K072023 · Ellipse A/S · GEX · Sep 12, 2007 · General, Plastic Surgery
Device Facts
Record ID
K072023
Device Name
ELLIPSE JUVIA, MODEL 9EJU7465
Applicant
Ellipse A/S
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 12, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Ellipse Juvia is intended for use in dermatology and plastic surgery for treatment of: - Skin Resurfacing - Wrinkles, Rhytides, and Furrows - Acne Scars
Device Story
The Ellipse Juvia is a CO2 laser system (10,600 nm) used in dermatology and plastic surgery for skin resurfacing and scar/wrinkle treatment. The system consists of a laser console, an optical fiber beam delivery system, and a scanner. The physician operates the device via a foot switch to deliver pulsed laser energy to the skin. The scanner manipulates the beam into pre-specified patterns to create light spots on the skin. By delivering controlled thermal energy, the device facilitates skin resurfacing, which can improve the appearance of wrinkles, rhytides, furrows, and acne scars. The device is intended for professional use in a clinical setting.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and comparison of technical specifications (wavelength, power, spot size, scanning parameters) against predicate devices.
Technological Characteristics
CO2 laser system; 10,600 nm wavelength; 0.11-20W power output; 635nm aiming beam (max 5mW); fiber-based beam delivery; scanner-based spot pattern delivery; foot switch activation; Class II device.
Indications for Use
Indicated for skin resurfacing, treatment of wrinkles, rhytides, furrows, and acne scars in dermatology and plastic surgery patients.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Lumenis UltraPulse Encore Carbon Dioxide Surgical Laser and Delivery Device (K022060)
SLIM Evolution Family of CO2 Lasers and Delivery Device Accessories (K063001)
Submission Summary (Full Text)
{0}------------------------------------------------
K072023
## Ellipse A/S
SEP 1 2 2007
# Ellipse Juvia laser system
This 510(k) summary is submitted in accordance with the requirements of 21 CFR Part 807.87(h) and Part 807.92.
#### A. Contact information and device identification:
| Date of the summary: | 20 July 2007 |
|-----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted by/manufacturer: | Ellipse A/S<br>Agern Alle 11<br>2970 Hoersholm, Denmark<br>Tel: + 45 4576 8808<br>Fax: + 45 4517 6851 |
| Contact person: | Ole Kofod |
| Device Trade Name: | Ellipse Juvia. |
| Device Model number: | 9EJU7465. |
| Common Name: | Laser treatment system. |
| Classification name: | Laser surgical instrument for use in general and plastic surgery and in<br>dermatology (per 21 CFR Part 878.4810). |
| Device classification: | Class II. |
| Product code: | GEX |
| Predicate devices legally<br>marketed to which Ellipse<br>A/S claims equivalence: | • Lumenis UltraPulse Encore Carbon Dioxide Surgical Laser and<br>Delivery Device (K022060) manufactured by Lumenis, Inc., Santa<br>Clara, CA, USA.<br>(Laser surgical instrument for use in general and plastic surgery<br>and in dermatology (per 21 CFR Part 878.4810)).<br>• SLIM Evolution Family of CO2 Lasers and Delivery Device Ac-<br>cessories (K063001) manufactured by Lasering S.r.l, Modena, It-<br>aly.<br>(Laser surgical instrument for use in general and plastic surgery<br>and in dermatology (per 21 CFR Part 878.4810)). |
#### B. Description of Ellipse Juvia system:
The Ellipse Juvia system comprises the following major parts:
- A laser console containing a CO2 laser module capable of providing a laser beam having a . wavelength of 10,600 nm.
- A scanner that is intended to manipulate and place a pulsed beam received from the laser . console in a pre-specified pattern on the skin being treated.
- An optical fiber providing a beam path from the laser to the scanner. .
{1}------------------------------------------------
## C. Intended Use of Ellipse Juvia system:
Ellipse Juvia is intended for use in dermatology and plastic surgery for treatment of:
- . Skin Resurfacing
- · Wrinkles, Rhytides, and Furrows
- Acne Scars
# D. Comparison of Ellipse Juvia to predicate devices:
| Issue/data com-<br>pared | Ellipse Juvia | SLIM<br>(Lasering S.r.l) | Ultrapulse Encore<br>(Lumenis) |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA clearance /<br>status | Being submitted (this<br>submission) | K063001 | K022060 |
| Indications | Skin resurfacing,<br>wrinkles, rhytids, and<br>furrows,<br>acne scars | Skin resurfacing,<br>treatment of furrows and<br>wrinkles,<br>acne scars,<br>and others | Skin resurfacing,<br>wrinkles, rhytids, and fur-<br>rows,<br>acne scars,<br>and others |
| Technology | The system comprises:<br>a) A laser console con-<br>taining a CO2 laser<br>module<br>b) A scanner for pro-<br>ducing a pattern of<br>light spots on the skin<br>c) a beam delivering<br>system connecting the<br>laser console and the<br>scanner. | The system comprises:<br>a) A laser console contain-<br>ing a CO2 laser module<br>b) A scanner for producing<br>a pattern of light spots on<br>the skin<br>c) a beam delivering sys-<br>tem connecting the laser<br>console and the scanner. | The system comprises:<br>a) A laser console contain-<br>ing a CO2 laser module<br>b) A scanner for producing<br>a pattern of light spots on<br>the skin<br>c) a beam delivering sys-<br>tem connecting the laser<br>console and the scanner. |
| Length of beam<br>delivering system | 165cm | 130cm | 150 / 180 cm depending on<br>model |
| Type of beam de-<br>livering system | Fiber providing full<br>freedom of movement | Articulated arm with 340°<br>of freedom | Articulated arm with 360°<br>of freedom |
| Wavelength | 10,600nm | 10,600nm | 10,600nm |
| Max power | 0.11-20W | 0.1-15W, 0.2-30W, 1-50W<br>depending on the model | 1-60W |
| Minimum scanner<br>spot size | Ø300µm | Ø400µm | Ø1300µm |
| Max power den-<br>sity (computed as<br>max power di-<br>vided by mini-<br>mum scanner spot<br>size) | 20W / Ø300µm =<br>28.6kW/cm2 | 15W / Ø400µm =<br>15.0kW/cm2<br>30W / Ø400µm =<br>30.0kW/cm2<br>50W / Ø400µm =<br>50kW/cm2 | 60W/ Ø1300µm =<br>6kW/cm2 |
| Aiming beam | 635nm, max 5mW | 635nm, max. 2mW | 635nm, max 5mW |
| Scanning speed<br>(light spots on the<br>skin per second) | 0.3-100 Hz | 1-10,000 Hz | 1-1,000 Hz |
| Time for a full<br>scan | In the range of 1 sec –<br>Actual time is depend-<br>ing on scan pattern<br>chosen | In the range of 1 sec – Ac-<br>tual time is depending on<br>scan pattern chosen | In the range of 1 sec – Ac-<br>tual time is depending on<br>scan pattern chosen |
| Beam activation | Foot switch | Foot switch | Foot switch |
{2}------------------------------------------------
### Conclusion:
Ellipse Juvia applications and indications are evaluated to be within the scope of the previously cleared devices. The same counts for the essential treatment parameters, the protective conditions for the skin during treatment, and the working conditions of the physician.
Based on this side-by-side comparison of the overall performance characteristics of the predicate devices under consideration Ellipse A/S concludes that no significant differences exist. The Ellipse Juvia should not raise any new issues of safety and effectiveness and is judged to be substantially equivalent to the mentioned predicate devices.
| (Signature) | |
|--------------|--|
| Ole Kofod | |
| (Typed Name) | |
| 20-July-2007 | |
| (Date) | |
# (Premarket Notification 510(k) Number)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The text is in all caps and is evenly spaced around the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ellipse A/S % Ole Kofod QA/RA Manager Agern Alle 11 DK-2970 Hørsholm Denmark
SEP 1 2 2007
Re: K072023 Trade/Device Name: Ellipse Juvia Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: August 31, 2007 Received: September 7, 2007
Dear Ole Kofod:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indicle for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the 1 c. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misburanding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, IDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other manufon wour not not hot act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ole Kofod
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-1000.
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to 100x) marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkerson
Director
Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Notification
Device Name: Ellipse Juvia
Indications for Use:
Ellipse Juvia is intended for use in dermatology and plastic surgery for treatment of:
- Skin Resurfacing .
- Wrinkles, Rhytides, and Furrows .
- Acne Scars .
Concurrence of CDRH, Office valuation (ODF Prescription Use X Counter Use Division of General, Restoration Format 1-2-96) (Per 21CFR 801.109) and Neurological Devices 23 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.