K141507 · Syneron Medical, Ltd. · GEI · Dec 8, 2014 · General, Plastic Surgery
Device Facts
Record ID
K141507
Device Name
TWO SKIN TREATMENT SYSTEM
Applicant
Syneron Medical, Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Dec 8, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The eTwo Skin Treatment System is intended for dermatological procedures. The Sublative RF applicator is indicated for dermatological procedures requiring ablation and resurfacing of the skin, and for the treatment of facial wrinkles. At higher energy levels greater than 62 mJ/pin. the Sublative RF applicator is limited to skin types I-IV. The Sublime applicator is indicated for non-invasive wrinkles treatment. At higher energy levels greater than 100 J/cm3, the Sublime applicator is limited to skin types I-IV.
Device Story
Mobile dermatological system; utilizes two applicators for skin treatment. Sublative RF applicator delivers bipolar radiofrequency (RF) via multi-electrode pin array to heat skin, causing ablation and resurfacing. Sublime applicator combines infrared (IR) light and bipolar RF to bulk-heat dermis, stimulating collagen for wrinkle reduction. Operated by clinicians in professional settings. Output is thermal energy delivery to target tissue. Healthcare providers use device to improve skin appearance and reduce wrinkles. Benefits include non-invasive or minimally invasive skin rejuvenation with demonstrated safety and efficacy profiles.
Clinical Evidence
Prospective multi-center study of 72 subjects (mean age 49, mostly female/Caucasian, Fitzpatrick types II-III). Primary endpoint: wrinkle reduction at 12 weeks post-treatment using Fitzpatrick Wrinkle and Elastosis Scale. Results: 71% of subjects achieved wrinkle reduction; 77% showed at least 1-unit improvement on Fitzpatrick Scale; 91% showed overall facial appearance improvement. Safety: 6.9% (5/72) device-related adverse events, all mild/moderate and resolved. No serious adverse events reported.
Technological Characteristics
Mobile system; bipolar RF and IR light sources. Sublative RF uses multi-electrode pin arrays (including 44-pin tip). Sublime uses IR light and bipolar RF. Electrical safety/EMC per IEC 60601-1 and 60601-1-2. Sterilization provides SAL 10^-6. Software controls energy delivery parameters.
Indications for Use
Indicated for dermatological procedures, skin ablation/resurfacing, and treatment of facial wrinkles. Sublative RF applicator (energy >62 mJ/pin) and Sublime applicator (energy >100 J/cm2) limited to Fitzpatrick skin types I-IV.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 8, 2014
Syneron Medical Ltd. c/o Ms. Janice Hogan Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, PA 19103
Re: K141507
Trade/Device Name: eTwo Skin Treatment System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical, Cutting & Coagulation & Accessories Regulatory Class: Class II Product Code: GEI Dated: November 4, 2014 Received: November 4, 2014
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
510(k) Number (if known) K141507
Device Name
eTwo Skin Treatment System
Indications for Use (Describe)
The eTwo Skin Treatment System is intended for dermatological procedures.
The Sublative RF applicator is indicated for dermatological procedures requiring ablation and resurfacing of the skin, and for the treatment of facial wrinkles. At higher energy levels greater than 62 mJ/pin. the Sublative RF applicator is limited to skin types I-IV.
The Sublime applicator is indicated for non-invasive wrinkles treatment. At higher energy levels greater than 100 J/cm3, the Sublime applicator is limited to skin types I-IV.
Type of Use (Select one or both, as applicable)
Z Prescription Use (Part 21 CFR 801 Subpart D) Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number. "
FORM FDA 3881 (1/14)
Page 1 of 1 FDA
PSC Publishing Services (301) 443-6740 EF
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page
□ Over-The-Counter Use (21 CFR 801
{3}------------------------------------------------
#### 510(k) SUMMARY Syneron Medical Ltd.'s eTwo Skin Treatment System
#### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Syneron Medical Ltd. P.O.B 550 Industrial Zone Tavor Building, Yokneam Illit 20692, Israel
Tel: +972 (0) 732-442-200 Fax: +972 (0) 732-442-202
Contact Person: Ruthie Amir, MD, Global Vice President of Clinical and Regulatory Affairs
Date Prepared: December 5, 2014
#### Name of Device
eTwo Skin Treatment System
#### Common or Usual Name
Electrosurgical cutting and coagulation device and accessories
### Classification Name/Product Code
Classification: 21 CFR 878.4400, Electrosurgical cutting and coagulation device and accessories
Product Codes: GEI (Electrosurgical, cutting & coagulation & accessories), OUH (Skin resurfacing RF applicator)
#### Predicate Devices
Syneron eTwo Skin Treatment System (K110672) EndyMed Medical, Ltd.'s Glow (K120513, K083461, K101510)
#### Intended Use / Indications for Use
The eTwo Skin Treatment System is intended for dermatological procedures.
The Sublative RF applicator is indicated for dermatological procedures requiring ablation and resurfacing of the skin, and for the treatment of facial wrinkles. At higher energy levels greater than 62 mJ/pin, the Sublative RF applicator is limited to skin types I-IV.
{4}------------------------------------------------
The Sublime applicator is indicated for non-invasive wrinkles treatment. At higher energy levels greater than 100 J/cm², the Sublime applicator is limited to skin types I-IV.
# Device Description
The eTwo Skin Treatment System is a mobile system combining two existing technologies and applications for dermatological procedures.
The Sublative RF applicator delivers bipolar radiofrequency (RF) to the skin surface via an array of multi-electrode pin tips. The applicator delivers bipolar RF energy to the skin, which results in heating of demarcated spots of skin to temperatures that result in ablation and resurfacing of the skin that comes in contact with the multi-electrode pins.
The Sublime applicator uses a combination of an infrared (IR) light source and bipolar radiofrequency (RF) to bulk heat the dermis and affect dermal collagen in order to treat wrinkles. The bulk heating of the dermal layers refers to the gradual and gentle accumulation of heat with each additional pass performed.
# Technological Characteristics
The eTwo System has generally the same technological characteristics as cleared in the predicate eTwo System, including device design, material, and energy source, except the subject eTwo System includes an additional 44-pin electrode tip and allows for higher RF energy levels to be delivered. The Sublime applicators of the eTwo Systems provide IR light and RF energy to the treatment area, whereas the non-ablative applicator of the EndyMed predicate provides RF energy only. The ablative applicators of the eTwo System and its predicates deliver RF energy to the treatment area. The hardware of the subject eTwo remains the same as previously cleared in the prior iteration of the device system.
### Performance Data
### Software
Software verification and validation was performed to evaluate the minor software changes made to support the increase in RF energy for the Sublime and Sublative RF applicators and to support the additional tip configuration (44 pin). Results showed that the Sublime and Sublative RF applicators with the eTwo Skin Treatment System function as intended.
# Electrical Safety and Electromagnetic Compatibility
The eTwo System has been evaluated for electromagnetic compatibility and electrical safety testing per applicable standards of IEC 60601-1-2 and 60601-1, and all results were passing.
{5}------------------------------------------------
# Histological Evaluations
Histological evaluation of skin tissues treated with the Sublime and Sublative applicators examined the impact of each applicator on skin tissue under the treated area, and demonstrated that the device performs as intended.
# Sterilization Validation
Sterilization validation study results demonstrated that the sterilization method for the eTwo System continues to provide a minimum sterility assurance level of 10°, including for the additional 44-pin tip.
# Clinical Study
In addition, the eTwo System treatment was evaluated in a prospective multi-center clinical study of 72 subjects through 24 weeks of follow up after the end of treatment. Four follow up visits (1 week, 6 weeks, 12 weeks after treatment) were scheduled for subjects receiving 1 treatment session, and five follow up visits (1 week, 6 weeks, 12 weeks, 18 weeks, and 30 weeks after first treatment) were scheduled for subjects receiving 2 treatment sessions. The study evaluated the higher energy levels and the additional 44-pin tip of the subject eTwo System through assessments by blinded evaluators, investigators, and subjects. Subjects' results at each follow up visit were compared to baseline to determine the amount of improvement with treatment. The primary endpoint of the study was to evaluate the reduction in wrinkles in the treated areas at 12 weeks following the last treatment session compared to baseline, based on photographic assessments by three independent blinded evaluators using the Fitzpatrick Wrinkle and Elastosis Scale (Degree of Elastosis categorized from class I (mild) to class The study also evaluated the safety of the treatment and recorded the III (severe)). number and type of any adverse event during the study. Additional assessments included the Global Aesthetic Improvement scale and patient satisfaction.
The majority of the enrolled population was female, Caucasian, with Fitzpatrick skin type II or III. Mean age was 49 years. The study population presented to the study with a range in Fitzpatrick Wrinkle and Elastosis scores and skin quality.
The pivotal study of the eTwo System demonstrated that device treatment achieves wrinkle reduction under the proposed parameters for a majority (71%) of the study subjects, meeting the primary study endpoint for effectiveness (50%). Results of secondary endpoint analyses showed improvement in wrinkles as observed in blinded evaluation, by investigators, and as perceived by subjects themselves over time. At the primary endpoint visit (12 weeks post treatment end), investigator assessments determined that 77% of the subjects were rated as improved from baseline (i.e., at least 1 unit improvement in Fitzpatrick Scale) and 91% of the subjects showed an overall improvement in facial appearance, further confirming the favorable results reported in
{6}------------------------------------------------
blinded evaluation. In addition, high levels of patient satisfaction were reported by the subjects, where the majority of subjects (≥63%) reported being satisfied or very satisfied with treatment throughout follow up. Treatment-associated discomfort or pain immediately following treatment generally resolved within a week.
At the proposed higher RF energy levels and with the additional 44-pin tip, the eTwo treatment continued to demonstrate a favorable safety profile with 6.9% (5/72 subjects) device-related adverse events throughout the study. All of the device-related adverse events were mild to moderate, and each event completely resolved during the study (most resolved within a few hours or days). There were or serious adverse events in the study and no subjects discontinued from the study due to adverse events. A few subjects reported hyperpigmentation or hypopigmentation following treatment, which were also recognized as treatment-associated responses. In total, 11 subjects reported any changes in pigmentation or texture. Of these subjects, the events resolved for 6 subjects (most within a few days). For the remaining 5 subjects, results showed that the noted event was not associated with a negative clinical impact on the patient. Therefore, the clinical study data demonstrate that the eTwo System performs as intended with a favorable safety and effectiveness profile.
### Substantial Equivalence
The eTwo System has the same intended use as the predicate devices. The eTwo System indications for use are the same as the predicate eTwo device, and similar to those of the EndyMed Glow predicate. The eTwo System presents similar technological characteristics and principles of operation as the predicate devices. The key technological differences between the current and predicate eTwo Systems are in terms of the increased RF energy level and the addition of an electrode pin tip size. The proposed eTwo treatment parameters have been evaluated in clinical testing, and results demonstrate that the device continues to perform as intended. The EndyMed predicate similarly presents higher levels of RF energy. Although the EndyMed device differs in terms of the energy type delivered (RF only) compared to the eTwo Systems (RF only and IR combined with RF), this difference does not raise new types of safety or effectiveness questions given that the predicate eTwo utilizes the same combination of energies. Further, the underlying mechanisms of all three devices are consistent to achieve the same intended therapeutic effect for wrinkles treatment. Therefore, the eTwo System has satisfied the criteria for substantial equivalence.
# Conclusions
Syneron's eTwo System has been evaluated in nonclinical testing, and results demonstrated the device performance is appropriate for its intended use. The eTwo System presents the same intended use and similar indications for use, technological characteristics, and principles of operation as its predicate devices. Therefore, the eTwo System is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.