K120116 · Thermedical · GEI · Oct 1, 2012 · General, Plastic Surgery
Device Facts
Record ID
K120116
Device Name
SERF ABLATION SYSTEM SERF ABLATION NEEDLE
Applicant
Thermedical
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 1, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SERF Ablation System is intended for the coagulation and ablation of soft tissue during percutaneous, laparoscopic and intraoperative surgical procedures.
Device Story
Monopolar electrosurgical system for soft tissue coagulation/ablation; utilizes RF energy and saline infusion. System console houses RF generators, fluid pump, and user interface; connects to single-use sterile ablation needle and return electrode pads. Operator programs power level, procedure time, saline infusion rate, and target saline temperature. Needle penetrates tissue under visual guidance; delivers RF energy and warm saline through shaft perforations to induce thermal necrosis. Used in clinical settings (OR/clinic) by physicians. System monitors voltage and current to control power delivery. Benefits include controlled thermal ablation of targeted soft tissue.
Clinical Evidence
Bench testing included bioburden/sterility, shelf-life, biocompatibility, electrical impedance, fluid leak, and console performance (flow rate, thermometry, pump response). Software validated for major concern level. Animal studies included in-vivo ablation lesion assessment, chronic lesion pathology, and explanted tissue comparison against two predicate systems. Non-significant risk human study conducted on return electrode pad thermal rise.
Indicated for coagulation and ablation of soft tissue in patients undergoing percutaneous, laparoscopic, or intraoperative surgical procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
AngioDynamics / RITA Model 1500 Electrosurgical RF Generator (K993944, K021329)
Covidien / ValleyLab Cool-Tip RF Generator, Cool-Tip RF System and Accessories (K984552, K052796, K053290)
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) Summary
K120116
PAGE 1 OF 3
#### Company: Thermedical 150 Bear Hill Road Waltham, MA 02451-1036 United States of America
OCT 1 2012
# FDA Establishment #: not yet assigned
Contact: Katharine M Stohlman, Vice President of Regulatory Affairs Email: k.stohlman@thermedical.com Office: +1 (781) 373-3688 x118 Cell: +1 (978) 764-5154 FAX: +1 (781) 373-2520
# Proprietary: Thermedical SERF Ablation System and Accessories
Common: System, Ablation, Radio-frequency and Accessories
Classification: General and Plastic Surgery, GEI, 21 CFR 878.4400
Date Prepared: September 05, 2012
Indications for Use of the SERF™ System
The SERF Ablation System is intended for the coagulation of soft tissue during percutaneous, laparoscopic and intraoperative surgical procedures.
#### Predicate Devices
The Thermedical SERF Ablation System is substantially equivalent to the following currently marketed devices:
- AngioDynamics / RITA Model 1500 Electrosurgical RF Generator - Class II -21CFR878.4400, which has been the subject of a cleared 510(k) with FDA log number K993944, K021329 and others.
- 트 Covidien / ValleyLab Cool-Tip RF Generator, Cool-Tip RF System and Accessories - Class II - 21CFR878.4400, which has been the subject of a cleared 510(k) with FDA log number K984552, K052796 and K053290.
#### Technological Characteristics
The SERF Ablation System and SERF Ablation Needle provide heat to unwanted tissue targeted for ablation and thermal necrosis. The SERF System has these main components:
- 1. SERF Ablation System console which houses the RF generators, fluid pump, and system user interface. The system is designed to comply with Electrical Safety (IEC60601-1), Electromagnetic Compatibility (IEC60601-1-2), and Particular Requirements for the Safety of High Frequency Surgical Equipment (IEC60601-2-2). Product testing was conducted to evaluate conformance to product specifications. Software is validated.
{1}------------------------------------------------
- 2. System user interface, an alphanumeric display for menus and messages and a control panel which displays the set-point of each parameter of the therapy.
- 3. SERF Ablation Needle, a single-use, sterile device intended to penetrate tissue to locate the point of therapy initiation based on visual guidance. The SERF Needle is electrically connected to the system and delivers RF energy and warm saline to the tissue through perforations in the shaft near the needle tip. All patient contacting materials are commonly used and have been tested for biocompatibility in this application.
- 4. A sterile single-use syringe for use with the SERF system is commercially available. The syringe mounts on the pump and delivers sterile saline through a fluid delivery tube to the Ablation Needle.
- 5. Return electrode pads and cables for use with the SERF System are commercially available, The SERF system is a monopolar electrosurgical device. The RF circuit is closed through a return path including the return electrode pads which are plugged into the SERF system.
The System is programmed for the desired power level, and procedure time by the operator. Saline infusion rate and target saline temperature are also programmed. The system controls power by monitoring the voltage and current.
#### Performance Testing
The SERF Ablation System and accessories have been extensively tested to demonstrate suitability for clinical use as follows:
- . Bioburden and sterility testing of the manufactured needle
- l Sterilization dose substantiation and process validation of the manufactured needle
- 트 Environmental control and package shelf life validation of the sterilized, single-use needle
- . Biocompatibility of the sterilized, needle against requirements of tissue contacting / communicating device
- I Needle performance characterization and function tests of electrical impedance and fluid leak
- 1 Software validation for software of Maior concern level
- B Electrical safety testing according to IEC 60601-1
- Electromagnetic compatibility testing according to IEC 60601-1-2
- = High frequency surgical equipment safety testing according to IEC 60601-2-2
- 시 System console performance characterization and function tests of fluid flow rate, thermometry accuracy, pump response, power display accuracy, tipping force, and system alarm thresholds.
- . Animal studies of ablation lesions performed at academic labs:
- 1) in-vivo study of ablation lesions under different system control settings,
- 2) chronic study of lesion pathology,
- 3) explanted tissue study of SERF system ablation compared with two predicate systems, and
{2}------------------------------------------------
K120116
PAGE 3 OF 3
4) thermal rise study of return electrode pads.
- Non-significant risk human study of thermal rise of return electrode pads
## Substantial Equivalence
The SERF Ablation System has the same intended use and indications for use as other previously 510(k) cleared RF ablation systems. The technological characteristics of the SERF Ablation System are also substantially similar to other previously cleared ablation systems. Although there are minor differences in technological characteristics between the SERF Ablation System and the predicates, these differences do not raise any new types of safety or effectiveness issues.
The SERF Ablation System is compared with the Covidien™ Cool-Tip™ and the AngioDynamics™ RITA™ predicate devices. Comparison of intended use, technological features, the therapy applicator, the control methodology, and the resulting ablative lession demonstrates equivalence as confirmed by performance testing.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
OCT 1 202
Thermedical % Janice M. Hogan Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, PA 19103
K120116 Re: Trade/Device Name: SERF Ablation System and SERF Ablation Needle Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: September 14, 2012 Received: September 14, 2012
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Ms. Janice M. Hogan
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Statement
510(k) Number: K120116
Device Name: Saline-enhanced Radiofrequency (SERF) Ablation System and Accessories
Indications for Use:
The SERF Ablation System is intended for the coagulation and ablation of soft tissue during percutaneous, laparoscopic and intraoperative surgical procedures.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Signature
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K120116
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.