K020847 · Photo Medex, Inc. · GEX · May 16, 2002 · General, Plastic Surgery
Device Facts
Record ID
K020847
Device Name
XTRAC EXCIMER LASER SYSTEM, MODEL AL7000
Applicant
Photo Medex, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
May 16, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use is targeted UVB phototherapy for psoriasis, vitiligo, atopic dermatitis, and non-autoimmune based leukoderma of affected skin.
Device Story
XTRAC Excimer Laser System (Model AL7000) is a self-contained, compact UV laser light source. It utilizes a XeCl gas mixture to generate targeted 308 nm ultraviolet light. The system comprises a keypad/display, fiberoptic delivery system, handpiece, and footswitch, all housed in a protective interlocked enclosure. Operated by clinicians in a medical setting, the device delivers targeted phototherapy to affected skin areas. The clinician selects the dose and targets the specific lesion. The output is a 308 nm laser beam used to induce repigmentation in conditions like vitiligo and leukoderma, or to treat inflammatory skin conditions. By providing localized UV exposure, the device aims to achieve therapeutic outcomes while minimizing exposure to healthy surrounding skin.
Clinical Evidence
Clinical study evaluated the 308nm excimer laser for treating post-resurfacing leukoderma. Subjects showed 50% to 75% or greater improvement in repigmentation. Supporting literature (Westerhof and Nieuweboer-Krobotova) compared narrow-band UV-B to PUVA for vitiligo, concluding twice-weekly treatment is safe and effective.
Technological Characteristics
XeCl excimer laser; 308 nm wavelength; fiberoptic delivery system; keypad/display interface; footswitch control; protective interlocked housing. Conforms to EN 60601 series, 21 CFR 1040.10 & 1040.11, and UL 2601 standards.
Indications for Use
Indicated for patients with psoriasis, vitiligo, atopic dermatitis, and non-autoimmune based leukoderma requiring targeted UVB phototherapy.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
XTRAC Excimer Laser System, model AL 7000 (K992914)
XTRAC Excimer Laser System, model AL 7000 (K003705)
XTRAC Excimer Laser System, model AL 7000 (K011382)
Submission Summary (Full Text)
{0}------------------------------------------------
620847 1/3
# 510(k) SUMMARY PhotoMedex, Inc. XTRAC Excimer Laser System, Model AL7000
### 1. GENERAL
PhotoMedex, Inc. . Submitter: 2431 Impala Drive Carlsbad, CA 92008
- Contact Person: Bob Rose . March 13, 2002 Date Prepared: ●
### 2. DEVICE NAME
- Classification Name: Laser surgical instrument for use in general and plastic . surgery and in dermatology (21 CFR §878.4810)
- Common or Usual Name: XeCl excimer laser .
- Trade or Proprietary Name: XTRAC Excimer Laser System, model AL7000 .
### 3. PREDICATE DEVICES
#### Ultraviolet Phototherapy Systems
- HOUVA II, Phototherapy System, National Biological Corporation, -510(k) number: K885026
- UviSol, Phototherapy System, National Biological Corporation, --510(k) number: K934808 DermaClear UV-B Phototherapy System, Coherent Star/Lumenis 510(k) number K011197
- Excimer Laser
- XTRAC Excimer Laser System, model AL 7000, PhotoMedex, Inc., -510(k) numbers: K992914, K003705, and K011382
{1}------------------------------------------------
20847 2/3
### 4. DEVICE DESCRIPTION
The XTRAC Excimer Laser System is a complete self-contained compact UV laser light source, which utilizes a XeCl gas mixture to generate an operator selected doseand target-specific ultraviolet light at wavelength of 308 nm. The laser system consists of a keypad and display, a fiberoptic delivery system, a handpiece and a footswitch. The laser is enclosed in a protective interlocked housing.
#### 5. INTENDED USE
The intended use is targeted UVB phototherapy for psoriasis, vitiligo, atopic dermatitis, and non-autoimmune based leukoderma of affected skin.
#### 6. SUBSTANTIAL EQUIVALENCE
The application of the XTRAC Excimer Laser System has been proven to be substantially equivalent to current legally marketed devices in the treatment of vitiligo via K003705. In that application, PhotoMedex demonstrated that narrow-band UV-B 311 alone had been shown to be just as effective in the treatment of vitiligo as PUVA (Westerhof and Nieuweboer-Krobotova). The authors in this study concluded, "the treatment of vitiligo with narrow-band UV-B twice weekly is a safe and effective treatment." Vitiligo, which is recognized as a cosmetic problem, is a form of leukoderma.
Current published data supports narrow-band UVB (utilizing ultraviolet lamp sources) to be effective in the regimentation of skin which has been affected by leukoderma such as the latent hypopigmentation, which is frequently experienced in complications resulting from carbon dioxide lasers used for skin resurfacing.
The intended use for the PhotoMedex XTRAC Excimer Laser System is within the scope of the predicate ultraviolet lamps. Both device types share the same methods and mechanisms (UVB light) to produce a result for the purpose of dermatological phototherapy.
The XTRAC Excimer Laser System, model AL7000 is currently market cleared for other skin conditions where UVB phototherapy has shown to be safe and effective via 510(k) numbers K992914, K003705, and K011382.
## XVI-3
<sup>4</sup> Westerhof W, Nieuweboer-Krobotova L. Treatment of vith UV-B radiation vs topical psoralen plus UV-A. Arch Dermatol 1997, 133:1525-1528.
{2}------------------------------------------------
020847 3/3
# 7. CLINICAL PERFORMANCE TESTING
Verification and validation to substantiate inclusion of the indication of UVB phototherapy for the treatment of leukoderma by the AL7000 is supported in a clinical study which utilized a 308nm excimer laser to successfully induce repigmentation on subjects. These subjects had experienced hypopigmentation (leukoderma) as a result of carbon dioxide laser resurfacing, and confirmed a 50% to 75% improvement or greater in treating leukoderma.
# 8. PRODUCT PERFORMANCE TESTING
Testing conducted on the XTRAC Excimer Laser System includes conformance to all relevant international EN 60601 series of standards, 21 CFR Part 1040.10 & 1040.11 Performance Standards for Light-Emitting Products and is a UL 2601 classified device.
#### 9. CONCLUSIONS
Based on the same intended use as UVB sources for phototherapy, the previously cleared technological characteristics of the XTRAC Excimer Laser (which are unchanged to support this additional indication), and the performance data, PhotoMedex believes that the XTRAC Excimer Laser System is substantially equivalent to the predicate devices.
## XVI-4
<sup>5</sup> Friedman. Paul MD, Geronemous, Roy MD, Use of the 308-nm Excimer Laser for Postresurfacing Leukoderma, ARCH DERMATOL, vol. 137, June 2001, 824-825.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 6 2002
Mr. Bob Rose Director of Regulatory Affairs and Quality Assurance Photo Medex, Inc. 2431 Impala Drive Carlsbad, CA 92008
Re: K020847
R020017
Trade/Device Name: XTRAC Excimer Laser System, Model AL7000 Regulation Number: 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: March 13, 2002
Received: March 15, 2002
Dear Mr. Rose:
We have reviewed your Section 510(k) premarket notification of intent to market the device wt nave reviewed your because is substantially equivalent (for the indications referenced above and nave actering only marketed predicate devices marketed in interstate 101 as suated in the encreated of the enactment date of the Medical Device Amendments, or to conninered prior to May 20, 1978, the excordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, mererere, mains of the Act include requirements for annual registration, listing of general volurold provider of of the startice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is exassinod (too as courols. Existing major regulations affecting your device can may be subject to back access Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Of R rate 077) wouldy systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
### Page 2 - Mr. Bob Rose
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
forCelia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): k020847
XTRAC Excimer Laser System, model AL7000 Device Name:
Indications for Use:
UVB Phototherapy for psoriasis, vitiligo, atopic dermatitis, and leukoderma
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter-Use _
Miriam C. Provost
Division of General, Restorative and Neurological Devices
510(k) Number II-2 K02089
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.