DIODENT MICRO DENTAL LASER SYSTEM, MODELS 810 AND 980
K063384 · Hoya Conbio, Inc. · GEX · Mar 16, 2007 · General, Plastic Surgery
Device Facts
Record ID
K063384
Device Name
DIODENT MICRO DENTAL LASER SYSTEM, MODELS 810 AND 980
Applicant
Hoya Conbio, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 16, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DioDent Micro 810/980 Dental Laser System is intended for incision excision, ablation, vaporization, and/or coagulation of oral soft tissue (including marginal and interdental gingival and epithelial lining of free gingiva). It is also intended for light activation for bleaching materials for teeth whitening, and laser assisted bleaching/whitening for teeth whitening.
Device Story
DioDent Micro 810/980 is a solid-state Gallium Aluminum Arsenide (GaAlAs) semiconductor diode laser. It emits invisible laser energy at 810nm or 980nm wavelengths. The system includes an autoclavable flexible treatment fiber and handpiece. Operated by a clinician in a dental setting, the device is activated via a footswitch. A color touch-screen display allows adjustment of laser output levels. The laser operates in continuous wave or pulse modes to perform soft tissue surgery (incision, excision, ablation, vaporization, coagulation) or to activate bleaching materials for teeth whitening. By providing precise tissue interaction, the device facilitates various oral surgical procedures and cosmetic whitening, potentially reducing patient discomfort and recovery time compared to traditional methods.
Clinical Evidence
No clinical data provided. Bench testing only.
Technological Characteristics
Solid-state Gallium Aluminum Arsenide (GaAlAs) semiconductor diode laser. Wavelengths: 810nm or 980nm. Delivery system: autoclavable flexible treatment fiber and handpiece. Operation: continuous wave or pulse mode. User interface: color touch-screen display. Energy source: electrical, footswitch-controlled.
Indications for Use
Indicated for incision, excision, ablation, vaporization, and hemostasis of oral soft tissue in patients requiring dental procedures such as biopsies, frenectomy, gingivectomy, implant recovery, pulpotomy, and sulcular debridement. Also indicated for light-activated teeth whitening.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
DioDent II Dental Laser System by HOYA ConBio (K050274)
Twilight Diode Laser System by BioLase Technologies (K991994)
Submission Summary (Full Text)
{0}------------------------------------------------
HOYA ConBio, Inc. Traditional 510(k) DioDent Micro 810/980
### MAR 1 6 2007
## 510k Summary
Submitter:
Contact:
Hoya ConBio, Inc. 47733 Fremont Blvd. Fremont, California 94538 Phone: 510-445-4500 Fax: 510-445-4550
Liza Burns Regulatory Consultant
November 7, 2006
Dental Diode Laser
79-GEX
Date Summary Prepared:
Device Trade Name:
Common Name:
Classification Name:
Classification Code:
Equivalent Device:
Device Description:
878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to
Instrument, surgical, powered, laser
DioDent Micro 810 Dental Laser System DioDent Micro 980 Dental Laser System
dioxide. (2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
cut, destroy, or remove tissue by light energy emitted by carbon
SIROLaser by Sirona Dental Systems, K053161 DioDent II Dental Laser System by HOYA ConBio, K050274 Twilight Diode Laser System by BioLase Technologies, K991994
The laser source of the DioDent Micro 810/980 is a solid-state Gallium Aluminum Arsenide (GaAlAs) semiconductor diode. It produces invisible laser energy at the 810-nanometer or 980nm wavelength. The delivery system consists of an autoclavable flexible treatment fiber threaded through a lightweight, autoclavable hand piece. Activation occurs when the operator enables the laser and presses the footswitch. Releasing the footswitch suspends laser treatment. The footswitch can function as an on/off switch. A color touch-screen display panel allows the operator to adjust or set laser output level with minimal effort.
Proprietary Information
K063384
{1}------------------------------------------------
The laser can operate in continuous wave or pulse mode.
| Intended Use: | The DioDent Micro 810/980 Dental Laser System is intended for<br>incision excision, ablation, vaporization, and/or coagulation of<br>oral soft tissue (including marginal and interdental gingival and<br>epithelial lining of free gingiva). It is also intended for light<br>activation for bleaching materials for teeth whitening, and laser<br>assisted bleaching/whitening for teeth whitening. |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison: | The DioDent II, the SIROLaser, and the Twilight<br>are equivalent in operating parameters, physical characteristics,<br>and intended uses. (NOTE: Of the equivalent devices, only the<br>DioDent II is cleared for teeth whitening intended uses). |
| Nonclinical Performance<br>Data: | None |
| Clinical Performance Data: | None |
| Additional Information: | None requested at this time. |
. .
:
:
:
:
-
-
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged around the upper portion of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Hoya ConBio, Inc. % Ms. Liza Burns Regulatory Consultant 47733 Fremont Boulevard Fremont, California 94538
MAR 1 6 2007
Re: K063384
Trade/Device Name: DioDent Micro 810 Dental Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: February 5, 2007 Received: February 9, 2007
Dear Ms. Burns:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Liza Burns
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Mellor
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Attachment 4: Indications for Use Statement
#### Indications for Use Statement
510(k) Number:
K063384
Device Name:
DioDent Micro 810™, DioDent Micro 980™
Indications for Use:
For the incision, excision, ablation, vaporization, and hemostasis of oral soft tissue.
Examples:
Excisional and incisional biopsies Exposure of unerupted teeth Fibroma removal Frenectomy and frenotomy Gingival troughing for crown impressions Gingivectomy Gingivoplasty Gingival incision and excision Hemostasis Implant recovery Incision and drainage of abscess Leukoplakia Operculectomy Oral papillectomies Pulpotomy Pulpotomy as an adjunct to root canal therapy Reduction of gingival hypertrophy Reduction of bacterial level (decontamination) and inflammation Soft tissue crown lengthening Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
Proprietary Information
Page 95
{5}------------------------------------------------
HOYA ConBio, Inc.
Traditional 510(k) DioDent Micro 810/980 Treatment of aphthous ulcers Vestibuloplasty Biopsy incision and excision Lesion (tumor) removal
For light activation for bleaching materials for teeth whitening For laser-assisted bleaching/whitening for teeth.
Prescription Use __ X (21 CFR 801 Subpart D)
OR
Over-the-Counter Use (21 CFR 801 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Posted November 13, 2003)
Page 1 of 1
wh
(Division Sign-Off) (Division Sign-On)
Division of General, Restorative,
Division of General, Devices Division of Secological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.