K123366 · Ellman International, Inc. · GEI · Jul 12, 2013 · General, Plastic Surgery
Device Facts
Record ID
K123366
Device Name
SURGITRON 4.0 DUAL RF S5-IEC
Applicant
Ellman International, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jul 12, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Surgitron 4.0 Dual RF S5 - IEC is intended to deliver electrosurgical output to perform cutting, coagulation, and hemostasis of soft tissues and to heat tissue in the non-ablative treatment of mild to moderate facial wrinkles and rhytids. The device is used with compatible Ellman International electrodes and handpieces which are used for these various applications in physician offices and surgical centers.
Device Story
Radiofrequency (RF) generator delivering energy at ~4 MHz for electrosurgical procedures; used in physician offices and surgical centers by clinicians. Input: user-selected waveform and power level via front panel push buttons; output controlled by foot or hand switch. Output: RF energy delivered to soft tissue via monopolar or bipolar electrodes/handpieces. Modes: cut, cut/coag (blend), hemo (coag), fulgurate, and bipolar. Clinical utility: tissue cutting, coagulation, hemostasis, and non-ablative wrinkle treatment. Benefits: precise tissue management, hemorrhage control, and aesthetic improvement.
Clinical Evidence
Bench testing only. Device tested for conformance to IEC 60601-1 and IEC 60601-2-2 standards. No clinical data presented.
Technological Characteristics
RF generator operating at ~4 MHz. Five energy delivery modes: cut, blend, hemo, fulgurate, bipolar. System includes RF generator, footswitch, handpieces, neutral plate, and cables. Complies with IEC 60601-1 and 60601-2-2.
Indications for Use
Indicated for non-ablative treatment of mild to moderate facial wrinkles/rhytids (skin phototypes I-IV); soft tissue cutting, coagulation, and hemostasis for procedures including snoring, uvulopalatoplasty, myringotomy, epistaxis, turbinate shrinkage, skin incisions, biopsy, cyst/abscess/tumor/verrucae/carcinoma/nevi/fistula/epithelioma management, cosmetic repairs, skin flaps, skin tags, blepharoplasty, epilation, and telangiectasia.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
#### K123366
#### 510(k) SUMMARY
# JUL 1 2 2013
## Ellman International - Surgitron 4.0 Dual RF S5 - IEC
#### Submission Correspondent and Owner:
| Name: | Ellman International |
|--------------------|----------------------------------------------------------------------------------------|
| Address: | 3333 Royal Avenue<br>Oceanside NY 11572-3625 |
| Phone: | 513.658.8960 |
| Facsimile: | 516.267.6750 |
| Contact: | Alison Sathe |
| Email: | asathe@ellman.com |
| Date Prepared: | October 25, 2012 |
| Trade Name: | Surgitron 4.0 Dual RF S5 - IEC |
| Common Name: | Electrosurgical, cutting & coagulation & accessories |
| Classification: | Class II, 21 CFR 878.4400, Product Code GEI |
| Predicate Devices: | Ellman International - Surgitron 120 IEC (Surgitron 4.0 Dual RF)<br>(K013255, K082834) |
#### Device Description:
The Surgitron 4.0 Dual RF S5 - IEC is a radiofrequency energy generator employed for a variety of electrosurgical procedures. The action is achieved by selection of the desired waveform and power level on the front panel. Selections are activated by push buttons and indicated on the digital display giving the operator feedback on device setting status. Device output is controlled via foot and/or hand switch. Monopolar and bipolar electrodes may be used with the device. The generator includes the following accessories:
- IEC Power Cord ●
- Dual Footswitch & Cable .
- IEC Foot controlled Handpiece ●
- . Bipolar Cable
- . Monopolar cables
- Disposable Neutral Plate .
- Multi-Button Finger switch Handpiece .
{1}------------------------------------------------
#### Indications for Use:
The Surgitron 4.0 Dual RF S5 - IEC is intended to deliver electrosurgical output to perform cutting, coagulation, and hemostasis of soft tissues and to heat tissue in the non-ablative treatment of mild to moderate facial wrinkles and rhytids. The device is used with compatible Ellman International electrodes and handpieces which are used for these various applications in physician offices and surgical centers.
The Surgitron 4.0 Dual RF S5 - IEC indications are the same as those cleared in K082835:
- Non-ablative treatment of mild to moderate facial wrinkles and rhytids for skin . . phototypes I-IV
- . Cutting: snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, turbinate shrinkage, skin incisions, biopsy, cysts, abscesses, tumors, cosmetic repairs, development of skin flaps, skin tags and blepharoplasty.
- . Blended Cutting and Coagulation: snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, turbinate shrinkage, skin tags, papilloma keloids, keratosis, verrucae, basal cell carcinoma, nevi, fistulas, epithelidma, cosmetic repairs, cysts, abscesses, and development of skin flaps.
- . Hemostasis: control of bleeding, epilation, telangiectasia.
- Fulguration: basal cell carcinoma, papilloma, cyst destruction, tumors, verrucae, . hemostasis.
- . Bipolar: pinpoint precise coagulation, pinpoint hemostasis in any field (wet or dry), snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment and turbinate shrinkage.
#### Technological Characteristics:
The Surgitron 4.0 Dual RF S5 - IEC system consists of: RF generator, footswitch and cable, handpieces, neutral plate and cables (bipolar and power cord). The device operates at approximately 4 MHz and is used with various electrodes and handpieces to deliver RF energy to tissue. The Surgitron 4.0 Dual RF S5 - IEC has five modes of energy delivery which operate over a range of power outputs: cut, cut/coag (Blend), hemo (Coag), fulgurate, and bipolar.
#### Performance Data
The Surgitron 4.0 Dual RF S5-IEC has been tested to IEC 60601-1 and 60601-2-2 standards. In all instances, the Surgitron 4.0 Dual RF S5 - IEC functioned as intended and in conformance with all applicable standards.
#### Substantial Equivalence
There are no unique applications, intended uses, materials or specifications presented herein. The Surgitron 4.0 Dual RF S5-1EC is as safe and effective as the Surgitron 4.0 Dual RF - IEC
{2}------------------------------------------------
predicate device. The Surgitron 4.0 Dual RF S5 – IEC has the same intended uses, indications and principles of operation and similar technological characteristics as the predicate. The minor technological upgrades to the Surgitron 4.0 Dual RF S5 – IEC raise no new issues of safety or effectiveness. Thus, the Surgitron 4.0 Dual RF S5 – IEC is substantially equivalent to the predicate.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines that suggest feathers or wings.
#### DEPARTMENT OF HEALTH & HUM AN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
July 12, 2013
Ellman International, Inc. % Ms. Alison Sathe Director of Regulatory and Clinical Affairs 3333 Royal Avenue Oceanside, New York 11572-3625
Re: K123366
Trade/Device Name: Surgitron 4.0 Dual RF S5 - IEC Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagniation device and accessories Regulatory Class: Class II Product Code: GEI Dated: June 12, 2013 Received: June 13, 2013
Dear Ms. Sathe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Ms. Alison Sathe
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/deliault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Mark N. Melkerson -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use Statement
510(k) Number (if known): K123366
Indications for Use: The Surgitron 4.0 Dual RF S5 - IEC is indicated for:
- Non-ablative treatment of mild to moderate facial wrinkles and rhytids for skin . phototypes I-IV
- . Cutting: snoring. submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control. epistaxis treatment. turbinate shrinkage. skin incisions, biopsy, cysts, abscesses, tumors. cosmetic repairs, development of skin flaps, skin tags and blepharoplasty.
- Blended Cutting and Coagulation: snoring, submucosal palatal shrinkage, traditional . uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, turbinate shrinkage, skin tags, papilloma keloids, keratosis, verrucae, basal cell carcinoma. nevi. fistulas, epithelidma, cosmetic repairs. cysts, abscesses, and development of skin flaps.
- Hemostasis: control of blecding, epilation, telangiectasia. .
- Fulguration: basal cell carcinoma. papilloma. cvst destruction, tumors. verrucae, . hemostasis.
- Bipolar: pinpoint precise coagulation, pinpoint hemostasis in any field (wet or dry), . snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment and turbinate shrinkage.
Prescription Use x (Part 21 C.F.R. 801 Subpart D) AND/OR
Over-The-Counter Use (21 C.F.R. 807 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF
Concurrence of CDRH, Office of Device Evaluation (ODE)
| DSD-DIVISION SIGN-OFF | Joshua C. |
|------------------------------|-----------|
| Division of Surgical Devices | Nipper -S |
| 510(k) Number: | K123366 |
Page 11 of 262
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.