Swyng® non-stick bipolar forceps, single-use are intended for use in electrosurgery for coagulation of tissue.
Device Story
Sutter Swyng non-stick bipolar forceps are single-use electrosurgical instruments; bayonet-style design with straight or angled tips. Device connects to an electrosurgical generator via a bipolar cable to deliver electrical energy for tissue coagulation. Used by clinicians in surgical settings. The device functions by applying high-frequency current to tissue held between the forceps tips. Non-stick properties facilitate tissue handling during coagulation. Benefits include effective hemostasis and reduced tissue adhesion during surgical procedures.
Clinical Evidence
No clinical data. Bench testing only. Testing included electromagnetic compatibility, electrical safety (IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2), mechanical strength, and thermal effects on tissue. Thermal testing compared coagulation performance against the predicate using histology and morphometric measurements, demonstrating equivalence.
Technological Characteristics
Materials: Silver and rhodium-plated aluminum tips/branches/handle, Polyamide (PA 11) insulation. Design: Bayonet-style, single-use. Energy: Electrosurgical (max 500 Vp). Sterilization: Gamma irradiation (SAL 10^-6). Standards: ISO 10993-1, ISO 11137, ISO 11607-1, IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2.
Indications for Use
Indicated for use in electrosurgery for coagulation of tissue in patients requiring such procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
February 10, 2020
Sutter Medizintechnik GmbH Mr. Ulrike Zeissler Manager, Regulatory Affairs Tullastrasse 87 79108 Freiburg Germany
Re: K193587
Trade/Device Name: Sutter Swyng non-stick bipolar forceps, single-use Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: December 17, 2019 Received: December 23, 2019
Dear Ulrike Zeissler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193587
#### Device Name
Sutter Swyng® non-stick bipolar forceps, single-use
Indications for Use (Describe)
Swyng® non-stick bipolar forceps, single-use are intended for use in electrosurgery for coagulation of tissue.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Section 807.92
| Date: | February 10, 2020 | |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Name:<br>Address: | Sutter Medizintechnik GmbH<br>Tullastrasse 87<br>79108 Freiburg<br>Germany |
| | Contact person:<br>Titel:<br>Telephone:<br>Fax: | Ulrike Zeissler<br>Manager Regulatory Affairs<br>+49 (0) 761 51551-14<br>+49 (0) 761 51551-30 |
| Product: | Trade Name:<br>Common Name:<br>Classification<br>name:<br>Product Code:<br>Regulation<br>Number:<br>Classification:<br>Classification<br>Panel: | Sutter Swyng® non-stick bipolar forceps,<br>single-use<br>Bipolar forceps<br>Electrosurgical, Cutting & Coagulation &<br>Accessories<br>GEI<br>CFR 21 § 878.4400<br>Class II<br>General and Plastic Surgery |
| Predicate<br>Device: | Predicate device to which Sutter Swyng® non-stick bipolar forceps,<br>single-use are claimed to be substantially equivalent is<br>manufactured by Medos International SARL as CODMAN<br>VersaTru™ Disposable Non-Stick Bipolar Forceps (K162469). | |
| Device<br>Description: | Sutter Swyng® non-stick bipolar forceps, single-use are<br>electrosurgical instruments. The bipolar forceps are provided with<br>bayonet-style handle design with straight or angled tips and<br>different total lengths. They are to be connected through an<br>appropriate bipolar cable with the bipolar output of an<br>electrosurgical generator. The electrodes are provided sterile and<br>are single-use instruments. | |
| Indications for<br>Use: | Swyng® non-stick bipolar forceps, single-use are intended for use<br>in electrosurgery for coagulation of tissue. | |
{4}------------------------------------------------
| Technological<br>Characteristics: | The table below provides a comparison between the subject device<br>and the predicate device | |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| | Sutter Swyng® non-<br>stick bipolar forceps,<br>single-use | Predicate device<br>K162469 |
| | Product Code | GEI |
| | Classification | Class II – 21 CFR<br>878.4400 |
| | Classification Name | Electrosurgical, Cutting<br>& Coagulation &<br>Accessories |
| | Forceps design | Bayonet Style |
| | Product Line | Standard, Slim |
| | Tip size | Ø 0.5 mm, 1.0 mm, 1.5<br>mm |
| | Material<br>• Tips, Branches, Handle<br>• Insulation | Silver and rhodium<br>plated aluminium<br>Polyamide (PA 11) |
| | Meets IEC 60601-1 | Yes |
| | Meets IEC 60601-2-2 | Yes |
| | Meets IEC 60601-1-2 | Yes |
| | Maximum peak<br>voltage | 500 Vp |
| | Meets ISO 10993-1 | yes |
| | Sterility Assurance<br>Level (SAL) | 10-6 |
| | Sterilization Method | Gamma irradiation |
| Non-Clinical<br>Performance<br>Data: | The following performance data has been obtained for the<br>substantial equivalence determination.<br><br>Bench Testing<br>Performance testing has been executed in line with the internal<br>R&D process and in compliance with the proposals and<br>recommendations of the FDA guidance: "Premarket Notification<br>[510(k)] Submissions for Electrosurgical Devices for General<br>Surgery" - Guidance for Industry and Food and Drug Administration<br>Staff, August 15, 2016.<br><br>In particular, tests were carried out with respect to following subject<br>areas: | |
| | Performance Data Bench Tests | |
| | Test | Conclusion |
| | Electromagnetic Compatibility and<br>Electrical Safety | Pass |
| | Mechanical strength and<br>functionality performance testing | Pass |
| | Thermal effects on tissue | Pass |
{5}------------------------------------------------
| <b>Electromagnetic Compatibility and Electrical Safety Testing</b><br>Electrical and electromagnetic tests were performed to demonstrate<br>that design specifications and performance requirements are met.<br>Compliance to the voluntary standards IEC 60601-1 (AAMI/ANSI<br>ES60601-1:2005), IEC 60601-2-2 and IEC 60601-1-2 has been<br>shown. | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Mechanical strength and functionality performance testing</b><br>Mechanical strength and functionality performance testing was<br>performed to demonstrate that design specification are met.<br>Mechanical stress tests showed that the design specification are<br>met. | |
| <b>Thermal effects on tissue</b><br>Thermal effects on tissue testing was performed to determine<br>thermal effects caused by the Sutter Swyng® non-stick bipolar<br>forceps, single-use at different power levels and application times in<br>comparison to the predicate device. Three different types of tissue<br>were used and tests were performed in triplicate. Visual comparison<br>as well as digital morphometric measurement using histology<br>showed equivalent coagulation performance of subject device and<br>predicate device. | |
| <b>Sterilization</b><br>Sutter Swyng® non-stick bipolar forceps, single-use are sterilized<br>by using a validated gamma irradiation cycle. The sterilization cycle<br>has been validated to ensure a sterility level of (SAL) 10-6 in<br>accordance with ISO 11137. | |
| <b>Shelf Life Testing</b><br>Shelf-life testing has been conducted in accordance ISO 11607-1.<br>The aging studies established that the Sutter Swyng® non-stick<br>bipolar forceps, single-use and packaging remain functional and<br>maintain sterility for up to 3 years. | |
| <b>Biocompatibility Testing</b><br>Biological evaluation and Biocompatibility testing has been<br>performed in compliance to ISO 10993-1 “Biological evaluation of<br>medical devices - Part 1: Evaluation and testing within a risk<br>management process”. | |
| <b>Animal Studies</b><br>No animal studies have been performed as appropriate verification<br>and validation of the subject device has been achieved based on<br>comparison to the predicate device and from results of the bench<br>testing, biocompatibility evaluation, and electrical / safety testing. | |
| <b>Clinical Studies</b><br>No clinical studies have been performed as appropriate verification and validation of the subject device has been achieved based on | |
| | comparison to the predicate device and from results of the bench<br>testing, biocompatibility evaluation, and electrical / safety testing. |
| <b>Conclusion:</b> | Sutter Swyng® non-stick bipolar forceps, single-use are<br>substantially equivalent to the predicate device since the intended<br>use, design, material and basic features are the same. The minor<br>differences raise no new issues of safety and effectiveness, as the<br>design differences have no effect on the performance, function or<br>intended use. |
| | |
{6}------------------------------------------------
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.