K051167 · Allmed Systems, Inc. · GEX · Jun 1, 2005 · General, Plastic Surgery
Device Facts
Record ID
K051167
Device Name
REVOLIX AND REVOLIX JR
Applicant
Allmed Systems, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 1, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The RevoLix and RevoLix Jr laser systems and its fiber optic delivery system are intended for use in surgical procedures and endoscopic and endoscopic intended for used in surgical procedures and experience coagulation and hemostasis of soft including excision, resection, ablation, vaporization, valuation, varian and tissue in use in medical specialites inoldung: Orelegy, Central Surgery, General Surgery.and Arthroscopy
Device Story
RevoLix and RevoLix Jr are 2.01-micron diode-pumped solid-state surgical laser systems. The device consists of a laser console with an internal computer control panel, display, footswitch, and silica fiber optic delivery system. The distal tip is guided by a handpiece or endoscopic/laparoscopic instrument. The laser energy is used for ablation, coagulation, dissection, and resection of soft tissue in various surgical settings, including OR and endoscopic suites. Operated by physicians, the laser provides precise tissue interaction for procedures across multiple specialties. The output allows surgeons to perform minimally invasive or open surgeries with controlled hemostasis, potentially reducing patient trauma and recovery time compared to traditional methods.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
2.01-micron diode-pumped solid-state laser system. Components include laser console, internal computer control, display, footswitch, and silica fiber optic delivery system. Designed for open, laparoscopic, and endoscopic applications. Class II device.
Indications for Use
Indicated for open, laparoscopic, and endoscopic surgical procedures requiring incision, excision, resection, ablation, vaporization, coagulation, and hemostasis of soft, mucosal, fatty, and cartilaginous tissue. Specialties include Urology (e.g., BPH, tumors, strictures), Gastroenterology (e.g., polyps, ulcers, calculi), Thoracic/Pulmonary (e.g., airway obstructions), Gynecology (e.g., fibroids, polyps), ENT (e.g., sinus surgery, tonsillectomy), Dermatology/Plastic Surgery (e.g., carcinomas, warts), General Surgery (e.g., cholecystectomy, mastectomy, appendectomy), and Arthroscopy/Orthopedic surgery (e.g., meniscectomy, discectomy, foraminoplasty). Contraindicated for use in the spine, except for specific percutaneous laser disc decompression/discectomy and foraminoplasty procedures.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Lumenis/Coherent Medical - VersaPulse Ho:YAG Trimedyne Omnipulse Max 80 watt
Submission Summary (Full Text)
{0}------------------------------------------------
JUN 1 - 2005
K051167 1/5
# Attachment V
# 510(k) Summary
# 1.General Information
AllMed Systems Inc. Submitter: 9232 Klemetson Drive Pleasanton CA 94588
> 925-468-0433 Phone:
925-399-5984 Fax
Peter Allen Contact Person
4" May 2005 Date Prepared
# 2. Names
| Device Name | Revolix Family of Laser Systems including the RevoLix<br>and RevoLix Jr |
|---------------------|-------------------------------------------------------------------------|
| Common Name | 2.01micron Laser System |
| Classification Name | Laser Surgical Instrument and accessories |
# 3. Predicate Device
Lumenis/Coherent Medical - VersaPulse Ho:YAG Trimedyne Omnipulse Max 80 watt
# 4. Product Description
The RevoLix and RevoLix Jr are diode pump solid state surgical laser system operating at a wavelength of 2.01 micron. The purpose of the laser is the ablation, coagulation, dissection and resection of soft tissue. The laser is designed for open surgery, laparoscopic and surgical applications in aqueous media. The
{1}------------------------------------------------
sme 7 2/5
laser power is delivered via standard silica laser fibers. The distal tip is guided by a handpiece or endoscopic/laparoscopic surgical instrument.
It consists of:
Laser Console with Internal Computer Control Panel and Display A fiber optic delivery system Footswitch
# 5. Indications for Use
#### Arthroscopy
Arthroscopy/orthodepic surgery (ablation, excision and coagulation of soft and cartilaginous tissue) in various small and large joints of the body, excluding the spine, including
Meniscectomy Plica removal Ligament and Tendon release Contouring and sculpting of articular surfaces Debridement of inflamed synovial tissue (synovectomy) Loose body debridement Chondromalacia and tears Lateral retinecular release Capsulectomy in the knee Chondroplasty in the knee
Disectomy including
Percutaneous vaporization of the L4-5 and L5-S1 lumbar discs of the vertebral spine; open and arthroscopic spine procedures: foraminotomy
### Urology
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
Urethral Strictures Bladder Neck Incisions (BNI) Ablation and resection of Bladder Tumors, Uretheral Tumors and Ureteral Tumors. Ablation of Benign Prostatic Hypertrophy (BHP), Transurethral incision of the prostate (TUIP) Laser Resection of the Prostrate (HoLRP) Laser Enuculeation of the Prostate (HoLEP) Laser Ablation of the Prostate (HoLAP) Condylomas
{2}------------------------------------------------
si167 3/5
# Lesions of external genitalia
# Gastroenterology
Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis ) including:
Appendectomy Polyps Biopsv Gail Bladder calculi Biliary/Bile duct calculi Ulcers Gastric ulcers Duodenal ulcers Non Bleeding Ulcers Pancreatitas Hemorrhoids Cholecystectomy Benign and Malignant Neoplasm Angiodysplasia Colorectal cancer Telangiectasias Telangiectasias of the Osler-Weber-Renu disease Vascular Malformation Gastritis Esophagitis Esophageal ulcers Varices Colitis Mallory-Weiss tear Gastric Erosions
### Thoracic and Pulmonary
Open and endoscopic thoracic and pulmonary surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue
Laryngeal Lesions Airway obstructions including carcinoma Polyps and Granulomas Palliation of obstructing carcinomas of the tracheobronchial tree
### Gynecology
Open and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis)
{3}------------------------------------------------
KOS1167
4/1
Intra-uterine treatment of submucous fibroids, benign endometrial polyps, and uterine septum by incision, excision, ablation and or vessel coagulation
Soft tissue excision procedures such as excisional conization of the cervix
# ENT
Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue) including:
- Endonasal/sinus Surgery Partial turbinectomy Polypectomy Dacryocystorhinostomy Frontal Sinusotomy Ethmoidectomy Maxillary antrostomy Functional endoscopic sinus surgery Lesions or tumors of the oral, nasal, glossal, pharyngeal and laryngeal Tonsillectomy Adenoidectomy
# Dermatology and Plastic Surgery
Incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft, mucosal, fatty and cartilaginous tissue, in therapeutic plastic, dermatologic and aesthetic surgical procedures including:
Basal Cell Carcinomas Lesions of skin and subcutaneous tissue Skin tags Plantar warts
### General Surgery
Open laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
> Cholecystectomy Lysis of adhesion Appendectomy Biopsy Skin incision Tissue dissection Excision of external tumors and lesions Complete or partial resection of internal organs, tumors and lesions Mastectomy
{4}------------------------------------------------
smos 1/5
- Hepatectomy Pancreatectomy Splenectomy Thyroidectomy Parathyroidectomy Herniorrhaphy Tonsillectomy Lymphadenectomy Partial Nephrectomy Pilonidal Cystectomy Resection of lipoma Debridement of Decubitus Ulcer Hemorrhoids Debridement of Statis Ulcer Biopsy
# 6. Rationale for Substantial Equivalence
The Revolix and RevoLix Jr laser system with fiber optic delivery devices share the same intended use, indications for use, similar design features and functional features and therefore are substantially equivalent to the Lumenis VersaPulse Laser
# 7. Conclusion
The Revolix and RevoLix Jr Laser System with fiber optic delivery devices were found to be substantially equivalent to similar currently marketed and predicate surgical laser systems and delivery devices.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a seal for the Department of Health and Human Services (HHS). The seal features an eagle-like symbol with three stylized lines forming its body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The text is in all capital letters and is evenly spaced around the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 1 - 2005
Mr. Peter N. Allen President AllMed Systems Incorporated 9232 Klemetson Drive Pleasanton, California 94588
Re: K051167
Trade/Device Name: RevoLix and RevoLix Jr Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: May 4, 2005 Received: May 5, 2005
Dear Mr. Allen:
We have reviewed your Section 510(k) premarket notification of intent to market the device wo nave roved above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use surfor to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Orth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Mr. Peter N. Allen
This letter will allow you to begin marketing your device as described in your Section 510(k) rms lotter with and hyour your finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you usen's ffice of Compliance at (240) 276-0115 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general micronal and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
R.R.
Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Number: k051167
RevoLix and RevoLix Jr Device Name:
Indications For Use:
The Revolix and RevoLix Jr laser systems and its fiber optic delivery system are The Revolix and Revolix of laser Systems and endoscopic and endoscopic intended for used in surgical procedures and experience coagulation and hemostasis of soft including excision, resection, abiation, raponiation, valuation, varian and tissue in use in medical specialites inoldung: Orelegy, Central Surgery, General Surgery.and Arthroscopy
Urology
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
Urethrai Strictures
Bladder Neck Incisions (BNI)
Bladder Neck molono (DRI)
Ablation and resection of Bladder Tumors, Uretheral Tumors and Ureteral Tumors.
Ablation of Benign Prostatic Hypertrophy (BHP),
Transurethral incision of the prostate (TUIP)
Laser Resection of the Prostrate (HoLRP)
Laser Enuculeation of the Prostate (HoLEP)
Laser Ablation of the Prostate (HoLAP)
Condylomas
Lesions of external genitalia
Gastroenterology
Open and endoscopic gastroenterology surgery (incision, excision, resection, Open and endooooplo gagulation and hemostasis ) including:
Appendectomy Polyps Biopsy Gall Bladder calculi
| Prescription Use | X |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
NEEDED)
of CDRH Office of Device Evaluation (ODF)Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 5
{8}------------------------------------------------
#### Kaf 1147 510(k) Number:
RevoLix and RevoLix Jr Device Name:
Indications For Use:
Biliary/Bile duct calculi Ulcers Gastric ulcers Duodenal ulcers Non Bleeding Ulcers Pancreatitas Hemorrhoids Cholecystectomy Benign and Malignant Neoplasm Angiodysplasia Colorectal cancer Telangiectasias Telangiectasias of the Osler-Weber-Renu disease Vascular Malformation Gastritis Esophagitis Esophageal ulcers Varices Colitis Mallory-Weiss tear Gastric Erosions
Thoracic and Pulmonary
Open and endoscopic thoracic and pulmonary surgery (incision, excision, Open und ondeooople workization, coagulation and hemostasis) of soft tissue
Laryngeal Lesions Airway obstructions including carcinoma Polvps and Granulomas r offpo and Grandising carcinomas of the tracheobronchial tree
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(Please do not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Carana. Page 2 of 5 12 8 239 11 11
Ka51167
{9}------------------------------------------------
| 510(k) Number: | ~051167<br>ﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮ |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
RevoLix and RevoLix Jr Device Name:
Indications For Use:
# Gynecology
Open and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis)
> Intra-uterine treatment of submucous fibroids, benign endometrial polyps, mira uterine septum by incision, excision, ablation and or vessel coagulation Coagulation
> Soft tissue excision procedures such as excisional conization of the cervix
#### ENT
Endoscopic endonasal surgery (incision, excision, resection, ablation, Endostoplo enabliation and hemostasis of soft tissue) including:
- Endonasal/sinus Surgery Partial turbinectomy Polypectomy Dacryocystorhinostomy Frontal Sinusotomy Ethmoidectomy Maxillary antrostomy Functional endoscopic sinus surgery Lesions or tumors of the orai, nasal, glossal, pharyngeal and laryngeal Tonsillectomy Adenoidectomy
# Dermatology and Plastic Surgery
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(Please do NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
| Page 3 of 5 |
{10}------------------------------------------------
510(k) Number:
RevoLix and RevoLix Jr Device Name:
Indications For Use:
Incision, excision, resection, ablation, vaporization, coagulation and hemostasis melsion, exolon, recosment rtilaginous tissue, in therapeutic plastic, of soft, maoooal, fax) and aesthetic surgical procedures including:
Basal Cell Carcinomas Lesions of skin and subcutaneous tissue Skin tags ﮨﮯ Plantar warts
### General Surgery
Open laparoscopic and endoscopic surgery (incision, excision, resection, Open raparooooooooon, coagulation and hemostasis) including:
> Cholecystectomy Lysis of adhesion Appendectomy Biopsv Skin incision Tissue dissection Excision of external tumors and lesions Excision of external carection of internal organs, tumors and lesions Mastectomy Hepatectomy Pancreatectomy Splenectomy Thyroidectomy Parathyroidectomy Herniorrhaphy Tonsillectomy Lymphadenectomy Partial Nephrectomy
| Prescription Use | X |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) | |
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
Recorative
Devices
| Page 4 of 5 |
|-------------|
|-------------|
K051167
AND/OR
{11}------------------------------------------------
#### Ko)1167 510(k) Number:
RevoLix and RevoLix Jr Device Name:
Indications For Use:
Pilonidal Cystectomy Resection of lipoma Debridement of Decubitus Ulcer Hemorrhoids Debridement of Statis Ulcer Biopsy
### Arthroscopy
Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue)
- Ablation of soft and cartilaginous tissue in Minimal Invasive Spinal Surgery including
- Surgery Including
Percutaneous Laser Disc Decompression/Discectomy Foraminoplasty
- Foraminoplasty
Ablation and coagulation of soft vascular and non vascular tissue in minimally invasive spinal surgery.
ズ Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--|
|--------------------------------------------------------|--|
(Urision Sign-Ori) libriston of Genoril. Reviewative cial Neurological Devices
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.