K124029 · Microline Surgical, Inc. · GEI · Mar 27, 2013 · General, Plastic Surgery
Device Facts
Record ID
K124029
Device Name
MICROLINE FLEXIBLE LIGATING SHEARS
Applicant
Microline Surgical, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Mar 27, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Microline Flexible Ligating Shears are intended for simultaneous cutting and cauterization of soft tissue during minimally invasive abdominal laparoscopic surgery. The device may also be used for cutting natural, non-metallic sutures during surgery.
Device Story
Microline Flexible Ligating Shears are sterile, single-patient-use handpieces designed for thermal ligation and division of soft tissue. The device features distal heating elements activated by a finger switch on the handpiece. It connects to a Microline Universal Power Supply, which allows the surgeon to control heating power to accommodate patient anatomy. The device is used in conjunction with an introducer, such as the TransEnterix SPIDER Single Port Device, which provides flexible instrument delivery tubes (IDTs) to facilitate multidirectional, triangulated access to the surgical field. The system is operated by a trained physician in a minimally invasive abdominal laparoscopic setting. By conducting heat to tissue compressed within the jaws, the device enables simultaneous cutting and cauterization, potentially reducing the need for multiple instruments and improving surgical efficiency.
Clinical Evidence
Bench testing only. Functionality and ligating performance were verified. System compatibility with the SPIDER device was confirmed. Sterilization validated per ISO 11137-2 (SAL 10^-6). Biocompatibility testing performed per ISO 10993-1. Electrical safety testing performed per IEC 60601-1.
Technological Characteristics
Thermal ligation via distal heating elements; finger-switch activation; powered by external Universal Power Supply (max 60W). Sterile, single-use handpiece; non-sterile reusable power supply. Compatible with SPIDER Single Port Device. Sterilization: Gamma irradiation (ISO 11137-2). Biocompatibility: ISO 10993-1 compliant.
Indications for Use
Indicated for patients undergoing minimally invasive abdominal laparoscopic surgery requiring soft tissue cutting and cauterization or cutting of natural, non-metallic sutures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
K124029
800 Cummings Center, Suite 166T Beverly, MA 01915, USA Tel: 978.922.9810 Fax: 978.922.9209 Web: www.microlinesurgical.com
# 510(k) Summary
MAR 2 7 2013
| Submitted By: | Microline Surgical |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Bill McCallum |
| Date Prepared: | February 19, 2013 |
| Proprietary Name: | Microline Flexible Ligating Shears |
| Classification Name: | Electrosurgical Cutting and Coagulation and Accessories<br>21 CFR § 878.4400<br>79 GEI |
| Predicate Devices: | This product is similar in design, composition, to and function to the:<br>Thermal Ligating Shears (K062257) cleared on October 10, 2006 and the<br>Spider Flexible Monopolar Surgical Instruments (K122299) cleared on<br>October 25, 2012. |
### Device Description and Technological Characteristics:
The Microline Flexible Ligating Shears system consists of the following:
Microline Flexible Ligating Shears Microline Universal Power Supply Single or multiple access port device such as the TransEnterix SPIDER Single Port.
The Microline Flexible Ligating Shears are designed to provide thermal ligation and division during minimally invasive abdominal laparoscopic surgery. The Microline Flexible Ligating Shears consist of a sterile single patient use handpiece. The device has heating elements at the distal tip which are activated by a finger switch located on the handpiece of the device. The Microline Universal Power Supply is designed to allow the surgeon control of the heating element power of the device in order to accommodate for individual patient anatomy. An instrument cord is attached to the handpiece and is plugged into the dedicated Microline Universal Power Supply.
The Microline Flexible Ligating Shears are designed to be used with an Introducer such as the SPIDER Single Port Device.
Image /page/0/Picture/11 description: The image shows a circular design. The design is composed of four curved shapes that resemble petals or lenses, arranged symmetrically around a central point. The shapes are enclosed within a circle, creating a pattern that looks like a stylized flower or a geometric emblem.
Image /page/0/Picture/12 description: The image contains two logos. The logo on the left is a circle with a checkmark inside, and the text "TÜV SÜD, SRQ" is written around the circle. The logo on the right has a crown on top of a checkmark, and the text "UKAS Product Certification 026" is written below the checkmark. Both logos appear to be certification marks.
{1}------------------------------------------------
The Microline Flexible Ligating Shears are supplied sterile in a Tyvek/Mylar sterilization pouch on a mounting card and thermoformed tray, placed in a fiberboard carton. The device is intended to be used by a trained physician for a single patient use during minimally invasive abdominal laparoscopic surgery.
The Microline Flexible Ligating Shears power cable is attached to the instrument handpiece and connects to the Microline Universal Power Supply. The Universal Power Supply is supplied nonsterile for reusable use outside the sterile field.
The Microline Flexible Ligating Shears incorporates Hi and Low heating modes that are used to cauterize and cut soft tissue. The heating elements in the tip are activated by a physician controlled finger switch located on the handpiece of the device. The Microline Flexible Ligating Shears is intended to provide general purpose cauterization, spreading, and grasping of soft tissue during minimally invasive or open surgical procedures.
## Principle of Operation:
The Microline Flexible Ligating Shears provide a path of entry to the surgical target by the SPIDER device's flexible extended IDTs (Instrument Delivery Tube), also called the SPIDER Single Port Device. Two (2) flexible channels known as IDTs (Instrument Delivery Tubes) are positioned horizontally to the operative plane and include extended flexible lumens to facilitate manipulation of surgical instruments, enabling control over operator determined distances. The IDTs facilitate x, y, and z motion to allow for a multidirectional and triangulated approach within the surgical field. This mimics the approach of standard laparoscopic surgery using multiple incisions.
The flexible IDTs provide 360 degrees of freedom at the distal tips. The right hand controls the right instrument and the left hand controls the left instrument. Two (2) rigid channels positioned vertically to the operative field enable the use of an endoscope and any other surgical instruments.
## Indications for Use:
The Microline Flexible Ligating Shears are intended for simultaneous cutting and cauterization of soft tissue during minimally invasive abdominal laparoscopic surgery. The device may also be used for cutting natural, non-metallic sutures during surgery.
## Performance Testing:
The Microline Flexible Ligating Shears have been tested for functionality and they have found to perform their intended functions for minimally invasive abdominal laparoscopic procedures. Additionally, testing has proven that they perform their ligating functions. The device has also been tested with the SPIDER® device and pasted all required testing for system compatibility.
The Microline Thermal Ligating Shears have been sterilized to a SAL of 1x10* using gamma irradiation. The gamma sterilization process was validated to the following: - Clause 9, Method VDmax -Substantiation of 25 kGy or 15 kGy as the sterilization dose in ISO 11137-2. Sterile barrier testing has
Image /page/1/Picture/12 description: The image shows a circular design. The design is composed of four leaf-like shapes arranged symmetrically around a central point. The leaves are oriented so that their pointed ends face outward, creating a pattern that resembles a four-petal flower or a stylized compass rose.
{2}------------------------------------------------
been performed and it has been shown that the sterile barrier has the capability of maintaining sterility during the required shelf life, and that the device itself maintains its performance requirements.
Biocompatibility testing was successfully conducted on the Microline Flexible Ligating Shears in accordance with ANSI/AAMI/ISO 10993-1. The Microline Flexible Ligating Shears falls into the category of external communicating device and the appropriate tests were performed as described in Table A.1 in Annex A of ANSI/AAMI/ISO 10993-1.
The Microline Flexible Ligating Shears were evaluated by an independent testing laboratory and found compliant to IEC 60601-1:1988 + A1:1999 + A2:1995 general requirements for basic safety, including all US, CAN National Differences.
As illustrated in the Predicate Comparison Table below, the Microline Flexible Ligating Shears do not incorporate any significant technological differences that effect safety and efficacy. Therefore clinical data was not necessary for evaluation of substantial equivalence. The verification and validation test results are sufficient to demonstrate safety and effectiveness when compared to the predicate devices. The minor design or material differences between The Microline Thermal Ligating Shears and the predicate devices have been evaluated and tested and found to present no new issues of safety and effectiveness.
| Characteristic | Starion Instruments<br>Corporation TLS3<br>Thermal Ligating Shears<br>510(k) K062257 (Predicate<br>device) | Microline Flexible Ligating<br>Shears (subject of this<br>submission) | Transenterix Spider Flexible<br>Monopolar Surgical<br>Instruments K122299<br>(Predicate device) |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | For the simultaneous<br>cutting and cauterization of<br>soft tissue during surgery,<br>and cutting of natural or<br>synthetic, non-metallic<br>sutures during surgery. | The Microline Flexible<br>Ligating Shears are intended<br>for simultaneous cutting and<br>cauterization of soft tissue<br>during minimally invasive<br>abdominal laparoscopic<br>surgery. The device may also<br>be used for cutting natural,<br>non-metallic sutures during<br>surgery. | The Spider Surgical Instruments<br>are intended for use on<br>minimally invasive abdominal<br>laparoscopic surgical procedures<br>for grasping, mobilizing,<br>dissecting, retracting, cutting,<br>cauterizing, ligating,<br>suction/irrigation and other<br>manipulation of tissues and<br>vessels during laparoscopic<br>surgery. |
| Principle of<br>Operation | Heat is conducted to tissue<br>compressed within the jaws<br>via heating elements in the<br>instrument tip. | Same | Monopolar high frequency<br>current passes from active jaw<br>tips of instrument to cut a<br>coagulate tissue at the immediate<br>surgical site, flowing through the<br>patient to an adhesive electrode<br>return pad connected to the<br>monopolar electrosurgical<br>generator. |
| Energy Source | Electrical current from<br>Starion Instruments<br>Universal Power Supply<br>(UPS) | Same | Electrical high frequency current<br>and voltage from a third party<br>monopolar electrosurgical unit<br>(ESU). |
## Predicate Comparison Table:
Image /page/2/Picture/6 description: The image contains the text "Confidence, simply delivered." The text is written in a bold, sans-serif font. The text is black and is set against a white background. The text appears to be a slogan or tagline for a company or product.
Image /page/2/Picture/7 description: The image shows a circular design composed of four leaf-like shapes arranged symmetrically around a central point. Each leaf shape has a curved outline and tapers to a point at both ends. The overall impression is that of a stylized flower or a geometric pattern with organic elements. The image is in black and white.
Image /page/2/Picture/8 description: The image contains two logos. The logo on the left is a circle with a checkmark inside, and the letters "SGS" are written below. The logo on the right is a checkmark with Arabic text below. Both logos appear to be certification marks.
{3}------------------------------------------------
| Method of<br>Actuation | Finger switch | Same | Same |
|--------------------------------------------------------------------|-----------------------------------------------------------------------------------|-----------------------------------------------------------------|--------------------------------------------------------------|
| Sterility | Gamma radiation, sterility<br>assurance level (SAL) is $10^6$ | Same | Same |
| Cleaning and re-<br>sterilization of<br>the reusable<br>hand piece | N/A<br>Sterile for single use only. | Same | Same |
| Primary Sterile<br>Barrier | Tyvek/Mylar pouch | Same | Same |
| Biocompatibility | All patient contact<br>materials meet<br>biocompatibility test per<br>ISO 10993-1 | Same | Same |
| Energy Used<br>and/or<br>Delivered | Maximum Output - 60W | Same | Monopolar high frequency<br>output. |
| Rigidity/Flexibil<br>ity | Rigid, not flexible | Functions with the<br>TransEnterix SPIDER Single<br>Port Device | Functions with the TransEnterix<br>SPIDER Single Port Device |
## Summary:
The information provided demonstrates that the Microline Flexible Ligating Shears is substantially equivalent to the Thermal Ligating Shears (K062257) and the Spider Flexible Monopolar Surgical Instruments (K122299) cleared on October 25, 2012, in function, intended use and indications for use.
Image /page/3/Picture/5 description: The image contains two certification logos. The logo on the left is the SGS certification mark, which features a stylized checkmark within a circular design. The logo on the right is the UKAS (United Kingdom Accreditation Service) certification mark, which includes a crown above a checkmark and the text "UKAS Accredited Management Systems 005".
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is a stylized image of an eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Microline, Surgical, Incorporated % Mr. Bill McCallum Regulatory Affairs Manager 800 Cummings Center, Suite 166T Beverly, Massachusetts 01915
March 27, 2013
Re: K124029 Trade/Device Name: Microline Flexible Ligating Shears Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: December 26, 2012 Received: December 28, 2012
Dear Mr. McCallum:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
#### Page 2 - Mr. Bill McCallum
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours, FOR
# Peter D. Rumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known): K124029
## Indications for Use
#### Device Name: Microline Flexible Ligating Shears
Indications for Use:
The Microline Flexible Ligating Shears are intended for simultaneous cutting and cauterization of soft tissue during minimally invasive abdominal laparoscopic surgery. The device may also be used for cutting natural, non-metallic sutures during surgery.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Joshua C. Nipper -S
For
(Division Sign-Off) Division of Surgical Devices 510(k) Number __K124029
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.