K180872 · VIOL Co., Ltd. · GEI · Oct 11, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180872
Device Name
SCARLET SRF
Applicant
VIOL Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 11, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SCARLET SRF is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
Device Story
SCARLET SRF is a radiofrequency (RF) electrosurgical system consisting of a main unit, handpiece with consumable tip, and footswitch. The device generates bipolar RF energy delivered to target tissue via penetrating needle electrodes in the consumable tip. The tip is held at right angles to the skin; RF energy creates an electro-thermal reaction for tissue coagulation and hemostasis. Used in clinical settings by healthcare professionals. The device is substantially equivalent to the INFINI Radiofrequency System (K121481).
Clinical Evidence
Preclinical animal study conducted on three Yucatan Mini-pigs to evaluate safety and effectiveness for dermal treatment and hemostasis. Histopathology and dermal Draize scores were monitored over 22 days. Results showed transient erythema resolving by day 3, with expected minimal necrosis and subsequent dermal fibrosis. Bench testing included sterilization validation (ISO 11135-1), shelf-life (ASTM F1980-07), biocompatibility (ISO 10993), and electrical safety/EMC (IEC 60601 series).
Technological Characteristics
Bipolar RF electrosurgical system. Consumable tips feature gold-plated surgical stainless steel needle electrodes. Depth range 0.5-3.5 mm. Spot size 10x10 mm. EO gas sterilization. Connectivity via footswitch/finger switch. Software is firmware-based. Complies with IEC 60601-1, IEC 60601-2-2, and EN 60601-1-2 standards.
Indications for Use
Indicated for dermatologic and general surgical procedures requiring electrocoagulation and hemostasis. Prescription use only.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
#### October 11, 2018
VIOl Co., Ltd. % Mina Joo Assistant Manager BT Solutions, Inc. 91-14 Seolleung-ro #402 Gangnam-gu Seoul, Korea 06150
Re: K180872
Trade/Device Name: SCARLET SRF Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: September 11, 2018 Received: September 13, 2018
Dear Mina Joo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Lo
20
-C
Long H. Chen -S 2018.10.11 07:58:57 -04'00'
for
Binita S. Ashar. M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K180872
Device Name SCARLET SRF
Indications for Use (Describe)
The SCARLET SRF is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of*
of conduct or sponsor, and a person is not required to respons
information unless it displays a currently valid OMB number." તિ
{3}------------------------------------------------
# 5. 510(k) Summary
# 1. General Information
| Applicant/Submitter: | VIOL Co., Ltd. |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Address: | C-808, 809, Bundang Technopark C,<br>744, Pangyo-ro, Bundang-gu, Seongnam-si,<br>Gyeonggi-do (13510), South Korea |
| | Tel) +82-31-8017-7893<br>Fax) +82-31-8016-7894 |
| Contact Person: | Mina Joo, BT Solutions, Inc. |
| Address: | 91-14 Seolleung-ro, #402<br>Gangnam-gu Seoul, 06150<br>Republic of Korea<br>Tel) +82.2.538.9140<br>Email) smanager@btsolutions.co.kr |
| Preparation Date: | September-5-2018 |
## 2. Device Name and Code
| Device Trade Name: | SCARLET SRF |
|----------------------|------------------------------------------------------|
| Common Name: | Radiofrequency System |
| Classification Name: | Electrosurgical, Cutting & Coagulation & Accessories |
| Product Code: | GEI |
| Regulation Number: | 878.4400 |
| Classification: | Class II |
| Review Panel: | General & Plastic Surgery |
# 3. Predicate Devices
Scarlet SRF is substantially equivalent to the following devices
| | Table 5.1 Predicate devices | |
|--|-----------------------------|--|
|--|-----------------------------|--|
| Applicant | Device Name | 510(k) Number |
|----------------------|------------------------------|---------------|
| Lutronic Corporation | INFINI Radiofrequency System | K121481 |
{4}------------------------------------------------
### Scarlet SRF 510(k) Summary
Table 5.2 Reference devices *
| Applicant | Device Name | 510(k) Number |
|-------------------------|------------------------|---------------|
| Sung Hwan E&B Co., Ltd. | VIVACE Electrosurgical | K150409 |
*This reference device is referred, of which similarity of having smaller number of non-insulated needles when compare to the same predicate device.
#### 4. Device Description
SCARLET SRF is basically an identical device with the 510(k)-cleared device CELFIRM™ (K172023).
The SCARLET SRF includes a system main device, a hand-piece equipped with a tip, and a footswitch. The RF signal is generated from the main device which is then delivered to the handpiece and then to consumable tip. The RF signal is delivered to the target tissue using penetrating needle electrodes in the tip. The tip is placed in light contact with the epidermis while the handpiece is being held at right angles to the target tissue. As the RF signal passes through the skin, it generates an electro thermal reaction which is capable of coagulating the tissue. Using the consumable tip, SCARLET SRF creates heat within the target skin tissue via needle electrodes from the consumable tip.
#### 5. Indications / Intended Use
The SCARLET SRF is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
#### 6. Technical Characteristics in Comparison to Predicate Devices
| | Predicate Device | Proposed Device |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) Number | K121481 | N/A |
| Manufacturer | Lutronic Corporation | VIOL Co., Ltd. |
| Device Name | INFINI Radiofrequency System | SCARLET SRF |
| Clearance Date: | Jun 25,2013 | N/A |
| Classification /<br>Regulation | Class II/21 CFR 878.4400<br>Electrosurgical, Cutting & Coagulation<br>& Accessories | Class II/21 CFR 878.4400<br>Electrosurgical, Cutting & Coagulation<br>& Accessories |
| Product Code | GEI | GEI |
| Intended Use/<br>Indications for<br>Use | The INFINI Radiofrequency System is<br>intended for use in dermatologic and<br>general surgical procedures for<br>electrocoagulation and hemostasis, and<br>the percutaneous treatment of facial<br>wrinkles. | The SCARLET SRF is intended for use<br>in dermatologic and general surgical<br>procedures for electrocoagulation and<br>hemostasis. |
| Intended for | Prescription Use | Prescription Use |
| RF Output Type | BiPolar | BiPolar |
{5}------------------------------------------------
#### Scarlet SRF 510(k) Summary
| Electrode<br>exposure<br>(Depth) | | |
|----------------------------------|----------------------------------------|----------------------------------------|
| Output Control | Foot switch | Foot / Finger switch |
| Electrode<br>Material | Gold plated (Surgical stainless steel) | Gold plated (Surgical stainless steel) |
| No. of usage | Single use | Single use |
| Sterilization | EO gas sterilization | EO gas sterilization |
| Spot size | $10 \times 10$ (mm) | $10 \times 10$ (mm) |
| | 0.5-3.5 mm (Unit: 0.1) | 0.5-3.5 mm (Unit: 0.25) |
## 7. Performance Data
SCALRET SRF is basically an identical device with the 510(k)-cleared device CELFIRM™.
Sterilization Validation, Shelf life Validation, Biocompatibility, Software Validation, EMC & Electrical Safety Test, Max Output Energy, and Animal Test were conducted on the subject device.
#### Sterilization Validation Test
To verify the sterility assurance level (10-6) for EtO sterilization and biological indicator (BI) overkill method was used in accordance to ISO 11135-1, ISO 10993-1, ISO 10993-7, ISO 11138-1, ISO 11138-2, ISO 11737-1, ISO 11737-2, AAMI TIR 15, and ISO 13485:2012.
#### Shelf life Validation Test
The tests to validate the shelf life of the device through the proposed shelf life were conducted using the accelerated aging method in accordance to ASTM F1980-07 (2011) Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. The referenced standards for the testing are ISO 11607-1, ISO 11607-2, ISO 11737-2, ASTM F1929, ASTM F 88, and USPNF.
#### Biocompatibility Test
The patient contacting component of the subject device is the Consumable Tip and the biocompatibility tests were in accordance with ISO 10993-5. ISO 10993-7. ISO 10993-10, ISO 10993-11, and USP 39 <151>.
#### Software Validation Test
The level of concern of the software (firmware) of the SCARLET SRF is moderate and the software validation tests were performed.
## EMC & Electrical Safety Test
{6}------------------------------------------------
#### Scarlet SRF
#### 510(k) Summary
The subject device has been evaluated for electromagnetic compatibility and electrical safety testing per applicable standards of IEC 60601-1:2005, IEC 6061-1-6:2010, IEC 60601-2-2:2009, EN 60601-1-2:2007, EN 6100-3-2:2006, and EN 6100-3-3:2008, and all results were passing.
#### Max Output Energy
The test on Max Output Energy per Electrode was performed on the subject device to evaluate the max energy per electrode. The test results were compared and evaluated with the standard values of maximum output energy per Electrode of the device. The test results met the pre-set criteria and validated the device performance specifications for max energy.
#### Preclinical Test
The study was designed to evaluate the safety and effectiveness of the subject device as a radiofrequency(RF) heating device to treat dermal conditions and hemostasis. Three (3) Yucatan Mini-pigs were used in this study. Animal 2162 was euthanized 3 h after treatment on Day 1;
Animal 2163 was euthanized on Day 4; and Animals 2164 was euthanized on Day 22. The various treatment sites were then harvested and stored in a 10% NBF prior to histopathology evaluation. During the in-life stage, mortality, clinical observations and body weights were monitored and recorded. In addition, dermal Draize scores were recorded within 3-4 h of treatment, and on Days 3, 5, 13, and 22 in surviving animals.
Treatment with the subject device resulted in transient very slight to well-defined erythema noted within 3-4 h of treatment but fully resolved by the next Draize evaluation on Day 3. In addition, minimal to mild necrosis was noted in animals euthanized on Days 1 and 5 with the animal euthanized on Day 22 showing dermal fibrosis at the treatment sites. There were no other application-related findings.
Test results including histology data demonstrated that the subject device is substantially equivalent to the predicate devices in the market for its intended use.
#### 8. Substantial Equivalence
The intended use of the SCARLET SRF is within the scope of the predicate device. SCARLET SRF, from both a design and clinical perspective, uses similar or identical technology as the cited predicate device and has the same intended uses. Based upon the predicted overall performance characteristics for the SCARLET SRF, VIOL Co., Ltd., believes that no significant differences exist in usage of its underlying technological principles between SCARLET SRF and the cited predicate device.
#### 9. Conclusions
Based on the information provided in this premarket notification. VIOL Co., Ltd. concludes that the SCARLET SRF is substantially equivalent to the predicate device as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.