K143654 · Covidien, LLC · GEI · Apr 3, 2015 · General, Plastic Surgery
Device Facts
Record ID
K143654
Device Name
Valleylab LS10 Generator
Applicant
Covidien, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
Apr 3, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Valleylab LS10, LS Series Single Channel Vessel Sealing Generator is an electrosurgical generator containing LigaSure vessel sealing technology. The vessel sealing function is indicated for use in sealing (fusing) vessels (including pulmonary) up to and including 7mm in diameter, tissue bundles, and lymphatics during general surgery including, but not limited to surgical specialties such as urologic, vascular, thoracic, gynecologic, plastic and reconstructive, and colorectal. Refer to each instrument's instructions for use (IFU) for additional indications, warnings, and specific contraindications. The LigaSure system has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use this function for these procedures
Device Story
Bipolar electrosurgical generator; uses LigaSure RF energy technology to ligate and divide vessels/tissue. Input: electrical resistance of tissue measured via compatible LigaSure instruments. Operation: microcontroller-based closed-loop control; monitors tissue properties; adjusts RF output to reform collagen/elastin; stops energy application upon seal completion. Output: permanent tissue seal; visual/audible status indicators. Used in OR/clinical settings by surgeons. Benefits: precise energy delivery; consistent vessel sealing; reduced thermal spread. Includes RFID for instrument identification and USB port for service/data exchange.
Bipolar electrosurgical generator; RF energy source. Microcontroller-based closed-loop control. Features: RFID for instrument identification, USB port for data/service, single LigaSure port, footswitch port. Dimensions smaller than predicate. Sterilization: N/A (generator). Software: IEC 62304 compliant.
Indications for Use
Indicated for sealing vessels up to 7mm, tissue bundles, and lymphatics in general surgery (urologic, vascular, thoracic, gynecologic, plastic/reconstructive, colorectal). Contraindicated for tubal coagulation for sterilization.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines suggesting hair or fabric. The profiles are arranged in a staggered formation, creating a sense of depth. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 3, 2015
Covidien Mr. Donald Henton Regulatory Affairs Manager 5920 Longbow Drive Boulder, Colorado 80301
Re: K143654
Trade/Device Name: Valleylab™ LS10, LS Series Single Channel Vessel Sealing Generator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: March 6, 2015 Received: March 9, 2015
Dear Mr. Henton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVIGES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K143654
Device Name
ValleylabTM LS10. LS Series Single Channel Vessel Sealing Generator (VLLS10GEN)
Indications for Use (Describe)
The Valleylab LS10, LS Series Single Channel Vessel Sealing Generator is an electrosurgical generator containing LigaSure vessel sealing technology.
The vessel sealing function is indicated for use in sealing (fusing) vessels (including pulmonary) up to and including 7mm in diameter, tissue bundles, and lymphatics during general surgery including, but not limited to surgical specialties such as urologic, vascular, thoracic, gynecologic, plastic and reconstructive, and colorectal.
Refer to each instrument's instructions for use (IFU) for additional indications, warnings, and specific contraindications.
The LigaSure system has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use this function for these procedures
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------|--|
| <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |
| <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for Covidien. The logo consists of a blue square with a white rectangle in the center and a light blue vertical line running through the center of the square. To the right of the square is the word "COVIDIEN" in blue, sans-serif font.
## 510(k) Summary
Date summary prepared: 12/12/2014
### 510(k) Submitter/Holder
Covidien 5920 Longbow Drive Boulder, CO 80301
#### Contact
Donald Henton Regulatory Affairs Manager Telephone: 303-530-6451 Fax: 303-530-6313 Email: donald.henton@covidien.com
#### Name of Device
| Trade Name: | Valleylab™ LS10, LS Series Single Channel Vessel Sealing Generator |
|----------------------|----------------------------------------------------------------------------------------------------|
| Catalog Numbers: | VLLS10GEN |
| Common Name: | Bipolar Electrosurgical Vessel Sealing Generator |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories (21 CFR § 878.4400, Class II, GEI). |
## Predicate Device
The Valleylab™ LS10, LS Series Single Channel Vessel Sealing Generator (VLLS10GEN) was compared to and found to be substantially equivalent to the following products of comparable type in commercial distribution:
| Trade Name: | ForceTriad™ Electrosurgical Generator |
|---------------------|---------------------------------------|
| Device Common Name: | Bipolar Electrosurgical Generator |
| Catalog Number: | ForceTriad |
| 510(k) Number: | K110268 (cleared 05/31/2011) |
| Manufacturer: | Covidien |
#### Device Description
Valleylab™ LS10, LS Series Single Channel Vessel Sealing Generator (VLLS10GEN) is a bipolar electrosurgical generator with LigaSure™ vessel sealing output mode. The generator is an electrically isolated, microcontroller-based device, incorporating closed-loop control for LigaSure™ mode implemented in the microcontroller firmware. It incorporates tissue sensing circuitry to constantly measure the electrical resistance of the tissue and instantaneously adjust the generator output to maintain the desired power.
{4}------------------------------------------------
The generator is used with compatible LigaSure™ instruments to ligate (seal) and divide (cut) vessels up to and including 7mm, tissue bundles, and lymphatics during the general surgical procedures using radio frequency (RF) energy. When a LigaSure™ instrument is applied to a vessel or tissue bundles, and RF energy is activated by the generator, the collagen and elastin in the tissues are reformed by heat and pressure to fuse vessel walls, thereby forming a permanent seal. The microprocessor within the generator monitors the tissue properties, stops the application of energy, and allows a brief period of cooling before indicating that the seal cycle is complete.
## Indications for Use
The Valleylab™ LS10, LS Series Single Channel Vessel Sealing Generator is an electrosurgical generator containing LigaSure™ vessel sealing technology.
The vessel sealing function is indicated for use in sealing (fusing) vessels (including pulmonary) up to and including 7mm in diameter, tissue bundles, and lymphatics during general surgery including, but not limited to surgical specialties such as urologic, vascular, thoracic, gynecologic, plastic and reconstructive, and colorectal.
Refer to each instrument's instructions for use (IFU) for additional indications, warnings, and specific contraindications.
The LigaSure™ system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use this function for these procedures
## Technological Characteristics
Valleylab™ LS10, LS Series Single Channel Vessel Sealing Generator (VLLS10GEN) has the equivalent LigaSure sealing performance with updated technological characteristics including improved electronics and microprocessors for energy delivery, simplified user interface, integrated radio frequency identification (RFID) technology, smaller overall dimensions, and the addition of a USB port as compared to the predicate, ForceTriad™. Valleylab™ LS10 generator incorporates LigaSure™ technology that provides precise energy delivery with electrode pressure on vessels to achieve a complete and permanent vessel seal.
Valleylab™ LS10 generator has one LigaSure™ port in the front panel allowing for connection of a single LigaSure" instrument, and provides a footswitch port in the rear panel to be used with compatible LigaSure™ single pedal footswitch. It provides visual and audible indicate the compatibility of instruments/footswitch and shows the status of seal cycle until the seal cycle is complete.
Valleylab™ LS10 generator is compatible with intelligent instruments that include a Radio Frequency Identification (RFID) tag and/or Aztec barcodes in the housing of the instrument cable plug. The RFID module is located above the LigaSure™ port of the generator, and is intended to identify the inserted LigaSure™ instrument and configure the generator with the data included in the RFID tag. The RFID tag on the instrument will be read by the proposed generator prior to scanning the barcode label upon detecting an instrument insertion.
Valleylab™ LS10 generator provides a USB Port used for remote service, data output, and data input. It can be connected with Covidien Valleylab 100 Exchange Platform to download the software for the purpose of generator configuration and data exchange.
## Performance
Evidence of safety and effectiveness were obtained from bench testing and preclinical testing.
Bench testing to support the intended use of this generator includes:
48
{5}------------------------------------------------
- Testing in accordance with IEC 60601-1 ●
- Testing in accordance with IEC 60601-2-2 ●
- Testing in accordance with IEC 60601-1-2
- Software verification and validation in accordance with IEC 62304
- Engineering and functional testing including system, mechanical, electrical and general functional testing
- Renal and pulmonary burst pressure
- . Thermal profile of the device including maximum external jaw temperature, external jaw cool down time, and maximum shaft temperature
Preclinical testing includes:
- Sealing and dividing vessels up to and including 7 mm ●
- . Ability to achieve hemostasis of tissue, vessels, tissue bundles
- Thermal spread
- Lymphatic burst pressure ●
- . Chronic animal study
## Conclusion
The results of the testing demonstrate that the proposed Valleylab™ LS10 generator operated as intended and is substantially equivalent to the predicate generator, ForceTriad TM
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.