K150112 · H.S Hospital Service S.P.A · GEI · Sep 10, 2015 · General, Plastic Surgery
Device Facts
Record ID
K150112
Device Name
HS AMICA DEVICES FAMILY
Applicant
H.S Hospital Service S.P.A
Product Code
GEI · General, Plastic Surgery
Decision Date
Sep 10, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Coagulation (thermoablation) of soft tissue. Not for use in cardiac procedures.
Device Story
HS AMICA devices family is an integrated system for soft tissue thermoablation using microwave (MW) or radiofrequency (RF) energy. System comprises a generator and disposable sterile applicators (probes). Three generator variants: AGN-H (MW and RF), AGN (MW only), and AGN-R (RF only). Physician selects energy mode; MW and RF operate in mutually exclusive fashion. Probes (coaxial antenna for MW; monopolar interstitial electrode for RF) are inserted into tissue to deliver electromagnetic energy. Generator includes electronic control unit for user command interpretation, energy module control, and feedback via front panel indicators. System uses automated cooling. Used in clinical settings by physicians. Output is controlled electromagnetic radiation causing tissue thermoablation. Benefits include precise tissue coagulation for therapeutic purposes.
Clinical Evidence
Bench testing only. Performance evaluated by measuring ablation dimensions (depth, width, length) in ex-vivo tissues (muscle, liver, lung, kidney) across various power settings and probe gauges. Results demonstrated equivalent ablation dimensions compared to predicate devices K053290 and K083157. Electrical safety (AAMI ANSI ES60601-1, IEC 60601-2-2), EMC (AAMI ANSI IEC 60601-1-2), and biocompatibility (ISO 10993) testing confirmed compliance.
Technological Characteristics
System uses MW (2450 MHz) and RF (450 KHz) energy. Materials: PTFE (MW probe), AISI 304 stainless steel coated with PET (RF probe). Features automated cooling system. Power output up to 200 W. Applicators are sterile, single-use, ethylene oxide sterilized. Software developed per IEC 62304. Connectivity includes front panel controls and external footswitch.
Indications for Use
Indicated for coagulation (thermoablation) of soft tissue. Not for use in cardiac procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, representing the department's focus on health and human well-being. The profiles are depicted in a simple, abstract manner.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 10, 2015
HS HOSPITAL SERVICE SPA % Maurizio Pantaleoni ISEMED srl Via Altobelli Bonetti 3/A, Imola 40026 Bologna Italy
Re: K150112
Trade/Device Name: HS AMICA devices family Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: January 14, 2015 Received: January 14, 2015
Dear Pantaleoni:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part
{1}------------------------------------------------
807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150112
Device Name HS AMICA DEVICES FAMILY
Indications for Use (Describe) Coagulation (thermoablation) of soft tissue. Not for use in cardiac procedures.
Type of Use (Select one or both, as applicable)
|X | Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### HS AMICA devices family
## V. 510(k) Summary
## HS AMICA devices family
This 510(k) Summary is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.
#### 1. General Information Submitter:
HS Hospital Service SpA. Via A. Vacchi, 23/25 04011 Aprilia (LT) - Italy tel.: 0039 06 9201961 fax: 0039 0692727871
Establishment Registration Number: 810312
Contact Person in Italy:
Maurizio Pantaleoni ISEMED srl Via Altobelli Bonetti 3/a 40026 Imola (BO) Italy Mob.phone: +39-348 4435155 Telephone: +39-0542 683803 Fax: +39-0542 698456 Email: regulatory@isemed.eu
Summary Preparation Date:
#### July 28, 2015
GEI
878.4400
||
2. Device
HS AMICA devices family
Electrosurgical cutting and coagulation device and accessories
Classification name
Product Code:
Device Name:
Regulation number:
Class:
#### 3. Predicate Device
HS AMICA devices family is substantially equivalent to the following devices:
{4}------------------------------------------------
#### HS AMICA devices family
| Applicant | Device name | 510(k) Number |
|---------------------------------|--------------------|---------------|
| H.S. HOSPITAL<br>SERVICE S.P.A. | HS AMICA | K083157 |
| VALLEYLAB | COOL-TIP RF SYSTEM | K053290 |
The HS AMICA devices family has to be used for the coagulation) of soft tissue; not for use in cardiac procedures.
The thermoablation is obtained by microwave- and radiofrequency-based technologies. Especially, the HS AMICA devices family is equivalent to the predicate device K083157 concerning the microwave performance and to K053290 with regard to the radiofrequency thermoablation.
#### 4. Device Description
HS AMICA devices family is an integrated system for thermoablation of tissues through controlled emission of non-ionizing electromagnetic radiations in the microwave and radiofrequency ranges.
Especially the HS AMICA family consists of 3 devices and relative accessories emitting:
- 1. Only microwaves (MW) or
- 2. Only radiofrequency waves (RF) or
- 3. Either microwaves or radiofrequency waves (not simultaneously)
The three devices emit the specific wavelength through their applicators (probes) that are inserted into the human body and deliver electromagnetic energy through their emitting tip, causing the thermoablation. Such functions are identical to the predicate devices and especially to the HS AMICA (K083157) concerning the microwave ablation and to the Cool-tip system (K053290) concerning the radiofrequency ablation.
Briefly, the HS AMICA devices family consists of two main components: 1) the generator; 2} the applicator (probe).
The generator of the electromagnetic energy is available in the following variants, depending on the specific electromagnetic emission:
- 1. AMICA GEN AGN-H. This variant is characterized by both the MW and RF modules that must be selected by the physician. When the MW function is selected, the RF emission is disabled and viceversa and thus the modules operate in mutually exclusive fashion
- 2. AMICA GEN AGN. This variant has only the MW module.
- 3. AMICA GEN AGN-R This variant has only the RF module.
In all cases, the AMICA GEN is provided with an electronic control unit (EC), for data exchange with the patient applied parts, the interpretation of user's commands through front panel controls or external peripherals (such as the footswitch for remote energy start/stop), the continuous update of the machine indicators, and the control of the energy generation modules MW and/or RF.
in light of the overall set of input values and upon proper verification of possible alerts or allows:
{5}------------------------------------------------
#### HS AMICA devices family
- enabling/disabling of MW or RF energy delivery (for the AMICA GEN AGN-H) through an . appropriate signal for switching the amplifier A or Age on/off
- fixing the MW/RF output level by adjusting the gain control voltage .
- generating messages and feedback signals through the front end indicators featured by AMICA-� GEN (all the variants)
The applicator (probe) is disposable and consists of the following variants, to be used for the MW or RF thermoablation:
- 1. AMICA PROBE. This variant is a microwave coaxial antenna of asymmetric dipole type allowing the emission of the MW energy
- 2. RF AMICA-PROBE. This variant is used for radiofrequency energy delivery through monopolar interstitial electrode.
HS AMICA devices family consists of non-sterile reusable electronic parts, and sterile disposable applied parts, identically to the predicate device HS AMICA (K083157).
The disposable components are supplied in sterile conditions (possibly kit with sterile disposable accessories) and they can not be re-sterilized.
The sterile components are:
- 1. Single use kit containing AMICA PROBE with its accessories
- 2. Single use kit containing RF AMICA PROBE with its accessories :
The sterilization method used for all the sterile components is Ethylene Oxide.
The method is exactly the same used and validated for the predicate device cleared in K083157.
The part of the HS AMICA devices family coming in contact with patient are : AMICA-PROBE and RF AMICA-PROBE. According to Table 1 of Blue Book Memo, G95-1 "Required biocompatibility Training and toxicology Profiles for Evaluation of Medical Devices", the contact between the devices and the patient is limited under 24 hours, usually less than 30 minutes.
Material used to manufacture the parts of AMICA-PROBE and RF AMICA-PROBE coming in contact with patient are:
- For AMICA-PROBE:
- Polytetrafluoroethylene (PTFE)
Such material is exactly the same material used in the previously cleared device HS AMICA (K083157) and proved to be biocompatible.
- . For RF AMICA-PROBE: - Stainless steel (AISI 304) coated with Polyethylene terephthalate (PET)
Such material has been proved to be biocompatible according to ISO 10993.
{6}------------------------------------------------
#### HS AMICA devices family
Identically to the predicate device HS AMICA (K083157), the indication for use of HS AMICA devices family is: "Coagulation (thermoablation) of soft tissue. Not for use in cardiac procedures"
The Indications for use of the predicate device K053290 is not identical to the subject device it includes either the ablation of tissue (identically to the subject device) or the laparoscopic coagulation (which is not provided with the subject device). Therefore the differences do not alter the intended therapeutic use of the device nor do they affect the safety and effectiveness of the subject device relative to the predicate.
#### ర్. Comparison of technological characteristics with the predicate devices.
The thermoablation is the technological principle for both the predicate devices. It consists of the delivery of microwave or radiofrequency energy to the target tissue, by means of dedicated applicators (probes). The subject device consists of three variants allowing the emission of both the microwave and the radiofrequency (not simultaneously). Concerning the microwave thermoablation the subject device may be found equivalent to the predicate HS AMICA (K083157), while for the radiofrequency thermoablation the subject device may be found equivalent to the predicate Cool-tip system (K053290),
At high level the subject and the predicate devices are based on the following same technological elements:
- Design structure: both the subject and the predicate devices consist of a generator provided with a cooling system and an applicator (coaxial microwave applicator and radiofrequency applicator) connected the generator
- · Power output: the subject device has a power output up to 200 W identically to the predicate K053290
- · Cooling system both the subject and the predicate device (K083157) provide an automated cooling system
- Time of treatment: both the subject and the predicate device K053290 allow up to 30 minutes of treatment
- . Needle temperature: both the HS AMICA devices family and the predicate devices K083157 and K053290 allow the detection of the needle temperature within equivalent range of temperature.
- · Microwave emission frequency: the subject as well as the predicate device K083157 emit at 2450 MHz
- Microwave emission management: the management of the microwave power is manually regulated . for both the subject and the predicate device K083157
- . Radiofrequency emission frequency: the emission frequency of the subject device (450 KHz) is equivalent to the emission frequency of the predicate device K053290 (480 KHz ±10%)
- . Radiofrequency emission management: both the subject and the predicate device K053290 allow manual and automatic management of the radiofrequency power delivery.
- Operative impedance range: both the subject and the predicate device K053290 adopts the same . operative emission range to control power delivery
- . Microwave applicator: both the subject and the predicate device (K083157) have an applicator equipped with stainlees steel needles.
- Radiofrequency applicator tipologies: both the subject and the predicate device K053290 use single . and cluster applicators.
- · Condition of use of applicators: both the subject and the predicate devices K053290 and K083157 use sterile and disposable applicators.
{7}------------------------------------------------
#### HS AMICA devices family
The following technological differences exsist between the subject and the predicate devices:
- Radiofrequency emission management-algorithms: specific algorithms are used by the subject and the predicate device K053290 to control the maximum power deliverable at the target site. Such control is based on the detection and control of the impedance of the tissue, in both the devices. However, the subject device has an additional control: a specific algorithms allows to control the operative temperature which is selected by the physician by decreasing the RF power delivery for not exceeding the selected temperature. Such feature does not raise new safety nor effectiveness issues and better yet it satisfies higher safety standards.
- Radiofrequency applicator: both the subject and the predicate device K053290 use applicator stainless steel needles. However the dimensional characteristics of the subject device are slightly different relative to the predicate device.
However in no way such differences affect the effectiveness nor the safety of the subject device as demonstrated by a specific performance test (see below).
#### 7. Performance Data
The following performance data were provided in support of the substantial equivalence determination
#### Electrical safety �
The subject device was tested for the electrical safety demonstrating that it complies with AAMI ANSI ES60601-1:2005 (19-5) and IEC 60601-2-2 (6-228).
Electromagnetic compatibilities:
The subject device was tested according to AAMI ANSI IEC 60601-1-2 (19-2). The results demonstrate the compliance with the standard.
Software verification and validation
The HS AMICA medical devices family has been developed and validated according to Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices requirements FDA - May 2005 and according to IEC 62304:2006 (13-8).
- . Biocompatibility
The biocompatibility evaluation for the subject device was conducted in accordance with Blue Book Memo, G95-1 "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing and the International standard ISO 10993-1 (2-156).
- . Performance test-bench
Two specific tests were conducted to verify the performance of the subject device compared to the predicate devices K053290 (concerning the radiofrequency ablation) and K083157 (concerning the microwave ablation). The effectiveness of the performance was evaluated through the measurement of the size of the obtained ablation (depth, width and length) on different ex-vivo tissues (muscle, liver, lung and kidney) when considering minimum, maximum and default power setting and all the available gauges and needles.
The results demonstrate that the ablations obtained with HS AMICA have equivalent dimensions compared to the predicate devices K053290 (RF ablation) and K083157 (MW ablation) proving that the performance of the HS AMICA device is at least as good as the performance of the predicate devices.
{8}------------------------------------------------
#### HS AMICA devices family
#### 8. Conclusions
The performance data including in particular the electrical safety, EMC, software verification and validation support the safety of the HS AMICA devices family and demonstrate that the subject device performs as intended in the specific use conditions. Also, the considerations and bench tests discussed above underline the equivalence of the subject device to the predicate device K083157 with regards to the microwave ablation and to the predicate device K053290 with particular regard to the radiofrequency-based thermoablation.
Therefore, on the basis of evidence discussed above, the HS AMICA devices family may be found substantially equivalent to the predicate devices (K083157 and K053290).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.