEX Dated: June 25, 2009 Received: June 30, 2009
Dear Mr. Rizoiu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
# Page 2 - Ms. Ioana M. Rizoiu
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801); please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Sincerely yours,
Mark N.Melkerson
Director
Division of Surgical, Orthopedic
Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
# 510(k) Number (if known):
# Device Name: Waterlase® C100
# Indications for Use:
# General Indications *
- 지 Class I, II, III, IV and V cavity preparation
- I Caries removal
- 트 Hard tissue surface roughening or etching
- 트 Enameloplasty, excavation of pits and fissures for placement of sealants
* For use on adult and pediatric patients
## Root Canal Hard Tissue Indications
- Tooth preparation to obtain access to root canal ■
- Root canal preparation including enlargement 트
- Root canal debridement and cleaning ■
### Root Canal Disinfection
- I Laser root canal disinfection after endodontic instrumentation
## Endodontic Surgery (Root Amputation) Indications
- li Flap preparation - incision of soft tissue to prepare a flap and expose the bone.
- Cutting bone to prepare a window access to the apex (apices) of the root(s). I
- 내 Apicoectomy - amputation of the root end.
- 보 Root end preparation for retrofill amalgam or composite.
- Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues (i.e., granulation . tissue) from around the apex
NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation
## Bone Surgical Indications
- Cutting, shaving, contouring and resection of oral osseous tissues (bone)
- Osteotomy
sseous tissues (bone)
Neil R. Oden for m
(Division Sign-Off Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K091922
{7}------------------------------------------------
## Laser Periodontal Procedures
- Full thickness flap 트
- Partial thickness flap 트
- 트 Split thickness flap
- Laser soft tissue curettage 트
- Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal l pocket
- Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining 피 and junctional epithelium
- Removal of granulation tissue from bony defects 트
- Sulcular debridement (removal of diseased, inflamed or necrosed soft tissue in the ■ periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
- . Osteoplasty and osseous recontouring (removal of bone to correct osseous defects and create physiologic osseous contours)
- Ostectomy (resection of bone to restore bony architecture, resection of bone for grafting, etc.) 트
- 미 Osseous crown lengthening
### Soft Tissue Indications including Pulpal Tissues*
Incision, excision, vaporization, ablation and coagulation of oral soft tissues, including:
- Excisional and incisional biopsies I
- I Exposure of unerupted teeth
- E Fibroma removal
- I Flap preparation - incision of soft tissue to prepare a flap and expose the bone.
- Flap preparation incision of soft tissue to prepare a flap and expose unerupted teeth (hard and soft tissue 치 impactions)
- l Frenectomy and frenotomy
- Gingival troughing for crown impressions 1
- I Gingivectomy
- 1 Gingivoplasty
- Gingival incision and excision i
- 1 Hemostasis
- Implant recovery 프
- Incision and drainage of abscesses 트
- Laser soft tissue curettage of the post-extraction tooth sockets and the periapical area during 트 apical surgery
- Leukoplakia . B
- Operculectomy 비
- 트 Operculectomy
Neil R.P. Oside by mka
(Division Sign-Off)
Division of Surgical. Orthopedic. and Restorative Devices
510(k) Number K091922
{8}------------------------------------------------
- Oral papillectomies
- Pulpotomy
- Pulp extirpation I
- Pulpotomy as an adjunct to root canal therapy I
- Root canal debridement and cleaning K
- I Reduction of gingival hypertrophy
- Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues I (i.e., granulation tissue) from around the apex
NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) nust be submitted to a qualified laboratory for histopathological evaluation.
- Soft tissue crown lengthening 트
- Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa 트
- 트 Vestibuloplasty
- * For use on adult and pediatric patient
Prescription Use _ X Use
AND/OR
Over-the-Counter
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
MulRedde Forman
Sign-Off Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K091922
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.