POLARIS LV, LVA APPLICATOR

K052324 · Syneron Medical, Ltd. · GEX · Sep 14, 2005 · General, Plastic Surgery

Device Facts

Record IDK052324
Device NamePOLARIS LV, LVA APPLICATOR
ApplicantSyneron Medical, Ltd.
Product CodeGEX · General, Plastic Surgery
Decision DateSep 14, 2005
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Polaris LV is intended for use in dermatology for treatment of vascular lesions.

Device Story

The Polaris LV is a dermatological device utilizing selective electromagnetic thermolysis to treat vascular lesions. It combines optical laser energy (780-980 nm) and radiofrequency (RF) energy. The device delivers energy through a light guide applicator to target tissues while sparing surrounding areas. The system operates in pulsed mode at a 1 Hz repetition rate. The modification increases optical fluence output up to 350 J/cm² alongside 100 J/cm³ RF energy. The device is intended for clinical use by healthcare professionals. By optimizing energy parameters, the device selectively damages target vascular structures, potentially benefiting patients by clearing vascular lesions.

Clinical Evidence

Bench testing only.

Technological Characteristics

Diode laser system combined with RF energy. Wavelength: 780-980 nm. Energy output: up to 350 J/cm² (optical) and 100 J/cm³ (RF). Pulse width: 100 msec. Repetition rate: 1 Hz. Delivery via light guide applicator. Pulsed operation mode.

Indications for Use

Indicated for the treatment of dermatological vascular lesions.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENES SYNERON MEDICAL Ltd. Polaris LV / LVA Applicator This summary of safety and effectiveness information is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92. Syneron Medical Ltd., Tavor Bld., Submitter: Industrial Zone Yokneam Illit, Israel Tel. +972.4.909-6200, Fax +972.4. 909-6202 Name of the Device: Polaris LV , LVA Applicator This is a Special 510(k) for the Polaris LV that was cleared Predicate Devices: under K030186. Device Description: The Polaris LV is a device that is used for treatment of vascular lesions. The Polaris LV treatment is based on the principle of selective (electromagnetic) thermolysis. According to this principle, parameters of optical and RF energy (spectrum, exposure duration and energy density) are chosen and optimized to selectively damage to the hair follicle without damaging the surrounding tissues. The Polaris LV is intended for use in dermatology for treatment of vascular lesions. The modifications to the Polaris LV do not affect the intended use or alter the fundamental scientific technology of the device. The only device modification is increasing the laser energy density output, still with in the range of predicate devices. There are no labeling changes that affect the intended use of the device. The device modifications raise no new issues of safety or effectiveness. August 22 20005 tuir Waldman Dr. Amir Waldman VP regulatory & clinical affairs Syneron Medical Ltd. {1}------------------------------------------------ Name of Manufacturer: Synergy Medical Ltd. Laser Model Name and Number: Polaris LV Laser Type: -(Circle all that apply)- Alexandrite, Argon, CO₂, Copper-Vapor, (Diode), Dye, Nd:YAG, Erb: Y Hol: YAG, Krypton, Ruby, KTP/532, Excimer, HENE, Accessory, Other_ Indications in this Application: Treatment for dermatological Vascular lesions. List of Examples: | <strong>Labels</strong> | <strong>Values</strong> | |-----------------------------|-------------------------| | FDA Document Control Number | K 052324 | | FDA Product Code | GEX | | Reviewer Computer Initials | SPB | | Date of Clearance Letter | | Basis of Approval: -(Circle all that apply)- Predicate Device (PD), Clinical Data (CD), Animal Data (AD Specifications (SPECS), Bench Test Data (BTD), Historical Informati (HI), Other Description of Laser: Operation Modes: -(Circle all that apply)- CW, (Pulsed), Q-Switched, Mode Locked, Contact, Free beam, Other Wavelength(s) in Nanometers: 780-980 Power/Energy Range (watts/joules): 350 J/cm² (optical) + 100 J/cm³ (RF) Pul Width: 100 msec Repetiti Rate: 1 Hz Delivery System(s Laser beam is transmitted through the light guide out of the applicator. Comments: Laser and RF Energy are combined for the treatment. This special 510(K) is for enhancing the optical fluence from 140 to 350 J/cm² {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three heads facing to the left, and two wavy lines below the eagle. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" is arranged in a circle around the eagle. SEP 1 4 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Dr. Amir Waldman VP Regulatory & Clinical Affairs Syneron Medical Ltd., Tavor Building Industrial Zone P.O.B. 550 Yokneam Illit, Israel 20692 Re: K052324 Trade/Device Name: Polaris LV, LVA Applicator Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX, GEI Dated: August 22, 2005 Received: August 25, 2005 Dear Dr. Waldman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Barbara Borlund for Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K US 2324 Device Name: Polaris LV, LVA Applicator Indications For Use: The Polaris LV is intended for use in dermatology for treatment of vascular lesions. Prescription Use AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Shatiav Buchup for WXM (Division Division of General, Restorative, and Neurological Devices **510(k) Number** K052324 Page 1 of
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...