LARES RESEARCH SUNLASE 800P LASER SYSTEM (TO BE SOLD AS POCKETPRO OR POCKETPRO S4)
K011960 · Lares Research, Inc. · GEX · Dec 21, 2001 · General, Plastic Surgery
Device Facts
Record ID
K011960
Device Name
LARES RESEARCH SUNLASE 800P LASER SYSTEM (TO BE SOLD AS POCKETPRO OR POCKETPRO S4)
Applicant
Lares Research, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 21, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SunLase 800P laser device is to provide the ability to perform intraoral soft tissue dental, surgery. The device is indicated for ablating, incising, excising, vaporization and coagulation of soft tissues using a contact, fiber-optic delivery system. The device will be used in the following areas: general and cosmetic dentistry, otolaryngology, oral maxillo-facial and general and plastic surgery. The following are the oropharangeal indications for use for which the device will be marketed: Excisional and incisional biopsies; Excision and vaporization of herpes simplex I and II; Exposure of unerupted teeth; Fibroma removal; Frenectomy and Frenotomy; Gingival troughing for crown impressions; Gingivectomy; Gingivoplasty; Gingival incision and excision; Hemostasis; Implant recovery; Incision and drainage of abscess; Laser Assisted Uvulopaletoplasty (LAUP) - This laser is effective for cutting, ablating, coagulating and removing oropharangeal soft tissue that has been diagnosed as anatomically abnormal or naturally occurring hypertrophic which has been identified and confirmed as being associated with chronic palatal snoring; Leukoplakia; Operculectomy; Oral Papillectomies; Pulpotomy and Pulpotomy as an adjunct to root canal therapy; Reduction of denture hyperplasia; Reduction of gingival hypertrophy; Removal of filling material such as gutta percha or resin as adjunct treatment during root canal therapy; Removal of post-surgical granulations; Selective ablation of enamel (First Degree Caries Removal); Soft tissue crown lengthening; Sulcular debridement or soft tissue curettage (removal of diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss, and tooth mobility); Tissue retraction for impression; Treatment of Apthous ulcers; Vestibuloplasty.
Device Story
The SunLase 800P (PocketPro/PocketPro S4) is a pulsed Nd:YAG laser system (1064 nm) used for intraoral soft tissue surgery. It utilizes a flashlamp-pumped Nd:YAG crystal rod within an optical resonant cavity to generate laser energy, which is delivered via a contact fiber-optic cable. A HeNe laser provides a visible red aiming beam. The device is operated by a clinician in a dental or surgical setting. The operator uses a keyboard to control energy levels, pulse widths, and repetition rates. The output is a laser beam used for cutting, ablating, coagulating, and vaporizing soft tissue. The device benefits patients by providing precise tissue management for various dental and oropharyngeal conditions, including snoring treatment via LAUP.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and established clinical use of predicate Nd:YAG and diode laser devices.
Technological Characteristics
Pulsed Nd:YAG laser (1064 nm); flashlamp-pumped crystal rod; HeNe aiming beam; contact fiber-optic delivery system; microprocessor-controlled settings for energy, pulse width, and repetition rate; software-based safety interlocks for fault conditions.
Indications for Use
Indicated for patients requiring intraoral soft tissue dental surgery, including biopsies, frenectomy, gingivectomy, LAUP for chronic snoring, and periodontal curettage. Applicable to general/cosmetic dentistry, otolaryngology, and oral/maxillofacial surgery.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
American Dental Technologies - PulseMaster™
American Dental Technologies - DioLase ST™
BioLase Technology, Inc. - TwiLite™
CeramOptec - Ceralas® D15
Dentek Laser Systems - LD-15™
OpusDent, Ltd. - Opus 10™
Premier Laser Systems - Aurora™
Xintec Corporation - Dentica™
Submission Summary (Full Text)
{0}------------------------------------------------
011960 '/
#### Lares Research Applicant: 295 Lockheed Ave. Chico, CA 95973 Contact Person: Thomas H. Louisell Tel: (530) 345-1767 x262 Fax: (530) 345-1870 Date Prepared: 17 May 2001 PocketPro or PocketPro S4 Device Trade Name: Nd:YAG Pulsed Laser Common Name: Instrument. Surgical, Powered, Laser (79-GEX) Classification Name: Legally Marketed American Dental Technologies - PulseMaster™ Predicate Devices: American Dental Technologies - DioLase ST™ BioLase Technology, Inc. - TwiLite™ CeramOptec - Ceralas® D15 Dentek Laser Systems - LD-15 ™ OpusDent, Ltd. - Opus 10TM Premier Laser Systems - Aurora ™ Xintec Corporation - Dentica TM The laser device produces an invisible treatment beam for surgery and a Device Description: visible aiming beam. The pulsed Nd:YAG treatment beam is an invisible beam of light (1064 nanometer wavelength); a Helium-Neon (HeNe) laser within the laser head console supplies the red aiming beam. The laser energy is developed inside an optical resonant cavity. A crystal rod of neodymium-doped Yttrium Aluminum Garnet (pulsed Nd:YAG) is placed between two mirrors. The rod is "pumped" by a flashlamp that excites the Neodymium ions in the YAG crystal. Once excited, stimulated emission of photons occurs. Laser energy exits the cavity through a partially transmitting mirror and enters the distal end of the contact fiber-optic cable at the aperture. The laser energy is transmitted though the fiber-optic cable and exits the proximal tip of the fiber. The device is controlled by microprocessor via keyboard access by the operator, which allows various settings for energy levels, pulse widths and repetition rates. Software controls preclude emission of laser energy in any of a number of fault conditions. Summary Rationale: The SunLase 800P is a free running Nd:YAG laser based on the same technology as the predicate Nd:YAG lasers listed above. In addition, the predicate diode laser devices listed above operate clinically in substantially equivalent fashion; their operational technology differs only by source of emission, pulse rate, and wavelength. Tissue effects by predicate Nd. YAG and diode devices exhibit identical physical characteristics. As such, the
indications for use and intended uses are the same. This consolidation of
clinical applications presents no new safety issues.
## 510(k) Summary of Safety and Effectiveness
SunLase 800P Dental Laser
{1}------------------------------------------------
11960 i/2
Intended Uses:
The following are the oropharangeal indications for use for which the device will be marketed:
- · Excisional and incisional biopsies
- · Excision and vaporization of herpes simplex I and II
- · Exposure of unerupted teeth
- Fibroma removal
- · Frenectomy and Frenotomy
- · Gingival troughing for crown impressions
- · Gingivectomy
- · Gingivoplasty
- · Gingival incision and excision
- · Hemostasis
- · Implant recovery
- · Incision and drainage of abscess
- · Laser Assisted Uvulopaletoplasty (LAUP)
- This laser is effective for cutting, ablating, coagulating and removing oropharangeal soft tissue that has been diagnosed as anatomically abnormal or naturally occurring hypertrophic which has been identified and confirmed as being associated with chronic palatal snoring.
- · Leukoplakia
- · Operculectomy
- · Oral Papillectomies
- · Pulpotomy and Pulpotomy as an adjunct to root canal therapy
- · Reduction of denture hyperplasia
- · Reduction of gingival hypertrophy
- · Removal of filling material such as gutta percha or resin as adjunct treatment during root canal therapy
- · Removal of post-surgical granulations
- · Selective ablation of enamel (First Degree Caries Removal)
- · Soft tissue crown lengthening
- · Sulcular debridement or soft tissue curettage (removal of diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss, and tooth mobility)
- · Tissue retraction for impression
- Treatment of Apthous ulcers
- · Vestibuloplasty
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features the department's seal, which includes a stylized caduceus symbol. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the seal.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 1 2001
Mr. Thomas H. Louisell Director, Regulatory Affairs and Quality Systems Lares Research, Inc. 295 Lockheed Avenue Chico, California 95973
Re: K011960
Roll 200 PocketPro S4® Regulation Number: 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: September 28, 2001 Received: October 5, 2001
Dear Mr. Louisell:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass surve in the encreativent of the enactment date of the Medical Device Amendments, or to conninered prior to ria) 2011-11-11 in accordance with the provisions of the Federal Food, Drug, de nees that have been require approval of a premarket approval application (PMA). and Oosmeter For ( 100) kes the device, subject to the general controls provisions of the Act. The r ou may, arorely mails of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 – Mr. Thomas H. Louisell
This letter will allow you to begin marketing your device as described in your Section 510(k) I ms letter whil anow you to begin manisting of substantial equivalence of your device to a legally premarket notification. The PDF micing of casion for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and II you desire specific au vice for your avitro diagnostic devices), please contact the Office of additionally 21 CFR Fall 607.10 for m. vial v. Secondition on the promotion and advertising of Compliance at (301) 594 1057. Themsel Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Tegulation citined, "Misormation on your responsibilities under the Act may be obtained from the Other general information on your respentional and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Hyk Qurdu
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### 510 (k) NUMBER: K011960
# DEVICE NAME: SunLase 800P Laser System (to be sold as "PocketPro®" or "PocketPro S4®"
### INDICATIONS FOR USE:
The SunLase 800P laser device is to provide the ability to perform intraoral soft tissue dental, The SunLase oool Tason downer is to pro surgery. The device is indicated for ablating, incising, general, oral maxilio-raolar and coatine soft tissues using a contact, fiber-optic delivery system. excising, vaponzation and ougalation oreas: general and cosmetic dentistry, otolaryngology, The device will be used in the following are the oropharangeal indications for use for which the device will be marketed:
- · Excisional and incisional biopsies
- · Excision and vaporization of herpes simplex I and II
- · Exposure of unerupted teeth
- Fibroma removal
- · Frenectomy and Frenotomy
- · Gingival troughing for crown impressions
- · Gingivectomy
- · Gingivoplasty
- · Gingival incision and excision
- · Hemostasis
- · Implant recovery
- · Incision and drainage of abscess
- · Laser Assisted Uvulopaletoplasty (LAUP)
- This laser is effective for cutting, ablating, coagulating and removing This labor to off tissue that has been diagnosed as anatomically abnormal or oropharangour our the which has been identified and confirmed as being associated with chronic palatal snoring.
- Leukoplakia
- · Operculectomy
- · Oral Papillectomies
- Pulpotomy and Pulpotomy as an adjunct to root canal therapy
- · Reduction of denture hyperplasia
- · Reduction of gingival hypertrophy
- · Reduction of filling material such as gutta percha or resin as adjunct treatment during root canal therapy
- Removal of post-surgical granulations
- · Selective ablation of enamel (First Degree Caries Removal)
- · Soft tissue crown lengthening
- Sulcular debridement or soft tissue curettage (removal of diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss, and tooth mobility)
- · Tissue retraction for impression
- · Treatment of Apthous ulcers
- · Vestibuloplasty
{5}------------------------------------------------
Page_2 of_2
Signature :
Thomas H. Fricolt
Name : Title : Date :
Thomas H. Louisell Director - Regulatory Affairs & Quality Systems 17 May 2001
Concurrence of CDRH, Office of Device Evaluation (ODE)
C Prescription Use (Per 21 CFR 801.109)
OR
Over-the-Counter-Use (Optional Format 1-2-96)
Styph Charly
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number k011960
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.