Ellusa Generator, Ellusa Bayonet Tip, Ellusa RF-Cobb Tip, Ellusa SP-Pack and Tips, Ellusa Tip Set
K202562 · Ellusa, LLC · GEI · Nov 2, 2020 · General, Plastic Surgery
Device Facts
Record ID
K202562
Device Name
Ellusa Generator, Ellusa Bayonet Tip, Ellusa RF-Cobb Tip, Ellusa SP-Pack and Tips, Ellusa Tip Set
Applicant
Ellusa, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
Nov 2, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ellusa Generator is intended to provide high frequency energy for resection, ablation, and coagulation of soft tissues and hemostasis of blood vessels during surgical procedures and is used to deliver RF energy via an assortment of surgical devices to cut and coagulate different kinds of tissue. The Ellusa Monopolar Tips are intended for use by a physician familiar with resection, dissection, incision, and hemostasis in soft tissue surgical procedures.
Device Story
The Ellusa Generator is a radiofrequency (RF) electrosurgical unit providing high-frequency energy for soft tissue resection, ablation, coagulation, and hemostasis. It operates by converting electrical energy into RF energy, delivered via monopolar or bipolar surgical tips. The operator selects waveforms and power levels via front-panel push buttons; digital displays provide status feedback and self-test monitoring. Output is activated by foot or hand switches. The device is used in surgical settings by physicians. The system includes various electrode shapes and sizes made of tungsten, brass, or stainless steel. The generator interlocks to prevent operation during failure states. By delivering predetermined RF energy, the device enables precise tissue cutting and coagulation, benefiting patients by facilitating surgical hemostasis and tissue management.
Clinical Evidence
Bench testing only. Verification and validation included electrical safety (AAMI ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-2), usability engineering (IEC 62366), sterilization validation (ISO 11135, 11737, 11138), biocompatibility (ISO 10993 series), shelf-life (ASTM F1980-07), and software validation (IEC 62304). Performance testing included energy output verification, Hipot testing, pull testing, and ex-vivo bovine tissue thermal effect comparisons against the predicate.
Technological Characteristics
RF electrosurgical generator and monopolar tips. Materials: surgical stainless steel, brass, tungsten, PFA, PTFE. Energy: 4.0 MHz (monopolar) and 1.7 MHz (bipolar) RF. Output: up to 170W. Connectivity: foot/hand switch activation. Sterilization: Gamma or Ethylene Oxide. Standards: AAMI ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-2, ISO 10993, ISO 11135, ISO 11737, ISO 11138, ASTM F1980-07.
Indications for Use
Indicated for use by physicians familiar with resection, dissection, incision, and hemostasis in soft tissue surgical procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
November 2, 2020
Ellusa, LLC Ms. Suzanne Lucas Sr. Regulatory Affairs Specialist 2473 Grand Avenue Baldwin, New York 11510
Re: K202562
Trade/Device Name: Ellusa Generator and Monopolar Tips
Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: August 31, 2020 Received: September 4, 2020
Dear Ms. Lucas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202562
Device Name Ellusa Generator and Monopolar Tips
## Indications for Use (Describe)
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
The Ellusa Generator is intended to provide high frequency energy for resection, ablation, and coagulation of soft tissues and hemostasis of blood vessels during surgical procedures and is used to deliver RF energy via an assortment of surgical devices to cut and coagulate different kinds of tissue.
The Ellusa Monopolar Tips are intended for use by a physician familiar with resection, incision, and hemostasis in soft tissue surgical procedures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "ellusa" in a stylized font. The word is in black and is set against a white background. There are three curved lines to the left of the word, which appear to be a design element. The font is sans-serif and has rounded edges.
# 510(k)SUMMARY (As required by 21 CFR 807.92(a)) (K202562)
October 28, 2020
# Submitter's Information (807.92(a)(1))
Company Name and Address: Ellusa, LLC 2473 Grand Avenue Baldwin, NY 11510 Phone: (516) 866-0001 www.ellusa.com
Establishment Registration 3016087306
#### Contact Information:
Ms. Suzanne Lucas Sr. Regulatory Affairs Specialist Phone: (516) 866-0001 Email: slucas@ellusa.com
## Device Information (807.92(a)(2))
Trade Name: Ellusa Generator and Monopolar Tips
#### Common/Usual Name
Electrosurgical unit and accessories, electrosurgical cutting and coagulation device and accessories
#### Classification Name and Regulation
Electrosurgical Cutting and Coagulation Device and Accessories; 21 CFR 878.4400
#### Classification Panel
General and Plastic Surgery Device
#### Device Class/Product Code
FDA Classification: Class II FDA Product Code: GEI
{4}------------------------------------------------
# Predicate Device (807.92(a)(3))
Soniquence RF Generator & Monopolar Electrodes (K183611)
## Use of Guidance Documents
The content provided within this 510k follows the FDA guidance document "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery" issued on March 9, 2020.
## Device Description (807.92(a)(4))
The Ellusa Generators are radiofrequency (RF) electrosurgical generators. It is a compact source of high-frequency RF energy employed for cutting and coagulation of soft tissue in a variety of surgical procedures. This action is achieved by front panel selection of waveforms and power levels. The subject device is designed to enable the performance of monopolar and bipolar functions for cutting, coagulation, and hemostasis in conjunction with separately-marketed electrosurgical accessories and ancillary equipment which have been developed for mutually- exclusive use with the Ellusa line of generators.
All selections are effected through push buttons and lamps, which give the operator feedback of status. The power level for each mode is indicated by front panel digital displays which also show the status of self-test and monitoring. The display is interlocked with controls to prevent operation when FAIL is displayed. The final output power control is made through foot and /or hand switches.
Ellusa Electrodes are used with the Ellusa Generator. The electrical power operating at radio frequency (RF) is transferred to tissue at the surgical site. The time-varying voltage produced by RF electrical power source yields a predetermined electrosurgical effect, such as tissue cutting or coagulation. The Ellusa family of electrodes is available in various shapes and sizes depending on the need of the surgeon. The devices are available in classic alloys (Tungsten, brass, stainless steel).
These devices are designed to comply with international safety standards including applicable IEC series electrical safety standards.
## Intended Use:
The Ellusa Generator is intended to provide high frequency energy for resection, ablation, and coagulation of soft tissues and hemostasis of blood vessels during surgical procedures and is used to deliver RF energy via an assortment of surgical devices to cut and coagulate different kinds of tissue.
The Ellusa Monopolar Tips are intended for use by a physician familiar with resection, dissection, incision, and hemostasis in soft tissue surgical procedures.
{5}------------------------------------------------
## Comparison of Technological Characteristics
The subject devices are substantially equivalent to the predicate devices based on a comparison of physical and performance characteristics. There are no technological differences between the subject device as compared to the predicate devices, and the indications for use are identical to the predicate devices.
## Comparison of Generator models
| Characteristic | Soniquence Generator<br>K183611<br>PREDICATE | Ellusa Generator<br>Type 1<br>Proposed Device | Ellusa Generator<br>Type 2<br>Proposed Device | Ellusa Generator<br>Type 3<br>Proposed Device | |
|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Photo | Image: Soniquence Generator K183611 PREDICATE | Image: Ellusa Generator Type 1 Proposed Device | Image: Ellusa Generator Type 2 Proposed Device | Image: Ellusa Generator Type 3 Proposed Device | |
| Model<br>numbers<br>available | (Type 1 models)<br>IEC6-SESP120, IEC6-IESP170,<br>IEC6-IENSP170, IEC6-SESP170,<br>IEC6-SENSP170<br>(Type 2 models)<br>IEC6-IESU120, IEC6-IENSU120,<br>IEC6-SENSU120, IEC6-<br>SESU120, IEC6-IESU170, IEC6-<br>IENSU170, IEC6-SESU170,<br>IEC6-SENSU170<br>(Type 3 models)<br>IEC6-IESV120, IEC4-SEN120,<br>IEC6-SENSV120, IEC6-<br>SESV120, IEC6-IESV170, IEC6-<br>IENSV170, IEC6-SESV170,<br>IEC6-SENSV170 | IEC6-SEEP120<br>IEC6-IEEP170<br>IEC6-IENEP170<br>IEC6-SEEP170<br>IEC6-SENEP170<br>IEC4-ESM120<br>IEC4-ESP<br>IEC3A-ES30W | IEC6-IEEU120<br>IEC6-IENEU120<br>IEC6-SENEU120<br>IEC6-SEEU120<br>IEC6-IEEU170<br>IEC6-IENEU170<br>IEC6-SEEU170<br>IEC6-SENEU170 | IEC6-IEEV120<br>IEC4-SEN120<br>IEC6-SENEV120<br>IEC6-SEEV120<br>IEC6-IEEV170<br>IEC6-IENEV170<br>IEC6-SEEV170<br>IEC6-SENEV170<br>IEC4-EJSM120 | |
| Principle of<br>Operation | Conversion of electrical<br>energy to high-frequency<br>RF energy (monopolar and<br>bipolar) | Identical to predicate | Identical to predicate | Identical to predicate | |
| Characteristic | | Soniquence Generator<br>K183611<br>PREDICATE | Ellusa Generator<br>Type 1<br>Proposed Device | Ellusa Generator<br>Type 2<br>Proposed Device | Ellusa Generator<br>Type 3<br>Proposed Device |
| Modes of<br>operation /<br>mechanism of<br>action | Cut | 4.0 MHz<br>square wave<br>envelope<br>150W max.<br>output | All Identical to<br>predicate except<br>models<br>IEC4-ESM120, IEC4-<br>ESP, &<br>IEC3A-ES30W are max<br>120W output | Identical to predicate | All Identical to<br>predicate except<br>model IEC4-EJSM120<br>is max 120W output |
| | Blend | 4.0 MHz<br>square wave<br>envelope<br>110W max.<br>output | All Identical to<br>predicate except<br>models<br>IEC4-ESM120, IEC4-<br>ESP, &<br>IEC3A-ES30W are max<br>90W output and the<br>mode is called<br>Cut/Coag | Identical to predicate | All Identical to<br>predicate except<br>model IEC4-EJSM120<br>is 90W output and<br>the mode is called<br>Cut/Coag. |
| | Hemo | 4.0 MHz<br>square wave<br>envelope<br>60W max.<br>output | Identical to predicate | Identical to predicate | Identical to predicate |
| | Bipolar | 1.7 MHz CW<br>sinusoidal<br>wave<br>40W max<br>output | Identical to predicate<br>except models<br>IEC4-ESM120, IEC4-<br>ESP, &<br>IEC3A-ES30W mode is<br>called Bipolar Hemo<br>(minor change in<br>nomenclature only,<br>function and<br>performance stay the<br>same) | Identical to predicate | All Identical to<br>predicate except<br>model IEC4-EJSM120<br>is Fulgurate<br>(Monopolar) and 4.0<br>MHz<br>Sinusoidal wave<br>40W max. output |
| | Bipolar<br>Turbo | 1.7 MHz<br>square wave<br>envelope<br>170W max.<br>output | All Identical to<br>predicate except<br>models<br>IEC4-ESM120, IEC4-<br>ESP, &<br>IEC3A-ES30W are max<br>120W output | Identical to predicate | All Identical to<br>predicate except<br>model IEC4-EJSM120<br>is max120W output<br>and mode is Bipolar |
| Output<br>frequency | | 4 MHz (monopolar)<br>1.7 MHz (bipolar) | Identical to predicate | Identical to predicate | Identical to predicate |
| Maximum<br>power output | | 170W | All Identical to<br>predicate except<br>models<br>IEC4-ESM120, IEC4-<br>ESP, &<br>IEC3A-ES30W | Identical to predicate | All Identical to<br>predicate except<br>model IEC4-EJSM120<br>is 120W |
| | Soniquence Generator<br>K183611<br>PREDICATE | Ellusa Generator<br>Type 1<br>Proposed Device | Ellusa Generator<br>Type 2<br>Proposed Device | Ellusa Generator<br>Type 3<br>Proposed Device | |
| Characteristic | | | | | |
| | | IEC3A-ES30W are<br>120W | | | |
| Voltage<br>(peak-to-peak) | 1,200V | Identical to predicate | Identical to predicate | Identical to predicate | |
| Power<br>activation<br>control | Footswitch and/or<br>fingerswitch | Identical to predicate | Identical to predicate | Identical to predicate | |
| Electrical<br>safety and<br>EMC testing<br>standards | AAMI ANSI ES60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2 | Identical to predicate | Identical to predicate | Identical to predicate | |
| Applied part<br>type | CF | All Identical to<br>predicate except<br>models<br>IEC4-ESM120,<br>IEC4-ESP, &<br>IEC3A-ES30W are<br>type BF | Identical to predicate | All Identical to<br>predicate except<br>model IEC4-EJSM120<br>is type BF | |
{6}------------------------------------------------
{7}------------------------------------------------
# Comparison of Monopolar Tips
| Predicate Information | Soniquence Monopolar Electrodes<br>(K183611) | Ellusa Monopolar Tips |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | PREDICATE | PROPOSED DEVICE |
| Intended Use | The Soniquence Monopolar Electrodes are<br>intended for use by a physician familiar with<br>resection, dissection, incision, and hemostasis<br>in soft tissue surgical procedures. | The Ellusa Monopolar Tips are intended for<br>use by a physician familiar with resection,<br>dissection, incision, and hemostasis in soft<br>tissue surgical procedures. |
| Product Code | GEI | Identical to predicate |
| Device Classification | Class II | Identical to predicate |
| Functions | Cutting/coagulation | Identical to predicate |
| Energy type/ operating<br>principle | Monopolar | Identical to predicate |
| Design Specifications | Integrated sterile components; handle, cable,<br>connector, shaft, monopolar electrode | Identical to predicate |
| Shaft forms | Straight, Bayonet, angled, and curved | Identical to predicate |
| Shaft malleability | Fixed and malleable models | Identical to predicate |
| Shaft lengths | 1.9 – 60 cm | Identical to predicate |
| Shaft diameter | 1/16" | Identical to predicate |
| Tip Configurations | Loop (round, diamond), ball, and blade (needle,<br>fine wired, spatula) | Identical to predicate |
{8}------------------------------------------------
| Material Composition | | Identical to predicate |
|------------------------------------------------|---------------------------------------------------------------------------|------------------------|
| Handle | Molded Plastic | |
| | Surgical Stainless Steel/Brass | |
| Shaft | Surgical Stainless Steel/ Tungsten/Brass | |
| Electrode | Surgical stainless steel / PFA insulation | |
| Patient contacting<br>materials | Medical grade stainless steel, brass, tungsten,<br>Loctite, PFA, and PTFE | Identical to predicate |
| Coating | None | Identical to predicate |
| Biocompatibility | ISO 10993 | Identical to predicate |
| Sterilization Methods | Gamma or Ethylene Oxide | Identical to predicate |
| Expiration Dating | Yes | Identical to predicate |
| Packaging | Sterile, Blister Tray<br>Sterile, Peel Pouch | Identical to predicate |
| Activation Method | Fingerswitch/Footswitch | Identical to predicate |
| Manual Controls | No, all power settings established on generator | Identical to predicate |
| Min., Max., Default<br>output Values (Wattage) | 0-170 | Identical to predicate |
| Cable Length (m) | 3 | Identical to predicate |
| Temperature Probe | No | Identical to predicate |
| Cooling Function | None | Identical to predicate |
# Summary of Technological Characteristics
The Ellusa Generator and Monopolar Tips are equivalent in intended use, technological characteristics, operating principle, and technical specifications to the predicate device Soniquence Generator and Monopolar Electrodes (K183611).
The differences of minor changes in nomenclature of operating modes does not affect the function or performance of the devices. In addition, the reduced output power on some models does not introduce new risks or affect the safety and effectiveness of the devices.
# Non-Clinical Testing (807.92(b)(1))
The Ellusa Generator and Monopolar Tips will be manufactured in accordance with the design control requirements of 21 CFR 820.30. Appropriate non- clinical verification and validation activities were planned and conducted in order to address identified risks and ensure the safety and effectiveness of the device. The following tests were successfully performed:
- . AAMI ANSI ES60601-1 - Medical Electrical Equipment, Part I: General requirements for basic safety and essential performance
{9}------------------------------------------------
- . IEC 60601-1-2 - Medical Electrical Equipment, Part I-2: General requirements for basic safety and essential performance. Collateral Standard: Electromagnetic Compatibility
- IEC 60601-2-2 - Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
- IEC 62366 Medical devices Application of usability engineering to medical devices. .
## Sterilization of Monopolar Tips:
- . ISO 11135 - Sterilization of health-care products - ethylene oxide - requirements for the development, validation and routine control of a sterilization process for medical devices
- . ISO 11737-1 - Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on products
- . ISO 11737-2 - Sterilization of medical devices -- Microbiological methods -- Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process /
- . ISO 11138-1 - Sterilization of health care products - Biological indicators - Part 1: General requirements
## Biocompatibility Testing:
- . ISO 10993-1 - Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process
- . ISO 10993-5 - Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-7 - Biological evaluation of medical devices - part 7: ethylene oxide sterilization residuals
- . ISO 10993-10 - Biological evaluation of medical devices — Part 10: Tests for irritation and skin sensitization
- ISO 10993-11 - Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
## Shelf Life Testing:
- . ASTM F1980-07 - Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
## Performance Testing:
- . Software verification and validation in accordance with IEC 62304 and the FDA Final Guidance document titled "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
- . Energy output verification of each of the five modes and comparison to predicate.
{10}------------------------------------------------
- . Thermal effect comparison for representative modes of the subject device and predicate on ex-vivo bovine tissue.
- . A design verification /validation was conducted to validate the design and functionality of the generator and Monopolar Tip accessories. Tests included output power test, Hipot test, pull test, lateral heat spread in tissue test, evaluation of temperature profile, and final test. All tests passed their applicable acceptance criteria.
# Conclusion (807.92(b)(3))
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the similarity to the predicate devices in terms of technology, performance, and indications for use, Ellusa, LLC concludes that the Ellusa Generator and Monopolar Tips are substantially equivalent to and as safe and effective as the predicate devices described above. The only difference between the subject devices of this 510k and the predicate devices is a rebranding with the Ellusa name. The technology remains the same.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.