K150993 · Nico Corporation · GEI · Apr 23, 2015 · General, Plastic Surgery
Device Facts
Record ID
K150993
Device Name
NICO TRIOwand
Applicant
Nico Corporation
Product Code
GEI · General, Plastic Surgery
Decision Date
Apr 23, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
Disposable device designed to be used in soft tissue surgical procedures that require slow to rapid fluid evacuation and low energy output for the coagulation of tissue.
Device Story
NICO TRIOwand is a surgeon-controlled, disposable catheter-type device for soft tissue surgery. It connects to third-party electrosurgical generators to deliver bipolar coagulation via two stationary insulated electrodes. The device features four lumens: two for electrode irrigation drip, one for aspiration, and one for surgical field irrigation. Aspiration is managed via a sliding valve on the handle. Used in clinical settings, the device allows surgeons to simultaneously coagulate tissue, evacuate fluids, and flush the surgical site. By combining these functions into a single instrument, it simplifies surgical workflows, potentially reducing procedure time and improving visibility in the surgical field.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including biocompatibility (ISO 10993-1), tensile testing, electrical safety (IEC 60601-1, IEC 60601-2-2), EMC (IEC 60601-1-2), ex-vivo animal study, longevity testing, usability/human factors analysis, design validation/simulated use, sterility validation, and shelf-life/packaging validation.
Technological Characteristics
Catheter-type device with 4 lumens. Materials: PEBAX, stainless steel (direct contact); Tygon, polycarbonate, ABS, PC/ABS, silicone, Dymax glue (indirect). Bipolar electrosurgical energy source. 4mm diameter, 14cm length, 30-degree pre-bent shaft. Sliding valve aspiration control. Terminally sterilized via Gamma radiation. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2.
Indications for Use
Indicated for soft tissue surgical procedures requiring fluid evacuation and low energy coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines suggesting hair or clothing. The profiles are arranged in a way that creates a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 23, 2015
NICO Corporation Mr. Jay Dittman Quality and Regulatory Director 250 East 96th Street, Suite 125 Indianapolis, Indiana 46240
Re: K150993
Trade/Device Name: NICO TRIOwand Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: April 8, 2015 Received: April 15, 2015
Dear Mr. Dittman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150993
Device Name NICO TRIOwand
Indications for Use (Describe)
Disposable device designed to be used in soft tissue surgical procedures that require slow to rapid fluid evacuation and low energy output for the coagulation of tissue.
Type of Use (*Select one or both, as applicable*)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5.0 510(K) SUMMARY
# NICO TRIOwand
21 CFR §807.92 Date Prepared: 06 March 2015
510(k) Number: _
| Submitter/Manufacturer | NICO Corporation<br>250 E. 96th Street, Suite 125<br>Indianapolis, IN 46240 |
|------------------------|----------------------------------------------------------------------------------------------|
| Primary Contact: | Jay Dittman<br>Quality and Regulatory Director<br>Office: 317.660.7118 |
| Trade Name | NICO TRIOwand |
| Common/Usual Name | Electrosurgical, cutting & coagulation & accessories |
| Classification | 21 CFR §878.4400 (Class II) |
| Product Code | GEI |
| Predicate Devices | Kirwan Bipolar Suction Coagulator; K960455<br>SilverGlide Non-Stick Bipolar Forceps; K992931 |
### Device Description
The NICO TRIOwand is a surgeon controlled device that enables the delivery of bipolar coagulation via third-party electrosurgical generators. Coagulation occurs between two stationary insulated electrodes which are continuously drip irrigated during use. Second, the TRIOwand provides user controlled delivery of aspiration/vacuum for the evacuation of fluids from the surgical site. Third, the TRIOwand provides the user the ability to deliver irrigation for flushing the surgical field.
This submission covers a single configuration of the device:
- . NICO TRIOwand - 4mm diameter, 14cm length, pre-bent (~30°)
The direct patient contacting materials consist of PEBAX® and stainless steel. The indirect contacting materials are Tygon®, polycarbonate, ABS, PC/ABS, silicone, and Dymax glue.
#### Intended Use
Disposable device designed to be used in soft tissue surgical procedures that require slow to rapid fluid evacuation and low energy output for the coagulation of tissue.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for NICO Corporation. The logo is in a sans-serif font, with the letters N, I, and C in black and the letter O in gold. Below the word NICO is the word CORPORATION in a smaller font, also in black. The logo is simple and modern.
# Comparison to Predicates
The NICO TRIOwand is substantially equivalent to the combination of Kirwan's Bipolar Suction Coagulator (K960455) and Stryker's SilverGlide Non-Stick Bipolar Forceps (K992931). The subject device and the Kirwan device are both used for soft tissue surgical procedures that require slow to rapid fluid evacuation and low energy output for the coagulation of tissue. The Stryker SilverGlide predicate is also intended for electrocautery, as well as grasping and manipulation of tissue. The TRIOwand is not designed for grasping or manipulation of tissue, and this additional feature of the Stryker predicate is not critical to the primary electrocautery function of subject device.
## Technological Characteristics
The following table compares the subject device and predicate devices.
| Feature | NICO TRIOwand | Kirwan Bipolar Suction<br>Coagulator | Stryker SilverGlide<br>Non-Stick Bipolar Forceps |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) # | TBD | K960455 | K992931 |
| Intended Use/<br>Indications | Disposable device designed to<br>be used in soft tissue surgical<br>procedures that require slow<br>to rapid fluid evacuation and<br>low energy output for the<br>coagulation of tissue. | <i>The Bipolar Suction<br/>Coagulator</i> is a disposable<br>device, designed to be used in<br>soft tissue surgical procedures<br>that require slow to rapid fluid<br>evacuation and low energy<br>output for the coagulation of<br>the tissue i.e. neurosurgery,<br>endoscopic, sinusoidal | Intended to facilitate grasping<br>and manipulation of soft<br>tissue and blood vessels and<br>provide electrocautery in<br>surgical procedures. |
| Principles of<br>Use | Catheter-type device which<br>includes four connections: 1)<br>bipolar coagulation 2)<br>aspiration 3) irrigation drip<br>for electrodes 4) irrigation<br>pathway for flushing the<br>surgical field. Aspiration<br>lumen used for removal of<br>tissues/fluids from the<br>surgical site. Irrigation drip<br>lumens constantly bath the<br>electrodes. Additional lumen<br>exists to irrigate the surgical<br>field. | Catheter-type device which<br>includes two connections: 1)<br>bipolar coagulation 2)<br>aspiration. Aspiration occurs<br>through center lumen of<br>device for the removal of<br>tissues/fluids from the<br>surgical site. Coagulation<br>occurs at/between two tubular<br>coaxial electrodes which are<br>located at the tip of the device<br>adjacent to the aspiration<br>lumen. | Forceps-type device which<br>includes two connections:<br>1) Bipolar coagulation<br>2) Irrigation drip for<br>electrodes. Irrigation drip<br>lumens constantly bathe<br>electrodes. |
| Fundamental<br>Technology | Delivery of bipolar<br>cauterization | Identical | Identical |
| Feature | NICO TRIOwand | Kirwan Bipolar Suction Coagulator | Stryker SilverGlide Non-Stick Bipolar Forceps |
| Design | Coagulation occurs<br>at/between two individual<br>wire electrodes which are<br>located at the tip of the device<br>adjacent to the aspiration<br>lumen.<br><br>Cavity (or annulus) around<br>each electrode throughout the<br>length of the device enables<br>the continuous flow of saline<br>(drip from bag).<br><br>Irrigation lumen exists for the<br>manual delivery of fluids such<br>as saline to the surgical field.<br><br>Sliding valve control over<br>aspiration port on handle | Coagulation occurs<br>at/between two tubular<br>coaxial electrodes which are<br>located at the tip of the device<br>adjacent to the aspiration<br>lumen.<br><br>No means for continuous flow<br>of saline at/around electrodes<br><br>No lumen or means of<br>delivery of fluids<br><br>No sliding valve, aspiration<br>controlled via extent of<br>fingertip coverage over<br>aspiration port | Coagulation occurs<br>at/between the tips of the<br>forceps as they are squeezed<br>within proximity of one<br>another.<br><br>Lumens within handle of<br>forceps enable the continuous<br>flow of saline (drip from bag).<br><br>No means of delivering<br>irrigation to clear the surgical<br>field<br><br>No aspiration capabilities |
| Materials | Stainless Steel, Pebax,<br>Polycarbonate | Unknown | Stainless Steel |
| Biocomp. | Externally Communicating<br>Device in Direct Contact with<br>Tissue/Bone/Dentin, Limited<br>Duration - Demonstrated | Identical | Identical |
| Cross<br>Sectional<br>Analysis | Total of 4 lumens: 2 for<br>electrodes/irrigation drip, 1<br>for aspiration, 1 for surgical<br>site irrigation | Total of 1 lumen for<br>aspiration | Forceps configuration with a<br>small lumen contained within<br>one of the forceps for delivery<br>of saline drip. |
| Shaft (dia.) | 4 mm | 5.5 mm | NA – no diameter (Forceps) |
| Lengths | 1) 14 cm (angled shaft) | 1) 14.2 cm (angled tip) | 1) 19cm<br>2) 22cm<br>3) 25cm |
| Config./<br>Components | Terminally sterilized, single-<br>use catheter-like device | Identical | Reprocessed, reusable<br>forceps-type device |
| Shipping | 5 per box | Identical | 1 per box |
| Usage | Single Patient Use | Identical | Reusable device |
| Sterilization | Gamma | Identical | Onsite steam sterilization |
| Handheld/<br>manual<br>operation and<br>placement | Yes | Identical | Identical |
| Placement | Via scope or unaided | Identical | Identical |
| Handle | ~14cm | ~7cm | No defined handle, forceps-<br>style device |
| Aspiration | Fingertip controlled w/ the aid<br>of a "slider" component<br>which moves over a slot | Fingertip controlled via slot<br>which is occluded to produce<br>aspiration | NA – no aspiration |
#### Table 1: Technological Characteristic Comparison
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for NICO Corporation. The letters "N," "I," and "C" are in black, and the "O" is in gold. The word "CORPORATION" is written in smaller, black letters below the letters "NICO."
## Table 1: Technological Characteristic Comparison (continued)
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for NICO Corporation. The logo consists of the word "NICO" in a serif font, with the "O" being a filled-in circle in a golden color. Below the word "NICO" is the word "CORPORATION" in a smaller, sans-serif font.
The TRIOwand has the same fundamental technology as the predicate devices – they all deliver bipolar energy for the purpose of coagulation of tissue. Like the Kirwan predicate, the TRIOwand enables the delivery of aspiration for evacuation of fluids. Like the secondary predicate, the TRIOwand enables the delivery of irrigation drip to the electrodes. The technological differences do not raise new questions of safety or effectiveness and where applicable the nonclinical testing provides adequate means to assess the effects of the combined characteristics of the subject device as compared to the predicates.
## Nonclinical Testing
The following tests were performed and demonstrate that the subject device met applicable design and performance requirements, and supports a determination of substantial equivalence:
- Biocompatibility per ISO 10993-1 .
- o Cytotoxicity. Sensitization, and Irritation
- Tensile Testing
- General Electrical Safety Testing per IEC 60601-1 ●
- EMC Testing per IEC 60601-1-2 ●
- High Frequency Surgical Equipment Testing per IEC 60601-2-2
- Ex-Vivo Animal Study for Comparison to Predicates
- Predicate Device Testing for Comparison to Predicates
- Longevity Testing
- Usability/Human Factors Analysis
- Design Validation/Simulated Use ●
- Sterility Validation
- . Shelf Life and Packaging Validation
## Conclusion
The subject device and the two predicate devices are equivalent in terms of intended use and technological considerations. Risk assessments, biocompatibility, electrical safety, bench testing. design validation, and compliance with recognized standards have demonstrated that any differences do not raise new questions of safety or effectiveness. Therefore, the conclusion drawn from the nonclinical activities demonstrate that the NICO TRIOwand is as safe, as effective, and performs as well as or better than the legally marketed predicate Kirwan and Stryker devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.