K202880 · Apyx Medical Corporation (Formerly Bovie Medical Corporation · GEI · Jan 4, 2021 · General, Plastic Surgery
Device Facts
Record ID
K202880
Device Name
J-Plasma Precise FLEX Handpiece
Applicant
Apyx Medical Corporation (Formerly Bovie Medical Corporation
Product Code
GEI · General, Plastic Surgery
Decision Date
Jan 4, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The J-Plasma Precise® FLEX is used for the delivery of radiofrequency energy and/or helium gas plasma (J-Plasma energy) for electrosurgical cutting, coagulation, and ablation of soft tissue. It is intended for use with a grasper during minimally invasive surgical procedures. The J-Plasma Precise® FLEX is compatible with Apyx Medical electrosurgical generators.
Device Story
J-Plasma Precise FLEX is a sterile, single-use monopolar electrosurgical handpiece; used with compatible Apyx Medical generators. Input: RF energy and helium gas. Operation: Generator delivers RF energy to electrode; helium gas flows over energized electrode to create helium plasma for tissue cutting, coagulation, or ablation. Output: Helium plasma energy. Used in minimally invasive surgery; operated by surgeon via footswitch. Features extendable blade (max 5mm) and introducer for tip control. Output affects tissue thermal state; provides surgical precision. Benefits: Enables controlled soft tissue treatment in minimally invasive settings.
Clinical Evidence
Bench testing only. Ex-vivo study on muscle, liver, and kidney tissue compared thermal effects (depth/lateral spread) of subject and predicate devices across clinically relevant power/flow settings. Results demonstrated equivalence in tissue effects. Electrical safety and EMC testing confirmed compliance with IEC 60601-1 and 60601-2-2 standards. Biocompatibility established per ISO 10993-1.
Technological Characteristics
Monopolar electrosurgical handpiece; ceramic nozzle tip; extendable blade (max 5mm). Energy: RF and helium gas plasma. Power: up to 40W; gas flow: 1-5 LPM. Connectivity: Proprietary 6-pin electrical/1-pneumatic connector. Sterilization: Single-use sterile. Materials: Biocompatible per ISO 10993. Standards: ANSI/AAMI/IEC 60601-1, 60601-2-2, 60601-1-2.
Indications for Use
Indicated for soft tissue electrosurgical cutting, coagulation, and ablation during minimally invasive surgical procedures using a grasper.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is in blue and consists of the letters "FDA" and the words "U.S. Food & Drug Administration".
January 4, 2021
Apyx Medical Corporation (formerly Bovie Medical Corporation) Topaz Kirlew Vice President, Quality Assurance & Regulatory Affairs 5115 Ulmerton Road Clearwater, Florida 33760-4004
Re: K202880
Trade/Device Name: J-Plasma Precise FLEX Handpiece Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: September 25, 2020 Received: September 28, 2020
Dear Topaz Kirlew:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K202880
Device Name J-Plasma Precise® FLEX (APYX-500BF)
Indications for Use (Describe)
The J-Plasma Precise® FLEX is used for the delivery of radiofrequency energy and/or helium gas plasma energy) for electrosurgical cutting, coagulation of soft tissue. It is intended for use with a grasper during minimally invasive surgical procedures.
The J-Plasma Precise® FLEX is compatible with Apyx Medical electrosurgical generators.
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for Apyx Medical. The logo is primarily orange, with the word "Apyx" in a stylized font. A small blue triangle is incorporated into the "A" of "Apyx." Below "Apyx" is the word "MEDICAL" in blue, smaller, block letters.
## K202880
| 1. | General Information | | |
|----|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | Submitted by: | Apyx Medical Corporation (formerly Bovie Medical)<br>5115 Ulmerton Road<br>Clearwater, Florida 33760-4004<br>United States of America | |
| | Establishment Registration #: | 3007593903 | |
| | Submitter FAX Number: | (727) 322-4465 | |
| | Applicant Contact: | Lauren Tiller, Senior Regulatory Affairs Specialist III<br>5115 Ulmerton Road<br>Clearwater, Florida 33760-4004<br>United States of America<br>Phone: (727) 803-8550<br>Email: Lauren.Tiller@apyxmedical.com | |
| | Correspondent Contact: | Dr. Topaz Kirlew, MBA, Vice President, Quality<br>Assurance & Regulatory Affairs<br>5115 Ulmerton Road<br>Clearwater, Florida 33760-4004<br>United States of America<br>Phone: (727) 803-8617<br>Email: Topaz.Kirlew@apyxmedical.com | |
| | Date Prepared: | December 4, 2020 | |
| | Trade Names (Model Number): | J-Plasma Precise® FLEX Handpiece<br>(APYX-500BF) | |
| | Common Name: | Electrosurgical Handpiece | |
| | Classification: | Class II per 21CFR 878.4400 - Electrosurgical<br>Cutting and Coagulation Device and Accessories<br>Product Code: GEI | |
| | Predicate Device(s): | Predicate Device<br>Bovie J-Plasma Precise® Flex (K170777) | |
#### 2. Indications for Use
The J-Plasma Precise® FLEX is used for the delivery of radiofrequency energy and/or helium gas plasma (J-Plasma energy) for electrosurgical cutting, coagulation, and ablation of soft tissue. It is intended for use with a grasper during minimally invasive surgical procedures.
The J-Plasma Precise® FLEX is compatible with Apyx Medical electrosurgical generators.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for Apyx Medical. The logo is primarily orange, with the word "Apyx" in a stylized font. The "A" in "Apyx" has a small, dark blue triangle in the lower-left corner. Below "Apyx" is the word "MEDICAL" in a bold, dark blue font, with a thin, dark blue line underneath.
## K202880
#### 3. Device Description
The J-Plasma Precise® Flex Handpiece is a sterile, single use electrosurgical (monopolar) device intended to be used in conjunction with compatible generators for the delivery of radiofrequency ("RF") energy and/or helium gas plasma for electrosurgical cutting, coaqulation, and ablation of soft tissue. It is intended for use with a grasper during minimally invasive surgical procedures. The handpiece has one configuration, model # APYX-500BF. The compatible Generators operate at an adjustable power of up to 40 W (expressed as 0-100% where 100% is 40 W) and provide an adjustable helium gas flow of 1-5 LPM. Radiofrequency energy is delivered to the handpiece by the generator and used to energize the electrode. When helium gas is passed over the energized electrode, a helium plasma is generated for soft tissue cutting, coagulation or ablation.
The J-Plasma Precise® Flex Handpiece has an extendable electrode (blade) in the tip to generate helium plasma. The tip is controlled by the user via graspers to actuate (extend and retract) the blade. The handpiece comes with an Introducer that allows for more control of the tip when introducing and removing the distal end to and from the cannula. The handpiece has a flexible, seamless cable that can be advanced with graspers through the cannula during minimally invasive surgical procedures. The handpiece is activated by the footswitch. The J-Plasma Precise® Flex System with all components and accessories is depicted in Figure 1 below:
Image /page/4/Picture/8 description: The image shows a medical device called J-Plasma. The device consists of a generator (ESU), a helium tank with a regulator, a footswitch, a grounding pad, and a handpiece with an introducer and cable. The generator has a display showing the numbers 8, 20, and 40.
Figure 1: J-Plasma Precise® FLEX System
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for Aryx Medical. The logo features the word "Aryx" in orange, with a small blue triangle replacing the "A". Below the word "Aryx" is the word "MEDICAL" in blue. The logo is simple and modern, and the colors are eye-catching.
## K202880
### 4. Technological Characteristics
The J-Plasma Precise® Flex Handpiece is a modification to the predicate device that was cleared in K170777 (Bovie J-Plasma Precise® Flex Handpiece). There are no changes to intended use, indications for use, operational principles, energy type, tissue treatment effects, performance specifications, electrosurgical modes, output power waveforms, maximum power settings, and plasma settings that change the energy delivered to the tissue.
The J-Plasma Precise® Flex Handpiece, as compared to the predicate device (K170777), has only the following differences in technological characteristics:
- . Blade actuation, extension, and retraction is now in the distal end of the device
- Blade configuration will now extend a maximum of 5 mm .
- Seamless, working length of the cable
- . Addition of the Introducer, as a tip gasping aid/accessory
- . Updated One-Wire Programmability for compatibility with the upgraded APYX-JS3/RS3 (Upgrades cleared under K192867)
The J-Plasma Precise® Flex Handpiece with its components is pictured below in Figures 2 and 3:
Image /page/5/Figure/14 description: The image shows a medical device with several labeled parts. The device includes a cable connected to a plug, which leads to an introducer. At the end of the introducer, there is a blade and a tip with grasping features, indicating its use in surgical or medical procedures.
Figure 2: J-Plasma Precise® FLEX (Subject Device)
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Apyx Medical. The logo is comprised of the word "Apyx" in orange, with a blue triangle replacing the "A". Below the word "Apyx" is the word "MEDICAL" in blue. The "TM" symbol is in the upper right corner of the logo.
Image /page/6/Picture/5 description: The image shows a medical device with labels pointing to different parts. The top part of the device is labeled as the "Introducer" and the "Tip". The bottom part of the device is labeled as the "Secondary Grasping Location", "Blade", and "Operative Grasping Location".
Figure 3: J-Plasma Precise® FLEX Introducer, Tip Configuration, and Grasping Features (Subject Device)
Results of verification and validation testing demonstrate that the J-Plasma Precise® Flex Handpiece is equivalent in safety and effectiveness to the predicate device, as described below.
### 5. Performance Data
#### a. Bench Testing
All testing protocols performed on the subject device, J-Plasma Precise® Flex Handpiece, were derived from the risk assessment in accordance with ISO 14971 which evaluated the safety and effectiveness of the design modification in accordance with Apyx Medical's design and development procedures. Both the predicate and the modified subject device are designed by the same manufacturer, Apyx Medical (formerly Bovie Medical). The test methodology and acceptance criteria were developed from the same standards and following Apyx Medical's Design and Development procedures, in compliance with the Design and Development requirements of 21 CFR 820.30. The testing performed are summarized in table below.
### b. Tissue Thermal Effect Equivalency Test
An ex-vivo study was conducted on several tissue types (i.e., muscle, liver, kidney) to compare the tissue thermal effect of the J-Plasma Precise® Flex and the predicate device. The quantitative results of this testing demonstrated that the tissue effects the subject device is equivalent to the predicate device's effect over a range of power and flow settings that are clinically relevant. This ex-vivo testing is summarized in the table below.
{7}------------------------------------------------
Image /page/7/Picture/2 description: The image shows the logo for Apyx Medical. The logo is primarily orange, with the word "Apyx" in a stylized font. A small, dark blue triangle is positioned to the left of the "A" in "Apyx." Below "Apyx," the word "MEDICAL" is written in a bold, dark blue font.
## K202880
| Test | Objective | Result |
|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Mechanical<br>Verification &<br>Functionality | Verify Mechanical functionality of the J-<br>Plasma Precise® Flex | Mechanical functionality<br>requirements were all met. |
| Tissue Effect<br>(Plasma<br>Activation) | Verify that the tissue effects for Plasma<br>Activation are similar between the<br>subject device and the Predicate device<br>(K170777) using 3 different tissue types<br>(Muscle, Kidney & Liver) | The depth and lateral spread<br>(i.e. average length and width)<br>of thermal tissue effects were<br>demonstrated to be equivalent<br>between the subject device<br>and Predicate device across<br>all tissue types and J-Plasma<br>energy settings. |
| Tissue Effect<br>(Monopolar<br>Coagulation) | Verify that the tissue effects of the<br>subject device and the Predicate device<br>(K170777) are the same for Monopolar<br>Coagulation using 3 different tissue<br>types (Muscle, Kidney & Liver) | The depth and lateral spread<br>(i.e. average length and width)<br>of thermal tissue effects were<br>demonstrated to be equivalent<br>between the subject device<br>and Predicate device across<br>all tissue types and Monopolar<br>Coagulation energy settings. |
### c. Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on the J-Plasma Precise® Flex. The handpiece complies with the ANSI/AAMI/IEC 60601-1:2005/(R)2012 and A1:2012, and AAMI/ANSI/IEC 60601-2-2:2017 standards for safety and the AAMI/ANSI/IEC 60601-1-2:2014 (4th Edition) standard for EMC. The J-Plasma Precise® Flex was determined to be in conformance with these standards.
### 6. Biocompatibility Testing
The biocompatibility evaluation for the J-Plasma Precise® Flex Handpiece was conducted in accordance with the June 16, 2016 FDA Guidance, "Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and the ISO 10993-1 Fifth edition 2018-08 standard. "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process".
The J-Plasma Precise® Flex Handpiece (subject device) is an external communicating device with an indirect blood path contact for a duration of less than 24 hours. Biocompatibility has been established per ISO 10993 guidelines for this category. Biocompatibility verification testing was satisfactorily conducted for the subject device.
{8}------------------------------------------------
Image /page/8/Picture/2 description: The image shows the logo for Apyx Medical. The logo is orange and blue. The word "Apyx" is in orange, with a blue triangle in place of the letter "A". Below the word "Apyx" is the word "MEDICAL" in blue.
# K202880
### 7. Substantial Equivalence
| Feature /<br>Characteristic | Bovie J-Plasma Precise Flex<br>(Predicate Device)<br>K170777 | J-Plasma Precise FLEX<br>(Subject Device) |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Bovie J-Plasma Precise® Flex<br>Handpiece is used for the delivery<br>of radiofrequency energy and/or a<br>helium gas plasma for<br>electrosurgical cutting, coagulation,<br>and ablation of soft tissue. It is<br>intended for use with grasping<br>instruments during minimally<br>invasive surgical procedures.<br>The Bovie J-Plasma Precise Flex<br>Handpiece is compatible with<br>Bovie Ultimate Generators BVX- | The J-Plasma Precise® FLEX is<br>used for the of radiofrequency<br>energy and/or helium gas<br>plasma (J-Plasma energy) for<br>electrosurgical cutting,<br>coagulation, and ablation of soft<br>tissue. It is intended for use with<br>a grasper during minimally<br>invasive surgical procedures.<br>The J-Plasma Precise® FLEX is<br>compatible with Apyx Medical<br>electrosurgical generators. |
| Procedures | 200H and BVX-200P.<br>Laparoscopic or minimally invasive<br>surgical procedures | Same |
| Energy Source | Apyx Generator | Same |
| Energy Type | Monopolar RF energy via Helium<br>Plasma (helium facilitates the use<br>of low current RF waveform) | Same |
| Output | Monopolar | Same |
| Plasma settings | Maximum 40 watts, 1-5 lpm gas<br>flow | Same |
| Plasma Activation<br>Tissue Effect | Same | Same |
| Monopolar Coagulation<br>Mode | Yes | Same |
| Monopolar Coagulation<br>Settings | Maximum 120 watts, 1-5 lpm gas<br>flow | Same |
| Monopolar Coagulation<br>Tissue Effect | Same | Same |
| System Components | The Bovie J-Plasma Precise Flex<br>System consists of:<br>- Apyx Medical (formerly Bovie<br>Medical) Generator<br>- Disposable handpiece<br>- Foot pedal<br>- Power cord<br>- Gas regulator<br>- Gas cylinder | Same |
| User Interface | Footswitch activation and<br>generator buttons and displays. | Same |
| Connector | Company Proprietary Plug<br>Connector involving 6 electrical | Same |
| Feature /<br>Characteristic | Bovie J-Plasma Precise Flex<br>(Predicate Device)<br>K170777 | J-Plasma Precise FLEX<br>(Subject Device) |
| | pins and 1 pneumatic connection<br>in the center to provide helium gas | |
| Shaft Design | Flexible | Same |
| Shaft Outer Diameter | 8 mm | Same |
| Total Distance from Tip<br>to Plug | 3,251.2 mm total distance from tip<br>to the plug | Same |
| Tip Configuration | Extendable Blade or Needle | Extendable Blade |
| Blade Extension | Maximum 10 mm | Maximum 5 mm |
| Blade Width x Thickness | 1.6 mm x 0.4 mm | 1.8 mm x 0.4 mm |
| Tip Protector | Ceramic nozzle | Same |
| Blade Actuation | Proximal Actuation | Distal Actuation |
| Device Activation | Footswitch activation only,<br>no hand activation | Same |
| Grasping Feature | Yes | Same |
| Tip grasping<br>aid/accessory | None | Introducer |
| Generator Compatibility | Compatible Bovie Medical<br>Electrosurgical Generators | Compatible Apyx Medical<br>(formerly Bovie Medical)<br>Electrosurgical Generators |
| One-Wire Proximal Chip | ETO Sterilizable | Same |
| Biocompatibility<br>Materials | All patient contacting materials are<br>biocompatible | Same |
| Sterility | Single-use disposable sterile<br>handpieces | Same |
| Electrical Safety | Complies with IEC Electrical Safety<br>Standards:<br>IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2 | Same |
{9}------------------------------------------------
## K202880
The subject device has the same indications for use, operational principle, and technological characteristics as the predicate device. The conclusions drawn from the bench and pre-clinical testing demonstrate that the J-Plasma Precise® Flex Handpiece is substantially equivalent in terms of safety and effectiveness to the legally marketed predicate device for the same indications. Functional and mechanical verification testing demonstrate that the J-Plasma Precise® Flex Handpiece performs as intended in the specified use conditions.
#### 8. Conclusion
The subject device, J-Plasma Precise® FLEX, has the same intended use, indications for use, operational principles, energy type, tissue treatment effects, performance specifications, electrosurgical modes, output power waveforms, maximum power settings, and plasma settings as the originally designed predicate device (Bovie J-Plasma Precise® Flex, K170777).
{10}------------------------------------------------
Image /page/10/Picture/2 description: The image shows the logo for Apyx Medical. The logo is comprised of the word "Apyx" in orange, with a blue triangle in place of the letter "A". Below the word "Apyx" is the word "MEDICAL" in blue. There is a thin blue line below the word "MEDICAL".
K202880
The technological differences between the subject and predicate device are the transfer of the blade actuation to the tip to create a seamless design of the cable, a reduction of the maximum blade length from 10 mm to 5 mm, the addition of an introducer as a tip grasping aid, and an updated one-wire programmability for compatibility with the upgraded APYX-JS3/RS3 (these generator upgrades were cleared under K192867). The subject device and the predicate device operate in a similar manner with the same range of energy settings; the modifications explained above accommodates the users' preferences and enhances the stability of the device. There are no differences that would raise new or different questions regarding safety or effectiveness, as the subject device and the predicate device operate in a similar manner with the same range of tissue treatment energy parameters.
The subject device's safety and performance have been confirmed by the results of the performance bench testing and the design change has gone through the design controls process and verification and validation testing to demonstrate that the device is as safe and effective as the predicate device (Bovie J-Plasma Precise® Flex, K170777).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.