FOCUS MEDICAL NATURALASE LP AND NATURALIGHT FAMILY OF PRODUCTS
K041829 · Focus Medical, LLC · GEX · Sep 7, 2004 · General, Plastic Surgery
Device Facts
Record ID
K041829
Device Name
FOCUS MEDICAL NATURALASE LP AND NATURALIGHT FAMILY OF PRODUCTS
Applicant
Focus Medical, LLC
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 7, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
NaturaLight IPL System and the NaturaLase LP+ System when used in IPL (intense pulsed light) Mode is indicated for: Dermatology/Piastic Surgery: The removal of unwanted hair from skin types I-VI, and to effect stable long-term, or permanent, hair reduction. The treatment of benign pigmented lesions, including lentigines, nevi, melasma, and cafe-au-lait. treatment of vascular lesions, including port wine stains, hemangiomas, angiomas, telangiectasias, rosacea, facial and leg veins. The Focus Medical NaturaLase LP Laser System and the NaturaLase LP+ System when used in Laser Mode is indicated for: General Surgical Applications: Incision, excision, coagulation, hemostasis, vaporization, and/or ablation of soft tissue in dermatology/plastic surgery, endoscopic/laparoscopic general surgery, gastroenterology, general surgery, gynecology, head and neck/ otorhinolaryngology (ENT), neurosurgery, oculoplastics, orthopedics, pulmonary surgery, thoracic surgery, and urology. Dermatology/Plastic Surgery: Coagulation and hemostasis of benign vascular lesions such as port wine stains, hemangiomas, warts, telangiectasia, rosacea, venus lake, leg and spider veins. In addition, the NaturaLase LP is intended for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café au lait macules, schorrheic keratoses, nevi chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or blue/black tattoos) and plaques. The Natural ase I.P Laser is also indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. Cutting, incision, excision, hemostasis, coagulation, vaporization and ablation of soft tissue in dermatology and plastic surgery. The NaturaLase LP Laser is also indicated for the treatment of facial wrinkles and wrinkles such as, but not limited to, periocular and periorbital wrinkles. The NaturaLase LP Laser is indicated for the removal of unwanted hair, for the stable long tem, or permanent, hair reduction through sclective targeting of melanin in hair follicles, and for the treatment for pscudofolliculitis barbae (PFB). The NaturaLase LP Laser is indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. The Natural asc LP Laser is indicated for use on all skin types (Fitzpatrick I - VI) including tanned skin. Endoscopic/Laparoscopic Surgery: The NaturaLase Ll' Laser is also indicated for use in a variety of surgical procedures in several surgical specialties. These include, but not limited to, cholccystectomy, appendectorny, vagotomy, and pyloromyomotomy where its abilities to incise, excise, coagulate, vaporize, or ablate soft tissue may be applied. Gastroenterology: Tissue ablation and hemostasis in the gastrointestinal tract; esophagea' neoplasic obstruction including squamous cell carcinoma and adenocarcinoma, gastrointestina! hemostasis including varices, esophagitis, esophageal ulcer, Mallory-Weiss tear, gastric angiodysplasia, stomal ulcers, non-bleeding ulcers, gastric erosions, ulcer. gastrointestinal tissue ablation including benign and malignant neoplasm, angiodysplasia, polyps, ulcer, colitis, and hemorrhoids. General Surgery: Incision, excision, vaporization, ablation, and hemostasis of soft tissue general surgery applications, skin incisions, tissue dissection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors, lesions, tissue ablation, vessel coagulation, tonsillectomy, and hemorrhoids. Gynecology: Treatment of menorhagia by photocoagulation of the endometrial lining of the uterus, ablation of endometrial implants and/or peritoneal adhesions, soft tissue excisional procedures such as excisional conization of the cervix, intrauterine gynecologic procedures where cutting, ablation, and/or vessel coagulation may be indicated including submucous fibroids, benign endometrial polyps, and uterine septum. Head and Neck/Otorhinolaryngology (ENT): Tissue incision, excision, ablation, and vessel hemostasis. Hemostasis During Surgery: Adjunctive coagulation and hemostasis (bleeding control) during surgery in endoscopic (e.g. laparoscopic) and open procedures. Neurosurgery: Hemostasis for pituitary tumor, meningioma, hemagioblastoma, AVMs. glioblastoma, astrocytoma, oligodendroglioma. Oculoplastics: Incision, excision, vaporization, and/or coagulation of tissues in oculoplastic procedures such as operations on the lacrimal system, operation on the eyelids, removal of biopsy or orbital tumors, enucleation on eyeball, exteneration of orbital contents. Orthopedics: Cutting, ablation, and/or hemostasis of intra-articular tissue in orthopedic surgical and arthroscopic applications. Pulmonary Surgery: Palliative treatment of benign and malignant pulmonary airway obstructions, including squamous cell carcinoma, adenocarcinoma, carcinoid, benign tumors, granulomas, and benign strictures. Thoracic Surgery: Cutting (incision/excision), coagulating, and vaporization of soft tissue. Thoracic applications including, but not limited to, isolation of vessels for endarterectomy and/or by-pass grafts, wedge resections, thoractomy, formation of pacemaker pockets; vaporization, coagulation, incision/excision, debulking, and ablation of lung tissue (thoracoscopy). Urology: All applications including superficial urinary bladder tumors, invasive bladder carcinoma, urethral strictures and lesions of the external genitalia (including condyloma acuminata).
Device Story
System comprises NaturaLase LP (Nd:YAG laser), NaturaLight (IPL), and NaturaLase LP+ (combination). Device delivers light energy to target soft tissue or melanin in hair follicles; principle of operation involves selective photothermolysis for dermatological applications and thermal ablation/coagulation for surgical procedures. Operated by physicians in clinical/surgical settings. Output is controlled light energy; healthcare providers use device to perform surgical incisions, hemostasis, or dermatological treatments. Benefits include precise tissue management, reduced bleeding during surgery, and effective hair/lesion removal.
Clinical Evidence
Bench testing only.
Technological Characteristics
Nd:YAG laser and Intense Pulsed Light (IPL) system; energy source: laser/light; form factor: clinical treatment system; intended for soft tissue surgery and dermatological applications; skin types I-VI.
Indications for Use
Indicated for patients of all skin types (Fitzpatrick I-VI) requiring soft tissue surgery (incision, excision, ablation, coagulation, hemostasis) across multiple specialties including dermatology, plastic surgery, gastroenterology, gynecology, ENT, neurosurgery, orthopedics, pulmonary, thoracic, and urology; also indicated for hair removal, treatment of benign pigmented/vascular lesions, and facial wrinkle reduction.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
K041829 1 of 4
# Section 10 510(k) Sun mary
This 510(k) summary of safety and effectiveness for the Focus Medical NaturaLase LP and NaturaLight Family of Products is submitted in accordance with the requirements of SMDA 1990 and follows Office of Device Evaluation guidance concerning the organization and content of a 510(k) summary.
Applicant: Focus Medical
SEP - 7 2004
Address: Focus Medical, LLC 23 Francis J. Clarke Circle Bethel, CT 06801
Contact Person: Mr. John B. Lee, Jr. President Focus Medical
Telephone: 203-730-8885
Preparation Date: June 2004 (of the Summary)
Device Name: NaturaLase LP and NaturaLight Family of Products
Common Name: Nd:YAG and Intense Pulsed Light treatment system
Classification: Laser surgical instrument for use in general and plastic surgery and in dermatology.
Class II medical device 21CFR 878.4810 Product Code: GEX Panel: 79
Predicate devices: Palomar Starlux Pulsed Light System (K033549) Lumcnis Family of Intense Pulsed Light (IPL) and IPL/Nd: YAG Laser Systems (K030527) I umenis IPI. Quantum Family of Products (K020839) Focus Medical NaturaLase LP (K031828)
{1}------------------------------------------------
NaturaLase LP and NaturaLight Family of Products 510(k) Submission
K041829 2 of 4
Device description:
The Focus Medical NaturaLase LP and NaturaLight Family of Products consists of the NaturaLase LP Nd: YAG Laser System, the NaturaLight IPL System and the NaturaLase LP+ a simple combination of the NaturaLase LP and the NaturaLight.
Indications for Use:
## NaturaLight Indications for Use
NaturaLight IPL System and the NaturaLase LP+ System when used in IPL (intense pulsed light) Mode is indicated for:
Dermatology/Piastic Surgery:
The removal of unwanted hair from skin types I-VI, and to effect stable long-term, or permanent, hair reduction.
The treatment of benign pigmented lesions, including lentigines, nevi, melasma, and cafe-au-lait. treatment of vascular lesions, including port wine stains, hemangiomas, angiomas, telangiectasias, rosacea, facial and leg veins.
## NaturaLase LP Indications For Use
The Focus Medical NaturaLase LP Laser System and the NaturaLase LP+ System when used in Laser Mode is indicated for:
## General Surgical Applications:
Incision, excision, coagulation, hemostasis, vaporization, and/or ablation of soft tissue in dermatology/plastic surgery, endoscopic/laparoscopic general surgery, gastroenterology, general surgery, gynecology, head and neck/ otorhinolaryngology (ENT), neurosurgery, oculoplastics, orthopedics, pulmonary surgery, thoracic surgery, and urology.
## Dermatology/Plastic Surgery:
Coagulation and hemostasis of benign vascular lesions such as port wine stains, hemangiomas, warts, telangiectasia, rosacea, venus lake, leg and spider veins. In addition, the NaturaLase LP is intended for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café au lait macules, schorrheic keratoses, nevi chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or blue/black tattoos) and plaques. The Natural ase I.P Laser is also indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
Cutting, incision, excision, hemostasis, coagulation, vaporization and ablation of soft tissue in dermatology and plastic surgery.
The NaturaLase LP Laser is also indicated for the treatment of facial wrinkles and wrinkles such as, but not limited to, periocular and periorbital wrinkles.
The NaturaLase LP Laser is indicated for the removal of unwanted hair, for the stable long tem, or permanent, hair reduction through sclective targeting of melanin in hair follicles, and for the treatment for pscudofolliculitis barbae (PFB).
{2}------------------------------------------------
The NaturaLase LP Laser is indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. The Natural asc LP Laser is indicated for use on all skin types (Fitzpatrick I - VI) including tanned skin.
#### Endoscopic/Laparoscopic Surgery:
The NaturaLase Ll' Laser is also indicated for use in a variety of surgical procedures in several surgical specialties. These include, but not limited to, cholccystectomy, appendectorny, vagotomy, and pyloromyomotomy where its abilities to incisc, excise, coagulate, vaporize, or ablate soft tissue may be applied.
#### Gastroenterology:
Tissue ablation and hemostasis in the gastrointestinal tract; esophagea' neoplasic obstruction including squamous cell carcinoma and adenocarcinoma, gastrointestina! hemostasis including varices, esophagitis, esophageal ulcer, Mallory-Weiss tear, gastric angiodysplasia, stomal ulcers, non-bleeding ulcers, gastric erosions, ulcer. gastrointestinal tissue ablation including benign and malignant neoplasm, angiodysplasia, polyps, ulcer, colitis, and hemorrhoids.
#### General Surgery:
Incision, excision, vaporization, ablation, and hemostasis of soft tissue general surgery applications, skin incisions, tissue dissection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors, lesions, tissue ablation, vessel coagulation, tonsillectomy, and hemorrhoids.
#### Gynecology:
Treatment of menorhagia by photocoagulation of the endometrial lining of the uterus, ablation of endometrial implants and/or peritoneal adhesions, soft tissue excisional procedures such as excisional conization of the cervix, intrauterine gynecologic procedures where cutting, ablation, and/or vessel coagulation may be indicated including submucous fibroids, benign endometrial polyps, and uterine septum.
#### Head and Neck/Otorhinolaryngology (ENT):
Tissue incision, excision, ablation, and vessel hemostasis.
#### Hemostasis During Surgery:
Adjunctive coagulation and hemostasis (bleeding control) during surgery in endoscopic (e.g. laparoscopic) and open procedures.
#### Neurosurgery:
Hemostasis for pituitary tumor, meningioma, hemagioblastoma, AVMs. glioblastoma, astrocytoma, oligodendroglioma.
#### Oculoplastics:
Incision, excision, vaporization, and/or coagulation of tissues in oculoplastic procedures such as operations on the lacrimal system, operation on the eyelids, removal of biopsy or orbital tumors, enucleation on eyeball, exteneration of orbital contents.
{3}------------------------------------------------
NaturaLase LP and NaturaLight Family of Products 510(k) Submission
K041829 4 of 4
#### Orthopedics:
Cutting, ablation, and/or hemostasis of intra-articular tissue in orthopedic surpical and arthroscopic applications.
#### Pulmonary Surgery:
Palliative treatment of benign and malignant pulmonary airway obstructions, including squamous cell carcinoma, adenocarcinoma, carcinoid, benign tumors, granulomas, and benign strictures.
## Thoracic Surgery:
Cutting (incision/excision), coagulating, and vaporization of soft tissue. Thoracic applications including, but not limited to, isolation of vessels for endarterectomy and/or by-pass grafts, wedge resections, thoractomy, formation of pacemaker pockets; vaporization, coagulation, incision/excision, debulking, and ablation of lung tissue (thoracoscopy).
#### Urology:
All applications including superficial urinary bladder tumors, invasive bladder carcinoma, urethral strictures and lesions of the external genitalia (including condyloma acuminata).
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, with three parallel lines forming a stylized staff and a pair of intertwined snakes.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP - 7 2004
Mr. John B. Lee, Jr. President Focus Medical, LLC 23 Francis J Clarke Circle Bethel, Connecticut 06801
Re: K041829
Trade/Device Name: NaturaLase LP and NaturaLight Family of Products Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: July 7, 2004 Received: July 12, 2004
Dear Mr. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can he found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Mr. John B. Lee, Jr.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known):K041829
Device Name:
Indications For Use:
NaturaLight IPL System (intense pulsed light) is indicated for: The removal of unwanted hair from skin types I-VI, and to effect stable long-term, or permanent, hair reduction. The treatment of benign pigmented lesions, including
lentigines, nevi, melasma, and cafe-au-lait. treatment of vascular lesions, including port wine stains, hemangiomas, telangiectasias, rosacea, facial and leg veins.
NaturaLase LP Indications For Use
The Focus Medical NaturaLase LP Laser System and the NaturaLase LP+ System when used in Laser Mode is indicated for:
#### General Surgical Applications:
Incision, excision, coagulation, hemostasis, vaporization, and/or ablation of soft tissue dermatology/plastic surgery, endoscopic/laparoscopic general surgerv. in gastroenterology, general surgery, gynecology, head and neck/ otorhinolaryngology (ENT), neurosurgery, oculoplastics, orthopedics, pulmonary surgery, thoracic surgery, and urology.
× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
Sion Sign-Off Division of General, Restorative, and Neurological Devices
Page 1 of 4
**510(k) Number** K04/829
{7}------------------------------------------------
Continuation for Indications for Use on Form FDA 3514 Section G Page 2
## Dermatology/Plastic Surgery:
Coagulation and hemostasis of benign vascular lesions such as port wine stains, hemangiomas, warts, telangiectasia, rosacea, venus lake, leg and spider veins. In addition, the NaturaLase LP is intended for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café au lait macules, seborrheic keratoses, nevi chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or blue/black tattoos) and plaques. The NaturaLase LP Laser is also indicated for pigmented lesions to reduce lesion size,
for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
Cutting, incision, excision, hemostasis, coagulation, vaporization and ablation of soft tissue in dermatology and plastic surgery.
The NaturaLase LP Laser is also indicated for the treatment of facial wrinkles and wrinkles such as, but not limited to, periocular and periorbital wrinkles.
The NaturaLase LP Laser is indicated for the removal of unwanted hair, for the stable long tem, or permanent, hair reduction through selective targeting of melanin in hair follicles, and for the treatment for pseudofolliculitis barbae (PFB).
The NaturaLase LP Laser is indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
The NaturaLase LP Laser is indicated for use on all skin types (Fitzpatrick I - VI) including tanned skin.
## Endoscopic/Laparoscopic Surgery:
The Natural.ase LP Laser is also indicated for use in a variety of surgical procedures in several surgical specialties. These include, but not limited to, cholecystectomy, appendectomy, vagotomy, and pyloromyomotomy where its abilities to incise, excise, coagulate, vaporize, or ablate soft tissue may be applied.
Over-The-Counter Use AND/OR Prescription Use X (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Meriam C. Provost
(Division Six, O.C.)
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
Page 2 of 4
**510(k) Number** K041829
{8}------------------------------------------------
#### Gastroenterology:
Tissue ablation and hemostasis in the gastrointestinal tract; esophageal neoplastic obstruction including squamous cell carcinoma and adenocarcinoma, gastrointestinal hemostasis including varices, esophageal ulcer, Mallory-Weiss tear, gastric ulcer, angiodysplasia, stomal ulcers, non-bleeding ulcers, gastric erosions, gastrointestinal tissue ablation including benign and malignant neoplasm, angiodysplasia, polyps, ulcer, colitis, and hemorrhoids.
#### General Surgery:
Incision, excision, vaporization, ablation, and hemostasis of soft tissue general surgery applications, skin incisions, tissue dissection, excision of external turnors and lesions, complete or partial resection of internal organs, tumors, lesions, tissue ablation, vessel coagulation, tonsillectomy, and hemorrhoids.
#### Gynecology:
Treatment of menorrhagia by photocoagulation of the endometrial lining of the uterus. ablation of endometrial implants and/or peritoneal adhesions, soft tissue excisional procedures such as excisional conization of the cervix, intrauterine gynecologic procedures where cutting, ablation, and/or vessel coagulation may be indicated including submucous fibroids, benign endometrial polyps, and uterine septum.
#### Head and Neck/Otorhinolaryngology (ENT):
Tissue incision, excision, ablation, and vessel hemostasis.
#### Hemostasis During Surgery:
Adjunctive coagulation and hemostasis (bleeding control) during surgery in endoscopic (e.g. laparoscopic) and open procedures.
#### Neurosurgery:
Hemostasis for pituitary tumor, meningioma, hemagioblastoma, AVMs. glioblastoma, astrocytoma, oligodendroglioma.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Muriam C. Provost
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
**510(k) Number** K041829
{9}------------------------------------------------
#### Continuation for Indications for Use on Form FDA 3514 Section G Page 4
#### Oculoplastics:
Incision, excision, vaporization, and/or coagulation of tissues in oculoplastic procedures such as operations on the lacrimal system, operation on the eyelids, removal of biopsv or orbital tumors, enucleation on eyeball, exteneration of orbital contents.
#### Orthopedics:
Cutting, ablation, and/or hemostasis of intra-articular tissue in orthopedic surgical and arthroscopic applications.
## Pulmonary Surgery:
Palliative treatment of benign and malignant pulmonary airway obstructions, including squamous cell carcinoma, adenocarcinoma, carcinoid, benign tumors, granilo mas, and benign strictures.
## Thoracic Surgery:
Cutting (incision/excision), coagulating, and vaporization of soft tissue. Thoracic applications including, but not limited to, isolation of vessels for endarterectomy and/or by-pass grafts, wedge resections, thoractomy, formation of pacemaker pockets, vaporization, coagulation, incision/excision, debulking, and ablation of lung tissue (thoracoscopy).
#### Urology:
All applications including superficial urinary bladder tumors, invasive bladder carcinoma, urethral strictures and lesions of the external genitalia (including condyloma acuminata).
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provot (Division Sign-Off) Division of General, Restorative. and Neurological Devices
Page 4 of 4
**510(k) Number** k04/821
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.