Smoke Evacuation Fingerswitch, Smoke evacuation system with electrosurgical pencil
K200372 · Modern Medical Equipment Manufacturing, Ltd. · GEI · Nov 30, 2020 · General, Plastic Surgery
Device Facts
Record ID
K200372
Device Name
Smoke Evacuation Fingerswitch, Smoke evacuation system with electrosurgical pencil
Applicant
Modern Medical Equipment Manufacturing, Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Nov 30, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is used to conduct electrosurgical current from an electrosurgical generator to target tissue in electrosurgical procedures for monopolar cutting and coagulation, with the additional function of evacuation to facilitate the removal of surgical smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system.
Device Story
Single-use electrosurgical pencil with integrated smoke evacuation; includes fingerswitch (PS947) or system (CE809) configurations. Device connects to electrosurgical generator (via 3-pin connector) and external smoke evacuation system (via 22mm EVA tubing). Surgeon operates buttons on handle to toggle cutting/coagulation; smoke is evacuated through suction tip/nosecone. Fingerswitch model features adjustable suction tip position along shaft for improved access in cavities. Compatible with standard electrosurgical generators (up to 4.5-5kVp). Facilitates smoke removal from surgical site to improve visibility and reduce inhalation of surgical plume. Used in clinical settings by surgeons.
Clinical Evidence
Bench testing only. Performance evaluated using porcine tissue (kidney, liver, muscle) to compare thermal effects and thermal zone sizes between subject and predicate devices. Safety verified via electrical safety (ANSI/AMMI ES60601-1, IEC 60601-2-2), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993-1), and sterile barrier integrity (ISO 11607-1, ASTM F1980-16).
Technological Characteristics
Materials: ABS housing, PVC cable, K-resin nosecone, EVA tubing, stainless steel electrode with Teflon coating. Energy: High-frequency electrical current (max 300W, 4.5-5kVp). Dimensions: 2.36mm electrode diameter. Connectivity: 3-pin active connector, 22mm vacuum tubing. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Standards: ISO 11135, IEC 60601-1, IEC 60601-2-2, ISO 10993.
Indications for Use
Indicated for patients undergoing monopolar electrosurgical procedures requiring cutting and coagulation, where surgical smoke evacuation is desired.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 30, 2020
Modern Medical Equipment Manufacturing, Ltd. Jerry Cheung Assistant Regulatory Manager Flat A, 11/F., Mai Wah Ind. Bldg., 1-7 Wah Sing Street Kwai Chung N.T. Hong Kong, China
Re: K200372
Trade/Device Name: Sterile and Non-sterile Single Use Smoke Evacuation Fingerswitch, Nosecone and Smoke Evacuation System with Electrosurgical Pencil Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: January 10, 2020 Received: February 14, 2020
Dear Jerry Cheung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA
has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K2000372
#### Device Name
Sterile and Non-sterile Single Use Smoke Evacuation Fingerswitch, Nosecone and Smoke Evacuation System with Electrosurgical Pencil
#### Indications for Use (Describe)
The device is used to conduct electrosurgical current from an electrosurgical generator to target tissue in electrosurgical procedures for monopolar cutting and coagulation, with the additional function of evacuation to facilitate the removal of surgical smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;"> <span style="text-decoration: overline;"></span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 1. 510(k) Owner<br>Name: | Modern Medical Equipment Manufacturing Limited |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Modern Medical Equipment Mfg., Ltd.<br>Flat A, 11/F., Mai Wah Ind. Bldg.,<br>1-7 Wah Sing Street,<br>Kwai Chung, N.T., Hong Kong, China. |
| Telephone: | (852) 2420 9068 |
| Fax: | (852) 2481 1234 |
| Contact person: | Mr. Jerry Cheung |
| Date of preparation: | 26 Nov, 2020 |
| 2. Device<br>Name of Device: | Sterile and Non-sterile Single Use Smoke Evacuation<br>Fingerswitch, Nosecone and Smoke Evacuation<br>System with Electrosurgical Pencil |
| Trade or proprietary name: | Sterile and Non-sterile Single Use Smoke Evacuation<br>Fingerswitch, Nosecone and Smoke Evacuation<br>System with Electrosurgical Pencil |
| Common or usual name: | Sterile and Non-sterile Single Use Smoke<br>Evacuation Pencil |
| Classification name: | Electrosurgical Cutting & Coagulation<br>Device & Accessories |
| Classification Panel: | General & Plastic Surgery |
| Product Code: | GEI |
| Class: | II |
#### 3. Predicate device
MEDTEK PLUMEPEN Integrated Smoke Evacuation Pencil with 510(k) number K103375.
## 4. Device description
There are two main series of smoke evacuation pencil.
- (1) Smoke Evacuation Fingerswitch (PS947)
{4}------------------------------------------------
The device includes the active handle which connects with the active electrode. The diameter of the mating part is 2.36mm+/-0.02mm. The buttons on the active handle control cutting and coagulation, the yellow side activates the cutting function and the blue side activates the coagulation function. The active handle's end is connected to an insulation cord with an active connector. The EVA tubing at the rear of the active handle then leads to a filter and pump system that facilitates the removal of harmful "surgical smoke" from the site of operation.
The device is compatible with the socket of generator accommodating the standard 3-pin active connector, the output voltage setting of generator not larger 4.5kVp, compatible cable connection of grounding pad to the generator and all smoke evacuators with an ID22mm EVA tubing connector.
In addition to these familiar features and functions, the switching controls are mounted onto an active handle that moves freely up and down the shaft of the device when the device is in vertical position (or moves freely forward and backward when the device is in horizontal position), when the switching control moves freely on the shaft, distance between the front end of the handle and the controls is reduced or increased that works like extending and retracting the suction tip on general smoke evacuation, allowing the surgeon a greater degree of control and removing the need for additional accessories to extend the length of the device when working in cavities and other areas with restrictive access. Furthermore, the electrode can be exchanged with 63.5mm, 69mm and 120mm long electrodes in diameter 2.36±0.2mm with corresponding nosecone for further extending the length of suction tip for smoke evacuation.
## (2) Smoke evacuation system with electrosurgical pencil (CE809)
The device includes electrosurgical pencil with electrode and smoke evacuation handle connected with EVA tubing with ID22mm tubing connector. Electrosurgical pencil (PD646 or PD631) with 69mm long electrodes in diameter 2.36±0.2mm, can be put into the handle that functions as smoke evacuation pencil.
The device is compatible with the socket of generator accommodating the standard 3-pin active connector, the output voltage setting of generator not larger 5kVp, compatible cable connection of grounding pad to the generator and all smoke evacuators with an ID22mm EVA tubing connector.
The products, PS947 and CE809 do not have technology difference.
## 5. Indication for use
The device is used to conduct electrosurgical current from an electrosurgical generator to target tissue in electrosurgical procedures for monopolar cutting and coagulation, with the additional function of evacuation to facilitate the removal of surgical smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system.
{5}------------------------------------------------
## 6. Technological characteristics
Single Use Smoke Evacuation Fingerswitch and Smoke Evacuation System with Electrosurgical Pencil integrates two technologies of electrosurgery and smoke evacuation. These two devices have substantially equivalent construction and performance as the predicate devices.
#### 7. Substantial Equivalence
The technological characteristics and performance testing of the subject and predicate devices are substantially equivalent. The following table shows the comparisons in more detail information among the subject devices and the predicate device.
| Compared<br>Items | Proposed Device | Proposed Device | Predicate Device<br>(K103375) | Comments on<br>difference |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | Single Use Smoke<br>Evacuation<br>Fingerswitch<br>(PS947)<br>The device is<br>intended to be used<br>as the active<br>monopolar electrode<br>in an electrosurgery<br>generator system for<br>cutting and<br>coagulation, with the<br>additional function to<br>facilitate the removal<br>of surgical smoke<br>generated by<br>electrosurgery when<br>used in conjunction<br>with an effective<br>smoke evacuation<br>system | Single Use Smoke<br>evacuation system<br>with electrosurgical<br>pencil (CE809)<br>The device is<br>intended to be used<br>as the active<br>monopolar electrode<br>in an electrosurgery<br>generator system for<br>cutting and<br>coagulation, with the<br>additional function to<br>facilitate the removal<br>of surgical smoke<br>generated by<br>electrosurgery when<br>used in conjunction<br>with an effective<br>smoke evacuation<br>system | MEDTEK<br>PLUMEPEN<br>Integrated Smoke<br>Evacuation Pencil<br>The PLUMEPEN is<br>designed for general<br>electrosurgical<br>applications,<br>including cutting and<br>coagulation, and for<br>removing smoke<br>generated by<br>electrosurgery when<br>used in conjunction<br>with an effective<br>smoke evacuation<br>system. The pencil<br>enables the operator<br>to remotely conduct<br>an electrosurgical<br>current from the<br>output connector of<br>an electrosurgical<br>unit to the operative<br>site for the desired<br>surgical effect. | Same |
| | | | | |
| Indication for<br>use | The device is used<br>to conduct<br>electrosurgical<br>current from an<br>electrosurgical<br>generator to target<br>tissue in<br>electrosurgical<br>procedures for<br>monopolar cutting<br>and coagulation,<br>with the additional<br>function of<br>evacuation to<br>facilitate the removal<br>of surgical smoke<br>generated by<br>electrosurgery when<br>used in conjunction<br>with an effective<br>smoke evacuation<br>system. | The device is used<br>to conduct<br>electrosurgical<br>current from an<br>electrosurgical<br>generator to target<br>tissue in<br>electrosurgical<br>procedures for<br>monopolar cutting<br>and coagulation,<br>with the additional<br>function of<br>evacuation to<br>facilitate the removal<br>of surgical smoke<br>generated by<br>electrosurgery when<br>used in conjunction<br>with an effective<br>smoke evacuation<br>system. | To remove smoke<br>plume form the<br>surgical site and<br>remotely conduct an<br>electrosurgical<br>current from the<br>output connector of<br>an electrosurgical<br>unit to the target<br>tissue for the<br>desired surgical<br>effect | Same |
| Regulation<br>number | 21 CFR 878.4400 | 21 CFR 878.4400 | 21 CFR 878.4400 | Same |
| Product code | GEI | GEI | GEI | Same |
| OTC or<br>prescription | For prescription use | For prescription use | For prescription use | Same |
| Energy<br>delivery | High frequency<br>electrical<br>current/energy | High frequency<br>electrical<br>current/energy | High frequency<br>electrical<br>current/energy | Same |
| Technology | Monopolar<br>electrosurgery and<br>use of vacuum<br>source to capture<br>smoke | Monopolar<br>electrosurgery and<br>use of vacuum<br>source to capture<br>smoke | Monopolar<br>electrosurgery and<br>use of vacuum<br>source to capture<br>smoke | Same |
| Device mode | Cut and coag. | Cut and coag. | Cut and coag. | Same |
| Design | Smoke evacuation | Smoke evacuation | Electrosurgical | Similar<br>construction and<br>provide the same<br>technology to<br>achieve target<br>function.<br>Extending the<br>suction tip and |
| | fingerswitch is | system with pencil | pencil is integrated | |
| | produced from that | includes pencil is | with smoke | |
| | electrosurgical | combined with the | evacuation with | |
| | pencil is integrated | specified suction | extendable suction | |
| | with smoke | handle that performs | tip A variety of | |
| | evacuation with | as those of suction | length of electrode | |
| | switching control | evacuation pencil. | be used with | |
| | moving on the shaft<br>of device in order to<br>perform as<br>extending or<br>retracting suction<br>tip. Nosecone can<br>be added to further<br>extend the suction<br>tip. | | extendable suction<br>tip. | electrode is only<br>for facilitating<br>operation in<br>deeper surgical<br>site |
| Structure of<br>Pencil<br>- Housing<br>- Cable | | | | |
| - Housing<br>- Cable | - ABS<br>- PVC | - ABS<br>- PVC | - ABS<br>- PVC | The material<br>used and<br>structure among |
| -Nosecone /<br>Suction tip | - K-resin | - K-resin | - K-resin | the purposed<br>and predicate<br>devices are very |
| -Suction<br>tubing | - EVA | - EVA | - EVA and PVC | similar and do<br>not raise safety<br>and<br>effectiveness<br>issues because<br>those were<br>tested according<br>to IEC test and<br>biocompatibility<br>requirements |
| -Switching | - Push button, rocker | - Push button, rocker | - Push button | Same function |
| -Electrode tip<br>style | -Blade | -Blade | -Blade | Same |
| -Electrode<br>Material | -Stainless steel | -Stainless steel | - Stainless steel | Same |
| - Length | - 50mm, 63.5mm, 69<br>mm, and 120mm. | - 69 mm | -69mm or longer | Similar, longer<br>lengths do not<br>affect cut and<br>coag.<br>performance but<br>only facilitates<br>the electrode tip<br>go to deeper<br>surgical site. |
| | | | | |
| - Diameter | -2,36 mm | -2,36 mm | ~2.36mm | Same |
| - Conductive<br>end | -15.5mm | -15.5mm | -15.5mm | Same |
| - Tip width | - 2.3mm | - 2.3mm | - 2.3mm | Same |
| - Tip depth | ~0.5mm | ~0.5mm | ~0.5mm | Same |
| - Insulation<br>material | - PTFE Shrink Wrap | - HIPS | - HIPS | The proposed<br>electrode passes<br>the required tests<br>according to<br>ISO10993,<br>IEC60601-1 and<br>IEC60601-2-2 so<br>there will not any<br>issue for safety. |
| - Electrode<br>coating | -Teflon one coat | -Teflon one coat | - Teflon one coat | Same |
| -Rated<br>accessory<br>voltage | - 4.5kVp | - 5kVp | - 5kVp | Similar, does not<br>affect<br>performance |
| Max.<br>operation<br>power | -300W | -300W | -300W | Same |
| Sterile | EO sterile SAL 10-6 | EO sterile SAL 10-6 | EO sterile SAL 10-6 | Same purpose,<br>EO sterilization is<br>validated<br>according to ISO<br>11135 |
| Shelf life | 3 years | 3 years | 3 years | Same |
| Electrical<br>safety | Comply with<br>dielectric strength in<br>with accordance<br>ES60601-1,<br>IEC60601-1-2 &<br>IEC60601-2-2 | Comply with<br>dielectric strength in<br>with accordance<br>ES60601-1,<br>IEC60601-1-2 &<br>IEC60601-2-2 | Comply with<br>dielectric strength in<br>with accordance<br>AAMI HF-18 | Same kinds of<br>safety<br>requirements |
| Biocompatibil<br>ity | Comply with<br>ISO10993 | Comply with<br>ISO10993 | Comply with<br>ISO10993 | Meet to<br>biocompatibility<br>requirements so<br>it does not raise |
| |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.