K123061 · Bioeconeer, Inc. · GEI · Aug 13, 2013 · General, Plastic Surgery
Device Facts
Record ID
K123061
Device Name
ACE MONOPOLAR ATTACHMENT
Applicant
Bioeconeer, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Aug 13, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ACE Monopolar Attachment is intended to be used with the compatible ERBE Monopolar Disposable Electrosurgical Pencil (model: No.20190-109) for coagulation and cutting of soft tissue when used in conjunction with compatible ERBE Electrosurgical Generator (ESU) Systems.
Device Story
ACE Monopolar Attachment is a single-use stainless steel electrode tip; designed for use with compatible ERBE Monopolar Disposable Electrosurgical Pencils and ERBE Electrosurgical Generator (ESU) Systems. Device functions by delivering High Frequency (HF) energy from the ESU through the electrode tip to the target soft tissue; enabling coagulation and cutting. Used in clinical settings by healthcare professionals. Output is thermal energy for tissue effect; directly controlled by the surgeon via the electrosurgical pencil. Benefits include precise soft tissue management during surgical procedures. Device is provided sterile via gamma irradiation.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993-1, -5, -10), electrical safety (IEC 60601-1, -1-2, -2-2), and performance testing (arcing, charring, thermal spread, dropping, and pulling tests). Results demonstrated compliance with pre-defined acceptance criteria.
Technological Characteristics
Material: Stainless steel electrode tip, Nylon insulation. Energy: High Frequency (HF) electrosurgical. Dimensions: 2.35mm shaft. Connectivity: Compatible with ERBE Monopolar 3-pin bovie receptacles. Sterilization: Gamma irradiation. Standards: ISO 10993-1/5/10, IEC 60601-1/1-2/2-2, ISO 11137-2, ASTM D7334-08.
Indications for Use
Indicated for coagulation and cutting of soft tissue in patients requiring electrosurgical procedures, when used with compatible ERBE electrosurgical pencils and generator systems.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
BioEconeer Inc. 510(k) Notification
5.4
ACE Monopolar Attachment
AUG 1 3 2
# K123061
# 510(k) Summary
5.1 Type of Submission: Traditional
5.2 Preparation Date: Sep 28, 2012
5.3 Revised Date: August 09 , 2013
> Submitter: BioEconeer Inc. Address: 10355 Norwich Avenue, Cupertino, CA 95014, U.S.A Phone: +886-989-103981 Fax: 408-252-6996 Contact: Tony ChaoFu Chang Establishment Registration Number: N/A
#### ર્સ્ટ Identification of the Device:
| Proprietary/Trade name: | ACE Monopolar Attachment |
|-------------------------|------------------------------------------------------|
| Common Name: | Electrosurgical, Cutting & Coagulation & Accessories |
| Classification Name: | Electrosurgical, Cutting & Coagulation & Accessories |
| Device Classification: | II |
| Regulation Number: | 878.4400 |
| Panel: | General & Plastic Surgery |
| Product Code: | GEI |
# 5.6 Identification of the Predicate Device:
| Predicate Device Name: | GEIGER DISPOSABLE ELECTROSURGICAL | |
|-----------------------------------------|-----------------------------------|--|
| | ELECTRODE | |
| Manufacturer: | GEIGER MEDICAL TECHNOLOGIES, INC. | |
| 510(k) Number or Clearance Information: | K994075 | |
{1}------------------------------------------------
ACE Monopolar Attachment
#### 5.7 Intended Use and Indications for Use of the subject device.
The ACE Monopolar Attachment is intended to be used with the compatible ERBE Monopolar Disposable Electrosurgical Pencil (model: No.20190-109) for coagulation and cutting of soft tissue when used in conjunction with compatible ERBE Electrosurgical Generator (ESU) Systems.
#### 5.8 Device Description
The ACE Monopolar Attachment is used with compatible ERBE Electrosurgical Generator (ESU) Systems which deliver High Frequency (HF) energy through the electrode tip of the ACE Monopolar Attachment for coagulation and cutting of soft tissue. The ACE Monopolar Attachment is an electrode tip, made of stainless steel, and compatible with marketed plastic insulation handle. The compatible ERBE Electrosurgical Unit (ESU) to be used must have a Monopolar 3-pin bovie receptacle. The Monopolar Attachment is provided single-use.
#### 5.9 Non-clinical Testing
A series of preclinical physical, mechanical and biocompatibility tests were performed to assess the safety and effectiveness of the ACE Monopolar Attachment. The safety tests were conducted in vitro and in vivo in accordance with ISO 10993-1:2009, Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process, ISO 10993-5: 2003, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity, ISO 10993-10:2002/Amd. 1:2006(E), Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity AMENDMENT 1, IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995, IEC 60601-1-2:2007, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests (Edition 3), IEC 60601-2-2:2009, Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgery equipment and high frequency surgical accessories, ISO 11137-2:2006, Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose, and ASTM D7334-08 Standard Practice for Surface Wettability of Coatings, Substrates and Pigments by Advancing Contact Angle Measurement.
{2}------------------------------------------------
BioEconeer Inc. 510(k) Notification ACE Monopolar Attachment
The Performance Tests were conducted on the subject device and the predicate device for the ex vivo test and mechanical test items listed below:
Ex vivo test: Arcing Test, Charring Test, Thermal Spread Test Mechanical test: Dropping Test, Pulling Test
All the test results demonstrate the safety and performance of ACE Monopolar Attachment meets the requirements of its pre-defined acceptance criteria and intended uses.
### 5.10Safety and Effectiveness
.
The results of the non-clinical testing demonstrate that the ACE Monopolar Attachment is substantially equivalent to the predicate devices.
#### 5.11Substantial Equivalence Determination
The ACE Monopolar Attachment submitted in this 510(k) file is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the cleared GEIGER Disposable Electrosurgical Electrode which is the subject of K994075. Differences between the devices cited in this section do not raise any new issues of safety or effectiveness.
| Item | Predicate Device<br>(GEIGER Disposable<br>Electrosurgical Electrode) | Proposed Device<br>(ACE Monopolar Attachment) |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Similarity | | |
| Classification | Class II | Class II |
| Code or Federal Regulations | 878.4400 | 878.4400 |
| Prescription Medical<br>Devices | YES | YES |
| Intended Use | The Geiger Disposable<br>Electrosurgical Electrode is<br>intended to be utilized for basic<br>non-sterile electrosurgical<br>procedures. Examples of non-sterile<br>procedures include the removal of | The ACE Monopolar<br>Attachment is intended to be<br>used with the compatible ERBE<br>Monopolar Disposable<br>Electrosurgical Pencil (model:<br>No.20190-109) for coagulation |
{3}------------------------------------------------
BioEconeer Inc. 5 10(k) Notification
## ACE Monopolar Attachment
| | | moles, warts and skin tags. The<br>electrodes are a standard 3/32" in<br>diameter and will fit the majority of<br>electrosurgical generators and<br>handpieces in the marketplace. | and cutting of soft tissue when<br>used in conjunction with<br>compatible ERBE<br>Electrosurgical Generator (ESU)<br>Systems. |
|------------------------------------------|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Material of electrode tip | | Stainless Steel | Stainless Steel |
| Dimension | | Shaft: 3/32" | Shaft: φ 2.35mm |
| Safety Standards | | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2<br>ISO 11137-2<br>ASTM D7334-08 | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2<br>ISO 11137-2<br>ASTM D7334-08 |
| Performance Standards | | Not applicable | Not applicable |
| Comparative<br>Performance<br>Test Items | <i>Ex vivo</i> | Arcing Test<br>Charring Test<br>Thermal Spread Test | |
| | Mechanical | Dropping Test | |
| | | Pulling Test | |
| Differences | | | |
| Material of insulation | | polystyrene | Nylon |
| Sterilization | | non-sterile | Gamma irradiation (Single Use) |
The differences between the subject device and predicate device are on the material of insulation and sterilization. The subject device has tested on safety and performance tests and the test results were complied with the test requests. Therefore, the differences of subject device and predicate device didn't raise any problems of safety or effectiveness. The ACE Monopolar Attachment is substantially equivalent to the predicate device in design, operation, intended use, method of preparation, and performance claims.
{4}------------------------------------------------
BioEconeer Inc. 5 10(k) Notification ACE Monopolar Attachment
# 5.12Conclusion
After analyzing bench tests. electrical safety testing data, it can be concluded that ACE Monopolar Attachment is substantially equivalent to the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an abstract human figure with outstretched arms, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the figure. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
BioEconeer Incorporated % Michael Lee President AcmeBiotechs Company, Ltd. No.45, Minsheng Road, Danshui District New Taipei City, Taiwan 251
August 13, 2013
Re: K123061
Trade/Device Name: ACE Monopolar Attachment Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: July 19, 2013 Received: July 26, 2013
Dear Mr. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Michael Lee
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm.
Sincerely yours,
Image /page/6/Picture/7 description: The image shows the name "Peter De Rumm -S" in a bold, sans-serif font. The letters are black against a white background, creating a high contrast. The letters "D" and "R" have a decorative pattern within them, adding a unique visual element to the text. The overall impression is clean and legible, with a touch of artistic flair in the letter design.
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
BioEconeer Inc. 510(k) Notification ACE Monopolar Attachment 510(k) Number:K123061
Indications for Use
510(k) Number (if known): K123061
Device Name: ACE Monopolar Attachment
Indications for Use:
The ACE Monopolar Attachment is intended to be used with the compatible ERBE Monopolar Disposable Electrosurgical Pencil (model: No.20190-109) for coagulation and cutting of soft tissue when used in conjunction with compatible ERBE Electrosurgical Generator (ESU) Systems.
Prescription Use _ X (Pari 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
1 Page I of __
| DSD-DIVISION SIGN-OFF | Joshua C. Nipper - |
|------------------------------|--------------------|
| Division of Surgical Devices | S |
| 510(k) Number: K123061 | |
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.