MicroFrance Monopolar and Bipolar Electrosurgical Instruments
K210942 · Integra Microfrance · GEI · May 14, 2021 · General, Plastic Surgery
Device Facts
Record ID
K210942
Device Name
MicroFrance Monopolar and Bipolar Electrosurgical Instruments
Applicant
Integra Microfrance
Product Code
GEI · General, Plastic Surgery
Decision Date
May 14, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The electrosurgical instruments are intended to remove tissue and/or control bleeding.
Device Story
MicroFrance Bipolar and Monopolar Electrosurgical Instruments consist of forceps, probes, knives, suction tubes, hooks, elevators, and picks. Devices connect to an external electrosurgical generator and cord to deliver electrical current to distal tips for tissue cutting and coagulation. Used in laparoscopic and open surgical procedures by surgeons. Reusable instruments available in unibody or modular designs. Output is thermal energy for tissue effect. Benefits include effective tissue removal and hemostasis during surgery.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by bench testing, including lesion size comparison, thermal effects on tissue (heart, liver, kidney), physical characterization, dimensional verification, functional testing, electrical/mechanical testing, and biocompatibility (ISO 10993-5, -10, -11, -17, -18).
Technological Characteristics
Reusable electrosurgical instruments (forceps, probes, knives, etc.). Materials: Blue Polyamide (Rislan ESY BLUE 7414) insulation, Polypropylene (Propylux) bipolar connector. Energy: High-frequency electrical current from external generator. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 60601-2-18. Sterilization: Steam.
Indications for Use
Indicated for removal of tissue and/or control of bleeding in patients undergoing laparoscopic or open surgery.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 14, 2021
Integra MicroFrance c/o Ms. Malena Zammetti Regulatory Affairs Specialist II Integra LifeSciences Corporation 1100 Campus Rd. Princeton, NJ 08540
Re: K210942
Trade/Device Name: MicroFrance Monopolar and Bipolar Electrosurgical Instruments Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI, Dated: March 29, 2021 Received: March 30, 2021
Dear Ms. Zammetti:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K210942
#### Device Name
MicroFrance® Bipolar and Monopolar Electrosurgical Instruments
Indications for Use (Describe)
The electrosurgical instruments are intended to remove tissue and/or control bleeding.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Integra. The word "INTEGRA" is written in gray, block letters. Below the word "INTEGRA" is the phrase "LIMIT UNCERTAINTY" in smaller, gray letters. To the right of the word "INTEGRA" are several green squares of varying sizes arranged in a column.
# 510(k) Summary
| I. Submitter Information | |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter | Integra LifeSciences Corporation<br>1100 Campus Rd.<br>Princeton, NJ, USA 08540<br><br>On behalf of:<br>Integra MicroFrance<br>Le Pavillon, Saint Aubin Le Monial Allier, 03160<br>France<br>Contact: Olivier Doizon, Manager, Quality<br>+33 (0) 470 67 98 0<br>olivier.doizon @ integralife.com |
| U.S. Contacts: | Primary Contact:<br>Ms. Malena Zammetti<br>Regulatory Affairs Specialist II<br>717-818-8774<br>malena.zammetti @ integralife.com<br><br>Secondary Contact:<br>Ms. Jocelyn Raposo<br>Director, Regulatory Affairs<br>508-813-7015<br>Jocelyn.Raposo @ integralife.com |
| Establishment Registration<br>Number: | 9680837 |
| Date 510(k) Summary Prepared: | May 14, 2021 |
| II. Device | |
| Trade or Propriety Name: | MicroFrance® Bipolar and Monopolar Electrosurgical<br>Instruments |
| Common or Usual Name: | Electrosurgical cutting and coagulation accessories |
| Classification Name: | Electrosurgical, Cutting & Coagulation & Accessories |
| Classification Panel: | General and Plastic Surgery |
| Regulation: | Class II, 21 CFR 878.4400 |
| Product Code: | GEI |
{4}------------------------------------------------
## III. Predicate Device
The predicate devices for this submission are the MicroFrance Electrosurgical Instruments, K993655, which were cleared on 12/21/1999.
## IV. Device Description
The MicroFrance® Bipolar and Monopolar Electrosurgical Instruments consist of forceps, probes, knives, suction tubes, hooks, elevators and picks used for laparoscopic access and open surgery. These electrosurgical instruments are reusable and available in unibody (one-piece) and dismantlable (modular) designs. They are used as part of an electrosurgical system consisting of a generator and cord attached to the proximal end of the devices to provide power and deliver electrical current from the generator to the distal tips of the devices.
### V. Indications for Use
The electrosurgical instruments are intended to remove tissue and/or control bleeding.
### VI. Comparison of Characteristics with the Predicate Devices
The MicroFrance® Bipolar and Monopolar Electrosurgical Instruments are substantially eqivalent to the predicate devices, the Micro-France Electrosurgical Instruments. The subject devices have the same indications for use, operating principles, clinical utility and similar design specifications and materials as the predicate devices. To comply with the latest electrical safety and EMC standards, material changes and minor componenet changes were made. The componenet changes include minor dimensional changes to subassemblies to ensure proper final assembly of the subject devices and to maintain the same finished dimensional specifications as the predicate devices. The table below provides a comparison between the subject devices and the predicate devices.
| Comparison of the Predicate and Subject Device | | | |
|---------------------------------------------------------|---------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| | Predicate Device:<br>Micro-France<br>Electrosurgical<br>Instruments, Various<br>(K993655) | Subject Device:<br>MicroFrance® Bipolar and<br>Monopolar Electrosurgical<br>Instruments<br>(This Submission) | Difference and Justification |
| FDA Product Code | GEI | Same as predicate | No difference |
| Classification | Class II - 21 CFR 878.4400 | Same as predicate | No difference |
| Classification Name | Electrosurgical cutting and<br>coagulation device and<br>accessories | Same as predicate | No difference |
| Indications for Use | The electrosurgical<br>instruments are used to<br>remove tissue and/or control<br>bleeding. | Same as predicate | No difference |
| Type | Various dismantlable<br>(modular) and unibody<br>bipolar and monopolar<br>instruments | Same as predicate | No difference |
| Device Sterility | Non-Sterile | Same as predicate | No difference |
| Reusable | Yes | Same as predicate | No difference |
| Reprocessing<br>Methods (cleaning<br>and sterilization) | Manual and Automated<br>cleaning and<br>Steam sterilization | Same as predicate | No difference |
| Design | Multiple bipolar and<br>monopolar instrument | Changes to the insulation<br>coating materials and thickness; | The performance, electrical<br>safety and EMC test results |
{5}------------------------------------------------
| | designs as cleared per<br>K993655 | changes to the monopolar pin<br>connector and bipolar screw set<br>alignment; and Minor<br>dimensional changes to<br>subassembly components to<br>accommodate increase in<br>thickness of Rislan insulation<br>coating. | demonstrate that the subject<br>devices do not raise any new<br>questions of safety and are<br>substantially equivalent to the<br>predicate devices. |
|-----------|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Materials | Materials as cleared per<br>K993655 | Materials remain the same as the<br>predicate device except for the<br>insulation coating and the<br>bipolar connector material<br><br>Insulation coating: From Rislan®<br>ES BLUE 7413 M<br>(Blue Polyamide) to Rislan®<br>ESY BLUE 7414<br>(Blue Polyamide)<br><br>Bipolar connector: From PEEK<br>(Polyether ether ketone) to<br>Propylux® (Polypropylene)<br><br>Monopolar Connector:<br>Connector pin is partially coated<br>with Rislan® ESY BLUE 7414 | The performance, electrical<br>safety, EMC, and<br>biocompatibility test results<br>demonstrate that the subject<br>devices do not raise any new<br>questions of safety and<br>effectiveness and are<br>substantially equivalent to the<br>predicate devices. |
| Packaging | Packaging types as cleared<br>per K993655 | Same as predicate | No difference |
## VII. Performance Data
The following bench, electrical safety, electromagnetic compatibility (EMC), and biocompatibility testing has been performed in support of the substantial equivalence determination.
| Performance Bench Testing | | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test | Test Method Summary | Results |
| Lesion Size | To demonstrate the subject devices,<br>monopolar and bipolar, are effective<br>for their intended use by evaluating<br>coagulation performance. | Results from side-by-side testing of the predicate<br>and subject devices prior to reprocessing showed<br>no significant difference in the lesion size.<br>supporting substantial equivalence between the<br>subject and predicate devices. Test results of the<br>subject device after 150 reprocessing cycles<br>demonstrates that the subject device continues to<br>meet user requirements. |
| Thermal Effects on Lesion<br>Study | To characterize the performance of the<br>subject devices by measuring the<br>typical lesion size (length, width and<br>depth) they generate on a range of<br>tissue densities ( heart, liver and<br>kidney) at three power settings:<br>minimal, default and maximal. | All test samples, at various power settings, on<br>three different tissue types showed coagulation<br>was effectively applied by the subject devices,<br>supporting substantial equivalence between the<br>subject and predicate devices. |
| Physical Characterization | Perform a visual inspection of the<br>physical characteristics for defects. | No defects were observed on the subject devices<br>during the visual inspection of the physical<br>characteristics, supporting substantial<br>equivalence between the subject and predicate<br>devices. |
| Dimensional Verification | Measure the device attributes and<br>confirm they are within defined | All measured dimensions are within defined<br>tolerances and specifications for the subject |
| | tolerances and specifications. | devices, supporting substantial equivalence<br>between the subject and predicate devices. |
| Functional Testing | Verify key functionality of the subject<br>devices remains similar to the<br>predicate devices. | All test samples passed the acceptance criteria<br>for key functionality of the subject devices,<br>supporting substantial equivalence between the<br>subject and predicate devices. |
| Electrical and Mechanical<br>Testing | Verify no failures occur during<br>electrical and resistivity testing of the<br>subject devices.<br>Measure the applied closing forces are<br>similar for the subject and predicate<br>devices. | All test samples passed the acceptance criteria<br>for electrical, resistivity and applied closing<br>force testing of the subject devices, supporting<br>substantial equivalence between the subject and<br>predicate devices. |
| Biocompatibility Testing | | |
| Test | Test Method Summary | Results |
| Cytotoxicity Study using the<br>ISO Elution Method<br>(1X MEM)<br>ISO 10993-5 | To evaluate the potential cytotoxic<br>effect of a test article extract using an<br>in vitro mammalian cell culture. | All products tested passed the acceptance criteria<br>demonstrating that the devices are biocompatible<br>and therefore are substantially equivalent to the<br>predicate devices. |
| ISO Guinea Pig Maximization<br>Sensitization Test<br>ISO 10993-10 | To evaluate the potential of the test<br>article extracts to cause delayed<br>dermal contact sensitization in the<br>guinea pig. | All products tested passed the acceptance criteria<br>demonstrating that the devices are biocompatible<br>and therefore are substantially equivalent to the<br>predicate devices. |
| ISO Intracutaneous Study in<br>Rabbits<br>ISO 10993-10 | To evaluate the potential of test article<br>extracts to induce local dermal<br>irritation following intracutaneous<br>injection in rabbits. | All products tested passed the acceptance criteria<br>demonstrating that the devices are biocompatible<br>and therefore are substantially equivalent to the<br>predicate devices. |
| Acute Systemic Toxicity<br>ISO 10993-11 | To evaluate the acute systemic toxicity<br>of the test article extracts following<br>intravenous or intraperitoneal injection<br>in mice. | All products tested passed the acceptance criteria<br>demonstrating that the devices are biocompatible<br>and therefore are substantially equivalent to the<br>predicate devices. |
| Chemical Characterization of<br>Extractables<br>ISO 10993-18 | To perform a chemical<br>characterization to identify and<br>quantitate the extractables and/or<br>leachables that may be released from<br>the test articles. | Extractables were identified and quantified, any<br>extractables above the Quantification Limit (QL)<br>underwent a toxicological risk assessment and<br>were deemed to be clinically acceptable. |
| Toxicology Risk Assessment<br>ISO 10993-17 | To assess extractables identifed during<br>chemical characterization testing. | Based on the chemical results evaluated in the<br>toxicological risk assessments, the risk of<br>induced toxicity during clinical use of the subject<br>devices had been deemed to be clinically<br>acceptable. |
{6}------------------------------------------------
## Bench Testing
Performance bench testing was conducted in alignment with the FDA's guidance document, "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery" issued March 9, 2020, and included the following testing: lesion size, thermal effects on lesion study, physical characterization, dimensional, verification, functional, electrical and mechanical testing. Lesion size testing of the predicate and subject devices prior to reprocessing showed no significant difference in the lesion size between the subject and predicate devices. Test results of
{7}------------------------------------------------
the subject device after 150 reprocessing cycles demonstrates that the subject device continues to meet user requirements. Thermal effects testing was performed using three tissue types (heart, liver and kidney). All test samples, at various power settings, on the three different tissue types showed coagulation was effectively applied by the subject devices. No defects were observed on the subject devices during the visual inspection of the physical characteristics. The dimensional verification confirmed that all measured dimensions are within defined tolerances and specifications for the subject devices. All functional testing samples passed the acceptance criteria for key functionality of the subject devices. The electrical and mechanical test samples passed the acceptance criteria for electrical, resistivity and applied closing force. These bench test results demomstrate substantial equivalence between the subject and predicate devices.
## Electrical safety and electromagnetic compatibility (EMC)
EMC and electrical safety testing were conducted in accordance with IEC 60601-1-2:2014, Edition 4.0, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests, ANSI/AAMI/ES60601-1:2005/ (R)2012 and A1:2012, C1:2009/ (R)2012 and A2:2010/(R)2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD), AAMI/ANSI/IEC 60601-2-2:2017, Edition 6.0, Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories, IEC 60601-1-6:2013, Edition 3.1, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability, and IECO 60601-2-18:2009, Edition 3.0, Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment. The MicroFrance® Bipolar and Monopolar Electrosurgical Instruments passed all EMC and electrical safety testing.
## Software Verification and Validation Testing
The MicroFrance® Bipolar and Monopolar Electrosurgical Instruments do not contain software.
### Biocompatibility Testing
Biocompatibility testing was conducted in accordance with ISO 10993-1:2018. Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process and FDA's guidance documents, Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process" issued September 4, 2020 and Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery issued March 9, 2020.
### Animal Studies
No animal studies were performed, as appropriate verification of the changes was achieved based on the comparison to the predicate devices and from the results of the bench testing. biocompatibility evaluation, electromechanical compatibility (EMC), and electrical safety testing.
### Clinical Studies
No clinical studies were performed, as appropriate verification and validation of the changes was
{8}------------------------------------------------
achieved based on the comparison to the predicate devices and from the results of the performance testing, biocompatibility evaluation, electromechanical compatibility (EMC), and electrical safety testing.
### VII. Conclusions
The proposed MicroFrance® Bipolar and Monopolar Electrosurgical Instruments are identical in intended use, indications, operating principles, technology, and clinical utility compared to the predicate devices. The changes to the subassembly dimensional specifications and material specifications have been verified and validated. The test results demonstrate that the subject devices do not raise new questions of safety and effectiveness and are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.