K202988 · Novoxel , Ltd. · GEI · Feb 25, 2021 · General, Plastic Surgery
Device Facts
Record ID
K202988
Device Name
Tixel System
Applicant
Novoxel , Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Feb 25, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Tixel® System is intended for dermatological procedures requiring ablation and resurfacing of the skin.
Device Story
Tixel System is a thermo-mechanical fractional skin treatment device for dermatological ablation and resurfacing. Device consists of a console, handpiece, and interchangeable tips (81-pin standard or 24-pin small). Tips feature biocompatible titanium square pyramids on a gold-plated copper base, heated by an underlying flat ceramic element. Clinician selects treatment parameters (pulse duration 5-16ms, speed, distance) via console touchscreen. Handpiece delivers heat via direct conduction through discrete, non-overlapping pyramids contacting skin, creating a matrix of coagulation sites and thermal necrosis. Used in clinical settings by trained professionals. Output is localized thermal energy; clinical benefit is controlled skin resurfacing and ablation. System tracks pulse count per tip to ensure performance integrity.
Clinical Evidence
Bench testing and in vivo porcine study. Porcine study (n=5) compared Tixel to Venus Viva predicate across low, medium, and high settings. Primary endpoints: lesion depth, width, inflammatory reaction, and healing (edema, necrosis, hemorrhage). Results: Tixel and predicate demonstrated similar immediate skin changes and histological lesion profiles. Average thermal lesion depth was similar at medium and high settings. Bench testing confirmed electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and software verification.
Technological Characteristics
Thermo-mechanical fractional skin treatment. Materials: biocompatible titanium pyramids, gold-plated copper base, ceramic heating element. Energy: resistive heating via electrical console. Dimensions: 81-pin or 24-pin arrays. Connectivity: console-to-handpiece via umbilical. Sterilization: dry heat (ISO 20857). Software: embedded controller for pulse management.
Indications for Use
Indicated for dermatological procedures requiring ablation and resurfacing of the skin in patients requiring such treatment.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Venus Concept USA Ltd., Venus Viva SR Device (K150161)
Reference Devices
CONMED Altrus® Thermal Tissue Fusion System (K101534)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "FDA U.S. FOOD & DRUG" on the top line and "ADMINISTRATION" on the second line.
February 25, 2021
Novoxel Ltd. % Janice Hogan Partner Hogan Lovells. US LLP 1735 Market Street, Floor 23 Philadelphia, Pennsylvania 19103
Re: K202988
Trade/Device Name: Tixel System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: January 28, 2021 Received: January 28, 2021
Dear Janice Hogan:
We have reviewed your Section 5100k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202988
Device Name
Tixel® System
Indications for Use (Describe)
The Tixel® System is intended for dermatological procedures requiring ablation and resurfacing of the skin.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
[ ] Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY Novoxel's Tixel® Svstem K202988
### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Novoxel. Ltd. 43 Hamelaha Street Netanya, Israel 4250573 Phone: + 972 9 7739074 Facsimile: +972 9 793 9953 Contact Person: Ronen Shavit, CTO
#### Date Prepared:
February 24, 2021
#### Name of Device
Tixel® System
#### Common or Usual Name
Thermo-mechanical fractional skin treatment device
#### Classification Name
21 CFR 878.4400, Class II, Product Code: GEI
### Predicate Devices
Venus Concept USA Ltd., Venus Viva SR Device (K150161) (Predicate Device)
CONMED Altrus® Thermal Tissue Fusion System (K101534) (Reference device)
#### Intended Use / Indications for Use
The Tixel® System is intended for dermatological procedures requiring ablation and resurfacing of the skin.
#### Device Description
The Tixel is a thermo-mechanical fractional skin treatment device that is designed to perform ablative fractional skin treatment is achieved by transfer of energy in the form of heat to the skin to create coagulation sites.
The treatment is applied through an operating Tip that consists of 81 (standard tip) or 24 (small tip) biocompatible titanium square pyramidal shape assembled over gold plated copper base that are heated by an underlying flat ceramic heating element. The desired skin treatment is achieved by defining the speed and distance at which the Tip contacts and pushes the skin and the amount of pulses performed. Based on the treatment parameters selected the pyramids contact the skin surface in 81 (or 24 for small tip) discrete, non-overlapping areas and by the transfer of heat, a matrix of coagulation sites and thermal necrosis is generated.
{4}------------------------------------------------
Tixel transfers heat to the tissue by direct conduction to target tissue in a localized manner via the discrete non-overlapping pyramids. The Tixel Tip contacts the skin for a very short period. defined by pulse duration that varies between 5 and 16ms.
The Tip is mounted directly above the heating element in the Handpiece. The Tip and the heating element are mounted on a titanium alloy assembly. The Tip assembly structure is rigidly connected to the motor shaft (also known as forcer) by a polymer component. The polymeric components provide both thermal and electrical insulation between the heating element and the motor shaft. The Handpiece is attached to the Tixel Console by an umbilical tube. The console touch screen is used to control the system parameters.
### Technological Characteristics
Both the Tixel and its predicate device transfer energy through a handpiece attached to a system console, where the output is controlled by the clinician operating the device to achieve the desired effect. The tips of the Tixel and the predicate device are comparable in physical design, i.e., they consist of a geometrical array of pins or pyramids. The size of the tip is comparable between the two devices. All patient contacting materials for both the subject and predicate device have been found to be biocompatible. The Tixel and its predicate device deliver non-invasive fractional skin treatments through similar methods of application (handpiece/Tip applied to the area with treatment parameters controlled through the console). For both devices, pulses are delivered to the target location using an array of pins or pyramids.
The predicate device utilizes fractional radiofrequency pulses emitted from the tip to achieve thermolysis, whereas the Tixel delivers fractional thermal energy through heat conduction. This difference does not result in different types of safety or effectiveness questions because the same key questions are raised by both methods of energy application. Both devices are intended to deliver a localized thermal effect with limited heat delivered to adjacent tissue, in a fractional pattern. Both energy delivery methods result in local heating of the skin to cause thermally induced tissue coagulation and ablation.
In addition, using metallic elements to deliver thermal energy has been previously cleared for other electrosurgical cutting and coagulation devices, such as the CONMED Altrus® reference device. For both the Tixel and the CONMED Altrus, the energy is delivered to the handpiece and increases the temperature of the heating elements by means of resistive heating.
# Biocompatibility
The Tixel is intended for treatment of soft tissue and dermal application (e.g., skin resurfacing treatment). The only patient contacting components of the Tixel include the tip and the handle. According to ISO-10993-1 Biological Evaluation of Medical Devices, and the FDA's Biocompatibility Guidance, Tixel is a surface device (≤ 24 hours contact duration) in contact with skin. Therefore, cytotoxicity, sensitization, and irritation testing have been conducted. All tests showed passing results. In addition, surface chemical analysis was performed to assess for residue buildup or chemical change over a lifetime of use. The results showed that no diffusion of the underlying materials could be detected.
### Cleaning/Disinfection/Shelf life
The Tixel is a reusable device. The user is required to clean and disinfect the device after each use. The Tip undergoes high level disinfection in accordance with ISO 20857 via dry heat, which has been shown to achieve complete kill of at least 10° Bacillus atrophaeus spores. Cleaning validation has been conducted in accordance with ASTM F3208-18: Standard Guide
{5}------------------------------------------------
for Selecting Test Soils for Validation of Cleaning Methods for Reusable Medical Devices, and was successfully passed.
The Tixel is designed to limit the number of pulses per a single Tip to 100,000 pulses. The number of pulses for each handpiece is stored in the handpiece controller. Shelf life testing following real life operation showed that there was no evidence of deterioration of the handpieces following 250,000 cycles.
### Performance Data
Electrical safety and electromagnetic compatibility (EMC) testing for Tixel was conducted by an independent test laboratory in accordance with IEC 60601-1. Medical electrical equipment. Part 1: General requirements for basic safety and essential performance and IEC 60601-1-2. Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests.
Software verification testing was conducted, and the testing results were found acceptable for software release.
Bench testing demonstrated that Tixel performs according to specifications and functions as intended.
# Animal Study
An in vivo study using a porcine model was conducted to evaluate performance of the Tixel device in comparison to the predicate Venus Viva device. The study comprised five female domestic pigs, each subjected to multiple topical exposures of fractional thermal pulses via Tixel or fractional radiofrequency via the predicate device, at different treatment settings (Low, Medium, and High). Macroscopic evaluation of the activation sites was performed, and the activation sites were harvested at the end of the 14-days study period. Harvested samples were processed histologically and evaluated for depth and width of the lesions, inflammatory reaction, and healing process (i.e. edema, necrosis, hemorrhage and polynuclear infiltration).
Tixel-treated and predicate-treated animals demonstrated similar immediate skin changes. Histopathology demonstrated that the Tixel created a focus of ablation over a wedge-shaped dermal area of collagen coagulation. Data analysis showed increased energy levels led to proportionately greater thermal lesions. The average thermal lesion depth of the Tixel was similar to the Venus Viva predicate at the medium and high settings the similarity of the treatment effects.
Taken together, the animal study results show that the Tixel displays a similar treatment effect compared to the predicate.
# Conclusions
The Tixel has the same intended use, technological characteristics, and principles of operation as its predicate device. The technological differences between the Tixel and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the differences in technological characteristics compared to the predicate do not adversely impact performance. Therefore, the Tixel is substantially equivalent to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.