K132797 · U&I Corporation · GEI · Jul 30, 2014 · General, Plastic Surgery
Device Facts
Record ID
K132797
Device Name
L'DISQ
Applicant
U&I Corporation
Product Code
GEI · General, Plastic Surgery
Decision Date
Jul 30, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The L'DISQ® is indicated for coagulation and ablation of disc material to treat symptomatic patients with contained herniated discs.
Device Story
L'DISQ® is a plasma-based radiofrequency electrosurgical device used for coagulation and ablation of disc material. It consists of a bipolar tip (available in 1.0 mm or 0.7 mm thicknesses), manipulator, and support components. The device is used to treat symptomatic patients with contained herniated discs. It functions by delivering high-frequency energy to remove disc lesions. The device is intended for single use and is operated by a clinician in a surgical setting. The clinician uses the device in conjunction with an electrosurgical power generator to perform the procedure. By removing disc material, the device aims to alleviate pain associated with disc herniation.
Clinical Evidence
Bench testing only. Testing included nucleus pulposus removal efficacy and electrode durability. Results met all acceptance criteria, demonstrating performance equivalent to the predicate device.
Technological Characteristics
Plasma-based radiofrequency electrosurgical device. Components include bipolar tip (1.0mm or 0.7mm), manipulator, polyimide tube, and RF connector. Energy source: high-frequency electrosurgical generator. Single-use, sterile device. No software or algorithm-based processing described.
Indications for Use
Indicated for coagulation of disc material in symptomatic patients with contained herniated discs.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for U&I Corporation, with the letters "U&i" in a stylized font on the left and the word "CORPORATION" in bold letters on the right. Above the word "CORPORATION" is the date "JUL 3 0 2014". The text is all in black and the background is white.
## 5. 510(k) Summary
| Manufacturer: | U & I Corporation<br>20, Sandan-ro 76beon-gil(Rd), Uijeongbu-si<br>Gyeonggi-do, 480-859, Korea<br>Gyeong-Je Kwon, Regulatory Affairs Specialist |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | U & I Corporation<br>20, Sandan-ro 76beon-gil(Rd), Uijeongbu-si<br>Gyeonggi-do, 480-859, Korea |
| Sponsor Contact: | Gyeong-Je Kwon, Regulatory Affairs Specialist |
| Date Prepared: | February 21, 2014 |
| Device Name: | Trade Name: L'DISQ® |
| Common Name: | Electrosurgical Device and Accessories |
| Classification Name: | Electrosurgical Cutting and Coagulation Device and<br>Accessories (GEI), per 21 CFR 878.4400 |
| Product Code: | GEI |
| Predicate Devices: | ArthroCare® Perc TMDC/DLR SpineWand™<br>(K010811, K030954, K053447, K061259) |
#### Description of Device:
The L'DISQ® is a plasma electrode which is 1.0 mm(PD01, PD02, PD03, PD04, PD05, PD07 and PC01) or 0.7 mm(PC02) thickness based on the high frequency electrosurgical system. L'DISQ® removes the disc lesions that are causing pain by a plasma based radiofrequency device.
L'DISQ® is a single use device and there are 3-type of designs (L'DISQ, L'DISQ C, L'DISQ C.7) according to the level of spine treated. When using the L'DISQ® the users should also study carefully "Operation Manual" of the electrosurgical power generator as well as "Instructions for Use" and "Procedure Technique" of L'DISQ®
L'DISQ®
**U&i** CORPORATION.
{1}------------------------------------------------
K132797 Page 2 of 3
1. L'DISQ (PD01, PD02, PD03, PD04, PD5, PD07): Device for lumbar disc
- L'DISQ consists of seven components:
- (1) BIPOLAR TIP 1.0MM (CAP + INSULATOR + RING + POLYMER TUBE + WIRE 0.2mm + COPPER WIRE)
- (2) TIP SUPPORT (25mm, 45mm, 15mm, 9mm, 35mm)
- (3) TIP PROTECTION
- (4) POLYIMIDE TUBE
- (5) RF CONNECTOR (4PIN)
- (6) MANIPULATOR
- (7) SPRING STOPPER
- 2. L'DISQ C (PC01): Device for cervical disc
- L'DISQ C consists of six components:
- (1) L'DISQ C BIPOLAR TIP (CAP + INSULATOR + RING + POLYMER TUBE-C +
- WIRE 0.2mm + COPPER WIRE)
- (2) TIP SUPPORT-C 17mm
- (3) POLYIMIDE TUBE
- (4) L'DISQ C MANIPULATOR
- (5) SPRING STOPPER
- (6) MANIPULATOR CABLE
- 3. L'DISQ C.7 (PC02): Device for cervical disc
- L'DISQ C.7 consists of four components:
- (1) L'DISQ C.7 BIPOLAR TIP (CAP+INSULATOR+EDM WIRE)
- (2) TIP SUPPORT
- (3) L'DISQ C.7 MANIPULATOR
- (4) MANIPULATOR CABLE
#### Intended Use:
The L'DISQ® is indicated for coagulation and ablation of disc material to treat symptomatic patients with contained herniated discs.
#### Substantial Equivalence:
In establishing substantial equivalence to the predicate device, L'DISQ® evaluated the indication for use, components, performance, approaching method, energy source, and sterilization methods of those systems.
**CORPORATION**
{2}------------------------------------------------
#### 1. Performance Testing
To evaluate the performance of L'DISQ®, we conducted following tests.
(1) Removal test for nucleus pulposus
(2) Durability test for electrode
The test result met all acceptance criteria and verifies that performance of the L'DISQ® is substantially equivalent to the predicate device, Perc™DC/DLR SpineWand™ (ArthroCare).
#### 2. Conclusion
The data and information provided in this submission support the conclusion that the L'DISQ® is substantially equivalent to its predicate device with respect to indications for use, components, performance, approaching method, energy source and sterilization method and other technological characteristics.
**U&i** CORPORATION
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized symbol resembling a human figure with outstretched arms, accompanied by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The symbol is black, and the text is also black.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 30, 2014
U & I Corporation Gyeong-Je Kwon Regulatory Affairs Specialist 20, Sandan-ro 76beon-gil(Rd) Uijungbu-si, Gyeonggi-do 480-059 Korea
Re: K132797 Trade/Device Name: L'DISQ® Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: June 20, 2014 Received: June 23, 2014
Dear Kwon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require'approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Gyeong-Je Kwon
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regalation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K132797
Device Name
L'DISQ®
Indications for Use (Describe)
The L'DISQ® is indicated for coagulation of disc material to treat symptomatic patients with contained herniated discs.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FOR FOR FOR FOR FOR FOR FOR USE ONLY Concurrence of Center for Devices and Radiological Heatth (CDRH) (Signature)
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.