K170610 · Medtronic Advanced Energy · GEI · Apr 13, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170610
Device Name
PlasmaBlade T
Applicant
Medtronic Advanced Energy
Product Code
GEI · General, Plastic Surgery
Decision Date
Apr 13, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PlasmaBlade T is a monopolar, single use, sterile, disposable device intended for use with the AEX Generator. The device delivers RF energy concurrent with saling and coagulation of soft tissue and bone and RF energy for cutting and coagulation of soft tissue. It is intended for, but not limited to. General, Plastic and Reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, Gynecologic, Orthopaedic, Arthroscopic, Spinal, Thoracic, and Open abdominal surgery procedures. The PlasmaBlade T should not be used on small appendages or body parts, as in circumcision. The device is not intended for contraceptive tubal coagulation (permanent female sterilization).
Device Story
PlasmaBlade T is a sterile, single-use monopolar electrosurgical handpiece used with the AEX Generator. It features an enamel-coated insulated blade electrode, an uncoated circular electrode, and a telescoping shaft. The device connects to the AEX Generator for RF power and to a saline source via integrated pump header tubing. During surgery, the device delivers RF energy for cutting and coagulation of soft tissue, or RF energy combined with saline (Transcollation Technology) for hemostatic sealing of soft tissue and bone. Operated by surgeons in various surgical settings (OR), the device provides tactile control via three integrated handle buttons. The output allows for precise tissue resection and hemostasis, potentially reducing thermal damage compared to traditional electrosurgery. The device is disposable, eliminating the need for reprocessing.
Clinical Evidence
No clinical data. Evidence consists of bench testing (sterilization, transit, shelf life, biocompatibility, electrical safety/EMC, mechanical, activation) and in-vivo animal testing. Animal study confirmed thermal damage zones were comparable to predicate devices, meeting criteria from the FDA guidance for electrosurgical devices.
Technological Characteristics
Monopolar electrosurgical handpiece; enamel-coated insulated blade electrode with uncoated circular electrode; telescoping shaft; integrated controls; co-extruded cable for RF power and saline delivery. Uses RF energy from AEX Generator. Single-use, sterile, disposable. No software algorithm details provided; device functions as an accessory to the AEX Generator.
Indications for Use
Indicated for soft tissue and bone hemostatic sealing/coagulation and soft tissue cutting/coagulation in general, plastic, reconstructive, ENT, gynecologic, orthopaedic, arthroscopic, spinal, thoracic, and open abdominal surgeries. Contraindicated for small appendages (e.g., circumcision) and contraceptive tubal coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 13, 2017
Medtronic Advanced Energy Lydia Sakakeeny, Ph.D. Principal Regulatory Affairs Specialist 180 International Drive Portsmouth, New Hampshire 03801
Re: K170610 Trade/Device Name: PlasmaBlade T Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: February 28, 2017 Received: March 1, 2017
Dear Dr. Sakakeeny:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170610
Device Name PlasmaBlade T
#### Indications for Use (Describe)
The PlasmaBlade T is a monopolar, single use, sterile, disposable device intended for use with the AEX Generator. The device delivers RF energy concurrent with saling and coagulation of soft tissue and bone and RF energy for cutting and coagulation of soft tissue. It is intended for, but not limited to. General, Plastic and Reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, Gynecologic, Orthopaedic, Arthroscopic, Spinal, Thoracic, and Open abdominal surgery procedures.
The PlasmaBlade T should not be used on small appendages or body parts, as in circumcision. The device is not intended for contraceptive tubal coagulation (permanent female sterilization).
Type of Use (Select one or both, as applicable)> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Submitter: | Medtronic Advanced Energy<br>180 International Drive<br>Portsmouth, NH 03801 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Lydia Sakakeeny, PhD<br>Principal Regulatory Affairs Specialist<br>Phone: (603) 294-5482<br>Fax: (603) 742-1488<br>E-mail: lydia.sakakeeny@medtronic.com |
| Date Summary Prepared: | February 28, 2017 |
| Device Trade Name: | PlasmaBlade T |
| Common Name: | Electrosurgical Instrument |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories<br>(21 CFR 878.4400) |
| Product Code: | GEI |
| Predicate Device: | PlasmaBlade T and AEX Generator (K143175)<br>Aquamantys SBS 5.0 (K111732, K132974)<br>PlasmaBlade 3.0S (K093695) |
| Device Description: | The proposed single-use disposable accessory device, the PlasmaBlade<br>T monopolar handpiece, provides the hemostatic capabilities of the<br>Aquamantys SBS 5.0 Handpiece (Cleared under K111732, K132974)<br>and the cutting and coagulative capabilities of the PEAK PlasmaBlade<br>3.0S (Previously cleared under K093695). |
| | The proposed PlasmaBlade T Handpiece device is a monopolar, single<br>use, disposable device and is provided sterile. The devices are not<br>intended for reuse or resterilization. The PlasmaBlade T handpiece<br>consists of an enamel coated insulated blade electrode with an uncoated<br>circular electrode, an insulated telescoping shaft, handle with three<br>integrated controls, and a co-extruded cable assembly to provide both<br>power and saline. The seven-pin electrical connector is designed to be<br>plugged into the AEx Generator (Cleared under K143175). |
| | The proposed PlasmaBlade T Handpiece is a disposable device, that<br>when connected to the AEX Generator (K143175), uses monopolar RF<br>energy for the resection and coagulation of soft tissue and bone. The<br>proposed PlasmaBlade T Handpiece device provides similar resection<br>and coagulative effect to the predicate device, the PlasmaBlade 3.0S<br>Monopolar Handpiece (Cleared under K093695). |
| | In addition to the monopolar cutting and coagulation capabilities the<br>proposed PlasmaBlade T Handpiece device also uses monopolar RF<br>energy concurrent with saline delivery to provide a broader coagulative<br>effect; similar to the predicate Aquamantys SBS 5.0 Handpiece (Cleared<br>under K111732, K132974), this result is trademarked as Transcollation®<br>Technology by Medtronic. The proposed PlasmaBlade T Handpiece |
{4}------------------------------------------------
device also has a pump header tubing segment and saline bag spike tubing allowing the user to manually connect the device to saline source in the OR as well as the AEX Generator's peristaltic pump. This ensures the device is connected for saline delivery at the same time as the device is connected electrically to the proposed AEX Generator.
Indications for use: The PlasmaBlade T is a monopolar, single use, sterile, disposable device intended for use with the AEX Generator (K143175). The device delivers RF energy concurrent with saline for hemostatic sealing and coagulation of soft tissue and bone and RF energy for cutting and coagulation of soft tissue. It is intended for, but not limited to, General, Plastic and Reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, Gynecologic, Orthopaedic, Arthroscopic, Spinal, Thoracic, and Open abdominal surgery procedures.
> The PlasmaBlade T should not be used on small appendages or body parts, as in circumcision. The device is not intended for contraceptive tubal coagulation (permanent female sterilization).
### Technological Characteristics:
The predicates and PlasmaBlade T are intended for RF based resection and coagulation of soft tissue, as well as RF based hemostatic sealing concurrent with saline delivery for hemostatic sealing and coagulation of soft tissue and bone during various surgical procedures. The proposed devices share the same operational characteristics as the predicate platforms, comprised of a radio-frequency generator which supplies RF power to disposable electrode devices for electrosurgical procedures.
The proposed single-use disposable accessory device, the PlasmaBlade T monopolar handpiece, provides the hemostatic capabilities of the Aquamantys SBS 5.0 Handpiece (Cleared under K111732, K132974) and the cutting and coagulative capabilities of the PEAK PlasmaBlade 3.0S (Previously cleared under K093695).
#### Summary of Non-Clinical Testing:
The design and performance of the modified PlasmaBlade T were verified and validated through bench testing. Medtronic product development processes and the finalized FDA Guidance Document, Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery Guidance for Industry and Food and Drug Administration Staff issued on August 15, 2016, were utilized to determine applicable bench testing requirements.
Bench testing included the following:
- . Sterilization
- Transit ●
- Shelf Life ●
- . Biocompatibility
- Electrical Safety/EMC
- Mechanical Testing
- Activation Testing
{5}------------------------------------------------
| | In-vivo testing on an animal model was completed to confirm that the proposed device performed adequately to its predicates for its intended use. The testing met the criteria outlined in the FDA Guidance Document, "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery Guidance for Industry and Food and Drug Administration Staff" issued on August 15, 2016. The in-vivo testing determined the zone of thermal damage imparted on the tissue by the proposed devices was comparable to the predicates.<br>The bench and animal testing demonstrated that the performance and tissue effect of the proposed devices are substantially equivalent to that of the predicate devices and do not raise any concerns of device safety or efficacy. |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of Clinical Tests: | Clinical testing was not required for this product. |
| Conclusion: | The indications for use, technology and performance characteristics of the modified PlasmaBlade T are equivalent to the predicate devices' and therefore Medtronic Advanced Energy claims substantial equivalence to the predicate devices. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.