K152570 · Bovie Medical Corporation · GEI · Nov 2, 2015 · General, Plastic Surgery
Device Facts
Record ID
K152570
Device Name
Bovie J-Plasma Precise 360 Handpiece
Applicant
Bovie Medical Corporation
Product Code
GEI · General, Plastic Surgery
Decision Date
Nov 2, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Bovie J-Plasma Precise 360 Handpiece is used for the delivery of helium gas plasma for cutting, coagulation, and ablation of soft tissue during open and laparascopic surgical procedures. The Bovie J-Plasma Precise 360 Handpiece is compatible only with Bovie J-Plasma generators.
Device Story
Sterile, single-use electrosurgical accessory; used with Bovie J-Plasma generators. Device channels helium gas over energized inner electrode to create plasma stream for soft tissue cutting, coagulation, and ablation. Features include retractable needle or blade tip, 30-degree angled shaft, and rotating knob for tip orientation; available in 33cm and 45cm lengths. Used in OR by surgeons during open or laparoscopic procedures (via 5mm trocar). Rotating angled tip improves visualization and positioning at surgical site. Benefits include precise tissue effect and enhanced access to target tissue.
Clinical Evidence
Bench testing only. Testing included mechanical/electrical verification, plasma stream characterization, performance evaluation on various tissue types, usability validation, shipping studies, shaft rotation effects, biocompatibility (ISO 10993), and shelf-life testing. All tests met performance requirements.
Indicated for cutting, coagulation, and ablation of soft tissue during open and laparoscopic surgical procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is the logo for the U.S. Department of Health and Human Services. The logo features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around a central emblem. The emblem consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare, with three human profiles incorporated into the design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 2, 2015
Bovie Medical Corporation Mr. Brian Kunst Vice President, Regulatory Affairs and Quality Assurance 5115 Ulmerton Road Clearwater, Florida 33760
Re: K152570
Trade/Device Name: Bovie J-Plasma Precise 360 Handpiece Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: September 4, 2015 Received: September 10, 2015
Dear Brian Kunst:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the
{1}------------------------------------------------
electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 . and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Joshua C. Nipper -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152570
Device Name Bovie J-Plasma Precise 360 Handpiece
#### Indications for Use (Describe)
The Bovie J-Plasma Precise 360 Handpiece is used for the delivery of helium gas plasma for cutting, coagulation, and ablation of soft tissue during open and laparascopic surgical procedures.
The Bovie J-Plasma Precise 360 Handpiece is compatible only with Bovie J-Plasma generators.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.2222 4.11108L5.77778 11.5555L2.77778 8.55548" stroke="#000000" stroke-linecap="round" stroke-linejoin="round" stroke-width="1.5"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for the brand "Bovie". The logo is written in a cursive font and is a light blue color. The logo is slanted upwards from left to right. There is a registered trademark symbol to the right of the word.
### 510(k) SUMMARY – K152570
### (Submitted As Required per 21 CFR 807.92)
### GENERAL INFORMATION:
| Submitter Name: | Bovie Medical Corporation |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration<br>Number: | 3007593903 |
| Submitter Address: | 5115 Ulmerton Road<br>Clearwater, Florida 33760-4004<br>United States of America |
| Submitter Telephone Number: | (727) 803-8617 |
| Submitter FAX Number: | (727) 322-4465 |
| Contact Person: | Brian Kunst<br>Vice President, Regulatory Affairs and Quality<br>Assurance<br>5115 Ulmerton Road<br>Clearwater, Florida 33760-4004<br>United States of America |
| Contact email: | brian.kunst@boviemed.com |
| Date Prepared: | October 30, 2015 |
| <b>DEVICE IDENTIFICATION:</b> | |
| Proprietary Name: | Bovie J-Plasma Precise 360 Handpiece |
| Common Name: | Electrosurgical Generator Accessory |
| Classification Name: | Electrosurgical Cutting and Coagulation Device and<br>Accessories |
| Model Numbers: | Multiple |
| Classification: | 21CFR 878.4400; Class II; Product Code GEI |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Bovie" in a stylized, cursive font. The word is rendered in a bright blue color. A horizontal line underlines the word, adding to the cursive effect. A registered trademark symbol is present to the right of the word.
Legally Marketed Predicate Device(s):
510(k) Number: K151325
Predicate Device Name: Bovie J-Plasma Handpiece Manufacturer: Bovie Medical Corporation
## INTENDED USE/INDICATIONS
The Bovie J-Plasma Precise 360 Handpiece is used for the delivery of helium gas plasma for cutting, coagulation, and ablation of soft tissue during open and laparascopic surgical procedures.
The Bovie J-Plasma Precise 360 Handpiece is compatible only with Bovie J-Plasma generators.
## DEVICE DESCRIPTION
The Bovie J-Plasma Precise 360 Handpiece is a sterile, single-use electrosurgical accessory intended to be used in conjunction with Bovie J-Plasma electrosurgical generators for the delivery of helium gas plasma for cutting, coagulating and ablating of soft tissue.
The Bovie J-Plasma Precise 360 Handpiece is available with a retractable needle tip, or with a retractable cutting blade tip for excising tissue. Both tip configurations serve as electrodes to generate helium plasma. A 30 degree angled shaft tip provides improved visibility during use, and a rotating knob is used to orient the angled shaft tip to the optimal position. The device is available with 33cm and 45cm shaft lengths. The device can be used in laparoscopic procedures using a standard 5mm trocar.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a medical device with a handle and a long, thin probe extending from it. The handle is white with blue accents, and the probe is black with a slight bend at the end. The device is likely used for minimally invasive surgical procedures, allowing doctors to access and treat areas inside the body without making large incisions. The brand name "Bovie" is visible in the upper left corner of the image.
The primary differences between the proposed J-Plasma Precise 360 Handpiece and the predicate J-Plasma Handpiece are:
- The J-Plasma Precise 360 Handpiece is available in 33cm and 45cm shaft lengths and the predicate J-Plasma Handpiece is available in 15cm, 33cm, and 45cm shaft lengths
- The J-Plasma Precise 360 Handpiece has a rotating knob that controls an angled shaft tip . which provides improved visibility and optimal positioning at the surgical site. The angled shaft tip uses a medical-grade which is not present in the predicate device. The rotating knob utilizes material which is present on the predicate device. The predicate J-Plasma Handpiece has a straight shaft tip.
The J-Plasma Precise 360 has the same intended use and energy source as the predicate device. There are no technological differences, no changes to the principle of operation or the method of application, and no changes to sterilization methods. Both devices require a Bovie J-Plasma electrosurgical generator and channel helium gas flow over an energized inner electrode to create a plasma stream.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Bovie" in a stylized, cursive font. The word is a bright, light blue color. The font is slanted slightly upwards from left to right. There is a registered trademark symbol to the right of the word.
## PERFORMANCE TESTING
Performance testing to assure that the Bovie J-Plasma Precise 360 meets performance requirements was performed and is summarized in the following table:
| Test | Objective | Reports |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------|--------------------|
| Mechanical Verification<br>and Functionality | Verify the mechanical functionality of the<br>device. | VR-1318<br>VR-1324 |
| Electrical Verification | Verify the electrical functionality and safety of<br>the device. | VR-1318 |
| Plasma Characteristics | Measure effect of generator settings on the<br>plasma stream, plasma stream characterization | VR-1324 |
| Performance Evaluation | Confirm device performance on<br>various tissue types. | VR-1324 |
| Usability Evaluation | To verify and validate the usability of the J-<br>Plasma Precise 360 as it relates to safety. | VP-1387 |
| Shipping Study | Verify the device meets requirements after<br>shipping and handling. | VP-1384 |
| Evaluation of Rotation<br>Effects | Evaluation of Shaft Rotation Effects on Plasma<br>stream for Bovie J-Plasma Precise 360<br>Handpieces | VR-1383 |
| Biocompatibility Testing | Verify material safety | VR-1335 |
| Shelf Life Testing | Confirm product and package performance at<br>labeled shelf life | VP-1326 |
The Bovie J-Plasma Precise 360 Handpiece, and the predicate Bovie J-Plasma Handpiece, were designed in accordance with the following standards:
| International Standard | Description |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC-60601-1, Edition 3.1 | Medical Electrical Equipment - Part 1: General Requirements<br>For Safety |
| IEC 60601-1-2:2007 | Medical electrical equipment - Part 1-2: General requirements<br>for basic safety and essential performance - Collateral Standard:<br>Electromagnetic disturbances - Requirements and tests |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Bovie" in a stylized, cursive font. The word is rendered in a light blue color. The "B" is capitalized and slightly larger than the other letters, with a curved flourish extending from the bottom of the "B" and underlining the rest of the word. A registered trademark symbol is located to the right of the "e".
| IEC-60601-2-2 : 2009 | Particular requirements for the safety of high frequency surgical<br>equipment |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ANSI / AAMI / ISO 10993-1:2009 | Biological Evaluation of Medical Devices - Part 1: Evaluation<br>and Testing |
| ANSI/AAMI/ISO 11135-1: 2007 | Sterilization of Healthcare Products - Ethylene Oxide - Part 1:<br>Requirements for the development, validation, and routine<br>control of a sterilization process for medical devices. |
| ANSI / AAMI / ISO 10993-7:2008 | Biological Evaluation of Medical Devices – Part 7: Ethylene<br>Oxide Residuals |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "Bovie" in a stylized, cursive font. The word is in a bright blue color, and there is a registered trademark symbol to the right of the word. The font is thick and has a flowing, handwritten appearance.
# 510(k) SUMMARY
# SUBSTANTIAL EQUIVALENCE
| Feature/<br>Characteristic | Bovie J-Plasma Precise 360<br>Current Submission | Bovie J-Plasma Handpiece<br>(Predicate K151325) |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications for Use | The Bovie J-Plasma Precise 360 is<br>used for the delivery of helium gas<br>plasma for cutting, coagulation, and<br>ablation of soft tissue during open and<br>laparascopic surgical procedures.<br><br>The Bovie J-Plasma Precise 360 is<br>compatible only with Bovie J-Plasma<br>generators. | The Bovie J-Plasma Handpiece is used<br>for the delivery of helium gas plasma for<br>cutting, coagulation, and ablation of soft<br>tissue during open and laparascopic<br>surgical procedures.<br><br>The Bovie J-Plasma Handpiece is<br>compatible only with Bovie J-Plasma<br>generators. |
| Energy Type | Helium gas plasma | Helium gas plasma |
| Output | Monopolar | Monopolar |
| User Interface | Pistol grip | Pistol grip |
| Shaft Design | Angled | Straight |
| Shaft Rotation | Yes | No |
| Shaft Working Lengths | 33cm and 45cm | 15cm, 33cm, 45cm |
| Shaft Configuration | Angled | Straight |
| Shaft Outer Diameter | 5mm | 5mm |
| Tip Configuration | Blade, Needle | Blade, Needle |
| Blade Extension | Maximum 10mm | Maximum 10mm |
| Blade Width x Thickness | 0.4mm x 0.08mm | 0.4mm x 0.08mm |
| Compatibility | Only with Bovie J-Plasma Generators | Only with Bovie J-Plasma Generators |
| Connector | Bovie Proprietary | Bovie Proprietary |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "Bovie" in a stylized, cursive font. The word is rendered in a bright, light blue color. A registered trademark symbol is located to the right of the word.
# CONCLUSION
There is no difference between the J-Plasma Precise 360 Handpiece and the legally marketed predicate device Bovie J-Plasma handpiece described in K151325 in terms of intended use, principle of operation, and the technology used for device performance. All features of the Bovie J-Plasma Precise 360 were subjected to verification testing. There is no new technology and no difference that would raise new or different questions of safety or efficacy. The angled shaft tip and rotating knob components are mechanical in nature and do not affect the technology of the device. The purpose of the angled tip and rotating shaft is to allow better positioning of the tip and better visualization of the target tissue. The devices exhibit substantially equivalent performance during bench testing, including equivalent depth of tissue effect.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.