K181852 · Teleflexmedical, Inc. · GEI · Oct 12, 2018 · General, Plastic Surgery
Device Facts
Record ID
K181852
Device Name
Percuvance Percutaneous Surgical System
Applicant
Teleflexmedical, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 12, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Percutaneous Surgical System is a platform consisting of a percutaneous surgical set and accessories used as a means to penetrate soft tissue to access certain areas of the abdomen. The devices' Tool Tips or attachments are provided separately from the percutaneous Shaft and are introduced to the site via a traditional conduit such as a trocar once they are attached extracorporeally to the Shaft tip. A Seal Bridge is used in conjunction with the trocar to maintain insufflation during extracorporeally exchange. Once inside the abdomen, the surgical set is used to grasp, manipulate, cut, cauterize, and deliver Hem-o-lok® ligating clips to soft tissues during laparoscopic surgery.
Device Story
Micro-laparoscopic platform; consists of reusable handles (ratcheted/non-ratcheted) and interchangeable tool tips (scissors, graspers, dissectors, cautery, clip applier) attached to a shaft. Surgeon operates device during laparoscopic surgery. Initial access via introducer tip; tool tips exchanged extracorporeally through a trocar using a seal bridge to maintain insufflation. Monopolar energy used for cautery/cutting. Clip applier delivers Hem-o-lok ligating clips. System facilitates minimally invasive surgical manipulation and hemostasis. Benefits include reduced trocar site trauma via micro-laparoscopic access.
Clinical Evidence
Bench testing only. Includes electrosurgical safety testing (IEC 60601-1), biocompatibility (ISO 10993-1), cytotoxicity, and shelf-life stability testing. No clinical data presented.
Technological Characteristics
Monopolar electrosurgical system; rated accessory voltage 1000VPeak. Components: reusable handles, single-patient-use shafts (29cm/36cm), and tool tips. Materials evaluated per ISO 10993-1. Sterilization: Gamma irradiation (sterile components) and steam sterilization (reusable handles). Compliant with IEC 60601-1:2005 (A1:2012).
Indications for Use
Indicated for penetrating soft tissue to access abdominal areas; used to grasp, manipulate, cut, cauterize, and deliver Hem-o-lok ligating clips during laparoscopic surgery. Contraindicated for contraceptive coagulation of fallopian tissue (though may be used for hemostasis post-transection), use when endoscopic techniques are contraindicated, and specific Hem-o-lok clip contraindications (fallopian tubal occlusion, renal artery ligation during donor nephrectomies).
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration" in a stacked format.
October 12, 2018
Teleflex Medical, Inc. Kim Campbell Regulatory Affairs Specialist 3015 Carrington Mill Blvd Morrisville, North Carolina 27560
Re: K181852
Trade/Device Name: Percuvance Percutaneous Surgical System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI, GCJ, GDO Dated: September 19, 2018 Received: September 21, 2018
Dear Kim Campbell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn
(http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long H. Chen -S 2018.10.12 10:32:25 -04'00'
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K181852
Device Name
Percuvance(TM) Percutaneous Surgical System
Indications for Use (Describe)
The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver Hem-o-lok® ligating clips to soft tissue during laparoscopic surgery.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
## Percuvance™ Percutaneous Surgical System
## A. Name, Address, Phone, and Fax Number of Applicant
Teleflex Medical, Incorporated 3015 Carrington Mill Boulevard Morrisville, NC 27560 USA Phone:919-433-4918 Fax: 919-433-4996
## B. Contact Person
Kim Campbell Regulatory Affairs Specialist
## C. Date Prepared
September 19th, 2018
## D. Device Name
## Trade Name
Percuvance™ Percutaneous Surgical System
### Common Name
| Primary: | Electrosurgical, Cutting and Coagulation and Accessories |
|------------|---------------------------------------------------------------------|
| Secondary: | Laparoscope, General and Plastic Surgery<br>Applier, Surgical, Clip |
### Classification Regulation
| Primary: | 21 CFR 878.4400 |
|------------|------------------------------------|
| Secondary: | 21 CFR 876.1500<br>21 CFR 878.4800 |
## Product Code
| Primary: | GEI |
|------------|----------|
| Secondary: | GCJ, GDC |
{4}------------------------------------------------
## Classification Name
| Primary: | Electrosurgical Cutting and Coagulation Device and Accessories |
|------------|-------------------------------------------------------------------------|
| Secondary: | Endoscope and Accessories<br>Manual Surgical Instrument for General Use |
## Classification
Class II
## Panel
General & Plastic Surgery
#### E. Device Description
Teleflex Medical's Percuvance™ Percutaneous Surgical System is a micro-laparoscopic platform that comprises fourteen (14) unique components, which are combined in various configurations to create a multifunctional set of instruments for laparoscopic procedures. In accordance with IEC 60601-1:2005 (A1:2012), the Percuvance™ system is classified as a BF active accessory with a rated accessory voltage of 1000VPeak-
System components include two reusable Handles (Ratcheted and Non-Ratcheted), which are manipulated by the surgeon and connect to a Shaft, which is available in two lengths (29 cm and 36 cm), affords various Tool Tips (or End Effectors) to be attached in order to perform basic surgical manipulations. The Percuvance™ Handles, Shafts, and Tool Tips are not compatible or interchangeable with components from other percutaneous systems.
Initial access to the surqical site is achieved with the Introducer Tool Tip attached to the Shaft. Once inside the patient, the Introducer Tool Tip is extracorporealized through a pre-inserted, central trocar and is then exchanged for one of the other Tool Tips, which include Scissors, Gripper Grasper, Johans Grasper, Maryland Dissector, Hook Cautery, Spatula Cautery and Clip Applier. No ligating clips are provided with the Percuyance™ Percutaneous Surgical System: however, the Clip Applier Tool Tip is compatible with Teleflex M/L Hem-o-lok® ligating clip (SKU 544230).
Finally, the Seal Bridge, which is available in two sizes (5 mm and 12 mm), is used to protect the trocar seal and to aid in maintaining insufflation when it used in conjunction with a trocar to facilitate the extracorporeal exchange of Tool Tips.
| Model Number | Component<br>Description | Reusability<br>Status | Sterility Status | Intended to<br>Apply Monopolar<br>Energy? |
|--------------|--------------------------|------------------------|----------------------------------------|-------------------------------------------|
| PCVINT3 | Introducer Tool Tip | Single-Patient-<br>Use | Provided Sterile -<br>Gamma Irradiated | No |
| PCVSC5 | Scissors Tool Tip | Single-Patient-<br>Use | Provided Sterile -<br>Gamma Irradiated | Yes |
{5}------------------------------------------------
| Model Number | Component<br>Description | Reusability<br>Status | Sterility Status | Intended to<br>Apply Monopolar<br>Energy? |
|--------------|---------------------------------------|------------------------|--------------------------------------------------------------------------|--------------------------------------------------------------------|
| PCVGG5 | Traumatic Gripper<br>Grasper Tool Tip | Single-Patient-<br>Use | Provided Sterile -<br>Gamma Irradiated | Yes |
| PCVJG5 | Atraumatic Johans<br>Grasper Tool Tip | Single-Patient-<br>Use | Provided Sterile -<br>Gamma Irradiated | Yes |
| PCVMD5 | Maryland Dissector<br>Tool Tip | Single-Patient-<br>Use | Provided Sterile -<br>Gamma Irradiated | Yes |
| PCVHK5 | Hook Cautery Tool Tip | Single-Patient-<br>Use | Provided Sterile -<br>Gamma Irradiated | Yes |
| PCVSPT5 | Spatula Cautery Tool<br>Tip | Single-Patient-<br>Use | Provided Sterile -<br>Gamma Irradiated | Yes |
| PCVHCA5 | Clip Applier Tool Tip | Single-Patient-<br>Use | Provided Sterile -<br>Gamma Irradiated | No |
| PCVSH3 | 29 cm Shaft | Single-Patient-<br>Use | Provided Sterile -<br>Gamma Irradiated | Yes |
| PCVSHL3 | 36 cm Shaft | Single-Patient-<br>Use | Provided Sterile -<br>Gamma Irradiated | Yes |
| PCVNRH | Non-Ratcheted Handle | Reusable | Provided Non-Sterile -<br>Intended to be Steam<br>Sterilized by Hospital | Yes |
| PCVRH | Ratcheted Handle | Reusable | Provided Non-Sterile -<br>Intended to be Steam<br>Sterilized by Hospital | Yes |
| PCVSB5 | 5 mm Seal Bridge | Single-Patient-<br>Use | Provided Sterile -<br>Gamma Irradiated | N/A - Not<br>connected to the<br>Handle and Shaft<br>configuration |
| PCVSB12 | 12 mm Seal Bridge | Single-Patient-<br>Use | Provided Sterile -<br>Gamma Irradiated | N/A – Not<br>connected to the<br>Handle and Shaft<br>configuration |
#### Indications for Use F.
The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver Hem-o-lok® ligating clips to soft tissue during laparoscopic surgery.
#### Contraindications G.
The following are contraindications of the Percuvance™ Percutaneous Surgical System:
- . The monopolar active Tool Tips are not intended for contraceptive coagulation of fallopian tissue, but may be used to achieve hemostasis following transection of the fallopian tube.
{6}------------------------------------------------
- Do not resterilize single patient use components. These components are provided sterile and are intended for use in a single procedure. Discard after use.
- The device is not intended for use when endoscopic techniques are generally contraindicated.
The following are contraindications of the M/L Hem-o-lok® ligating clip (SKU 544230), which is an accessory to the Percuvance™ system:
- Hem-o-lok® ligating clips are not intended for use as a fallopian contraceptive tubal occlusion device.
- . Hem-o-lok® ligating clips are contraindicated for use in ligating the renal artery during laparoscopic donor nephrectomies.
#### Substantial Equivalence H.
The Percuvance™ Percutaneous Surgical System is substantially equivalent to the following predicate system.
| Predicate Device | Manufacturer | 510(k) No. | Date Cleared |
|----------------------------------------------|------------------------|------------|-----------------|
| Percuvance ™ Percutaneous<br>Surgical System | Teleflex Medical, Inc. | K153063 | April 8th, 2016 |
Additionally, the technology of Teleflex Medical's Percuvance™ Clip Applier Tool Tip is substantially equivalent to that of the M/L Hem-o-lok® manual, endoscopic clip appliers cleared in 510(k) K133202. Like the M/L Hem-o-lok® manual, endoscopic clip appliers, the Percuvance™ Clip Applier Tool Tip is compatible with Teleflex Medical's M/L Hem-o-lok® ligating clip (SKU 544230), which was also cleared in 510(k) K133202.
| Reference (Accessory) Device | Manufacturer | 510(k) No. | Date Cleared |
|---------------------------------|------------------------|------------|------------------|
| Hem-o-lok® Ligating Clip System | Teleflex Medical, Inc. | K133202 | December 3, 2013 |
#### l. Comparison to Predicate Device
The proposed Percuvance™ Percutaneous Surgical System is substantially equivalent to the predicate system with respect to technology, intended use, indications for use and functional characteristics. The proposed modifications to Percuvance™ Percutaneous System does not introduce any new issues of safety and effectiveness.
## J. Materials
All patient contacting materials are equivalent to the predicate Percuvance™ Percutaneous Surgical System and have been evaluated in accordance with ISO 10993-1:2009 (C2010) and FDA Guidance: Use of International Standard ISO 10993 (published June 16, 2016), according to their nature and duration of contact.
{7}------------------------------------------------
# K. Technological Characteristics
A comparison of the technological characteristics of the proposed Percuvance™ Percutaneous Surgical System and the predicate system has been performed. The results of this comparison demonstrate that the proposed system utilizes substantially equivalent technology as the predicate system.
| Comparative<br>Characteristics | Predicate Device (K153063) | Proposed Device<br>Percuvance™ |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| | Percuvance™ Percutaneous Surgical System | Percutaneous<br>Surgical System |
| Indications for Use | The Percutaneous Surgical System with 5mm<br>attachments is indicated for the means to<br>penetrate soft tissue to access certain areas of the<br>abdomen. The system is used to grasp,<br>manipulate, cut, cauterize and deliver Hem-o-lok®<br>ligating clips to soft tissue during laparoscopic<br>surgery. | Equivalent |
| Intended Use | The Percutaneous Surgical System is a platform<br>consisting of a percutaneous surgical set and<br>accessories used as a means to penetrate soft<br>tissue to access certain areas of the abdomen.<br>The devices' Tool Tips or attachments are<br>provided separately from the percutaneous Shaft<br>and are introduced to the site via a traditional<br>conduit such as a trocar once they are attached<br>extracorporeally to the Shaft tip. A Seal Bridge is<br>used in conjunction with the trocar to maintain<br>insufflation during extracorporeally exchange.<br>Once inside the abdomen, the surgical set is used<br>to grasp, manipulate, cut, cauterize, and deliver<br>Hem-o-lok® ligating clips to soft tissues during<br>laparoscopic surgery. | Equivalent |
| Contraindications | The monopolar active Tool Tips are not intended<br>for contraceptive coagulation of fallopian tissue,<br>but may be used to achieve hemostasis following<br>transection of the fallopian tube. (system<br>contraindication)<br>Do not resterilize single patient use components.<br>These components are provided sterile and are<br>intended for use in a single procedure. Discard<br>after use. (system contraindication)<br>The device is not intended for use when<br>endoscopic techniques are generally<br>contraindicated. (system contraindication) | Equivalent |
| Comparative<br>Characteristics | Predicate Device (K153063)<br>Percuvance™ Percutaneous Surgical System | Proposed Device<br>Percuvance™<br>Percutaneous<br>Surgical System |
| | Hem-o-lok® ligating clips are not intended for use<br>as a fallopian contraceptive tubal occlusion device.<br>(accessory ligating clip contraindication) | |
| | Hem-o-lok® ligating clips are contraindicated for<br>use in ligating the renal artery during laparoscopic<br>donor nephrectomies. (accessory ligating clip<br>contraindication) | |
| Intended Procedure | Laparoscopic | Equivalent |
| Reusability Status | Handles are reusable; All other components are<br>single patient use and disposable | Equivalent |
| Sterility Status | Handles are provided non-sterile and are intended<br>to be sterilized by end user prior to use; All other<br>components are provided sterile | Equivalent |
| Sterilization Method | Non-sterile (reusable) components to be steam<br>sterilized by end user; Sterile components<br>sterilized by gamma irradiation using method<br>VDmax | Equivalent |
| Energy | Monopolar | Equivalent |
| Rated Accessory<br>Voltage | 1000 VPeak | Equivalent |
| Connector Type | Handle Post | Equivalent |
| Electrosurgical<br>Testing | CB Scheme (in compliance with IEC 60601-1<br>series 2nd and 3.1 Editions) | Equivalent |
| Handle | Yes | Equivalent |
| Shaft | Yes (2.9 mm) | Equivalent |
| Introducer Tip | Yes | Equivalent |
| Tool Tips | Yes (5 mm) | Equivalent |
| Graspers | Yes | Equivalent |
| Maryland<br>Dissectors | Yes | Equivalent |
| Scissors | Yes | Equivalent |
| Cautery Tips | Yes (Spatula and Hook) | Equivalent |
| Clip Applier | Yes | Equivalent |
| Seal Bridges | Yes | Equivalent |
| Exchange<br>Technology | Exchanged extracorporeally | Equivalent |
{8}------------------------------------------------
{9}------------------------------------------------
## L. Performance Data
Comprehensive bench testing was completed on the modifications identified below to ensure the system performed equivalently (or better) to the predicate system. Existing design, usability validations and packaging (ship) testing, performed on the predicate device still adequately covers the proposed device. Stability testing was repeated on the sterile components and their packaging to support the legacy extended 3 year shelf life according to aging methods and test methods previously submitted in the predicate 510(k) in support of the initial 1 year shelf life. Though the majority of the electrosurgical safety testing submitted for the system is still valid, certain electrosurgical safety tests were repeated at a third party test house to support the Shaft metal lock tube change.
Biocompatibility data submitted for the Percuvance system still adequately represents the proposed system, as no new patient contacting materials were introduced. However, as a result of Shaft's green ink band modification, which is pad printed on the Shaft's proximal knob, a full biocompatibility panel was completed. Additionally, cytotoxicity testing was performed on the Shaft metal lock tube change. The additional biocompatibility and cytotoxicity tests were performed as a conservative measure, as both modifications impacted parts of the device that are not intended to have any patient contact.
## M. Conclusion
Based upon the testing presented throughout the submission and in this 510(k) Summary, Teleflex Medical's Percuvance™ Percutaneous Surgical System is substantially equivalent in to the predicate device cleared to market via 510(k) K153063. The modifications to the proposed Percuvance™ Percutaneous Surgical System do not introduce any new issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.