The Aquamantys™ Minimally Invasive Sealer (MIS) FLEX and Aquamantys™ Minimally Invasive Sealer (MIS) Flex Mini are sterile, single-use bipolar electrosurgical devices intended to be used in conjunction with a qualified Pump Generator for delivery of Radio-frequency (RF) energy and saline for hemostatic sealing and coagulation of soft tissue and bone at the operative site. They are intended for, but not limited to orthopaedic, neurosurgical, spine, endoscopic procedures, abdominal and thoracic surgery, and epidural vein sealing during surgery. These devices are not intended for contraceptive tubal coagulation (permanent female sterilization).
Device Story
The Aquamantys MIS FLEX and MIS FLEX Mini are single-use, bipolar electrosurgical handpieces designed for use in surgical settings (OR). They utilize Transcollation Technology, which delivers RF energy and saline irrigation simultaneously to achieve hemostatic sealing and coagulation of soft tissue and bone. The device features a malleable shaft, dual electrode tips, and dual saline delivery ports. It connects to a qualified Medtronic Pump Generator (AEX or Aquamantys) via a three-pin electrical connector. A handpiece button activates both RF energy and saline flow. Surgeons use the device to control bleeding during procedures; the saline irrigation helps manage thermal effects on tissue. The device provides a surgical tool for improved hemostasis, potentially reducing blood loss and improving surgical site visibility.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including biocompatibility (ISO 10993), electrical safety (IEC 60601-1, IEC 60601-2-2), mechanical testing, and performance testing. Performance was validated via a cadaver study for anatomical access and comparative in-vivo porcine and ex-vivo tissue studies against the predicate device to assess thermal effects.
Technological Characteristics
Sterile, single-use, bipolar electrosurgical device. Features malleable shaft, dual electrode tips, and dual saline delivery ports. Energy source: RF energy via external Pump Generator. Standards: ISO 10993 (biocompatibility), IEC 60601-1 (general safety), IEC 60601-2-2 (HF surgery equipment), ISO 11737-1, ISO 11135 (sterilization).
Indications for Use
Indicated for hemostatic sealing and coagulation of soft tissue and bone in patients undergoing orthopaedic, neurosurgical, spine, endoscopic, abdominal, or thoracic surgery, and epidural vein sealing. Contraindicated for contraceptive tubal coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{1}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 7, 2017
Medtronic Advanced Energy Heather Valley Principal Regulatory Affairs Specialist 180 International Drive Portsmouth, New Hampshire 03801
Re: K170381
Trade/Device Name: Aquamantys MIS FLEX and Aquamantys MIS FLEX Mini Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: August 18, 2017 Received: August 21, 2017
Dear Heather Valley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{2}------------------------------------------------
related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/9 description: The image shows the name "Jennifer R. Stevenson -S3" in a large, bold font. The text is arranged vertically, with "Jennifer R." on the top line and "Stevenson -S3" on the bottom line. The background is a light blue color, and the letters are black.
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170381
#### Device Name
Aquamantys(TM) Minimally Invasive Sealer (MIS) FLEX Aquamantys(TM) Minimally Invasive Sealer (MIS) FLEX Mini
#### Indications for Use (Describe)
The Aquamantys™ Minimally Invasive Sealer (MIS) FLEX and Aquamantys™ Minimally Invasive Sealer (MIS) Flex Mini are sterile, single-use bipolar electrosurgical devices intended to be used in conjunction with a qualified Pump Generator for delivery of Radio-frequency (RF) energy and saline for hemostatic sealing and coagulation of soft tissue and bone at the operative site. They are intended for, but not limited to orthopaedic, neurosurgical, spine, endoscopic procedures, abdominal and thoracic surgery, and epidural vein sealing during surgery. These devices are not intended for contraceptive tubal coagulation (permanent female sterilization).
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Medtronic logo. The logo consists of a circular symbol on the left and the word "Medtronic" in blue on the right. The circular symbol contains three figures in different poses, arranged in a circular pattern.
MIS FLEX
Traditional 510(k) Premarket Notification
# 510(k) Summary
| Submitter: | Medtronic Advanced Energy<br>180 International Drive<br>Portsmouth, NH 03801 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Heather Valley, RAC<br>Principal Regulatory Affairs Specialist<br>Phone: (603) 294-5447<br>Fax: (603) 742-1488<br>E-mail: Heather.a.valley@medtronic.com |
| Date Summary Prepared: | September 5, 2017 |
| Device Trade Name: | Aquamantys™ Minimally Invasive Sealer (MIS) FLEX and<br>Aquamantys™ Minimally Invasive Sealer (MIS) FLEX Mini |
| Common Name: | Electrosurgical Instrument |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories<br>(21 CFR 878.4400) |
| Regulatory Class: | Class II |
| Product Code: | GEI |
| Predicate Device: | Aquamantys SBS 5.0 (K132974) |
| Device Description: | MIS FLEX and MIS FLEX Mini are single-use, disposable, bipolar<br>devices. The MIS FLEX and MIS FLEX Mini employ radio-frequency<br>(RF) energy and saline irrigation (termed Transcollation Technology™)<br>for hemostatic sealing and coagulation. The devices are equipped with<br>dual electrode tips while saline and electrical lines exit the opposite end<br>of the handpiece from the electrodes. The handpiece is equipped with an<br>on-off button that simultaneously activates both RF and saline flow. A<br>saline fluid delivery line is provided with the device and includes a<br>section of pump tubing and drip chamber or spike. The three-pin<br>electrical connector is designed to be plugged into either of the FDA<br>cleared AEX (K143175) or Aquamantys (K052859) Pump Generators.<br>The MIS FLEX devices feature a fully malleable shaft, altered saline<br>delivery and dual ports for saline delivery (two for each electrode). The<br>distal end of the device features small, rounded electrode geometry.<br>The MIS FLEX and MIS FLEX Mini differ by the length of the shaft. The<br>MIS FLEX is designed for transsphenoidal and MIS Spine procedures<br>while the MIS FLEX Mini is designed for craniotomies with application in<br>posterial cervical and other spinal procedures. |
| Indications for Use: | The Aquamantys™ Minimally Invasive Sealer (MIS) FLEX and<br>Aquamantys™ Minimally Invasive Sealer (MIS) Flex Mini are sterile,<br>single-use bipolar electrosurgical devices intended to be used in |
{5}------------------------------------------------
K170381 p 2/4
Traditional 510(k) Premarket Notification
conjunction with a qualified Pump Generator for delivery of Radiofrequency (RF) energy and saline for hemostatic sealing and coaqulation of soft tissue and bone at the operative site. They are intended for, but not limited to orthopaedic, neurosurgical, spine, endoscopic procedures, abdominal and thoracic surgery, and epidural vein sealing during surgery. These devices are not intended for contraceptive tubal coagulation (permanent female sterilization).
### Technological Characteristics:
The MIS FLEX devices are similar to the predicate devices as they are sterile, single use, bipolar electrosurgical devices used to provide hemostatic sealing and coagulation of soft tissue and bone at the operative site.
The three major differences between the subject and predicate device are the malleable shaft, altered saline delivery and electrode size. These new features do not impact the safety or effectiveness of the device.
### Summary of Non-Clinical Testing:
The design and performance of the MIS FLEX devices were verified and validated through bench testing. The following performance data were provided in support of the substantial equivalence determination:
- Biocompatibility ●
- Sterilization ●
- Shelf Life
- Electrical Safety ●
- Electromagnetic Compatibility ●
- Mechanical Testing
- Performance Testing .
### Biocompatibility Testing
The Biocompatibility evaluation for MIS FLEX was conducted in accordance with the following International Standard ISO-10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process", as recognized by FDA. The testing includes the following:
- InVitro Cytotoxicity ●
- . Acute System Toxicity
- Irritation or Intracutaneous Reactivity ●
- Sensitization ●
- . Pyrogenicity
The MIS FLEX devices are considered externally communicating devices (Tissue/Bone/Dentin) with limited exposure of less than 24 hours.
Image /page/4/Picture/24 description: The image shows a logo. The logo has a blue winged figure on the left. To the right of the figure is the word "Mewburn" in blue.
MIS FLEX
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the Medtronic logo. The logo consists of a blue circle on the left side with a stylized human figure inside. To the right of the circle is the word "Medtronic" in a bold, blue font. The logo is simple and modern, and the use of blue gives it a sense of trustworthiness and reliability.
MIS FLEX
Traditional 510(k) Premarket Notification
## Electrical Safety Testing
Electrical safety and EMC testing were conducted on MIS FLEX devices and were found to be in compliance with associated IEC 60601-1 and IEC 60601-2-2 standards.
## Performance Testing
Two studies were completed in support of the MIS FLEX 510(k). One study was limited to validating anatomical access (Cadaver Study). The second study was a comparative performance test, which was conducted in an in-vivo porcine model. Ex-vivo tissue testing was also completed.
The access study was conducted to ensure that the MIS FELX will be able to access key anatomy and meet customer needs for neurosurgical and spine procedures. The devices met the requirements for access.
The comparative performance testing assessed the thermal effect of the MIS FLEX devices in both the in-vivo and ex-vivo models against the predicate in accordance with the "Premarket Notification (510(k) Submissions for Electrosurgical Devices for General Surgery Guidance for Industry and Food and Drug Administration Staff". The subject devices were substantially equivalent.
Testing of the MIS FLEX and MIS FLEX Mini were conducted in accordance with the following FDA recognized standards:
| Recognition<br>Number | Standard | Title of Standard |
|-----------------------|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 19-4 | ES60601-1:20 X05/(R)2012 And<br>A1:2012 | C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated<br>Text) Medical Electrical Equipment - Part 1: General<br>Requirements For Basic Safety And Essential<br>Performance (IEC 60601-1:2005, MOD) |
| 9-64 | IEC 60601-2-2Ed 5.0 | Medical Electrical Equipment - Part 2-2: Particular<br>Requirements For The Basic Safety And Essential<br>Performance Of High Frequency Surgery Equipment<br>And High Frequency Surgical Accessories |
| 14-407 | 11737-1 Second Edition 2006-<br>04-01 | Sterilization Of Medical Devices - Microbiological<br>Methods - Part 1: Determination Of A Population Of<br>Microorganisms On Products [Including: Technical<br>Corrigendum 1 (2007)] |
| 14-479 | 11135:2014 | Sterilization Of Health Care Products - Ethylene Oxide<br>- Requirements For Development, Validation And<br>Routine Control Of A Sterilization Process For Medical<br>Devices. |
| 2-220 | 10993-1 Fourth Edition 2009-10-<br>15 | Biological Evaluation Of Medical Devices - Part 1:<br>Evaluation And Testing Within A Risk Management<br>Process [Including: Technical Corrigendum 1 (2010)] |
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Medtronic logo. The logo consists of a circular graphic on the left and the word "Medtronic" in blue on the right. The graphic contains three figures in different positions inside of a circle.
K170381 p 4/4
MIS FLEX
Traditional 510(k) Premarket Notification
| 2-235 | 10993-4 Second Edition 2002-<br>10-15 AMENDMENT 1 | Biological Evaluation Of Medical Devices - Part 4:<br>Selection Of Tests For Interaction With Blood<br>[Including: Amendment 1 (2006)]. |
|-------|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| 2-153 | 10993-5:2009/(R)2014 | Biological Evaluation Of Medical Devices - Part 5:<br>Tests For In Vitro Cytotoxicity |
| 2-174 | 10993-10 Third Edition 2010-08-<br>01 | Biological Evaluation Of Medical Devices - Part 10:<br>Tests For Irritation And Skin Sensitization |
| 2-176 | 10993-11 Second Edition 2006-<br>08-15 | Biological Evaluation Of Medical Devices - Part 11:<br>Tests For Systemic Toxicity |
Summary of Clinical Tests: Clinical testing was not required for this product.
Conclusion: The indications for use, technology and performance characteristics of the MIS FLEX and MIS FLEX Mini are equivalent to that of the predicate device and therefore Medtronic Advanced Energy claims substantial equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.