K091320 · Lutronic Corporation · GEX · Aug 19, 2009 · General, Plastic Surgery
Device Facts
Record ID
K091320
Device Name
SPECTRA SP II, SPECTRA DENTA II
Applicant
Lutronic Corporation
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 19, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Spectra DENTA II Laser System is indicated for use in soft tissue dental indications including periodontic procedures such as, but not limited to, removal of diseased or inflamed soft tissue in the periodontal pocket (sulcular debridement), vaporization, gingivectomy-removal of hyperplasias, gingivoplasty, papillectomy, vestibuloplasty, fibroma (nonmalignant tumor, mucosa, tongue), epulis, incision and excision, removal of soft tissue, cysts, and tumors, and laser assisted new attachment procedure (cementummediated periodontal ligament new attachment to the root surface in the absence of long junctional epithelium); Oral surgery such as frenectomy, frenum release, drainage (abscess), flap surgery, incisional and excisional biopsy, incision and excision of aphthous ulcers, incision of infection when used with antibiotic therapy, excision and ablation of benign and malignant lesions, oral cavity tumors and hemangiomas, salivary gland pathologies, preprosthetic gum preparation, leukoplakia; partial glossectomy, periodontal gum resection, homeostasis, operculectomy, and crown lengthening. The Spectra SP II Laser System is indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: dermatology, plastic surgery, podiatry, neurosurgery, gynecology, otorhinolaryngology (ENT), arthroscopy, (knee), and open endoscopic general surgery.
Device Story
Spectra DENTA II and Spectra SP II are CO2 laser systems (10.6um) for soft tissue surgery. System comprises console, articulated arm, footswitch, goggles, and interchangeable handpieces. Console houses DC-excited sealed-off CO2 tube, cooling system, power supply, and microcontroller. Operation modes include Continuous Wave, Char-Free, and Super Pulse. Used in clinical settings by physicians/dentists. Laser beam exits articulated arm into handpiece; final focus lens determines spot size at treatment plane. Output is infrared light beam for tissue ablation, vaporization, excision, incision, or coagulation. Clinical benefit: precise soft tissue management across multiple surgical specialties.
Clinical Evidence
No clinical data presented; bench testing only.
Technological Characteristics
CO2 laser (10.6um); DC-excited sealed-off gas tube; articulated arm delivery; Continuous Wave, Char-Free, and Super Pulse modes; console-based with microcontroller; interchangeable handpieces; Class II device.
Indications for Use
Indicated for soft tissue surgical procedures (ablation, vaporization, excision, incision, coagulation) in dental, dermatological, plastic, podiatric, neurosurgical, gynecological, ENT, and general surgical specialties. Patient population includes adults and children requiring soft tissue management for conditions ranging from benign/malignant lesions and tumors to periodontal disease, skin resurfacing, and various anatomical obstructions/pathologies.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Lutronic Corporation Spectra SP Laser System (K050254)
Lumenis Family of UltraPulse SurgiTouch CO2 Laser Systems (K030147)
Cynosure Smart US 20 D / UltraSpeed, and Smart Clinic Laser (K081181)
Submission Summary (Full Text)
{0}------------------------------------------------
91320 page (1) of (3)
# 510(k) Summary for the Lutronic Corporation Spectra DENTA II/Spectra SP II Laser Systems
This 510(k) Summary is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.
### 1. General Information
# AUG 1 9 2009
| Submitter: | Lutronic Corporation<br>#403-2,3,4, Ilsan Technotown<br>1141-1 Baeksok-Dong, Ilsan-Gu<br>Goyang-Si, Gyeonggi-Do, 410-722<br>Republic of Korea |
|------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
|------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
Maureen O'Connell O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, MA 01864 Telephone: 978-207-1245 Fax: 978-824-2541
Summary Preparation Date:
2. Names
Device Name:
Contact Person:
Classification Name:
Spectra DENTA II Laser System Spectra SP II Laser System
April 29, 2009
Laser Instrument, Surgical, Powered Product Code: GEX Panel: General & Plastic Surgery
### 3. Predicate Devices
The Spectra DENTA II and Spectra SP II Laser Systems are substantially equivalent to a combination of the Lutronic Corporation Spectra SP Laser System (K050254), the Lumenis Family of UltraPulse SurgiTouch CO2 Laser Systems (K030147), the Lumenis Compact 30C CO2 Lasers (K935563 and K963229), the MedArt Uni-Laser 450P CQ2 Laser System & Accessories (K991297), the OpusDent Ltd. Opus Duo EC (K021508) and the Cynosure Smart US 20 D / UltraSpeed, and Smart Clinic Laser (K081181).
### 4. Device Description
The Spectra DENTA II and Spectra SP II Laser Systems are identical CO2 laser systems supplied with different handpieces depending upon the indications for use. They consist of a self-contained console, an articulated arm delivery system, a footswitch and a pair of goggles. The Spectra DENTA II and Spectra SP II Laser Systems produce a beam of coherent infrared (10.6um) light and have Continuous Wave, Char-Free and Super Pulse operation modes.
{1}------------------------------------------------
K091320 Page 0 of 3
The main console is the heart of the Spectra DENTA II and Spectra SP II Laser Systems and contain the optical system (with DC-Excited Sealed-off Carbon Dioxide gas tube), cooling system, arm mount, micro-controller, and power supply. The main console also includes a key switch used to turn the power on and off, an emergency stop push button that quickly de-energizes the system in emergency situations, and the control panel. There are 4 casters in the console base for moving the system.
The laser beam exits an articulated arm into a handpiece where it is focused by the final focus lens contained in the handpiece to produce a spot size at the treatment focal plane. The handpiece components, when inserted in the beam delivery system, change the laser beam characteristics. Different handpieces are provided with the Spectra DENTA II and the Spectra SP II depending upon the desired indications for use.
### 5. Indications for Use
The Spectra DENTA II Laser System is indicated for use in soft tissue dental indications including periodontic procedures such as, but not limited to, removal of diseased or inflamed soft tissue in the periodontal pocket (sulcular debridement), vaporization, gingivectomy-removal of hyperplasias, gingivoplasty, papillectomy, vestibuloplasty, fibroma (nonmalignant tumor, mucosa, tongue), epulis, incision and excision, removal of soft tissue, cysts, and tumors, and laser assisted new attachment procedure (cementummediated periodontal ligament new attachment to the root surface in the absence of long junctional epithelium); Oral surgery such as frenectomy, frenum release, drainage (abscess), flap surgery, incisional and excisional biopsy, incision and excision of aphthous ulcers, incision of infection when used with antibiotic therapy, excision and ablation of benign and malignant lesions, oral cavity tumors and hemangiomas, salivary gland pathologies, preprosthetic gum preparation, leukoplakia; partial glossectomy, periodontal gum resection, homeostasis, operculectomy, and crown lengthening.
The Spectra SP II Laser System is indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: dermatology, plastic surgery, podiatry, neurosurgery, gynecology, otorhinolaryngology (ENT), arthroscopy, (knee), and open endoscopic general surgery.
#### Dermatology & Plastic Surgerv
The ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:
- . Laser skin resurfacing
- . Treatment of wrinkles, rhytids and furrows
- . Ablation and/or vaporization of soft tissue in dermatology and plastic surgery for the reduction, removal, and/or treatment of actinic keratosis, skin tags, solar/actinic elastosis, actinic cheilitis, lentigines, uneven pigmentation/dyschromia, acne scars, surgical scars, keloids, hemangiomas (including buccal hemangiomas), tattoos, telangiectasia, squamous and basal cell carcinoma, spider and epidermal naevi. xanthelasma palpebrarum, syringoma, and verrucae and seborrhoecae vulgares (warts); laser derm-ablation; and laser burn debridement.
### Dermatology, Plastic Surgery & General Surgery
Laser incision and/or excision of soft tissue in dermatology, plastic and general surgery, including the performance of blepharoplasty and for the creation of recipient sites for
{2}------------------------------------------------
K091320 page 3 of 3
hair transplantation, treatment of hemorrhoids, atheroma, cysts, abscesses, and all other soft tissue applications.
#### Podiatry
Laser ablation, vaporization, and/or excision of soft tissue in podiatry for the reduction, removal, and/or treatment of verrucae vulgares, and matrixectomy.
#### Otorhinolaryngology (ENT)
Laser incision, excision, ablation and/or vaporization of soft tissue in otorhinolaryngology for the treatment of choanal atresia, leukoplakia of larynx, nasal obstruction, UPP, rhinophyma, adult and juvenile papillomatosis polyps, rhinophyma and verrucae vulgares.
# Gynecology
Laser incision, excision, ablation and/or vaporization of soft tissue in gynecology for the treatment of cervical intraepithelial neoplasia, condyloma acuminata, leukoplakia (vulvar dystrophies) and vulvar and vaginal intraepithelial neoplasia.
#### Neurosurgery
Laser incision, excision, ablation and/or vaporization of soft tissue in neurology for the treatment of basal tumor-meningioma, posterior fossa tumors, peripheral neurectomy, and lipomas/large tumors.
6. Performance Data None presented.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with outstretched wings. The bird is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular pattern around the bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Lutronic Corporation % O'Connell Regulatory Consultants Ms. Maureen O'Connell Regulatory Consultant 5 Timber Lane North Reading, Massachusetts 01864
AUG 1 9 2009
Re: K091320
Trade/Device Name: Spectra DENTRA II Laser System Spectra SP II Laser System
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: Class II ---Product Code: GEX Dated: July 3, 2009 Received: July 6, 2009
Dear Ms. O'Connell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
# Page 2 - Ms. Maureen O'Connell
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free momber (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
For
Mark N. Melkerson
Dep Dir
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
| | Indications for Use |
|---------------------------|---------------------|
| 510(k) Number (if known): | K091320 |
| | |
Device Name: Spectra DENTA II Laser System
### Indications for Use:
The Spectra DENTA II Laser System is indicated for use in soft tissue dental indications including periodontic procedures such as, but not limited to, removal of diseased or inflamed soft tissue in the periodontal pocket (sulcular debridement), waporization, gingivectomy-removal of hyperplasias, gingivoplasty, papillectomy, vestibuloplasty, fibroma (nonmalignant trimor, mucosa, tongue), epulis, incision and excision, removal of soft tissue, cysts, and tumors, and laser assisted new attachment procedure (cementum-mediated periodontal ligament new attachment to the root surface in the absence of long junctional epithelium); Oral surgery such as frenectomy, frenum release, drainage (abscess), flap surgery, incisional and excisional biopsy, incision and excision of aphthous ulcers, incision of infection when used with antibiotic therapy, excision and ablation of benign and malignant lesions, oral cavity turnors and hemangiomas, saliyeny gland pathologies, preprosthetic gum preparation, leukoplakia; partial glossectomy, periodontal gum resection, homeostasis, operculectomy, and crown lengthening.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
X Prescription Use (Per 21 CFR 801.109) 96)
OR
Page 1 of 3
Over The Counter Use (Optional Format 1-2-
forman
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
510(k) Number K091320 17
{6}------------------------------------------------
| Indications for Use | |
|---------------------------|----------------------------|
| 510(k) Number (if known): | KU9132J |
| Device Name: | Spectra SP II Laser System |
Indications for Use:
The Spectra SP II Laser System is indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: dermatology, plastic surgery, podiatry, neurosurgery, gynecology, otorhinolaryngology (ENT), arthroscopy, (knee), and open endoscopic general surgery.
### Dermatology & Plastic Surgery
The ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:
- . Laser skin resurfacing
- . Treatment of wrinkles, rhytids and furrows
- Ablation and/or vaporization of soft tissue in dermatology and plastic surgery for the . reduction, removal, and/or treatment of actinic keratosis, skin tags, solar/actinic elastosis, actinic cheilitis, lentigines, uneven pigmentation/dyschromia, acne scars, surgical scars, keloids, hemangiomas (including buccal hemangiomas), tattoos, telangiectasia, squamous and basal cell carcinoma, spider and epidermal naevi, xanthelasma palpebrarum. syringoma, and verrucae and seborrhoecae vulgares (warts); laser derm-ablation; and laser burn debridement.
### Dermatology, Plastic Surgery & General Surgery
Laser incision and/or excision of soft tissue in dermatology, plastic and general surgery, including the performance of blepharoplasty and for the creation of recipient sites for hair transplantation. treatment of hemorrhoids, atheroma, cysts, abscesses, and all other soft tissue applications.
#### Podiatry
510(k) Number
Laser ablation, vaporization, and/or excision of soft tissue in podiatry for the reduction, removal, and/or treatment of verrucae vulgares, and matrixectomy.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| X Prescription Use (Per 21 CFR 801.109) | |
|-----------------------------------------|--|
|-----------------------------------------|--|
*Nul Re Ogle for mxm*
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
OR
| Over The Counter Use (Optional Format 1-2-) | |
|---------------------------------------------|--|
|---------------------------------------------|--|
Page 2 of 3
{7}------------------------------------------------
091320
### Otorhinolaryngology (ENT)
Laser incision, excision, ablation and/or vaporization of soft tissue in otorhinolaryngology for the treatment of choanal atresia, leukoplakia of larynx, nasal obstruction, UPP, rhinophyma, adult and juvenile papillomatosis polyps, rhinophyma and verrucae vulgares.
### Gynecology
Laser incision, excision, ablation and/or vaporization of soft tissue in gynecology for the treatment of cervical intraepithelial neoplasia, condyloma acuminata, leukoplakia (vulvar dystrophies) and vulvar and vaginal intraepithelial neoplasia.
#### Neurosurgery
Laser incision, excision, ablation and/or vaporization of soft tissue in neurology for the treatment of basal tumor-meningioma, posterior fossa tumors, peripheral neurectomy, and lipomas/large tumors.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
X Prescription Use (Per 21 CFR 801.109) તેર)
OR
Over The Counter Use (Optional Format 1-2-
96)
Nikola Tesla
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
Page 3 of 3
510(k) Number K091320 19
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.