K180372 · Rebound Therapeutics · GEI · Apr 9, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180372
Device Name
Aurora Evacuator
Applicant
Rebound Therapeutics
Product Code
GEI · General, Plastic Surgery
Decision Date
Apr 9, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AURORA Evacuator is a powered instrument with a handpiece intended for removal of soft tissue and fluids under direct visualization. Types of direct visualization may include laparoscopic, pelviscopic, endoscopic, percutaneous, and open. Applications include those when access to the surgical site is limited, such as Neurosurgical/Spinal and ENT/Otolaryngological.
Device Story
AURORA Evacuator is a sterile, single-use, handheld aspiration device for minimally invasive surgery. It consists of a stainless-steel suction wand with a distal aspiration window and an internal rotating whisk-shaped shaft, connected to a plastic handle containing a battery-powered motor and suction tubing. The device connects to an operating room vacuum source. The surgeon controls vacuum strength via a vent on the handle and activates the motor via a button to rotate the whisk. Soft tissue is drawn into the window by vacuum, cut by the rotating whisk, and transported through the wand and tubing to a waste receptacle. Used in OR settings by surgeons to remove soft tissue and fluids in limited-access sites. Benefits include efficient tissue removal in confined surgical fields.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1, IEC 60529-1), electromagnetic compatibility (IEC 60601-1-2), particulate testing (USP 36 <788>), sterilization validation (ISO 11135-1), packaging/shelf-life (ISTA 2A, ASTM F1980), and product performance validation using surrogate soft tissue and fluid materials.
Technological Characteristics
Materials: stainless-steel wand, plastic handle. Energy: battery-powered DC motor. Connectivity: connects to external OR vacuum source. Dimensions: wand lengths of 7cm, 10cm, 13cm. Sterilization: validated to SAL 10^-6. Standards: ISO 10993-1, IEC 60601-1, IEC 60529-1, IEC 60601-1-2, USP 36 <788>, ISO 11135-1, ISTA 2A, ASTM F1980.
Indications for Use
Indicated for removal of soft tissue and fluids during surgical procedures (laparoscopic, endoscopic, percutaneous, open) where access is limited, including neurosurgical, spinal, and ENT/otolaryngological applications.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 9, 2018
Rebound Therapeutics Ms. Jane Metcalf Vice President of Regulatory Affairs 13900 Alton Parkway, Suite 120 Irvine, California 92618
Re: K180372
Trade/Device Name: Aurora Evacuator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: February 7, 2018 Received: February 12, 2018
Dear Ms. Metcalf:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180372
Device Name AURORA Evacuator
Indications for Use (Describe)
The AURORA Evacuator is a powered instrument with a handpiece intended for removal of soft tissue and fluids under direct visualization. Types of direct visualization may include laparoscopic, endoscopic, percutaneous, and open. Applications include those when access to the surgical site is limited, such as Neurosurgical/Spinal and ENT/ Otolaryngological.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 1) SUBMITTER
Rebound Therapeutics Corporation 13900 Alton Parkway, STE 120 Irvine, CA 92618
Contact Person: Jeffrey J. Valko, President and CEO Telephone: (949) 633-2747 Email: jvalko@reboundtx.com
Contact Person: Jane Metcalf, Vice President of Regulatory Affairs Telephone: (949) 525-1493 Email: jmetcalf@reboundtx.com
Date Prepared: April 3, 2018
### 2) DEVICE
Name of Device: AURORA Evacuator Common or Usual Name: Aspirator Classification Name: Electrosurgical, Cutting & Coagulation & Accessories Regulatory Class: II Product Code: GEI
#### 3) PREDICATE DEVICE
Primary Predicate: NICO Myriad, 11GA, K161307 Secondary Predicate: ConMed Corporation, Frazier 8Fr, Class II, 510(k) Exempt
### 4) DEVICE DESCRIPTION
The AURORA Evacuator is provided sterile, for single-use only. It is a disposable, handheld, aspiration device designed to remove fluids and soft tissue during minimally invasive surgical procedures. The AURORA Evacuator consists of a stainless-steel suction wand connected to a plastic handle.
The stainless-steel suction wand is configured with an aspiration window, near its distal end, for fluid and soft tissue to enter. Internal to the suction wand is a shaft with a whisk-shaped configuration at its distal end. The whisk-shaped configuration is aligned with the aspiration window. The proximal end of the shaft is connected to a battery powered motor in the handle.
The interior of the plastic handle contains the battery powered motor, batteries, electrical circuitry and the suction tubing. The suction tubing connects the stainless-steel suction wand to operating room vacuum. On the exterior of the handle is a tear-dropped shaped vent for user control of vacuum strength, a green light that illuminates when power is available, and a button for user activation of the motor.
{4}------------------------------------------------
The AURORA Evacuator is shipped in a non-functional state that requires removal of a pull tab from the handle to initiate power to the device.
There are three (3) configurations of the AURORA Evacuator; AURORA Evacuator -13, AURORA Evacuator -10 and AURORA Evacuator -7. All configurations are fabricated from the same materials and electrical components. The only difference among the configurations is the working length of the wand component. The wand working lengths are 13cm, 10cm, and 7cm, respectively.
### 5) STATEMENT OF INTENDED USE
The AURORA Evacuator is a powered instrument with a handpiece intended for removal of soft tissue and fluids under direct visualization. Types of direct visualization may include laparoscopic, pelviscopic, endoscopic, percutaneous, and open. Applications include those when access to the surgical site is limited, such as Neurosurgical/Spinal and ENT/Otolaryngological.
### 6) SUMMARY OF TESTING
The following testing was conducted to demonstrate the safe and effective use of the AURORA Evacuator and its' substantial equivalence to the primary predicate.
- Biocompatibility Testing per ISO 10993-1 Cytotoxicity (MEM Elution), Sensitization . (Kligman Maximization), Irritation (Intracutaneous Injection), Systemic Toxicity (Systemic Injection, Material Mediated Pyrogenicity)
- Electrical Safety and Enclosure Protection per IEC 60601-1 and IEC 60529-1 ●
- Emissions and Immunity per IEC 60601-1-2
- Particulate Testing per USP 36 <788>:
- Sterilization per ISO 11135-1 to validate a SAL of 10-6
- Packaging and Shelf-life per ISTA 2A and ASTM F1980
- . Verification of product specifications including, materials, multiple bonds evaluations, physical characteristics, audible characteristics, rotational speeds, duration of use and operational temperatures
- Validation of product performance using surrogate soft tissue materials and surrogate fluid materials
# 7) SUMMARY TECHNOLOGICAL CHARACTERISTICS
The AURORA Evacuator has two (2) modes of operation: one (1) for aspiration and another for cutting. For aspiration, the AURORA Evacuator connects to an external vacuum source that is present in the operating room. For cutting, the AURORA Evacuator utilizes a rotating whisk that is internal to the wand.
To aspirate, the surgeon covers the suction vent on the device's handle and positions the aspiration window so that vacuum draws fluid and soft tissue through the window and into the wand. The aspirant travels through the wand, into the suction tubing and is deposited into the waste receptacle connected to the operating room vacuum-line.
To cut soft tissue, the surgeon pushes the motor activation on the exterior of the handle causing the whisk to rotate. Vacuum is used to draw soft tissue through the aspiration window and into the rotating whisk. The soft tissue is cut by the rotating whisk and transported to a collection
{5}------------------------------------------------
receptacle connected to the operating room vacuum-line. Soft tissue cutting can only occur inside of the wand.
# 8) COMPARISON TO PREDICATE
Comparing the AURORA Evacuator to the primary predicate shows that there are three main technological differences; vacuum source, energy source, and tissue cutting. The primary predicate uses an internally controlled vacuum source instead of an external vacuum source. Its energy source is from a wall outlet rather than direct current from internal batteries. Finally, the primary predicate cuts both soft tissue and solid tissue using a scissoring action. The AURORA Evacuator only cuts soft tissue and uses a rotating whisk rather than scissoring. All three of these differences have been analyzed and tested. Results show no new or different questions of safety and effectiveness.
### 9) CONCLUSION
Upon reviewing the performance data provided in this submission and comparing indicated use, design, materials, principle of operation and overall technological characteristics, the AURORA Evacuator has been determined, by Rebound Therapeutics Corporation, to be substantially equivalent to the primary predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.