K190422 · Viveve Medical · GEI · Jun 12, 2019 · General, Plastic Surgery
Device Facts
Record ID
K190422
Device Name
Viveve System 2.0
Applicant
Viveve Medical
Product Code
GEI · General, Plastic Surgery
Decision Date
Jun 12, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Viveve® 2.0 System is indicated for use in general surgical procedures for electrocoagulation and hemostasis.
Device Story
The Viveve 2.0 System is a monopolar radiofrequency (RF) electrosurgical device used for tissue coagulation and hemostasis. It consists of an RF generator, a handpiece, a footswitch, and sterile disposable treatment tips (5cm or 8cm). The system delivers RF energy to heat a specific volume of tissue beneath the surface while simultaneously delivering cryogen (R134a or 1234ze) to the treatment tip to cool and protect the surface tissue, creating a reverse thermal gradient. The device is operated by a clinician in a surgical setting. The generator provides the heating energy and incorporates a cooling module. The handpiece couples the energy to the tissue via the treatment tip. The footswitch controls RF energy activation. The system uses a touchscreen interface for user control. By heating submucosal tissue while protecting the mucosa, the device achieves coagulation/hemostasis. The 2.0 version features an updated touchscreen console, modified user interface, and a PAL-based operating system for improved technology compatibility.
Clinical Evidence
No human clinical data provided. Evidence consists of bench testing (electrical safety, EMC, software verification/validation, packaging, shelf-life) and three animal studies (ovine vaginal model). GLP studies (IV0311g, VI-OV-SAF-04) demonstrated that RF treatment at 90-120 J/cm² increases submucosal fibroblast activation and collagen density without mucosal erosion, necrosis, or hypertrophic scarring. Non-GLP study (VI-OV-SAF-05) confirmed the Viveve 2.0 System's temperature-time profile is substantially equivalent to the predicate device, with deep tissue heating (3-8 mm) and rapid dissipation, confirming safety and performance.
Technological Characteristics
Monopolar RF electrosurgical device; 6.78 MHz frequency; 240W max power; 25-120 Ω impedance range. Uses cryogen (R134a or 1234ze) for surface cooling. Components: RF generator, handpiece, footswitch, disposable tips. Software runs on a Platform Abstraction Layer (PAL) supporting UNIX/Linux environments. Sterilization: ETO. Packaging: Tyvek pouch. Connectivity: Touchscreen interface, optical power switch.
Indications for Use
Indicated for use in general surgical procedures for electrocoagulation and hemostasis.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 12, 2019
Viveve Inc. Suzon Lommel Sr. VP, Regulatory and Quality Affairs 345 Inverness Drive South, Building B, Suite 250 Englewood, Colorado 80109
Re: K190422
Trade/Device Name: Viveve 2.0 System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: February 15, 2019 Received: February 21, 2019
Dear Suzon Lommel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Jennifer R. Stevenson Acting Division Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190422
Device Name Viveve® 2.0 System
Indications for Use (Describe)
The Viveve® 2.0 System is indicated for use in general surgical procedures for electrocoagulation and hemostasis.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "VIVEVE" in a sans-serif font. The letters are a light green color. The word is all in uppercase letters. There is a small circle with an R in it to the right of the last E.
## SECTION 5
## 510(k) SUMMARY
#### 5.1 REGULATORY AUTHORITY
Safe Medical Devices Act of 1990, 21 CFR 807.92
#### 5.2 APPLICANT INFORMATION
- Viveve® Inc. Applicant: 345 Inverness Drive South Suite: B-250 Englewood, CO 80112
- Suzon Lommel Contact: Sr. VP, Regulatory and Quality Affairs slommel@viveve.com C: 408-645-4979 F: 720-696-8199
Date Prepared: February 15, 2019
#### 5.3 SUBJECT DEVICE INFORMATION
Trade Name: Viveve® 2.0 System Electrosurgical System Common Name: Product Code: GEI Classification Name: Electrosurgical Cutting and Coagulation Device and Accessories (21 CFR 878.4400) Device Panel: General Surgery/Restorative Device Device Classification: Class II
#### 5.4 PREDICATE DEVICE
Viveve® System (K180584)
#### 5.5 DEVICE DESCRIPTION
The Viveve 2.0 System utilizes monopolar radiofrequency (RF) energy to selectively heat a given volume of tissue beneath the surface, while cryogen is delivered to the inside of the treatment tip to cool the surface tissue. The generator delivers energy to the treatment tip to create an electric field under the treatment tip (electrode). The mechanism of action is the application of RF energy to the tissue causing coagulation and/or hemostasis.
The Viveve 2.0 System consists of four (4) primary components:
- . An RF Generator to provide the heating energy. The Generator incorporates the Cooling Module to supply coolant which provides the cooling energy.
- . A hand piece that couples the cooling and heating energy to the tissue through the treatment tip.
- A footswitch that allows the user to turn the RF Energy on or off.
- . 5cm or 8cm Sterile Disposable Treatment Tips.
{4}------------------------------------------------
Accessories include:
- . Coupling Fluid
- Cryogen ●
- Return Cable ●
- Return Pad ●
- Power Cord .
## 5.6 INDICATIONS FOR USE
The Viveve 2.0 System is indicated for use in general surgical procedures for electrocoagulation and hemostasis.
## 5.7 COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The technological characteristics of the subject device Viveve 2.0 System are substantially equivalent to the predicate device, Viveve System (K180584). The Viveve 2.0 System is an electrosurgical device that delivers radiofrequency (RF) energy to selectively heat a given area of tissue, while cryogen is delivered to the inside of the treatment tip to cool the surface tissue at the end of energy deposition. The application of RF energy causes the tissue to coagulate and/or become hemostatic.
This submission application confirms the continued conformance to applicable technical design specifications and performance requirements, including requirements associated with industry safety and performance standards.
## 5.8 BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE
The Viveve 2.0 System is substantially equivalent to the predicate device listed in K180584. The principle of operation between the predicate device and the subject device remain the same as do all output parameters to tissue, however the subject device includes the following design and technological/environmental specification modifications:
## 5.8.1 DESIGN MODIFICATIONS
## 5.8.1.1 RF Console (Generator)
- Overall footprint of the RF Console and corresponding case (weight, ● shape and size) is changed to allow for the upgrade to a touchscreen display panel
- RF Console Display screen is updated from standard display to ● touchscreen display
- RF Console Display incorporates an optical power switch
- RF Console Display has modified user interface screens and display graphics for ease of use
- RF Console connector panel is now located in the back to minimize inadvertent damage over the use life
{5}------------------------------------------------
#### 5.8.1.2 Footswitch
- Cover is added to the unit
- . Stylistic changes are incorporated in the overall design
#### 5.8.1.3 Handpiece
- Color is changed from purple to white for brand alignment ●
- Connector is changed from female to male to prevent connection to . Predicate system
#### 5.8.1.4 Cryogen
- R134a (1,1,1,2 tetrafluoroethane) or 1234ze (trans-1, 3, 3, 3-● Tetrafluoroprop-1-ene) cryogen can be used in the Viveve 2.0 System. 1234ze has been developed as a more environmentally friendly option.
# 5.8.2 SOFTWARE MODIFICATIONS
#### 5.8.2.1 Viveve RF Console Software
- Updated to most recent version for incorporation of changes impacting the ● RF Console
#### 5.8.2.2 Viveve RF Display Module Software
- Updated to most recent version for incorporation of changes impacting the ● RF Display Module
#### 5.8.2.3 Viveve RF Handpiece Software
- Updated to most recent version for incorporation of changes impacting the ● RF Handpiece
## 5.8.3 HARDWARE MODIFICATIONS
## 5.8.3.1 Operating System
- . Viveve 2.0 System operates on a PAL platform instead of PC/104 to allow for most up-to-date technology. The PAL is a Platform Abstraction Layer that is used in a variety of projects. The PAL allows for easy compatibility between many different flavors of UNIX/Linux, including AIX 6.1 and later, HP/UX 11.31 and later, Solaris 5.10 and later, and most versions of Linux as far back as RedHat 5.0, SuSE 10.1, and Debian 5.0.
## 5.8.4 LABELING MODIFICATIONS
## 5.8.4.1 Technical User Manual
- Updates include but are not limited to modified user interface graphics and ● nomenclature relative to the Viveve 2.0 System
## 5.8.4.2 Instructions for Use
- Updates include but are not limited to nomenclature relative to the Viveve ● 2.0 System There have been no changes related to clinical use other than the differences in the GUI
{6}------------------------------------------------
## 5.8.5 TECHNICAL/ENVIRONMENTAL SPECIFICATION MODIFICATIONS
#### 5.8.5.1 Environmental and Packaging Specifications
- . IEC60601. Electrostatic Discharge (ESD) and Voltage Dip are aligned with CMO's Quality Management System (QMS) Requirements
#### 5.8.5.2 RF Frequency
- Is modified to 6.78 MHz to allow for the appropriate function due to ● updated internal components (many predicate components are at end of life per the manufacturers) of the updated system with the same output to tissue
#### 5.8.5.3 Operation temperature
The operational temperature in 2.0 System is decreased to allow for increased product longevity while maintaining the same heat profile to tissue in the clinical environment.
#### 5.8.5.4 Storage pressure
- Range is now aligned with that outlined in packaging and environmental ● testing
#### 5.8.6 CONTRACT MANUFACTURER MODIFICATIONS (CMO)
- Sparton Medical Systems is added for closer proximity to Viveve ● headquarters to allow for increased engineering involvement
- . Cirtec Medical is now added for increased treatment tip sourcing options and the only manufacturer of the Viveve 2.0 Treatment Tip
All previously outlined modifications to the Viveve 2.0 System are discussed in further detailed in Section 12: Substantial Equivalence Discussion of this Premarket Notification. A comparison of the technical characteristics of Viveve 2.0 System are compared to those of the predicate device, Viveve System, in Table 5-1 below.
| Item | Viveve 2.0 System<br>(Subject Device) | Viveve System<br>(Predicate Device K180584) |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Subject device | K180584 |
| Legal Manufacturer | Viveve, Inc. | Viveve, Inc. |
| Contract Manufacturer | Sparton Medical Systems (Generator and Handpiece) Cirtec Medical (Treatment Tips) | Stellartech Research Corporation (All 3 system components) |
| Indication for Use | The Viveve 2.0 System is indicated for use in general | The Viveve System is indicated for use in general surgical |
| Item | Viveve 2.0 System | Viveve System |
| | (Subject Device) | (Predicate Device K180584) |
| | surgical procedures for<br>electrocoagulation and<br>hemostasis. | procedures for electrocoagulation<br>and hemostasis. |
| FDA Classification | Class II | Class II |
| CFR/Product Code | 21 CFR 878.4400/GEI | 21 CFR 878.4400/GEI |
| Invasiveness of Treatment | Non-invasive. Device<br>applies to the surface. | Non-invasive. Device applies to<br>the surface. |
| Principles of Operation | Radiofrequency (RF)<br>energy selectively heats a<br>given volume of tissue<br>beneath the surface, while<br>cryogen is delivered to the<br>inside of the Treatment Tip<br>to cool the surface tissue.<br>The Treatment Tip is placed<br>on the surface of the skin<br>and the internal tissues are<br>heated while the surface<br>tissue is protected. (Reverse<br>thermal gradient) | Radiofrequency (RF) energy<br>selectively heats a given volume<br>of tissue beneath the surface,<br>while cryogen is delivered to the<br>inside of the Treatment Tip to<br>cool the surface tissue. The<br>Treatment Tip is placed on the<br>surface of the skin and the<br>internal tissues are heated while<br>the surface tissue is protected.<br>(Reverse<br>thermal gradient) |
| Energy | RF | RF |
| Treatment Type | Monopolar | Monopolar |
| Main Input<br>(Input<br>voltage/Current/Frequency) | 100 - 120 Vac / 10A / 50/60 Hz<br>220 - 240 Vac / 5A / 50/60 Hz | 100 - 120 Vac / 10A / 50/60 Hz<br>220 - 240 Vac / 5A / 50/60 Hz |
| Maximum Power<br>(generator) | 240 Watts | 240 Watts |
| Operating Frequency | 6.78 MHz ±15%. | 6 MHz ±2%. |
| Voltage Waveform | 6.78 MHz continuous<br>sinusoidal waveforms | 6.0 MHz continuous sinusoidal<br>waveforms |
| Electrode Probe | Monopolar | Monopolar |
| Impedance Range | 25 - 120 Ω | 25 - 120 Ω |
| Tip | 5cm and 8cm Treatment<br>Tips | 5cm and 8cm Treatment Tips |
| Packaging | Tyvek pouch | Tyvek pouch |
| Sterility | ETO | ETO |
| Cooling Solution | Cryogen | Cryogen |
#### Table 5-1: Comparison of Technological Characteristics of Viveve 2.0 System and Viveve System (K180584)
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "viveve" in a sans-serif font. The letters are a light green color. The word is all lowercase.
# 5.9 PERFORMANCE DATA
Design verification testing, including bench performance, electrical safety/electromagnetic compatibility, software verification/validation, packaging and shelf-life studies, provided in
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "viveve" in a sans-serif font. The color of the text is a light green. The letters are all lowercase, and the word is written in a single line. The background is white.
the subject premarket notification demonstrate that the Viveve 2.0 System is substantially equivalent to the predicate device. Viveve System.
## 5.9.1 Performance Data
Design development and control is conducted in phases as described in the Sparton Medical Design Control Procedure 120924. Coordination of project activities is assigned to a Project Leader. A Design Development team is formed and includes members from appropriate groups in the organization. Team members are assigned responsibilities as appropriate for completion of the project. The design of the product is verified and validated to ensure design inputs are met. Copies of the Design Verification and Validation Test Reports are maintained in the Design History File and further detailed in Section 18.
The following are the documents to support the Design Verification and Validation testing of the Viveve 2.0 System.
| Document Title | Document Number |
|--------------------------|----------------------------|
| Design Control Procedure | Sparton 120924 |
| Animal Study Reports | Study Group C: 15 Dec 2010 |
| | Study Group D: 16 Dec 2010 |
| | Study Group E: 18 Jan 2011 |
| | Study Group F: 20 Jan 2011 |
| | Study Group G: 26 Jan 2011 |
| | Study Group H: 26 Apr 2011 |
| | Study Group I: 21 Apr 2011 |
## 5.9.1 Pre-Clinical Performance Data
Viveve conducted the following Animal Studies to prove safety and efficacy of the device:
- GLP Study IV0311g ●
- GLP Study VI-OV-SAF-04 (VIV-1702) ●
- Non-GLP Study VI-OV-SAF-05 (VIV-1801) ●
Study summaries are below:
## GLP STUDY IV0311g
The objective of GLP study IV0311g was to evaluate the safety of and tissue response after the use of a radiofrequency (RF) device intended to treat vaginal laxity in humans. Using an in-vivo ovine vaginal model, this study demonstrated how the Viveve System preserves the surface vaginal mucosal viability while raising the temperature of the submucosal tissues to approximately 50°C. The various means of heating tissue include laser, ultrasound, electrosurgical tools, and conduction of radio frequency (RF) electrical current through the tissue. Of these methods, the conduction of RF electrical current
{9}------------------------------------------------
through tissue was the most benign and practical means of gently heating the tissue to a temperature low enough to disrupt the tissue, but not destroy/denature the collagen or elastin.
A summary of the GLP animal study IV0311g is listed below.
- · Study Design: Blinded vaginal introitus study conducted at a single test facility.
- · Objective: The Good Laboratory Practices ovine vaginal introitus study was performed to evaluate the Viveve Vaginal Laxity RF Treatment and Accessories in a fashion that emulated its future clinical use.
- · Animal Number and Species: 25 non-nulliparous female ovine (Ovisaries)
- Purebred or black-faced Suffolk crossbreeds of various strains: This animal model was selected based on the similar anatomical structure as compared to the human.
- · Test System: In-vivo Suffolk ovine vaginal introitus model.
- · Control Treatment: Untreated (Group A, n=3) and surface cryogen delivery only (Group B, n=3) ovine.
- Group A consisted of 3 untreated sheep that were anesthetized for a time o period comparable to the treatment groups in this study. Group A sheep served as observational and biopsy controls. Biopsy samples were taken at post-treatment intervals of 7, 30, and 90 days.
- Group B consisted of 3 sheep serving as a sham control that were treated with O a minimum of RF energy (1-2 J/cm2) and cryogen cooling. Biopsy samples were taken at post-treatment intervals of 7, 30, and 90 days.
- Active Treatment: Viveve Vaginal Laxity RF Treatment and Accessories. ● Delivering 60, 90, 120, or 160 J/cm² with surface cryogen cooling and 5 treatment passes. (Groups C, F, & G, n=19). Group C consisted of 4 different treatment levels of 16 total sheep, with each group receiving an escalated treatment dose of RF energy starting at the 1:00 o'clock position and finishing at the 11:00 o'clock position with 5 repetitive treatments. Biopsy samples were taken at post-treatment intervals of 7, 30 and 90 days.
- 60 J/cm2: 4 sheep o
- 90 J/cm2: 4 sheep [These 4 sheep were transitioned over to Group F, see O below.l
- 120 J/cm2: 5 sheep O
- 160 J/cm2: 3 sheep O
Group F consisted of the 4 sheep from Group C that were treated with 90 J/cm². Of these 4 sheep, 3 sheep were necropsied at 90 days and 1 sheep necropsied at 180 days. This was updated in Amendment 12.
Group G consisted of 3 acute sheep treated with 90 J/cm² while monitoring temperature at 3 depth levels below the vaginal mucosa. The placement of the thermistors to monitor temperature were at depths of 3, 6 and 9 mm from the surface of the vaginal mucosa and within the 2 cm area of the treatment probe when placed on the vaginal mucosa.
Non-overlapping vaginal introitus biopsies were obtained on Days 7. 30, and 90. These biopsies were formalin-fixed, routine processed, and hematoxylin and eosin (H&E) and
{10}------------------------------------------------
Gömöri trichrome stained. In total, 96 in vivo treated and 15 control vaginal introitus biopsies were evaluated between 7- and 90-days post-treatment. Using routine light microscopy, the slides were evaluated by two blinded board-certified pathologists. The vaginal tissues were histologically evaluated for signs of thermal tissue injury, fibroblast activation, and qualitative collagen increases.
The study's pathology-based criteria for success were defined as the presence of increased submucosal fibroblast activation over control levels with:
- . No mucosal epithelial erosion and/or ulceration;
- No tissue necrosis regions;
- No granulation tissue foci; and/or .
- . No extensive or hypertrophic scar-like collagen increases.
The submucosa was assessed for the cytologic features of fibroblast activation. Fibroblasts were considered quiescent when they had condensed nuclear chromatin and scant cytoplasm. Fibroblasts were considered stimulated/activated when they had open chromatin and more abundant cytoplasm. Based on their prominence and cellular density, the stimulated fibroblasts were graded along the following spectrum:
- Minimal, if activated, fibroblasts were rare; ●
- Mild, if activated, fibroblasts were frequent, scattered and without clustering; .
- . Moderate, if activated, fibroblasts were frequent, focally clustered and with increased cellular density; or
- . Extensive, if activated, fibroblasts were frequent with multifocal clustering and increased cellular density.
The submucosa was also qualitatively assessed for increases in collagen content. If the collagen appeared increased, it was graded along the following spectrum:
- Minimal, if collagen appeared to have a slightly increased fiber density, maintained a fibrillary architecture, and had a normal cellular density;
- Mild, if collagen fibers appeared slightly courser or compact without a decrease in regional cellularity;
- . Moderate, if collagen fibers appeared moderately courser or compact with mild regional hypocellularity; or
- Extensive, if collagen fibers appeared densely compacted or hyalinized . (hypertrophic scar-like) with regional hypocellularity.
To facilitate statistical comparison of the Viveve Treatment doses with the controls, the descriptive fibroblast activation and collagen categories for each biopsy were converted to an ordinal, approximately interval type, scaled score as follows:
- Score 0 = Not identified ●
- Score 0.3 = Minimal ●
- Score 0.5 = Minimal to Mild .
- Score 1 = Mild or Minimal with focal mild ●
- Score 1.5 = Mild to Moderate ●
{11}------------------------------------------------
- Score 2 = Moderate or Mild with focal moderate .
- . Score 2.5 = Moderate to Extensive
- Score 3 = Extensive or Moderate with focal extensive
The individual biopsy results were used to calculate composite group means, standard deviations, medians, minimums and maximums for both the treated and control groups. Fibroblast activation and qualitative collagen scores were statistically compared using the non-parametric Wilcoxon rank sums method with significance of p≤0.05.
Results: In this study, the 90 J/cm² and 120 J/cm² Viveve Treatments significantly increased fibroblast activation over the baseline activation identified in the controls. This fibroblast activation was associated with an increase in submucosal collagen in Days 30 and 90 biopsies. In these biopsies, the mean collagen increase was approximately twice that of the controls. At Days 30 and 90, some controls showed focal minimal-mild qualitative collagen increases, consistent with background changes related to their prior vaginal manipulation and biopsies.
In support of the Viveve Vaginal Laxity RF Treatment mode of action and efficacy, the 90 J/cm² and 120 J/cm² treatments met the Criteria for Success with increased submucosal fibroblast activation and reasonable collagen increases. Mucosal erosion-ulceration, tissue necrosis, granulation tissue and excessive collagen increases (hypertrophic scarlike) were not identified in this study, consistent with an appropriate safety profile.
Summary: A series of ninety-six (96) treated and fifteen (15) control ovine vaginal introitus biopsies have been evaluated 7 to 90 days post-treatment. Over the 90-day study period, the 90 J/cm² and 120 J/cm² treatments met the pathology-based Criteria for Success with:
- . Significantly increased submucosal fibroblast activation without mucosal erosionulceration, tissue necrosis:
- . No granulation tissue or extensively increased (hypertrophic scar-like) collagen.
The 160 J/cm² energy level treatment showed similar changes that approached significantly increased fibroblast activation over time. The 60 J/cm² energy level treatment lacked sufficient fibroblast activation to meet the criteria.
## GLP Study VI-OV-SAF-04 (VIV-1702)
The objective of this acute GLP study was to evaluate the temperature-time profile and histopathology of vaginal tissue following radiofrequency (RF) energy administration in the ovine model. Using an in-vivo ovine vaginal model, this study demonstrated how the Viveve System preserves the surface vaginal mucosal viability while raising the temperature of the submucosal tissues to approximately 50°C. The various means of heating tissue include laser, ultrasound, electrosurgical tools, and conduction of radio frequency (RF) electrical current through the tissue. Of these methods, the conduction of RF electrical current through tissue was the most benign and practical means of gently heating the tissue to a temperature low enough to disrupt the tissue, but not destroy/denature the collagen or elastin.
A summary of the GLP animal study VI-OV-SAF-04 is provided below.
- Study Design: Vaginal introitus study conducted at a single test facility. ●
{12}------------------------------------------------
- Animal Number and Species: Six parous female ovine (Ovisaries) .
- . Test System: In-vivo Suffolk ovine vaginal introitus model.
- . Treatments: Two treatments were employed in this study.
The Clinical Treatment Protocol consisted of 4 sheep that were treated according to the standard clinical protocol (110 pulses, 5 passes around the vaginal introitus. Prior to treatment, fluoroptic temperature probes were placed at four positions in the in the introital region of the vaginal canal in the plane of energy flow (at the mucosal surface, two submucosal positions and one probe in the muscularis). Once the temperature probes were placed, baseline local tissue temperatures were taken for 3-5 minutes. While still under general anesthesia, each sheep was treated with the Viveve device at the introitus in a manner similar to the clinical setting and tissue temperatures measured over time during and after treatment (1-hour post-treatment or until temperature returned to baseline). Baseline temperature was defined as the tissue temperatures just prior to treatment. The core body temperature was also recorded prior to, during and post treatment, and changes were noted in the assessment of potential local tissue temperature baseline changes. Each sheep was euthanized, and necropsy performed.
Collection of vaginal wall segments (one section on the side of the vagina where the probes were located and one 180 degrees across the vagina in the treatment site) and perivaginal tissue (urethra, rectum, and bladder samples) were conducted. Tissues were preserved in 10% formalin and the vaginal/peri-vaginal tissues sent for histological analysis to assess any potential effects of treatment. Prior to treatment, fluoroptic temperature probes were placed at four positions in the introital region of the vaginal canal in the plane of energy flow (at the mucosal surface, two submucosal positions and one probe in the muscularis). Once the temperature probes were placed, baseline local tissue temperatures were taken for 3-5 minutes.
While still under general anesthesia, each sheep was treated with the Viveve device at the introitus in a manner similar to the clinical setting and tissue temperatures measured over time during and after treatment (1-hour post-treatment or until temperature returned to baseline). Baseline temperature was defined as the tissue temperatures just prior to treatment. The core body temperature was also recorded prior to, during and post treatment, and changes were noted in the assessment of potential local tissue temperature baseline changes. Each sheep was euthanized, and necropsy performed. Collection of vaginal walls (one section on the side of the vagina where the probes were located and one 180 degrees across the vagina in the treatment site) and peri-vaginal tissue (urethra, rectum, and bladder samples) were conducted. Tissues were preserved in 10% formalin and the vaginal/peri-vaginal tissues sent for histological analysis to assess any potential effects of treatment.
The Supra-Therapeutic Protocol consisted of 2 sheep that were treated with 110 pulses, but with 15 passes which is three times the clinical dose with no time lag between pulses. This treatment constituted a worst-case scenario in that more energy was delivered in a short duration due to no lag time between pulses which would be expected to cause the tissue temperature to rise to higher levels since the tissue would not have time to return to basal temperature before another pulse was delivered at the same location.
Results: Neither the Clinical Protocol nor the Supra-Therapeutic Protocol with the Viveve RF device resulted in temperatures that would be expected to cause cellular damage. This was confirmed by the histological evaluation. Maximum temperatures were well below temperatures that would pose safety concerns and were within the range that
{13}------------------------------------------------
could result in the desired cellular changes. Administration of radiofrequency from the Viveve system resulted in tissue temperature increases in deep tissue lavers, with approximately 3-8 mm seeming to be the target area. Heating did appear in the muscularis, but there was not deposition of heat in this area and the heat dissipated rapidly. Variations in temperature-time profiles appeared to be a result of bodytype/tissue characteristics that are unique to each ewe; but in all cases, there were no profile differences that would indicate a safety concern.
Conclusion: The objective of this acute GLP study was to evaluate the temperature-time profile and histopathology of vaginal tissue following radiofrequency (RF) energy administration in the ovine model. A total of 6 parous ewes were divided into 2 groups of 4 sheep in one and 2 sheep in the other; one group (N=4) receiving the Clinical Treatment protocol energy level and the other group (N=2) receiving the Supra-therapeutic energy administration protocol. Tissues were preserved in 10% formalin and the vaginal/perivaginal tissues sent for histological analysis to assess any potential effects of treatment. Tissue temperature data were reported in a sub-report which included graphical and tabular summaries for each location and treatment protocol.
Tissue temperature results demonstrated that during both the Clinical and Supratherapeutic procedures, treatment with the Viveve RF device did not result in temperatures that would be expected to cause cellular damage. This was confirmed by the histological evaluation. Maximum temperatures were well below temperatures that would pose safety concerns and yet were within the range that could result in the desired cellular changes. Administration of radiofrequency from the Viveve system resulted in tissue temperature increases in deep tissue layers, with approximately 3-8 mm seeming to be the target area. Heating did appear in the muscularis, but there was not deposition of heat in this area and the heat dissipated rapidly. Variations in temperature-time profiles appeared to be a result of body-type/tissue characteristics that are unique to each ewe: but in all cases, there were no profile differences that would indicate a safety concern. Under the conditions of this study, there were no histopathologic findings associated with radiofrequency (RF) energy administration at either planned Clinical levels or Supratherapeutic levels of RF treatment.
# Non-GLP Study VI-OV-SAF-05 (VIV-1801)
The objective of this acute non-GLP study was to evaluate the temperature-time profile of vaginal tissue following radiofrequency (RF) energy administration in the ovine model to validate the Viveve 2.0 System. Using an in-vivo ovine vaginal model, this study demonstrated how the Viveve 2.0 System preserves the surface vaginal mucosal viability while raising the temperature of the submucosal tissues to≤50°C. This study also demonstrates substantial equivalency between the Viveve 2.0 System and the currently cleared Viveve System. The various means of heating tissue include laser, ultrasound, electrosurgical tools, and conduction of radio frequency (RF) electrical current through the tissue. Of these methods, the conduction of RF electrical current through tissue was the most benign and practical means of gently heating the tissue to a temperature low enough to disrupt the tissue, but not destroy/denature the collagen or elastin.
A summary of the non-GLP animal study VI-OV-SAF-05 is provided below.
- Study Design: Vaginal introitus study conducted at a single test facility. ●
{14}------------------------------------------------
- Animal Number and Species: Three parous female ovine (Ovisaries) ●
- . Test System: In-vivo Suffolk ovine vaginal introitus model.
- Treatments: One treatment was employed in this study.
The Clinical Treatment Protocol consisted of 3 sheep that were treated according to the standard clinical protocol (110 pulses, 5 passes around the vaginal introitus). Prior to treatment, fluoroptic temperature probes were placed at four positions in the in the introital region of the vaginal canal in the plane of energy flow (at the mucosal surface, two submucosal positions and one probe in the muscularis), as per similar procedures used for Study VI-OV-SAF-04. Once the temperature probes were placed, baseline local tissue temperatures were taken for 3-5 minutes. While still under general anesthesia, each sheep was treated with the Viveve 2.0 System at the introitus in a manner similar to the clinical setting and tissue temperatures measured over time during and after treatment (30-min post-treatment or until temperature returned to baseline). Baseline temperature was defined as the tissue temperatures just prior to treatment. The core body temperature was also recorded prior to, during and post treatment, and changes were noted in the assessment of potential local tissue temperature baseline changes. Each sheep was euthanized post treatment.
Results: The Clinical Protocol with the Viveve 2.0 System resulted in temperatures that would not be expected to cause cellular damage and were substantially equivalent to the currently cleared Viveve System; temperature values were compared to the values measured in the VIV-OV-SAF-04 to determine substantial equivalency. Maximum temperatures were well below temperatures that would pose safety concerns and were within the range that could result in the desired cellular changes. Administration of radiofrequency from the Viveve 2.0 System resulted in tissue temperature increases in deep tissue layers, with approximately 3-8 mm seeming to be the target area. Heating did appear in the muscularis, but there was not deposition of heat in this area and the heat dissipated rapidly. Variations in temperature-time profiles appeared to be a result of body-type/tissue characteristics that are unique to each ewe; but in all cases, there were no profile differences that would indicate a safety concern.
Conclusion: The objective of this acute non-GLP study was to evaluate the temperaturetime profile and demonstrate substantial equivalency between the Viveve 2.0 System and the currently cleared Viveve System when treating vaginal tissue following radiofrequency (RF) energy administration in the ovine model. A total of 3 parous ewes were treated with the Clinical Treatment protocol energy level. Tissue temperature data were reported in a sub-report which included graphical and tabular summaries for each location and treatment protocol.
Tissue temperature results demonstrated that during the Clinical procedure, treatment with the Viveve 2.0 System did not result in temperatures that would be expected to cause cellular damage and were substantially equivalent to the Viveve System. Maximum temperatures were well below temperatures that would pose safety concerns and yet were within the range that could result in the desired cellular changes. Administration of radiofrequency from the Viveve 2.0 System resulted in tissue temperature increases in deep tissue layers, with approximately 3-8 mm seeming to be the target area. Heating did appear in the muscularis, but there was not deposition of heat in this area and the heat dissipated rapidly. Variations in
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the word "viveve" in a sans-serif font. The color of the text is a light green. The text is all lowercase.
temperature-time profiles appeared to be a result of body-type/tissue characteristics that are unique to each ewe; but in all cases, there were no profile differences that would indicate a safety concern.
The studies serve two purposes:
- 1. Prove equivalency to the predicate device and
- 2. Support future equivalency to an indication under evaluation (VI-15-01)
Study summaries are submitted in Section 19. The full studies are available in Attachment 1.
## 5.10 CONCLUSION
The design, technical characteristics, functionality, indications for use, and principle operation of the subject device Viveve 2.0 System remains unchanged from that of the predicate device, Viveve System (K180584). The proposed design modifications do not raise new questions of the safety or efficacy of the device and the intended use of the Viveve 2.0 System remains unchanged from that of the cleared predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.