K183284 · Jeisys Medical, Inc. · GEI · Jan 15, 2020 · General, Plastic Surgery
Device Facts
Record ID
K183284
Device Name
INTRAcel RF Microneedle System
Applicant
Jeisys Medical, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jan 15, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
INTRAcel RF Microneedle System is indicated for use in dermatological and general surgical procedures for electrocoagulation, hemostasis and the percutaneous fractional treatment of the skin in monopolar and bipolar mode.
Device Story
System comprises RF generator with user interface and disposable hand-held electrode insertion device with 49 microneedles. Operator (physician) uses device to deliver RF energy into skin tissue. In bipolar mode, energy flows between adjacent electrode pairs; in monopolar mode, energy flows between active microneedle electrode and external neutral electrode pad. Heat generated at active electrode causes tissue coagulation. Used in clinical settings for dermatological/surgical procedures. Output provides controlled thermal effect for hemostasis and skin treatment. Benefits include precise tissue coagulation and fractional skin treatment.
Clinical Evidence
No human clinical trials provided. Evidence consists of bench testing (electrical safety, EMC, performance, biocompatibility, sterilization, shelf-life) and an animal study (porcine model) evaluating wound healing response, collagen/HSP/MMP expression via H&E staining, HSP47 staining, and RT-PCR over 10 weeks.
Technological Characteristics
Electrosurgical unit; 1MHz frequency; 50W power; 90V output. Microneedle electrodes (SUS304) with ABS tips. Monopolar/bipolar modes. Single-use sterile electrode insertion device. EO gas sterilization. Connectivity via footswitch and cable. Complies with IEC 60601-1, IEC 60601-1-6, and IEC 60601-2-2.
Indications for Use
Indicated for dermatological and general surgical procedures requiring electrocoagulation, hemostasis, and percutaneous fractional skin treatment in monopolar and bipolar modes. No specific age or gender restrictions provided.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
INTRAcel Premium Fractional RF Micro needle (FRM) System (K153727)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 15, 2020
Jeisys Medical Inc. % Parul Chansoria Regulatory Consultant Elexes Medical Consulting 6494 Tralee Village Dr Dublin, California 94568
Re: K183284
Trade/Device Name: INTRAcel RF Microneedle System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: December 16, 2019 Received: December 16, 2019
Dear Parul Chansoria:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183284
Device Name INTRAcel RF Microneedle System
Indications for Use (Describe)
INTRAcel RF Microneedle System is indicated for use in dermatological and general surgical procedures for electrocoagulation, hemostasis and the percutaneous fractional treatment of the skin in monopolar and bipolar mode.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### I. SUBMITTER
Jeisys Medical Inc. 307, Daeryung Techno Town 8th Gamasan-ro 96, Geumcheon-Gu, Seoul, Republic of Korea Contact Person: Parul Chansoria Elexes Medical Consulting 6494 Tralee Village Dr Dublin, CA 94568, USA Telephone: 650-528-2445, 408-475-8091 E-mail: parul@elexes.com Summary Prepared: January 15, 2020 FDA Establishment Number: 3006985163
### II. DEVICE
Common/Usual Name: Electrosurgical coagulation device and accessories Trade Name: INTRAcel RF Microneedle System Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Classification Panel: General and Plastic Surgery Product Code: GEI Regulation Number: 21 CFR 878.4400
## III. PREDICATE DEVICE
The INTRAcel RF Microneedle System is substantially equivalent to the following cleared device:
| Company | Predicate<br>Priority | Product | 510(k) Number |
|---------------------------|-----------------------|-------------------------------------------------------------------|---------------|
| Jeisys<br>Medical<br>Inc. | Primary Predicate | INTRAcel Premium<br>Fractional<br>RF Micro needle (FRM)<br>System | K153727 |
{4}------------------------------------------------
#### IV. DEVICE DESCRIPTION
The INTRAcel RF Microneedle System is indicated for use in dermatological and general surgical procedures. INTRAcel RF Microneedle System comprises of two primary components; an RF Generator with User Interface Software and a disposable Electrode Insertion Device with an integrated cable.
RF Generator: RF energy is delivered from the RF Generator through the electrodes into the target tissue. Treatment is provided in bipolar mode and monopolar mode. In the monopolar mode, RF energy flows from the main unit and a patient loop is formed by pairing the active electrode (microneedle electrode) with the neutral electrode pad. Heat is not generated in a neutral electrode plate which has a low contact resistance but heat is generated in an active electrode which has a relatively large contact resistance and this heat momentarily causes tissue to heat up to cause coagulation. The bipolar RF energy is delivered between independent adjacent electrode pairs. The RF Generator is not disposable.
Electrode Insertion Device: The insertion device is a hand-held mechanical device that is used to insert the electrodes into the target tissue. It is supplied sterile, is for single patient use only and cannot be resterilized. The device has forty-nine electrodes. The device has a mechanism to deploy and retract the electrodes when actuated by the operator.
Cable: There is a cable with a connector that connects to the RF Generator. The RF Generator is the energy source for the system.
Accessories: Cables, neutral electrode pad, neutral electrode pad cable, footswitch, and power cord are also supplied with the system.
#### V. INDICATIONS FOR USE
INTRAcel RF Microneedle System is indicated for use in dermatological and general surgical procedures for electrocoagulation, hemostasis and percutaneous fractional treatment of the skin in monopolar and bipolar mode.
# TECHNOLOGICAL CHARACTERISTICS W.R.T. THE PREDICATE DEVICE VI.
The Indications for Use, key technological characteristics of the Electrosurgical Unit (ESU) and operating principle of the Subject Device are equivalent to those of the Predicate Device.
{5}------------------------------------------------
| Comparison with Primary Predicate | | | | |
|--------------------------------------------|-----------------------------|------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Parameter | | Subject Device | Predicate Device | Comments |
| Device Name | | INTRAcel RF Microneedle System | INTRAcel Premium Fractional RF Micro needle (FRM) system | |
| Electrosurgical Unit(ESU) | Monopolar or Bipolar | Monopolar and Bipolar | Bipolar | Different. Addition of Monopolar mode. |
| | Output frequency | 1MHz | 1MHz | Same as predicate |
| | Power Output | 50W | 50W | Same as predicate |
| | Voltage Output | 90V | 90V | Same as predicate |
| | Dimensions | 350mm(W)x<br>405mm(D)x<br>1080mm(H) | 350mm(W)x<br>x400mm(D<br>)x1080mm(<br>H) | Same as predicate |
| | Weight | 63Kg | 63Kg | Same as predicate |
| | Power Source/Input | AC 120V,<br>50-60Hz | AC 120V,<br>50/60Hz | Same as predicate |
| Active<br>accessory<br>(RF<br>Electrode) | Monopolar or<br>Bipolar | Monopolar and<br>Bipolar | Bipolar | Different. Addition<br>of Monopolar Mode |
| | Electrode<br>Type | Micro Needle | Micro<br>Needle | Same as predicate |
| | Physical<br>Dimensions | Depth of skin<br>Ablation:<br>0.5/0.8/2.0mm<br>Thickness:0.25<br>mm (Tip of the<br>needle:22") | Depth of<br>skin<br>ablation:0.5<br>/0.8/2.<br>0mm<br>Thickness:<br>0.25mm | Same as predicate |
| | Materials | Tip:ABS(sr-030<br>0)Needles:<br>SUS304 | Tip:ABS(sr<br>-0300)<br>Needles:S<br>US304 | Same as predicate |
| | Single-Use or<br>Reusable | Single-Use | Single-Use | Same as predicate |
| | Sterilization | EO gas | EO gas | Same as predicate |
| Neutral<br>electrode pad | Conductive or<br>Capacitive | Conductive<br>area:<br>250mm<br>x140mm<br>contact on the<br>skin | Not a<br>feature of<br>this device | Different. Addition<br>in the subject device<br>for the monopolar<br>mode. |
| | Single-Use or<br>Reusable | Single-Use | | |
| | Physical<br>Specification | 250mm(W)<br>x 140mm(L)<br>x0.5mm(H) | | |
| | Materials | ABS High<br>density<br>Polyethene<br>(material that is<br>patient<br>contacting) | | |
| Miscellaneous<br>accessory<br>(Footswitch) | Functions | For emitting RF<br>energy into<br>electrode | For<br>emitting<br>RF<br>energy into<br>electrode | Same as predicate |
| | Performance | Single pole,<br>single<br>throw | Single<br>pole, single<br>throw | Same as predicate |
| | Physical<br>Specification | Single pedal,<br>IPX8 | Single<br>pedal,<br>IPX8 | Same as predicate |
| | Materials | Steel and plastic | Steel and<br>plastic | Same as predicate |
# Comparison with Primary Predicate
{6}------------------------------------------------
{7}------------------------------------------------
The following differences exist between the Subject Device and Predicate Device:
- The Subject Device operates in both monopolar and bipolar mode but the primary predicate operates only in bipolar mode.
- The Neutral Electrode Pad is an addition to the Subject Device.
## VII. NON-CLINICAL STUDY
- 1.1. Non-Clinical Performance Data
The device's hardware and software development, verification, and validation have been carried out in accordance with FDA guidelines. The software was tested against the established Software Design Specifications for each of the test plans to assure that the device performs as intended. The Device Hazard Analysis was completed and risk control was
{8}------------------------------------------------
implemented to mitigate identified hazards. The test results demonstrate that all the acceptance criteria of each module and interaction of processes have been met. The Subject Device passed all testing which supports substantial equivalence and safe operation. The INTRAcel RF Microneedle System complies with the applicable voluntary standards for Electromagnetic Compatibility and Safety. The device passed all the electrical and safety testing according to national and international standards.
| Testing Type | Test Description | Test Result |
|-------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrical Safety<br>and<br>Electromagnetic<br>Compatibility<br>Testing | • IEC 60601-1:2012<br>• IEC 60601-1-6:2010,A<br>MD1:2013<br>• IEC 60601-2-2:2009 | The INTRAcel RF<br>Microneedle<br>System met all<br>acceptance criteria in<br>accordance with IEC<br>60601-1:2012,<br>IEC 60601-1-6:2010,<br>AMDA1:2013<br>and IEC<br>60601-2-2:2009. |
| Performance<br>Testing | • Performance Test<br>Report for ESU<br>• INTRAcel Tip<br>Count Test<br>• Verification<br>Performance of<br>INTRAcel Tip | The INTRAcel RF<br>Microneedle has<br>met the acceptance<br>criteria. |
| Biocompatibility<br>Testing | • Cytotoxicity<br>• Sensitization<br>• Intracutaneous<br>Reactivity | The microneedle<br>electrodes are<br>biocompatible. |
| Sterilization and<br>Shelf Life Testing | • Ethylene Oxide<br>Sterilization | The INTRAcel RF<br>Microneedle has |
| | Residuals Accelerated aging test | met the acceptance criteria. |
Tests Performed
{9}------------------------------------------------
#### VIII. SOFTWARE VERIFICATION AND VALIDATION TESTING
Software Verification and Validation Testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as "moderate" level of concern, since a failure or latent flaw in the software would not directly or indirectly result in serious injury or death to the patient or operator.
#### ANIMAL PERFORMANCE TESTING IX.
A study was conducted to see the wound healing response following FRM treatment for both human and porcine skin. Healing responses were observed by the time after Fractional Radiofrequency Microneedling (FRM) treatment at various energy levels. Biopsy was conducted to see the wound healing process immediately after the treatment, 2days, 14days, 28days, and 10 weeks post the treatment. H&E stain and HSP47 stain were conducted to see the changes in inflammatory cell, collagen. Also, the study has conducted RT-PCR (Reverse Transcription - Polymerase Chain Reaction) with the tissue biopsied from Micro-pig covering 10 weeks to see mRNA change of collagen, Heat Shock Proteins (HSPs), and matrix metalloproteinase (MMPs).
#### CONCLUSION X.
The INTRAcel RF Microneedle System is substantially equivalent to the Predicate Device in Indications for Use, key technological characteristics of the Electrosurgical Unit (ESU) and operating principle. Safety and performance testing was performed and Jeisys Medical Inc. has concluded that the Subject Device does not raise any new questions of safety and efficacy and is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.