K130027 · Cathay Manufacturing Corp. · GEI · Aug 22, 2013 · General, Plastic Surgery
Device Facts
Record ID
K130027
Device Name
DISPOSABLE NEUTRAL ELECTRODE GP202
Applicant
Cathay Manufacturing Corp.
Product Code
GEI · General, Plastic Surgery
Decision Date
Aug 22, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Indications for Use
This disposable neutral electrode for adult patients with conductive adhesive gel is used as neutral reference during electrosurgical procedures.
Device Story
Disposable Neutral Electrode (GP202) functions as an accessory to high-frequency electrosurgical devices; provides a return path for high-frequency current to prevent unwanted patient burns. Device consists of an electrode carrier, conductive gel (high molecular monomer, glycerinum, water), aluminum foil, sponge, anti-sticking film, and connecting wire. Used in clinical settings during electrosurgical procedures; operated by healthcare professionals. Device is applied to the patient's body to complete the electrical circuit. Benefits include safe dissipation of electrosurgical current, reducing risk of thermal injury.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety, mechanical performance, and biocompatibility verification against consensus standards (IEC 60601-1, IEC 60601-2-2, ISO 10993-5, ISO 10993-10).
Technological Characteristics
Materials: conductive gel (high molecular monomer, glycerinum, water), aluminum foil, sponge, PET carrier. Connectivity: wired connection to electrosurgical generator. Standards: IEC 60601-1 (electrical safety), IEC 60601-2-2 (electrosurgical devices), ISO 10993-5/10 (biocompatibility). Passive device; no software or active electronics.
Indications for Use
Indicated for adult patients as a neutral reference electrode during electrosurgical procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
| 1/3<br>510(k) Summary | |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Company: CATHAY MANUFACTURING CORP.<br>Address: No.328, Xishe Road, Maogang Town, Songjiang Area, Shanghai City, China<br>Contact Name: Ren Ren<br>Title: Assistant<br>Telephone: (86) 021- 57878328<br>E-mail: rubyfantasy@qq.com |
| Consultant: | Taizhou EBO information Technology Co., Ltd<br>Contact person: Zhang Hui<br>Telephone: (86) 010-59795724<br>Fax: (86) 010-52214567-5514<br>Email: itczhanghui@gmail.com<br>AUG 2 2 2013 |
| 1. Device information: | |
| Proprietary Name: | Disposable Neutral Electrode |
| Model: | GP202 |
| Regulation Description | Electrosurgical cutting and coagulation device and accessories |
| Product Code | GEI |
| Submission Type | 510(k) |
| Regulation Number | 21 CFR 878.4400 |
| Device Class | 2 |
| 2. Predicate Devices | |
K130027
K102372
ﻬﺎ
Trade/Device Name: OBS Disposable Electrosurgical Pads Models: GBS-Db, GBS-Dm Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories, Regulatory Class: Class II Product Code: GEI
#### Manufactured by:
Jiangmen City Xinhui BaiSheng Medical Equipment Co., Ltd.
### 3. Description of the device
The Disposable Neutral Electrode, is an accessory for high voltage high frequency electrosurgical devices, connect it to the body of the PATIENT and the HF SURGICAL
{1}------------------------------------------------
EQUIPMENT, it can perform it function for provide a return path for the HIGH FREQUENCY current and unwanted burns are avoided.
Construction --- The product consists of Electrode carrier and Clarity PET. It is packed in an al film bag.
The electrode is composed of conductive gel, aluminum foil, sponge, anti-sticking film and connecting wire. Conductive gel is composed of high molecular monomer, glycerinum and water, which is a gel, can be conducted in high frequency and has voltage-sensitive by polymerization.
#### 4. Indications for Use
This disposable neutral electrode for adult patients with conductive adhesive gel is used as neutral reference during electrosurgical procedures.
#### 5. Testing
Laboratory testing was conducted to validate and verify that the Disposable Electrosurgical Pencil met all design specifications, including electrical safety, EMC, biocompatibility, specification. Results of these tests demonstrate compliance to the requirements of all consensus standards
#### 6. Compared to the predicate device
Disposable Neutral Electrode has been carefully compared to legally marketed devices with respect to intended use, appearance, essential components, materials and performance specifications.
They are similar in intended use, appearance, essential components, materials and performance specifications. Although they may differ from the predicate devices in color and size, it won't affect safety and effectiveness of subject devices.
In addition, performance and safety testing have been done to validate the performance and safety of the device.
#### 7.0 Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence is as below:
Electrical safety test, Mechanical performance test, and Biocompatibility test have been done to demonstrate the safety and performance of subject devices. Tests was conducted in accordance with the "510(k) Guidance Document for General Surgical Electrosurgical Devices", which outlines safety and performance requirements.
The proposed device is equivalent to the identified predicate device with respect to technological characteristics and function. The device has been designed to comply with the applicable sections of ANSI/AAMI American National Standard for Electrosurgical Devices:
The International Standard for electrical medical device: IEC60601-1; the International
{2}------------------------------------------------
Standard for Electrosurgical Devices: IEC 60601-2-2; Biocompatibility: ISO 10993-5 and ISO 10993-10.
None of the test demonstrated and design characteristics that violated the requirements of the above mentioned standards or resulted in any safety hazards.
#### 8.0 Conclusions:
The comparison and validation results presented in this 510k notification, show that the GP202 Disposable Neutral Electrode is substantially equivalent to predicated devices and are safe and effective in their intended use.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
August 22, 2013
Cathay Manufacturing Corporation % Taizon EBO Information Technology Co., Ltd. Zhang Hui No. 328, Xishe Road, Maogang Town, Songhang Area Shanghai, China 201607
Re: K130027
Trade/Device Name: Disposable Neutral Electrode GP202 Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: July 15, 2013 Received: July 25, 2013
Dear Zhang Hui:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Zhang Hui
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Peter DJRuimm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K130027 Disposable Neutral Electrode Device Name:
Indications for Use:
This disposable neutral electrode for adult patients with conductive adhesive gel is used as neutral reference during electrosurgical procedures.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Joshua @ Mipper -S
(Division Sign-off) Division of Surgical Devices 510(k) Number K130027
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.