K211170 · Modern Medical Equipment Manufacturing, Ltd. · GEI · Oct 19, 2021 · General, Plastic Surgery
Device Facts
Record ID
K211170
Device Name
Single Use Irrigation bipolar cable
Applicant
Modern Medical Equipment Manufacturing, Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 19, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is intended to provide irrigation and energy simultaneously to standard bipolar forceps specifically designed for irrigation. The device is used with Malis bipolar coagulators or the compatible generators and the Malis CMC-II Irrigation Module or The Malis irrigation module 1000.
Device Story
Single-use irrigation bipolar cable; connects irrigation-capable bipolar forceps to electrosurgical generators and irrigation modules. Device delivers high-frequency electrical current for coagulation while simultaneously supplying saline irrigation to forceps tips. Used in clinical settings by surgical staff. Socket connector fits Codman-style forceps; fixed 2-pin plug or flying lead connectors interface with generators; tubing integrates with Malis irrigation modules. Facilitates precise tissue coagulation with concurrent irrigation to prevent tissue sticking and thermal damage; improves surgical efficiency and patient safety.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2), biocompatibility (ISO 10993-1), and sterile barrier integrity/shelf-life testing (ASTM F1980-16, ISO 11607-1).
Technological Characteristics
Single-use cable assembly combining irrigation tubing and bipolar electrical leads. Materials: Copper wire, PVC insulation, PC luer lock, ABS puncture utensil, PP pinch clamp. Rated voltage: 500Vp. Sterilization: Ethylene Oxide (EO). Connectivity: Mechanical interface to electrosurgical generators and irrigation modules. Complies with IEC 60601-1, IEC 60601-2-2, and IEC 60601-1-2 standards.
Indications for Use
Indicated for use with standard bipolar forceps designed for irrigation, in conjunction with Malis bipolar coagulators or compatible electrosurgical generators and Malis irrigation modules, to provide simultaneous irrigation and energy delivery during surgical procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Codman Integrated Irrigation Tubing and Bipolar Cord Set (K052449)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 19, 2021
Modern Medical Equipment Manufacturing, LTD. Philip Hung Management Representative Flat A, 11/F., Mai Wah Ind. Bldg., 1-7 Wah Sing Street, Kwai Chung, N.T. Hong Kong, China
Re: K211170
Trade/Device Name: Single Use Irrigation bipolar cable Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: September 14, 2021 Received: September 20, 2021
Dear Philip Hung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent ffor the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (0S) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211170
Device Name
Sterile Single Use Irrigation Bipolar Cable
#### Indications for Use (Describe)
The device is intended to provide irrigation and energy simultaneously to standard bipolar forceps specifically designed for irrigation. The device is used with Malis bipolar coagulators or the compatible generators and the Malis CMC-II Irrigation Module or The Malis irrigation module 1000.
Type of Use (Select one or both, as applicable)
| <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|------------------------------------------------------------------|-----------------------------------------------------------------|
|------------------------------------------------------------------|-----------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 1. 510(k) Owner<br>Name: | Modern Medical Equipment Manufacturing Limited | |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|--|
| Address: | Modern Medical Equipment Mfg., Ltd.<br>Flat A, 11/F., Mai Wah Ind. Bldg.,<br>1-7 Wah Sing Street,<br>Kwai Chung, N.T., Hong Kong, China. | |
| Telephone: | (852) 2420 9068 | |
| Fax: | (852) 2481 1234 | |
| Contact person: | Philip Hung | |
| Date of preparation: | 16 October 2021 | |
| 2. Device<br>Name of Device: | Sterile Single Use Irrigation Bipolar Cable | |
| Trade or proprietary name: | Sterile Single Use Irrigation Bipolar Cable | |
| Common or usual name: | Sterile Single Use Irrigation Bipolar Cord | |
| Classification name: | Electrosurgical Cutting & Coagulation<br>Device & Accessories | |
| Classification number: | 21 CFR 878.4400 | |
| Classification Panel: | General & Plastic Surgery | |
| Product Code: | GEI | |
| Class: | II | |
## 3. Predicate device
Codman Integrated Irrigation Tubing and Bipolar Cord Set with 510(k) number K052449.
## 4. Device description
The device is single use and is used to connect an irrigation bipolar forceps to electrosurgical generator and irrigation module for coagulation and delivering saline to the tip of forceps simultaneously.
The socket connector of the cable is to fit Codman irrigating bipolar forceps. The fixed 2-pin plug is connected to any compatible generators or the flying lead connectors are connected to Malis bipolar coagulators or any compatible generators. The tubing set of the cable can be used with the Malis CMC-II irrigation module or the Malis
{4}------------------------------------------------
irrigation module 1000.
# 5. Indication for use
The device is intended to provide irrigation and energy simultaneously to standard bipolar forceps specifically designed for irrigation. The device is used with Malis bipolar coagulators or the compatible generators and the Malis CMC-II Irrigation Module or the Malis irrigation module 1000.
## 6. Substantial Equivalence
The technological characteristics and performance testing of the subject and predicate devices are substantially equivalent. The following table shows the comparisons in more detail information among the subject devices and the predicate device.
| Compared<br>Items | Proposed Device<br>(K211170)<br>Single Use<br>Irrigation bipolar<br>cable (CD933) | Predicate Device<br>(K052449)<br>Coeman Integrated<br>Irrigation Tubing and<br>Bipolar Cord Set | Comments on<br>difference |
|-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Indication for<br>use | The device is<br>intended to provide<br>irrigation and<br>energy<br>simultaneously to<br>standard bipolar<br>forceps specifically<br>designed for<br>irrigation. The<br>device is used with<br>Malis bipolar<br>coagulators or the<br>compatible<br>generators and the<br>Malis CMC-II<br>Irrigation Module<br>or the Malis<br>irrigation module<br>1000. | The device is intended to<br>provide irrigation and<br>energy simultaneously to<br>bipolar forceps specifically<br>designed for irrigation.<br>They are intended for use<br>with Codman/Malis CMC-II<br>and the Codman/Malis<br>CMC III I.E.C. Irrigation<br>Modules and the<br>Codman/Malis Bipolar<br>Coagulators. | Same |
| Regulation<br>number | 21 CFR 878.4400 | 21 CFR 878.4400 | Same |
| Product code | GEI | GEI | Same |
| OTC or<br>prescription | For prescription<br>use | For prescription use | Same |
| Energy<br>delivery | High frequency<br>electrical<br>current/energy | High frequency electrical<br>current/energy | Same |
| Technology | For monopolar<br>electrosurgery | For monopolar<br>electrosurgery | Same |
| Design | Combined of<br>Irrigation tubing<br>with bipolar cable | Combined of Irrigation<br>tubing with bipolar cable | Same |
| Material of the<br>components | | | |
| -<br>Cable wire<br>with plug | Copper wire and<br>PVC insulation | Copper wire and PVC<br>insulation | Same |
| -<br>Tubing | PVC | PVC | Same |
| -<br>Luer lock | PC | PC | Same |
| -<br>Puncture<br>utensil | ABS | ABS | Same |
| -<br>Protector<br>for luer and<br>puncture<br>utensil | ABS | LDPE/PP | Similar,<br>provide the<br>same function |
| -<br>Pinch<br>clamp | PP | PP | Same |
| Rated<br>accessory<br>voltage | 500Vp | 500Vp-p | Both meets<br>IEC safety<br>requirements |
| Sterile | EO sterile | EO sterile | Same |
| Shelf life | 3 years | 3 years | Same |
| Electrical<br>safety and<br>EMC | Comply with<br>dielectric strength<br>in with accordance<br>IEC60601-1,<br>IEC60601-1-2 &<br>IEC60601-2-2 | Comply with dielectric<br>strength in with accordance<br>AAMI HF-18 | Same |
| Biocompatibility | Comply with<br>ISO10993 | Comply with ISO10993 | Meet<br>biocompatibility<br>requirements so<br>it does not raise<br>any safety issue<br>for |
{5}------------------------------------------------
{6}------------------------------------------------
# 7. Non-clinical Performance Testing Data
Validation and Verification testing was performed on device sterility in accordance with ISO 11135:2014.
The safety performance of the subject device passed all the testing according to internal requirements and international standards shown below to support the substantial equivalence of the subject device
- IEC 60601-1:2005+AMD1:2012 CSV Medical electrical equipment Part 1: -General requirements for basic safety and essential performance
- -IEC60601-2-2: 2017 (Fifth Ed), Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surqical equipment and high frequency surqical accessories.
- IEC60601-1-2:2014, Medical Electrical Equipment Part 1-2: General -Requirements for Basic Safety and Essential Performance - Collateral Standard. Electromagnetic Disturbances - Requirements and Tests
- -ISO10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- -ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
- ISO 11607-1:2019 Packaging for terminally sterilized medical devices — Part 1: Requirements for materials, sterile barrier systems and packaging systems
Furthermore, the packaging integrity and accelerated aging test were completed on the subject device to support the proposed shelf life.
## 8. Conclusion
The non-clinical testing demonstrates that the subject device, the Single Use Irrigation Bipolar Cable, is substantially equivalent as the predicate device (K052449).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.