MODIFICATION TO CUTERA ER:YSGG LASER HANDPIECE

K080530 · Cutera, Inc. · GEX · Aug 15, 2008 · General, Plastic Surgery

Device Facts

Record IDK080530
Device NameMODIFICATION TO CUTERA ER:YSGG LASER HANDPIECE
ApplicantCutera, Inc.
Product CodeGEX · General, Plastic Surgery
Decision DateAug 15, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Cutera Er: YSGG laser hand piece is designed for use in applications requiring the excision, incision, ablation, vaporization and coagulation of soft tissue. Dermatology and Plastic Surgery Indications include: treatment of wrinkles and skin resurfacing. Opthalmology Indications include: Incision, excision, vaporization and coagulation of tissue surrounding the eye and orbit. The Cutera Fractionated Er:YSGG laser hand piece is designed for use in dermatology for skin resurfacing and for coagulation.

Device Story

Cutera Er:YSGG laser handpiece is an optional accessory for Xeo and Solera Opus laser systems. Device emits 2790nm laser energy; water-cooled laser integrated into handpiece; utilizes computer-controlled scanner. Used in dermatology, plastic surgery, and ophthalmology clinics by physicians. Provides precise tissue interaction for resurfacing, coagulation, and excision. Benefits include controlled soft tissue management and aesthetic skin improvement.

Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and design similarity to predicate.

Technological Characteristics

2790nm Er:YSGG laser; water-cooled; computer-controlled scanner; handpiece-integrated laser source; Class II device.

Indications for Use

Indicated for soft tissue excision, incision, ablation, vaporization, and coagulation in dermatology, plastic surgery (wrinkle treatment, skin resurfacing), and ophthalmology (tissue surrounding eye and orbit). Fractionated version indicated for dermatology skin resurfacing and coagulation.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image contains handwritten text that appears to read "K esc 530". The text is written in a casual, cursive style, with varying stroke thicknesses. The characters are spaced somewhat unevenly, and the overall impression is that of a quick, handwritten note. ## Attachment 10 510(k) Summary for the Cutera Er:YSGG Laser System ### I. General Information AUG 1 5 2008 | Submitter: | Cutera, Inc.<br>3240 Bayshore Blvd<br>Brisbane, CA 94005 | |--------------------|----------------------------------------------------------| | Contact Person: | Kathy Maynor | | Telephone:<br>Fax: | 415-657-5586<br>415-330-2443 | Summary Preparation Date: February 23, 2008 ### II. Names | Device Proprietary Name: | Cutera Er:YSGG Laser Handpiece | |--------------------------|---------------------------------| | Classification Name: | Instrument, Powered, Laser, GEX | | Common Name: | Dermatology Laser | ### III. Predicate Devices - . K063867 Cutera Er:YSGG Laser Hand Piece #### IV. Product Description/Technological Characteristics The Cutera Er:YSGG Laser handpiece is an optional handpiece for the currently marketed Xeo and Solera Opus laser systems. The handpiece emits laser energy at a wavelength of 2790nm. The water cooled laser is located in the handpiece and utilizes a computer controlled scanner. ### V. Statement of Intended Use ## Cutera Er: YSGG Laser Hand Piece (K063867 - 4-8 mm spot size) The Cutera Er:YSGG Laser System is designed for use in applications requiring the excision, incision, ablation, vaporization and coagulation of soft tissue. For {2}------------------------------------------------ Dermatology and Plastic Surgery, indications include: treatment of wrinkles and skin resurfacing. For Opthalmology, indications include: Incision, excision, vaporization and coagulation of tissue surrounding the cyc and orbit. # Cutera Er: YSGG Laser Hand Piece (K080530) - Fractionated/Small Spot Size The Cutera Fractionated Er:YSGG laser hand piece is designed for use in dermatology for skin resurfacing and for coagulation. ### Rationale for Substantial Equivalence VI. The Cutera Er:YSGG Handpiece shares the same general indications for use as the currently marketed predicate devices, and does not raise any issues with safety and effectiveness. The Cutera Er:YSGG Handpiece is therefore substantially equivalent to the currently marketed predicate devices. ### VII. Safety and Effectiveness Information Technologically, the Cutera Er:YSGG Handpiece is substantially equivalent to the listed predicate devices. Therefore the risks and benefits for the Cutera Er:YSGG Handpiece are comparable to the predicate devices. Cutera therefore believes that there are no new questions of safety or effectiveness raised by the introduction of this device. ### VIII. Conclusion The Cutera Er: YSGG Handpiece was found to be substantially equivalent to currently marketed devices. The Cutera Er:YSGG shares similar indications for use, design features, and similar functional features as the currently marketed predicate devices. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Cutera, Inc. % Ms. Kathy Maynor VP of Regulatory/Quality 3240 Bayshore Boulevard Brisbane, California 94005 AUG 1 5 2008 Re: K080530 Trade/Device Name: Cutera Er: YSGG Laser Hand Piece Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: July 7, 2008 Received: July 17, 2008 Dear Ms. Maynor: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced a bove and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 -- Ms. Kathy Maynor forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desirc specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark M. Wilkerson Mark N. Melkerson Director Division of General. Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### Attachment 2 Indications For Use Statement K 080530 510(k) Number (if Known): Device Name: Cutera Er:YSGG Laser Hand Piece Indications for Use (K063867) – Er:YSGG Laser Hand Piece - 4-8 mm spot size: The Cutera Er: YSGG laser hand piece is designed for use in applications requiring the excision, incision, ablation, vaporization and coagulation of soft tissue. Dermatology and Plastic Surgery Indications include: treatment of wrinkles and skin resurfacing Opthalmology Indications include: Incision, excision, vaporization and coagulation of tissue surrounding the eye and orbit. Indications for Use (K080530) -Er: YSGG Laser Hand Piece - Fractionated/Small spot size The Cutera Fractionated Er:YSGG laser hand piece is designed for use in dermatology for skin resurfacing and for coagulation. Prescription Use __ V OR Over-The-Counter Use (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of G vision Sign-Cit) Division of General. Restorative, and Neurological Devices 510(k) Number K080530
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...