VISUALASE COOLED LASER APPLICATOR SYSTEM, LASER DIFFUSING FIBER, COOLING CATHETER SYSTEM

K053087 · Biotex, Inc. · GEX · Mar 1, 2006 · General, Plastic Surgery

Device Facts

Record IDK053087
Device NameVISUALASE COOLED LASER APPLICATOR SYSTEM, LASER DIFFUSING FIBER, COOLING CATHETER SYSTEM
ApplicantBiotex, Inc.
Product CodeGEX · General, Plastic Surgery
Decision DateMar 1, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Visualase Cooled Laser Applicator System (VCLAS) is comprised of a Laser Diffusing Fiber (LDF), a Cooling Catheter System (CCS), and a Bare Tip Fiber (BTF), The Laser Diffusing Fiber (LDF) and Cooling Catheter System (CCS) are indicated for use to necrotize or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vessels of the pericardial sac), dermatology, ear-nose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, neurosurgery, plastic surgery, pulmonology, radiology and urology for the wavelength range of 800nm through 1064nm. The Bare Tip Fiber (BTF) is intended for use in vaporization, cutting, ablation and coagulation of soft tissues with or without scopes or handpieces, for contact or noncontact surgery in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vessels of the pericardial sac), dermatology, ear-nose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, neurosurgery, plastic surgery, pulmonology, radiology and urology for the wavelength range of 800nm through 1064nm.

Device Story

VCLAS consists of Laser Diffusing Fiber (LDF), Cooling Catheter System (CCS), and Bare Tip Fiber (BTF). LDF/BTF transmit laser energy to tissue; LDF uses diffusing tip with light-dispersing particles in transparent matrix to create ellipsoidal coagulation zones. CCS provides optional cooling to tissue/fiber interface. BTF features flat, polished silica fiber tip for contact/non-contact ablation/cutting. Used by surgeons in various specialties; operates with external laser source (800-1064nm). Output is thermal energy for tissue destruction/modification. Benefits include controlled interstitial thermal therapy and precise soft tissue management.

Clinical Evidence

Bench testing only. Engineering studies demonstrated functional equivalence to predicate devices. Testing confirmed ellipsoidal coagulation area surrounding the diffusing region. Materials verified as biocompatible per international standards.

Technological Characteristics

Silica fiber optic cables; diffusing tip assembly (plastic tube with light-dispersing particles); cooling catheter (optional). Wavelength range 800nm-1064nm. Biocompatible materials. Mechanical/optical delivery system; no internal software or electronic sensing.

Indications for Use

Indicated for soft tissue necrosis, coagulation, vaporization, cutting, and ablation in patients undergoing surgery in cardiovascular thoracic (excluding heart/pericardial vessels), dermatology, ENT, gastroenterology, general, gynecology, head/neck, neurosurgery, plastic, pulmonology, radiology, and urology. Compatible with 800nm-1064nm laser wavelengths.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # MAR 1 2006 # 510(k) Summary In accordance with the Safe Medical Devices Act of 1990 and in compliance with 21CFR 807, the following serves as the 510(k) Summary information upon which the substantial equivalence determination is based. ## Contact Information | Submitter: | BioTex, Inc.<br>8058 El Rio St.<br>Houston, TX 77054 | |-----------------|------------------------------------------------------| | Phone: | 713.741.0111 | | Contact Person: | Ashok Gowda, PhD | | Date Prepared: | 11/01/2005 | ## Device Names | Trade Name: | Laser Diffusing Fiber (LDF)<br>Cooling Catheter System (CCS)<br>Bare Tip Fiber (BTF) | |-------------------|-----------------------------------------------------------------------------------------------| | Proprietary Name: | Visualase Cooled Laser Application System (VCLAS) | | Common Name: | Diffusing laser fiber or probe<br>Cooling catheter or sheath<br>Bare tip laser fiber or probe | | Classification: | Accessory to powered surgical laser instrument | #### Predicate Device SLT Diffuser™ Fiber (K010041) Indigo Bare-Tip™ Fiberoptic (K963081) #### Description of Device The VCLAS family consists of three components, the LDF, CCS, and BTF. The LDF and BTF transmit laser energy to the tissue situated at their distal regions. The LDF is comprised of a standard silica fiber optic cable affixed with a diffusing tip assembly. The diffusing tip assembly is comprised of a plastic tube that is filled with a transparent matrix in which light dispersing particles are embedded. The distal end of the tube is sealed with a higher concentration of scattering {2}------------------------------------------------ particles in a conical structure to prevent significant forward transmission of energy. The LDF may be used with a cooling catheter, in particular the compatible CCS, but such a catheter is not required. The CCS can be employed with or without coolant flow. The use of coolant provides cooling for the surfaces of the CCS in contact with both the tissue and the LDF. The BTF is comprised of a standard silica fiber optic cable identical to fiber optic used in the LDF. The distal end of the BTF is terminated in a flat, polished face and is designed for use in both contact and non-contact applications. #### Indications for Use The LDF and CCS are indicated for use to necrotize or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vessels in the pericardial sac), dermatology, ear-nose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, neurosurgery, plastic surgery, pulmonology, radiology, and urology, for wavelengths 800nm through 1064nm, The Bare Tip Fiber (BTF) is intended for use in vaporization, cutting, ablation and coagulation of soft tissues with or without scopes or handpieces, for contact or non-contact surgery in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vessels of the pericardial sac), dermatology, earnose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, neurosurgery, plastic surgery, pulmonology, radiology and urology for the wavelength range of 800nm through 1064nm. #### Comparison to Predicate Device The Laser Diffusing Fiber (LDF) and Cooling Catheter System (CCS) have been shown to be substantially equivalent to the SLT Diffuser™ fiber (K010041). The Bare Tip Fiber (BTF) has been shown to be substantially equivalent to the bare tip fiber marketed with the Indigo Diffuser Fiberoptic (K963081). Differences were determined to be minor and are each within the specifications listed by the predicate devices and does not raise any concerns regarding the overall safety and effectiveness of the device. ## Non-clinical Performance Tests: Engineering studies have demonstrated the substantial equivalence of the Laser Diffusing Fiber (LDF) to the SLT Diffuser Fiber Delivery Systems (K 010041). The studies concluded that the treated tissue displayed an ellipsoidal coagulation area surrounding the diffusing region of the laser diffusing fiber. In all instances, {3}------------------------------------------------ the fiber and lasers functioned as intended and performed in a manner similar to the predicate device when used in accordance with the labeled directions for use and specified indications. The materials used in the Visualase Laser Diffusing Fiber (LDF). Cooling Catheter System (CCS), and Bare Tip Fiber (BTF) were determined to be biocompatible according to appropriate international test standards. #### Conclusion BioTex has demonstrated the Visualase Laser Diffusing Fiber (LDF), Cooling Catheter System (CCS), and Bare Tip Fiber (BTF) are substantially equivalent to the predicate devices based on design, test results, and indications for use. {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 2006 MAR I BioTex, Inc. c/o Ashok Gowda, Ph.D. President 8058 El Rio Street Houston, Texas 77054 Re: K053087 Trade/Device Name: Visualase Cooled Laser Application System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: January 5, 2006 Received: January 20, 2006 Dear Dr. Gowda: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FIDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ Page 2 - Dr. Gowda This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation chiritled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, el.mde Mark N. Melkerson, M.S. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use Statement 510(k) Number (if known): Device Name: Visualase Cooled Laser Application System (VCLAS) Laser Diffusing Fiber (LDF) Cooling Catheter System (CCS) Bare Tip Fiber (BTF) Indications for Use: The Visualase Cooled Laser Applicator System (VCLAS) is comprised of a Laser Diffusing Fiber (LDF), a Cooling Catheter System (CCS), and a Bare Tip Fiber (BTF), The Laser Diffusing Fiber (LDF) and Cooling Catheter System (CCS) are indicated for use to necrotize or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vessels of the pericardial sac), dermatology, ear-nose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, neurosurgery, plastic surgery, pulmonology, radiology and urology for the wavelength range of 800nm through 1064nm. The Bare Tip Fiber (BTF) is intended for use in vaporization, cutting, ablation and coagulation of soft tissues with or without scopes or handpieces, for contact or noncontact surgery in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vessels of the pericardial sac), dermatology, ear-nose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, neurosurgery, plastic surgery, pulmonology, radiology and urology for the wavelength range of 800nm through 1064nm. # PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (O Division Sign-Off 600 Division of General, Restorative. and Neurological Devices > KOS 3087 510(k) Number_ K053087 Response Letter Page 5 of 60
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...